X-trodes System M

K232210 · X-Trodes · GWL · Feb 4, 2024 · Neurology

Device Facts

Record IDK232210
Device NameX-trodes System M
ApplicantX-Trodes
Product CodeGWL · Neurology
Decision DateFeb 4, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1835
Device ClassClass 2

Indications for Use

The X-trodes System M is intended for prescription use only in the home, healthcare facility, or clinical research environment to acquire, record, transmit and display physiological signals from adult patients. The X-trodes System M acquires, records, transmits, and displays electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG), and accelerometer and gyroscope signals. The X-trodes System M only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision, and reliability.

Device Story

X-trodes System M acquires physiological signals (EEG, EOG, ECG, EMG, accelerometer, gyroscope) via data acquisition unit connected to dry electrode array patches. Applied by technician or patient. Data recorded and transmitted to cloud; converted to European Data Format (EDF) for third-party software analysis. Used in home, healthcare facility, or clinical research environment. Device provides raw signal acquisition and display; no automated diagnostic analysis or interpretation performed. Healthcare providers use displayed signals for clinical decision-making. Benefits include non-invasive, multi-modal physiological monitoring in flexible settings.

Clinical Evidence

Case-controlled study of 55 adult subjects comparing X-trodes System M to reference device (X8). 20-minute duration in clinic. Primary endpoint: proportion of interpretable readings. Agreement proportions: ECG 89.36%, EEG 97.37%, EMG 96.15%, EOG 95.24%. All lower 98.75% CIs > 60%. Secondary endpoint: RMS value comparison confirmed consistency and reliability of measurements.

Technological Characteristics

Physiological signal amplifier with 16-bit A/D conversion, 4000 Hz/channel sampling rate, and 0.35-700 Hz frequency response. Input impedance ≥10 MΩ. Uses dry electrode arrays. Connectivity via Bluetooth Low Energy. Applied part type CF. Complies with IEC 60601-1, IEC 60601-2-25 (ECG), ISO/IEC 80601-2-26 (EEG), and IEC 60601-2-40 (EMG).

Indications for Use

Indicated for adult patients requiring acquisition, recording, transmission, and display of EEG, EOG, ECG, EMG, and accelerometer/gyroscope signals in home, healthcare, or clinical research settings.

