K Number
K143363

Validate with FDA (Live)

Date Cleared
2015-04-22

(149 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

LnK Spinal Fixation System is non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation system T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use: degenerative dis disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); deformities or curvatures (i.e., scoliosis, kyphosis); tumor, stenosis; and failed previous fusion (pseudoarthrosis).

The LnK Spinal Fixation System is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

In addition, the LnK Spinal Fixation System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

Device Description

This system is comprised of screws, set screws, rods, crosslink, Hook and connectors. The components of this system are manufactured by Titanium alloy (Titanium-6Aluminum-4Vanadium ELI, per ASTM F136) and CoCrMo alloy (Cobalt-28Chromium-6Molybdenum, per ASTM F1537). The screws are available from 4.0 to 8.5mm diameters with lengths ranging from 20-150mm. The purpose of this submission is to add Hooks to this system.

AI/ML Overview

The provided document describes a 510(k) premarket notification for the "Lnk Spinal Fixation System" and focuses on demonstrating its substantial equivalence to predicate devices, rather than a study on a new AI/software device that requires acceptance criteria and performance data in the typical sense of a diagnostic or assistive AI.

The "acceptance criteria" and "device performance" in this context refer to the mechanical and material properties of the spinal fixation system, compared against established standards and predicate devices.

Here's the information extracted and interpreted based on the content:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Bench Test Standard)Reported Device Performance (L&K Spinal Fixation System)
Static Compression Bending (ASTM F1717-10)Performs equivalently to predicates
Static TensionPerforms equivalently to predicates
Static TorsionPerforms equivalently to predicates
Dynamic Compression Bending (ASTM F1717-10)Performs equivalently to predicates
Gripping-Push Down (ASTM F1798)Performs equivalently to predicates
Material CompositionTitanium alloy (ASTM F136) and CoCrMo alloy (ASTM F1537)
SterilityNon-sterile

2. Sample sized used for the test set and the data provenance

The document does not specify the exact sample sizes (number of fixation devices tested) for each bench test. The provenance is "Bench testing," meaning laboratory-controlled mechanical tests, not human or patient data. Details on the specific lab or country of origin for these tests are not provided, other than the manufacturer being based in the Republic of Korea. These are non-clinical studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is a mechanical device, and "ground truth" is established by adherence to ASTM standards and comparison to predicate devices, not by expert consensus on clinical data.

4. Adjudication method for the test set

Not applicable, as this involves mechanical bench testing against established standards, not human expert adjudication of clinical outcomes or interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a spinal fixation system, not an AI/software device, and therefore no MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a spinal fixation system, not an AI/software device.

7. The type of ground truth used

The "ground truth" for the performance claims is defined by established ASTM standards for spinal implant testing (e.g., ASTM F1717-10, ASTM F1798) and the performance characteristics of legally marketed predicate devices. The goal is to demonstrate "substantial equivalence" in mechanical performance.

8. The sample size for the training set

Not applicable. This is not an AI/software device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/software device that requires a training set or its associated ground truth establishment.

Summary of the Study:

The study described is a non-clinical bench testing study designed to demonstrate the mechanical and material substantial equivalence of the "Lnk Spinal Fixation System" to predicate devices. The study involved subjecting the device components (screws, rods, hooks, connectors) to various mechanical tests, including static compression bending, static tension, static torsion, dynamic compression bending, and gripping-push down, in accordance with recognized ASTM standards (F1717-10 and F1798). The performance was then compared to that of the predicate devices. The conclusion was that the Lnk Spinal Fixation System performs equivalently to the predicates, sharing similar design, materials, and performance characteristics, thus supporting its substantial equivalence for the stated indications for use. No clinical data or AI performance studies were involved or required for this 510(k) submission.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of a human face in profile, with three overlapping profiles to represent the department's focus on health, human services, and the well-being of the nation. The seal is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 22, 2015

L&K Biomed Company, Limited Ms. Yerim An #201, 202 16-25, Dongbaekjungang-ro 16 beon-gil Giheung-gu, Yongin-si Gyeonggi-do, 446-916 Republic of Korea

Re: K143363

Trade/Device Name: LnK Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI, KWP, KWQ Dated: March 19, 2015 Received: March 23, 2015

Dear Ms. An:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143363

Device Name LnK Spinal Fixation System

Indications for Use (Describe)

LnK Spinal Fixation System is non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation system T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use: degenerative dis disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); deformities or curvatures (i.e., scoliosis, kyphosis); tumor, stenosis; and failed previous fusion (pseudoarthrosis).

