K Number
K043488

Validate with FDA (Live)

Date Cleared
2005-03-21

(94 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CD HORIZON® Spinal System is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

When used in a percutaneous, non-cervical, posterior approach with the SEXTANT instrumentation, the CD HORIZON® screws are intended for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system, CD HORIZON® components such as ECLIPSE® components are intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion.

The CD HORIZON® SPIRE Plate is posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1 – S1). It is intended for plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies; spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX™ Reconstruction System with the VERTEX™ rod connector. Refer to the VERTEX™ Reconstruction System Package Insert for a list of the VERTEX™ indications of use.

Device Description

The CD HORIZON® Spinal System consists of a variety of rods, hooks, screws, CROSSLINK® plates, staples, and other connecting components used to build a spinal construct. Instrumentation is also available to facilitate implantation of the device components.

The CD HORIZON® Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine. The CD HORIZON® Spinal System implant components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. If necessary, the CD HORIZON® Spinal System can be connected to the VERTEX™ Reconstruction System through a rod connector.

Certain implant components from other Medtronic Sofamor Danek spinal systems can be used with the CD HORIZON® Spinal System. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples and washers; GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK PLUS® bolts; and Medtronic Sofamor Danek Multi-Axial rods and screws.

CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORIZON® ECLIPSE® rods and screws are intended for anterior use only. However, for patients of smaller stature, CD HORIZON® 4.5mm rods and associated components may be used posteriorly.

The purpose of this 510(k) submission is to add modified 5.5mm and 6.35mm diameter cobalt-chromium-molybdenum rods, and set screws to the CD HORIZON® Spinal System.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the CD HORIZON® Spinal System based on the provided text.

This document describes a 510(k) submission for modifications to an existing spinal system, primarily focusing on adding new rod and set screw components. The "study" mentioned here refers to mechanical testing for substantial equivalence, not a clinical trial or algorithm performance study as typically understood for AI/software devices. Therefore, many of the requested points regarding AI/algorithm performance are not applicable.


Acceptance Criteria and Device Performance (Mechanical Testing)

Acceptance CriteriaReported Device PerformanceComments
Substantial Equivalence to Predicate DevicesDemonstrated substantial equivalence to predicate CD HORIZON® Spinal System components previously cleared in K031655, K041777, K040962, K040583, and K981676.The specific parameters (e.g., strength, fatigue life) used for judging substantial equivalence are not detailed in this summary for the mechanical testing, but the overall conclusion of equivalence is stated.

Study Details (Focusing on Mechanical Testing for Substantial Equivalence)

Since this is a submission for a physical medical device (spinal system components) and not an AI/software device, many of the typical AI-related study questions are not applicable.

  1. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not specified in the provided text. The testing would involve a sufficient number of samples of the newly introduced components (5.5mm and 6.35mm diameter cobalt-chromium-molybdenum rods and set screws) and predicate components to perform comparative mechanical characterization.
    • Data Provenance: The nature of the study (mechanical testing) indicates the "data" would be generated in a lab setting. The company, Medtronic Sofamor Danek, Inc., is based in the USA.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Not applicable. For mechanical testing, "ground truth" is established by engineering standards and validated testing protocols rather than expert human interpretation. The testing would be conducted by qualified engineers or technicians.
  3. Adjudication Method for the Test Set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 are for expert consensus in clinical contexts. Mechanical test results are quantitative and compared against predefined engineering specifications or predicate device performance.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    • Not applicable. This is a physical device, not an AI or imaging diagnostic tool. An MRMC study is irrelevant here.
  5. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done:

    • Not applicable. This device does not involve an algorithm.
  6. The Type of Ground Truth Used:

    • For mechanical testing, the "ground truth" would be established by engineering specifications, industry standards (e.g., ASTM, ISO for spinal implants), and the established performance characteristics of the predicate devices. The goal is to demonstrate that the new components perform equivalently or better in terms of mechanical properties (e.g., strength, fatigue, stiffness).
  7. The Sample Size for the Training Set:

    • Not applicable. This is not a machine learning model. There is no "training set."
  8. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As above, no training set is involved.

Summary of the 510(k) Submission:

The provided document K043488 is for a modification to an existing spinal implant system, the CD HORIZON® Spinal System. The primary purpose of this specific 510(k) was to add modified 5.5mm and 6.35mm diameter cobalt-chromium-molybdenum rods and set screws. The demonstration of safety and effectiveness for such modifications typically relies on mechanical testing to show that the new components are substantially equivalent to previously cleared predicate devices and meet the same performance standards. The FDA's clearance confirms this substantial equivalence.

{0}------------------------------------------------

K043488

CD HORIZON® Spinal System Summary of Safety and Effectiveness March 2005

Medtronic Sofamor Danek, Inc. USA I. Company: 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133

Richard W. Treharne, PhD Contact: Senior Vice President Regulatory Affairs

  • Proposed Proprietary Trade Name: CD HORIZON® Spinal System II.
  • Classification Name(s)/Product Code(s): Spinal Interlaminal Fixation and Spinal III. Intervertebral Fixation Orthosis and/or Pedicle Screw Spinal System (per 21 CFR Section 888.3050, 888.3060 and/or 888.3070). Product Codes: MNI, MNH, KWP, KWQ and NKB

Product Description IV.