Regulatory Classification

Identification

A physiological signal amplifier is a general purpose device used to electrically amplify signals derived from various physiological sources (e.g., the electroencephalogram).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ February 4, 2024 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, the FDA logo is displayed in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" written in a clear, sans-serif font. The overall design is clean and professional, reflecting the organization's role in public health and safety. X-Trodes % Donna-Bea Tillman Senior Consultant Biologics Consulting Group 100 Daingerfield Rd, Suite 400 Alexandria, VA 22314 Re: K232210 Trade/Device Name: X-trodes System M Regulation Number: 21 CFR 882.1835 Regulation Name: Physiological Signal Amplifier Regulatory Class: Class II Product Code: GWL, GXY, DPS, IKN Dated: January 4, 2024 Received: January 4, 2024 Dear Donna-Bea Tillman: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Patrick Antkowiak -S for Jay Gupta Assistant Director {2}------------------------------------------------ DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232210/S001 Device Name X-trodes System M ### Indications for Use (Describe) The X-trodes System M is intended for prescription use only in the home or healthcare facility to acquire, record, transmit and display physiological signals from adult patients. The X-trodes System M acquires, records, transmits, and displays electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG), and accelerometer and gyroscope signals. The X-trodes System M only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision, and reliability. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ #### 1. SUBMITTER INFORMATION | Applicant: | X-trodes | |------------|----------------------------------| | Contact: | Ziv Peremen, Ph.D | | Phone: | +972 54 2244811 | | Email: | Ziv@xtrodes.com | | Address | 4 Maskit St.<br>Herzliya, Israel | #### 2. CORRESPONDENT INFORMATION | Contact: | Donna-Bea Tillman, Ph.D. | |----------|----------------------------| | Title: | Senior Consultant | | Firm: | Biologics Consulting Group | #### 3. -DATE PREPARED: FEBRUARY 2, 2024 #### 4. DEVICE INFORMATION | Device Name: | X-trodes System M | |--------------------|---------------------------------------------------| | Common Name: | Reduced-Montage Standard<br>Electroencephalograph | | Regulation Number: | 21 CFR 882.1835 | | Regulation Name: | Physiological signal amplifier | |-----------------------------|--------------------------------| | Primary Product Code: | GWL | | Secondary Product:<br>Codes | GXY, DPS, IKN | | Regulatory Class: | Class II | #### 5. PREDICATE DEVICE INFORMATION | Primary EEG/EOG Predicate Device Name: | Quick-20m | |----------------------------------------|-----------| | 510(k) Number: | K203331 | | Manufacturer: | CGX, LLC | {5}------------------------------------------------ | ECG Predicate Device Name: | Electrocardiograph (SE-1200 Pro, SE-1201 Pro) | |----------------------------|-----------------------------------------------| | 510(k) Number: | K222902 | | Manufacturer: | Edan Instruments, Inc. | | EMG Predicate Device Name: | Focus EMG | | 510(k) Number: | K102610 | | Manufacturer: | TeleEMG, LLC | The predicate devices have not been subject to a design related recall. #### DEVICE DESCRIPTION 6. The X-trodes System M combines hardware, firmware, and software to acquire the following physiological signals: physiologic signal amplifier (EEG), electrooculography (EOG), surface electromyography (sEMG), electrocardiography (ECG), and accelerometer and gyroscope signals. It acquires physiological data through a data acquisition unit connected to electrode arrays patches, applied by a technician or patient to the patient. The data is recorded and transmitted to a cloud where it is converted to an EDF (European Data Format) format, suitable for analysis by third party software. #### 7. INDICATIONS FOR USE The X-trodes System M is intended for prescription use only in the home, healthcare facility, or clinical research environment to acquire, record, transmit and display physiological signals from adult patients. The X-trodes System M acquires, records, transmits, and displays electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG), and accelerometer and gyroscope signals. The