The LnK Spinal Fixation System is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

In addition, the LnK Spinal Fixation System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):

1.Submitter:Gook Jin KangL&K BIOMED Co., Ltd.#201, 202 16-25, Dongbaekjungang-ro 16 beon-gilGiheung-gu, Yongin-si, Gyeonggi-do, 446-916,KoreaPhone. 82-2-6717-1985FAX .82-2-6717-1989
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Contact Person: Yerim An

March 19, 2015 Date prepared:

Device Identification 2.

Trade NameLnK Spinal Fixation System
Common NameSpinal Fixation Appliances
Product CodeNKB, KWP, KWQ, MNH, MNI
Classification NameClass III
Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050Pedicle Screw Spinal System, 21 CFR §888.3070

3. Predicate or legally marketed devices which are substantially equivalent

4. Description of the Device

This system is comprised of screws, set screws, rods, crosslink, Hook and connectors. The components of this system are manufactured by Titanium alloy (Titanium-6Aluminum-4Vanadium ELI, per ASTM F136) and CoCrMo alloy (Cobalt-28Chromium-6Molybdenum, per ASTM F1537). The screws are available from 4.0 to 8.5mm diameters with lengths ranging from 20-150mm. The purpose of this submission is to add Hooks to this system.

{4}------------------------------------------------

5. Indications for Use

LnK Spinal Fixation System is non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation systems (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use:

  • degenerative disc disease (defined as discogenic back pain with degeneration of the disc ● confirmed by history and radiographic studies);
  • spondylolisthesis; ●
  • trauma (i.e., fracture or dislocation);
  • deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis);
  • tumor:
  • stenosis, and
  • failed previous fusion (pseudoarthrosis)

The LnK Spinal Fixation System is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the LS-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

In addition, the LnK Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis).The pedicle screws are limited to placement in T1-T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine.

6. Comparison of the technology characteristics of the device to predicate and legally marketed devices

SubmitterL&K BIOMEDStryker
Device NameLnK Spinal Fixation systemXia 3 Spinal System
510K #K071373(T), K083393 (T),K091291 (S)
RegulationNumber21 CFR 888.3050, 21 CFR 888.3060,21 CFR 888.307021 CFR 888.3050, 21 CFR 888.3060, 21 CFR 888.3070
ClassIIIIII
ProductCodeMNI,MNH,KWQ,KWP,NKBMNI,MNH,KWQ,KWP,NKB
MaterialASTM F136, ASTM F1537ASTM F136, ASTM F1537
SterilityNon-sterileNon-sterile
RodDia: 5.0/5.5/6.0mmLength:40~600mmDia:5.5 mm Length:30-600mm
ScrewDia:4.0~8.5mm Length:20-150mmDia: 4.0.-10.5mm Length: 20-100mm
HooksRamped Hook - 6.0 Rod(Narrow,Standard, Wide)General Hook - 6.0 Rod(Narrow,Standard, Wide)Pedicle Hook - 6.0 Rod(Narrow,Standard, Wide)Angled Hook - 6.0 Rod(Right, Left)Offset Hook - 6.0 Rod(Right, Left)Laminar Hook(Small Medium Large, Extended Body, Offset), (Narrow Blade,Standard Blade)Thoracic Laminar Hook (Small Medium Large, Extended Body,Offset)Offset Hook (Right, Left)Pedicle Hook (Small, Medium, Large)Transverse Process Hook (Right, Left)
IndicationLnK Spinal Fixation System is non-The Stryker Spine XIA® 3 Spinal System is intended for use in the

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image provides information on cervical and noncervical spinal fixation devices. Cervical spinal fixation devices are intended for use as posterior pedicle screw fixation systems (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). The LnK Spinal Fixation System is a pedicle screw system indicated for the treatment of severe Spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra. The XIA® 3 Spinal System is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients.

The LnK Spinal Fixation System shares technological characteristics similar to the predicate devices. These characteristics include similar design, the same materials, substantially equivalent performance characteristics and the same intended use.

7. Performance Data

Non-Clinical Performance and Conclusions:

Bench testing results demonstrate that LnK Spinal Fixation system performs equivalently to the predicates in static compression bending, static tension, static torsion, dynamic compression bending (in accordance with ASTM F1717-10) and gripping-push down (in accordance with ASTM F1798). Clinical Performance and Conclusions:

Clinical data and conclusions were not needed for this device.

8. Conclusion

The LnK Spinal Fixation System is substantially equivalent to the predicate device referenced above.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.