K Number
K222902

Validate with FDA (Live)

Date Cleared
2023-07-28

(308 days)

Product Code
Regulation Number
870.2340
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SE-1200 Pro&SE-1201 Pro 12-lead electrocardiographs are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiographs are only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.

Device Description

SE-1201 Pro&SE-1200 Pro (2 models: SE-1201 Pro and SE-1200 Pro) features a LCD touch screen, an operation panel, user-programmable reports, and the ability to operate on either battery or AC power. It is capable of simultaneous acquisition, display, and print of 12-lead ECG. It uses algorithm to generate measurements, data presentations, graphical presentations and interpretative statements. The record can be saved in flash memory or send to PC.

AI/ML Overview

The provided document is a 510(k) premarket notification for Electrocardiograph models SE-1200 Pro and SE-1201 Pro. It aims to demonstrate substantial equivalence to previously cleared predicate devices.

However, the document makes a clear statement regarding clinical data: "Clinical data: Not applicable." This indicates that no clinical study was conducted to prove the device meets specific acceptance criteria based on human performance or diagnostic accuracy. Instead, the submission relies on non-clinical data, including electrical safety, electromagnetic compatibility (EMC), and performance testing (bench).

Therefore, I cannot provide the information requested in your prompt regarding acceptance criteria, study details, expert involvement, or MRMC studies, as the submission explicitly states that clinical data was not applicable.

The device's performance demonstration relies on:

  1. Electrical safety and electromagnetic compatibility (EMC): Conformity to standards like ANSI AAMI ES 606601-1:2005/(R) 2012, IEC 60601-1-2:2014, and IEC 60601-2-25 Edition 2.0 2011-10.
  2. Performance testing (bench): Functional and system-level testing to validate accuracy specifications and compliance with relevant consensus standards (e.g., IEC 60601-2-25).
  3. Software Verification and Validation Testing: Documentation provided as recommended by FDA guidance for software in medical devices.

The conclusion states that "The bench testing data and software verification and validation demonstrate that Electrocardiograph SE-1201 Pro&SE-1200 Pro is substantially equivalent to the predicate devices." This substantial equivalence is based on the technical and non-clinical performance characteristics, not on a clinical effectiveness study involving human subjects or interpretation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

July 28, 2023

Edan Instruments, Inc. Milin Wu Regulatory Engineer #15 Jinhui Road, Jinsha Community, Kengzi Sub-District Pingshan District Shenzhen, Guangdong 518122, China

Re: K222902

Trade/Device Name: Electrocardiograph, model: SE-1200 Pro and SE-1201 Pro Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: June 28, 2023 Received: July 5, 2023

Dear Milin Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shruti N. Mistry -S

for

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222902

Device Name Electrocardiograph(SE-1200 Pro. SE-1201 Pro)

Indications for Use (Describe)

The SE-1200 Pro&SE-1201 Pro 12-lead electrocardiographs are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrocardiographs are only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recordiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92

1. Submitter:Edan Instruments, Inc.#15 Jinhui Road, Jinsha Community, Kengzi Sub-District, PingshanDistrict, Shenzhen, 518122 P.R.China.Tel: +86(0755) 84513592Fax: +86(0755) 26882223
Liu Yongying
Contact person:Preparing date:September 19, 2022
Trade Name: Electrocardiograph, model: SE-1200 Pro, SE-1201 Pro
2. Device name andclassification:Common/Usual Name: Electrocardiograph
Classification NameProduct code
21 CFR 870.2340 ElectrocardiographDPS
Regulatory Class: Class II
3. Predicate Device(s):1) Edan Instruments, iSE-1210 Electrocardiograph, cleared under K212278(Primary)2) Edan Instruments, SE-1202 Electrocardiograph, cleared under K210140(Reference)
4. Device Description:SE-1201 Pro&SE-1200 Pro (2 models: SE-1201 Pro and SE-1200 Pro) featuresa LCD touch screen, an operation panel, user-programmable reports, and theability to operate on either battery or AC power. It is capable of simultaneousacquisition, display, and print of 12-lead ECG. It uses algorithm to generatemeasurements, data presentations, graphical presentations and interpretativestatements. The record can be saved in flash memory or send to PC.
5. Indication for UseThe SE-1200 Pro&SE-1201 Pro 12-lead electrocardiographs are intended toacquire ECG signals from adult and pediatric patients through body surfaceECG electrodes. The electrocardiographs are only intended to be used inhospitals or healthcare facilities by doctors and trained healthcare professionals.The cardiogram recorded by the electrocardiograph can help users to analyzeand diagnose heart disease. However, the interpreted ECG with measurements

{4}------------------------------------------------

6. Predicate Device Comparison

Comparison to the predicate devices, the subject device has the same intended use, similar product design, same performance effectiveness, performance safety as the predicate device as summarized in the following tables:

ItemPrimary predicate device(iSE-1210)Proposed device(SE-1201 Pro&SE-1200 Pro)Comparison Result
K#K212278K222902
Intended UseThe iSE Series electrocardiograph isintended to acquire ECG signals from adultand pediatric patients through body surfaceECG electrodes. The electrocardiograph isonly intended to be used in hospitals orhealthcare facilities by doctors and trainedhealthcare professionals. The cardiogramrecorded by the electrocardiograph canhelp users to analyze and diagnose heartdisease. However, the interpreted ECGwith measurements and interpretivestatements is offered to clinicians on anadvisory basis only. The electrocardiographis capable of network communications andsupports the informatized management ofworkflows in hospital and healthcarefacilities.The SE-1200 Pro&SE-1201 Pro 12-leadelectrocardiographs are intended to acquireECG signals from adult and pediatricpatients through body surface ECGelectrodes. The electrocardiographs are onlyintended to be used in hospitals orhealthcare facilities by doctors and trainedhealthcare professionals. The cardiogramrecorded by the electrocardiograph can helpusers to analyze and diagnose heart disease.However, the interpreted ECG withmeasurements and interpretive statements isoffered to clinicians on an advisory basisonly.Different
AlgorithmSEMIPSEMIP or GlasgowDifferent
Number of electrodes1010Same
Safety Specifications
Safety StandardsIEC 60601-1:2005/A1:2012IEC 60601-1-2:2014IEC 60601-2-25:2011IEC 60601-1:2005/A1:2012IEC 60601-1-2:2014IEC 60601-2-25:2011Same
Anti-electric-shock typeClass I with internal power supplyClass I with internal power supplySame
Anti-electric-shockdegreeType CFType CFSame
Power Supply Specifications
Mains SupplyOperating Voltage = 100V-240V~Operating Voltage = 100V-240V~Same
Operating frequency = 50 Hz / 60 HzOperating frequency = 50 Hz / 60 Hz
Output Power: 19V, 2.53AOutput Power: 19V, 2.53A
Rated voltage = 15.2VRated voltage = 14.8 VDifferent
Rated capacity = 3550mAhRated capacity = 2500mAh
Built-in Lithium BatteryPackWhen the battery is fully charged, iSE canwork (without printing) continuously atleast 8 hours.100% recharge time: ≤ 5 hours90% recharge time: < 4 hoursWhen the battery is fully charged, theSE-1200 Pro&SE-1201 Proelectrocardiograph can work for ≥4 H inauto mode (printing a report once every 2minutes), the SE-1200 Pro&SE-1201 Proelectrocardiograph can work for≥2 H inmanual mode, and print 3*4+1R copies ofreports automatically ≥ 250 copies.The battery takes just under 2 hours tocharge from empty to 90%.
Recording
RecorderThermal dot-matrix recorderThermal dot-matrix recorderSame
Effective Width210mm210mmSame
ECG Unit
HR Range30 BPM ~ 300 BPM±1 BPM30 BPM ~ 300 BPM±1 BPMSame
A/D24 bits24 bitsSame
Sampling Frequency64,000 Hz64,000 HzSame
FrequencyResponse0.01 Hz ~ 350 Hz (-3 dB)0.01 Hz ~ 350 Hz (-3 dB)Same
Input Impedance≥100 MΩ(10 Hz)≥100 MΩ(10 Hz)Same
DC Offset Voltage±900 mV±960mV, ±5%Different
Noise≤12.5 μVp-p≤12.5 μVp-pSame
FilterAC Filter:50 Hz / 60 Hz / OffDFT Filter:0.01 Hz / 0.05 Hz / 0.32 Hz / 0.67 HzEMG Filter:AC Filter:50 Hz / 60 Hz / OffDFT Filter:0.01 Hz / 0.05 Hz / 0.32 Hz / 0.67 HzEMG Filter:Same
LOWPASS Filter:350 Hz / 300 Hz / 270 Hz / 150 Hz / 100 Hz / 75 HzLOWPASS Filter:350 Hz / 300 Hz / 270 Hz / 150 Hz / 100 Hz / 75 Hz
CMRR≥140 dB (AC filter on)≥123 dB (AC filter off)≥140 dB (AC filter on)≥123 dB (AC filter off)Same
Pacemaker Detection
Amplitude±500μV to ±700mV±500μV to ±700mV
Width30μs to 2.0ms30μs to 2.0msSame
Sampling Frequency80,000 /sec/channel, Rhythm Lead80,000 /sec/channel, Rhythm Lead
Connection
Wireless connectionWIFI, 4G, Bluetooth, NFCWIFI, 4G, Bluetooth,Different

Table 2-1: Comparison between SE-1201 Pro&SE-1200 Pro and iSE-1210 (primary predicate device)

{5}------------------------------------------------

{6}------------------------------------------------

Table 2-2: Comparison between SE-1201 Pro&SE-1200 Pro and SE-1202 (reference predicate device)

ItemReference predicate device(SE-1202)Proposed device(SE-1201 Pro&SE-1200 Pro)Comparison Result
K#K210140Current Submission
Intended UseThe SE-1202 12-lead electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.The SE-1200 Pro&SE-1201 Pro 12-lead electrocardiographs are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiographs are only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.Same
AlgorithmSEMIP or GlasgowSEMIP or GlasgowSame

As seen in the comparison tables, the subject and predicate devices have similar design features and performance specifications. The technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness.

7. Performance Data:

{7}------------------------------------------------

Non-clinical data:

Electrical safety and electromagnetic compatibility (EMC)

The SE-1201 Pro&SE-1200 Pro electrocardiograph were assessed for conformity with the relevant requirements of the following standards and found to comply:

  • ANSI AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009(R) 2012 and A2:2010(R) 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
  • IEC 60601-1-2:2014 (Fourth Edition) Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: electromagnetic disturbances - Requirements and tests.

Performance testing-Bench

Edan has conducted functional and system level testing to validate the performance of the results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards.

  • IEC 60601-2-25 Edition 2.0 2011-10 Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

Clinical data: Not applicable.

Summary

The non-clinical performance testing showed that the subject devices are as safe and as effective as the predicate device.

8. Conclusion

The bench testing data and software verification and validation demonstrate that Electrocardiograph SE-1201 Pro&SE-1200 Pro is substantially equivalent to the predicate devices.

§ 870.2340 Electrocardiograph.

(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).