X-trodes system M only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision, and reliability. ### COMPARISON OF INTENDED USE AND TECHNOLOGICAL 8. CHARACTERISTICS WITH THE PREDICATE DEVICE The table below compares the intended use and the technological characteristics of the subject device and predicate device. {6}------------------------------------------------ | | Subject Device | Primary Predicate<br>EEG/EOG | Predicate ECG | Predicate EMG | Comments | |-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | K232210 | K203331 | K222902 | K102610 | | | Applicant | Xtrodes Ltd. | CGX, LLC | Edan Instruments, Inc. | TeleEMG, LLC | | | Device Name | X-trodes System M | Quick-20m | Electrocardiograph (SE-<br>1200 Pro, SE-1201 Pro) | Focus EMG | | | Classification<br>Regulation | 21 CFR 882.1835<br>Physiological signal | 21 CFR 882.1835<br>Physiological signal | 21 CFR 870.2340<br>Electrocardiograph | 21 CFR 890.1375<br>Diagnostic | | | | amplifier<br>21 CFR 882.1320<br>Cutaneous Electrode<br>21 CFR 870.2340<br>Electrocardiograph<br>21 CFR 890.1375<br>Diagnostic<br>Electromyograph | amplifier<br>21 CFR 882.1320<br>Cutaneous Electrode | | Electromyograph | | | Product Code | GWL, GXY, DPS, IKN | GWL, GXY | DPS | IKN | | | Intended Use | The X-trodes System M is<br>intended for prescription<br>use only in the home,<br>healthcare facility, or<br>clinical research<br>environment to acquire,<br>record, transmit and<br>display physiological<br>signals from adult patients.<br>The X-trodes System M<br>acquires, records,<br>transmits, and displays<br>electroencephalogram<br>(EEG), electrooculogram<br>(EOG), electrocardiogram<br>(ECG), and/or<br>electromyogram (EMG),<br>and accelerometer and<br>gyroscope signals. The X-<br>trodes System M only | The Quick-20m is<br>intended to be used to<br>acquire<br>electroencephalograph<br>(EEG) and transmit it<br>wirelessly to a computer. | The SE-1200 Pro&SE-<br>1201 Pro 12-lead<br>electrocardiographs are<br>intended to acquire ECG<br>signals from adult and<br>pediatric patients through<br>body surface ECG<br>electrodes. The<br>electrocardiographs are<br>only intended to be used<br>in hospitals or healthcare<br>facilities by doctors and<br>trained healthcare<br>professionals. The<br>cardiogram recorded by<br>the electrocardiograph can<br>help users to analyze and<br>diagnose heart disease.<br>However, the interpreted<br>ECG with measurements | The Focus is intended for<br>use by a healthcare<br>provider to perform nerve<br>conductions and EMG<br>studies as an aid in the<br>evaluation of patients<br>with diseases of muscle<br>and nerves. The machine<br>can also use electrical<br>stimulus or sound<br>stimulus for evoked<br>potentials (EP) studies. | | | | Subject Device | Primary Predicate<br>EEG/EOG | Predicate ECG | Predicate EMG | Comments | | | acquires and displays<br>physiological signals, no<br>claims are being made for<br>analysis of the acquired<br>signals with respect to the<br>accuracy, precision, and<br>reliability. | | and interpretive<br>statements is offered to<br>clinicians on an advisory<br>basis only. | | | | Patient<br>Population | Adults | Adults | Adult and pediatric<br>patients | Adults | Same | | Environment of<br>Use | Home, Healthcare<br>facility, | Home or healthcare<br>facility setting | Hospitals or healthcare<br>facilities | Clinical environment. | The subject device and the<br>primary predicate device are<br>intended for use in either<br>healthcare facility or home<br>settings | | Signals Acquired | Forehead/head EEG, EOG,<br>EMG, ECG, Microphone,<br>Accelerometer, Gyroscope | EEG/EOG | ECG | EMG | The subject device is a<br>combination of the features<br>of the predicate devices | | Full scale Input<br>Range | ± 12.5 mV | ± 300 mV | No information in the<br>predicate 510(k)<br>Summary | No information in the<br>predicate 510(k)<br>Summary | See Discussion item 1 | | A/D conversion | 16 bits | 24 bits | 24 bits | 16 Bits | Same for EMG<br>See Discussion item 2 | | Sampling<br>Frequency (Rate) | 4000 Hz/Channel | 500 Hz/channel | 64,000 Hz | 200-80000 Hz | See Discussion item 3 | | Frequency<br>Response | 0.35 Hz ~ 700 Hz (-3 dB) | Linear between 0.1<br>and 100 Hz | 0.01 Hz ~ 350 Hz (-3 dB) | No information in the<br>predicate 510(k) | XTR frequency range<br>supersedes the predicates,<br>thus constitutes SE. | | Input Impedance | ≥10 MΩ | >100 MOhm (125 Hz) | ≥100 MΩ(10 Hz) | ≥100 MOhm<br>< 25pF | See Discussion item 4 | | DC Offset Voltage | +3000mV, -1000mV ±5%. | No information in the<br>predicate 510(k) | ±960mV, ±5% | No information in the<br>predicate 510(k) | See Discussion item 5 | | Noise | <0.9 µV RMS,<br>1-35 Hz<br>≤6 µVp-p 0.5-50Hz | <0.6 µV RMS,<br>1-30 Hz | ≤12.5 µVp-p | < 0. 