BAR PEDICLE SCREW SPINAL FIXATION SYSTEM, MODEL 2020

K061162 · Applied Spine Technologies, Inc. · MNH · Jul 26, 2006 · Orthopedic

Device Facts

Record IDK061162
Device NameBAR PEDICLE SCREW SPINAL FIXATION SYSTEM, MODEL 2020
ApplicantApplied Spine Technologies, Inc.
Product CodeMNH · Orthopedic
Decision DateJul 26, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The BAR Pedicle Screw Spinal Fixation System is intended for posterior non cervical pedicle screw fixation in the lumbo-sacral spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the lumbar or sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). In addition the BAR Pedicle Screw Spinal Fixation System, used as a pedicle screw fixation system is intended for: a. Patients receiving only autogenous bone graft. b. Patients having the device fixed or attached to the lumbar and sacral spine and having severe spondylolisthesis grade 3 or 4 at the fifth lumbar-first sacral (L5-S1) vertebral joint.

Device Story

Posterior pedicle screw spinal fixation system; consists of pedicle screws, rods, and connectors. Components manufactured from titanium (ASTM F-67), titanium alloy (ASTM F-136), and Cobalt Chromium Molybdenum alloy (ASTM F-1547). Features ball-and-socket connectors allowing adjustment during implantation; mechanism offloads pedicle screws and distributes stress across vertebral body; clamping mechanism secures rigid rods. Used by experienced spinal surgeons in clinical settings to stabilize spinal segments during fusion procedures. Benefits patients by providing immobilization for acute/chronic instabilities or deformities.

Clinical Evidence

Bench testing only. Testing performed per ASTM F1717; results demonstrate performance comparable to predicate devices. No clinical data provided.

Technological Characteristics

Materials: Titanium (ASTM F-67), titanium alloy (ASTM F-136), Cobalt Chromium Molybdenum alloy (ASTM F-1547). Components: Pedicle screws, rods, connectors with ball-and-socket joints and clamping mechanisms. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients requiring posterior non-cervical pedicle screw fixation in the lumbo-sacral spine for stabilization/immobilization as an adjunct to fusion. Conditions include degenerative spondylolisthesis with neurologic impairment, fracture dislocation, scoliosis, kyphosis, spinal tumor, failed previous fusion (pseudoarthrosis), and severe spondylolisthesis (grade 3 or 4) at L5-S1. Contraindications include conditions without significant mechanical instability or deformity requiring fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Bar Pedicle Screw Spinal Fixation System # 510(k) Summary JUL 26 2006 # July 25, 2006 | Submitter: | Applied Spine Technologies, Inc.<br>300 George St.<br>New Haven, CT 06511 | |------------|---------------------------------------------------------------------------| |------------|---------------------------------------------------------------------------| Contact Person: Michele Lucey Trade Name: BAR Pedicle Screw Spinal Fixation System Common Name: Posterior Pedicle Screw System Classification Name: Class II per 21 CFR Section 888.3070 Product Code: MNI/MNH Equivalent Devices: Dynesys® Dynamic Stabilization System (K031511) Manufactured and distributed by: Zimmer, Inc. > ISOBAR TTL Semi-Rigid Pedicle Screw Fusion System (K990118) SCIENT'X USA CD Horizon® Spinal System (K043488) Medtronic Sofamor Danek ## Device Description: The BAR Pedicle Screw Fixation System consists of pedicle screws, rods, and connectors. It can be used for single level fixation. All components are manufactured from titanium (ASTM F-67), titanium alloy (ASTM F-136), Cobalt Chromium Molybdenum alloy (ASTM F-1547. The system consists of pedicle screw anchors which are inserted into the pedicle, a connector that has a ball and socket feature that connects to the pedicle screws. The ball and socket feature which allows for adjustment during implantation. The motion allowed by the ball and socket serves to offload the pedicle screws and distribute the stress across the vertebral body. The connector also contains a clamping mechanism which clamps the rigid rod into place. {1}------------------------------------------------ ## Intended Use: The BAR Pedicle Screw Spinal Fixation System is intended for posterior non cervical pedicle screw fixation in the lumbo-sacral spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the lumbar or sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). In addition the BAR Pedicle Screw Spinal Fixation System, used as a pedicle screw fixation system is intended for: a. Patients receiving only autogenous bone graft. b. Patients having the device fixed or attached to the lumbar and sacral spine and having severe spondylolisthesis grade 3 or 4 at the fifth lumbar-first sacral (L5-S1) vertebral joint. ## Summary Nonclinical Tests: Testing was performed per ASTM F1717 and the results are comparable to the predicate devices. ## Basis for Substantial Equivalence: The BAR Pedicle Screw Spinal Fixation System has similar design characteristics, i.e. material, screw size, rod based, clamping connection mechanisms, and indications as the ISOBAR TTL Spinal System (k990119) and the Dynesys Dynamic Stabilization System (k031511). All three systems stabilize the spine to support fusion while allowing some motion or flexibility which serve to distribute loads across the segment being fused. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image is a black and white seal for the Department of Health and Human Services. The seal features a stylized eagle with three lines representing its body and wings. The eagle is facing right and appears to be in flight. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## JUL 26 2006 Applied Spine Technologies, Inc. % Ms. Michele Lucey Vice President, Regulatory Affairs and Quality Assurance 300 George Street, Suite 511 New Haven, Connecticut 06511 Re: K061162 Trade/Device Name: BAR Pedicle Screw Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNH, MNI Dated: June 13, 2006 Received: June 19, 2006 Dear Ms. Lucey: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Michele Lucey This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Hucher Lermer us Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K061162 Device Name: BAR Pedicle Screw Spinal Fixation System Indications for Use: ## Intended Use/Indications for Use The BAR Pedicle Screw Spinal Fixation System is intended for posterior non cervical pedicle screw fixation in the lumbo-sacral spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the lumbar or sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis). In addition the BAR Pedicle Screw Spinal Fixation System, used as a pedicle screw fixation system is intended for: a. Patients receiving only autogenous bone graft. b. Patients having the device fixed or attached to the lumbar and sacral spine and having severe spondylolisthesis grade 3 or 4 at the fifth lumbar-first sacral (L5-S1) vertebral joint. #### Warning The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the lumbar and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L551 vertebra, degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). The safety and effectiveness of these devices for any other conditions are unknown. #### Precautions The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient. Prescription Use (Part 21 CFR 801 Subpar lukin Lenu (w) (DIRECTOR Sign Of) (Division Sign-Off (21 CFR 801 Subpart C)) (PLEASE DO NOT WR **510(k) Number** k061162
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...