The CD HORIZON® Spinal System consists of a variety of rods, hooks, screws, CROSSLINK® plates, staples, and other connecting components used to build a spinal construct. Instrumentation is also available to facilitate implantation of the device components.

The CD HORIZON® Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine. The CD HORIZON® Spinal System implant components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. If necessary, the CD HORIZON® Spinal System can be connected to the VERTEX™ Reconstruction System through a rod connector.

Certain implant components from other Medtronic Sofamor Danek spinal systems can be used with the CD HORIZON® Spinal System. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples and washers; GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK PLUS® bolts; and Medtronic Sofamor Danek Multi-Axial rods and screws.

CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORIZON® ECLIPSE® rods and screws are intended for anterior use only. However, for patients of smaller stature, CD HORIZON® 4.5mm rods and associated components may be used posteriorly.

The purpose of this 510(k) submission is to add modified 5.5mm and 6.35mm diameter cobaltchromium-molybdenum rods, and set screws to the CD HORIZON® Spinal System.

{1}------------------------------------------------

V. Indications

The CD HORIZON® Spinal System is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with ronowing indications a confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

When used in a percutaneous, non-cervical, posterior approach with the SEXTANT instrumentation, the CD HORIZON® screws are intended for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system, CD HORIZON® components such as ECLIPSE® components are intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis. (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis). (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion.

The CD HORIZON® SPIRE Plate is posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (TI – S1). It is intended for plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies; spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX™ Reconstruction System with the VERTEX™ rod connector. Refer to the VERTEX™ Reconstruction System Package Insert for a list of the VERTEX™ indications of use.

Substantial Equivalence V.

Mechanical testing was performed and demonstrated that the subject components are substantially equivalent to predicate CD HORIZON® Spinal System components previously cleared in K031655 (SE 06/27/03), K041777 (09/03/04) and K040962 (5/14/04), K040583 (03/31/04) and K981676 (SE 01/28/99).

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's logo, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" are arranged in a circular pattern around the logo.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 1 2005

Richard Treharne, Ph.D. Senior Vice President Regulatory Affairs Medtronic Sofamor Danek 1800 Pyramid Place Memphis, Tennessee 38132

Re: K043488

R043460
Trade/Device Name: CD HORIZON® Spinal System Trade/Device Name: CD HOREDIN Spinal - Spinal - Spinal 21 CFR 888.3060, and 21 CFR 888.3050 Regulation Name: 21 OFF Society Spinal Intervertebral Body Fixation Orthosis, and Spinal Interlaminal Fixation Orthosis Regulatory Class: III Product Code: NKB, MNH, MNI, KWQ, and KWP Dated: February 22, 2005 Received: February 23, 2005

Dear Dr. Treharne:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your Section 910(s) pe device is substantially equivalent (for the indications felerenced above and nave determined by marketed predicate devices marketed in interstate for use stated in the encrosure) to regard cate of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, devices that have been recuire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, therefore, market the act include requirements for annual registration, listing of general controls provisions of the rist labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it n your device is classified (500 above) me. Existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulations of th may be subject to subli additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean 1 least be advised that I DTF 31.00am.co evice complies with other requirements of the Act that 1177 has made a acted regulations administered by other Federal agencies. You must or any i cachar statutes und regarants, including, but not limited to: registration and listing (21 comply with an the 11et 31et 31equirements(01); good manufacturing practice requirements as se CI N I art 6077, idooning (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Richard Treharne, Ph.D.

This letter will allow you to begin marketing your device as described in your Section 510(k) I ms letter will anow you to begin manceing your actively of your device to a legally premitation. The I Driving of our maills of basis on for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice ion your ac now to views note the regulation entitled, Colliact the Office of Compulation and et notification" (21CFR Part 807.97). You may obtain other Misoranding by reference to promabilities under the Act from the Division of Small general intornation on your responsionnee assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours,

Sincerely yours,

Mark A. Milliman

Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

(Division Sign-Off)

Division of General, Restorative,

and Neurological Devices

Page 1 of 1

December 2004

510(k) Number (if known):
510(k) Number:K043488
Device Name:CD HORIZON® Spinal System

Indications for Use:

The CD HORIZON® Spinal System is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; degenciation of the disocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

When used in a percutaneous, non-cervical, posterior approach with the SEXTANT instrumentation, the CD HORIZON® screws are intended for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degencration of the disc acgeneral by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or continued by instery and ristory atures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system, CD HORIZON® components such as ECLIPSE® components are intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, and/or lordosis), (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion.

The CD HORIZON® SPIRE Plate is posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1 -- S1). It is intended for plate fixation/attachment to spinous process for the not of the spassion in the following conditions: to spinous process for the parfood as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies; spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be in order to the VERTEX™ Reconstruction System with the VERTEX™ rod connector. Refer to the VERTEX™ Reconstruction System Package Insert for a list of the VERTEX™ indications of use.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

N/A