6 µVrms | See Discussion item 6 | | Filter | Analog: 0.35-700Hz | DC up to 131 Hz | AC Filter:<br>50 Hz / 60 Hz / Off | Low pass filter: | Analog filter is applied prior<br>to A/D conversion. | | | Subject Device | Primary Predicate<br>EEG/EOG | Predicate ECG | Predicate EMG | Comments | | | Digital display filters:<br>EEG: 0.32-35 Hz<br>EOG: 0.32-35 Hz<br>ECG: 0.32-70 Hz<br>EMG: 10-114 Hz | | DFT Filter:<br>0.01 Hz / 0.05 Hz / 0.32<br>Hz / 0.67 Hz<br>EMG Filter:<br>25Hz / 35Hz / 45Hz /<br>OFF<br>LOWPASS Filter:<br>350 Hz / 300 Hz / 270 Hz<br>/ 150 Hz / 100 Hz<br>/ 75 Hz | (-12dB/octave) 10 to<br>10,000<br>High pass filter:<br>(-6dB/octave) 0.05 to<br>3000 Hz | Data is delivered raw.<br>Modality Filters are applied<br>for Real-time display only. | | Notch filter | 50/60 Hz selectable | No information in the<br>predicate 510(k) | No information in the<br>predicate 510(k) | 50/60 Hz selectable | Filters are applied for Real-<br>time display only. | | CMRR | >95 dB at 60 Hz/50Hz | >90 dB at 60 Hz | ≥140 dB (AC filter on)<br>≥123 dB (AC filter off) | >100 dB | For EEG - Similar to<br>predicates<br>For ECG and EMG<br>See Discussion item 7 | | Connection | Bluetooth Low Energy | Wireless to receiver, USB<br>to computer | WIFI, 4G, Bluetooth | USB to computer | Similar to predicates | | Number Of<br>Possible<br>Electrodes | 16 | 20 Channels in<br>standard 10-20 head<br>map | 10 | 2 | Similar to predicates range.<br>Differ for each predicate for<br>the different<br>electrophysiological signal<br>application | | Type of<br>Electrodes | Dry | Dry | Body surface ECG<br>electrodes | Stainless steel electrodes.<br>No additional information<br>is supplied in the<br>predicate 510(k) summary | Similar to the EEG/EOG<br>predicate.<br>FDA has found dry<br>electrodes to be<br>substantially equivalent to<br>gel electrodes. | | Type Of Applied<br>Part | CF | BF | CF | No information in the<br>predicate 510(k) | Similar to predicates | ### Table 1: Comparison of Technological Features {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ #### 9. PREDICATE DEVICE DISCUSSION ITEMS # 1. Full Scale Input Range A typical adult human EEG signal is about 10 µV to 100 µV in, so even if we consider 10-fold increase to this range (1mV), it is well covered within the X-trodes device's inputs range of 12.5 mV. Any input range larger than that is unnecessary. For ECG, there is no information about the predicate device's input range. Therefore, we aim to establish that the X-trodes device has an input range suitable for ECG. "Normal ECG signals are time-varying signals with a small amplitude ranging from 10 µV to 5 mV" (cited from PubMed PMC7664289). The X-trodes device's input range of 12.5 mV is more than double this range. For EMG, there is no information about the predicate device's input range. Therefore, we aim to establish that the X-trodes device has an input range suitable for EMG. "The amplitude range of EMG signal is 0-10 mV (+5 to -5) prior to amplification" (cited from PubMed PMC1455479). X-trodes device's input range of 12.5 mV is more than double this range. Lastly, the results of the clinical study presented in Section 20 of the Original 510(k) submission demonstrate that the quality of the EEG/ECG/EOG/EMG waveforms is sufficient for clinical purposes. Taken together, these considerations are sufficient to demonstrate that the X-trodes device is substantially equivalent to the predicates in this aspect. # 2. A/D conversion For EEG, a typical adult human EEG signal is about 10 µV to 100 µV. If we consider 10fold this signal, it takes constitutes a portion of about 1000µV (1/300) of the predicate device's input range. For ECG, there is no information about the predicate devices input range. However, based on the DC offset of the predicate device, we assume it to be 960mV, to cater for handling this DC offset. Since the X-Trodes device is AC coupled, it has no need to digitize all this vast DC range and ADC conversion is applied only on the AC signal. A typical adult human ECG signal, which is about 5mV amplitude. It constitutes a portion of about 5mV/960mV (0.0166) of the assumed input range. This portion practically utilizes only 17 bits out of its 24-bit resolution, according to the following calculation: Quantization step is 960,000μV/(2^24-1)= 0.057μV. Digital representation of 5mV would be 5000uV/0.057μV=87,381 (binary 1 0101 0101 0101 0101) which is 17 bits. FDA has also cleared many ECG devices that have a 16-bit resolution. Lastly, the results of the clinical study presented in Section 20 of the Original 510(k) submission demonstrate that the quality of the EEG/ECG/EOG/EMG waveforms is sufficient for clinical purposes. Taken together, these considerations are sufficient to demonstrate that the X-trodes device is substantially equivalent to the predicates in this aspect. # 3. Sampling frequency (Rate) {10}------------------------------------------------ Sampling rate should be considered in conjunction with the frequency response and should be at least 220% of the max sampled frequency to practically comply with Nyquist's law of digitation. Therefore, the discussion should be whether the device has sufficient sampling rate to cover its frequency range. The X-Trodes device has a frequency range of 750Hz, which is broader than the predicate device's (350Hz). This frequency range can be accommodated by a sampling rate of 1.650Hz. Therefore, the X-Trodes device's sampling rate of 4KHz exceeds the required sampling rate and qualifies as substantially equivalent in this aspect. In addition, the results of the clinical study presented in Section 20 of the Original 510(k) submission demonstrate that the quality of the EEG/ECG/EOG/EMG waveforms is sufficient for clinical purposes. Taken together, these considerations are sufficient to demonstrate that the Xtrodes device is substantially equivalent to the predicates in this aspect. # 4. Input impedance The input impedance of the X-Trodes device complies with clause 201.12.4.103 of the IEC60601-2-25 ECG standard which states that "The input impedance shall be at least 2,5 MS2 within a d.c. offset voltage range of ±300 mV". In addition, FDA has cleared EEG devices such as the Digital NeuroPort Biopotential Signal Processing System(K202174) that also have an input impedance of ≥10 MΩ. Lastly, the results of the clinical study presented in Section 20 of the Original 510(k) submission demonstrate that the quality of the EEG/ECG/EOG/EMG waveforms is sufficient for clinical purposes. Taken together, these considerations are sufficient to demonstrate that the X-trodes device is substantially equivalent to the predicates in this aspect. # 5. DC offset The X-Trodes device is AC coupled, thus its DC offset limits are derived from the OV/UV protection devices which are set at +3.0v and -1 V. This DC offset range is broader than the predicate's DC offset range, rendering the XTR system more tolerant to DC offset, thus the Xtrodes device is substantially equivalent in this aspect. ## 6. Noise For ECG, the X-trodes device's noise is smaller than the predicate device's, thus constitutes substantial equivalence. For EEG, the X-trodes device's noise is slightly higher than the predicate device, however it relates to only a portion of the EEG spectrum. Furthermore, the overall noise performance over the full EEG spectrum of the X-trodes device, complies with the IEC60601-2-26 EEG standard as shown in the Bench Test Report. For EMG, the X-trodes device's noise is slightly higher than the predicate device's, however, it was demonstrated in the clinical trial that the measured EMG signal was of the required signal quality. Moreover, since there is no standard noise requirement for EMG, and since EMG signal's dynamic input range is similar to ECG, compliance with ECG noise standard can fulfill this requirement. The X-Trodes device complies with ECG noise requirement of 30 µ V peak-tovalley referred to the input (IEC60601-2-25). Therefore, the X-Trodes device is substantially equivalent in this aspect. {11}------------------------------------------------ ## 7. CMRR The X-Trodes device CMRR complies with the applicable provisions for both- IEC60601-2-26 EEG and IEC60601-2-25 ECG standards of CMRR > 90db, which constitutes adequate performance. ## Conclusion In conclusion, the subject X-trodes device has the same intended use as the proposed predicate devices, that is to acquire and display electrophysiological data. While there are technological differences, these differences do not raise different questions of safety and effectiveness, and therefore the predicate devices can serve as predicate devices for the X-trodes 510(k). {12}------------------------------------------------ #### 10. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING ### Biocompatibility Testing The only patient-contacting component of the X-trodes System M is the XTR Electrode patches. The Xtrodes System M is classified as a surface medical device, in contact with intact skin for less than 24 hours. The following biocompatibility testing was conducted: - Sensitization, ISO 10993-10:2021 - Irritation, ISO 10993-23:2021 - Cytotoxicity, ISO 10993-5:2009 ### Electrical Safety The following tests were conducted: IEC 60601-1:2005 (3rd ed) + CORR. 1:2006 + CORR.2:2007+A1:2012, Medical electrical equipment: Part 1: General requirements for basic safety and essential performance, including US deviations IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems IEC 60601-2-25:2011, Medical electrical equipment – Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs ISO/IEC 80601-2-26:2019, Particular requirements for the basic safety and essential performance of electroencephalographs IEC 60601-2-40:2016, Particular requirements for the basic safety and essential performance of electromyographs and evoked response equipment ### Electromagnetic Compatibility (EMC) The following tests were conducted: AAMI TIR69:2017, Risk management of radio-frequency wireless coexistence for medical devices and systems ANSI/IEEE C63.27:2017, Evaluation of Wireless Coexistence IEC 60601-1-2:2014+A1:2020/Ed. 4.1, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: electromagnetic disturbances – Requirements and tests ### Software Software was evaluated for a moderate level of concern as recommended in the 2005 FDA guidance document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. {13}------------------------------------------------ ### Performance Testing Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include: IEC 60601-1-6:2020. Medical electrical equipment - Part I-6: General requirements for basic safety and essential performance - Collateral standard: Usability Human factors testing in accordance with the FDA guidance document, Applying Human Factors and Usability Engineering to Medical Devices. ### SUMMARY OF CLINICAL TESTING 11. X-trodes conducted a study to evaluate the accuracy and consistency of the X-trodes System M compared to an FDA-cleared clinical electrophysiology device. Methods: X-trodes conducted a case-controlled study of 55 subjects who wore the subject and reference devices (X8) simultaneously. The test duration was approximately 20 minutes and was conducted in a clinic environment. There was a one minute preliminary screening evaluation to determine that electrode conductivity was established and then four minutes of data collection for each modality. Study subjects were adults who had been referred by either a neurologist for an electrophysiology test in the last 5 years. Results: The primary study endpoint was the proportion of interpretable readings of each ExG signal by the XTR that are equivalent to those of the reference device. The agreement proportion for the ECG was 89.36%, for the EEG 97.37%, for the EMG 96.15% and for the EOG 95.24%. All lower 98.75% confidence intervals are higher than 60%. Taking into consideration the high agreement proportions per signal modality and the high CI, it can be concluded that the primary end point was met successfully. For the secondary end point, each modality, the characteristic signal was quantified, and the RMS values were obtained by comparing the signals captured by the X-trodes and X8 devices. The results demonstrate the consistency and reliability of the measurements obtained from the devices and provide an indication of the level of agreement between them. Conclusions: The comparison between the X-trodes XTR and the predicate device demonstrates that the XTR satisfied the pre-specified performance criteria and thus can be found substantially equivalent to the X8 sleep profiler PSG. #### 12. CONCLUSION The results of the performance testing described above demonstrate that the X-trodes System is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...