K Number
K203331

Validate with FDA (Live)

Device Name
Quick-20m
Manufacturer
Date Cleared
2021-05-20

(189 days)

Product Code
Regulation Number
882.1835
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Quick-20m is intended to be used to acquire the electroencephalogram (EEG) and transmit it wirelessly to a computer.

Device Description

The Quick-20m is a wireless, battery-operated 10-20 montage EEG headset utilizing dry sensor technology. The headset provides an integrated approach to the wireless acquisition of EEG signals. A seated patient is free to exhibit natural movements while real-time data is collected. The Quick-20m includes advanced amplification and shielding to reject ambient electrical noise.

The headset obtains high-quality EEG with minimal scalp preparation. Patented mechanisms and a range of replaceable dry sensors align to various head shapes and sizes, maintaining sensor positions in a standard 10-20 layout. EEG channels are digitized with 24 bits of resolution at 500 Hz. The Quick-20m is suitable for general-purpose EEG.

AI/ML Overview

Regrettably, the provided text does not contain the information needed to answer your request about acceptance criteria and the study proving the device meets them.

The document is a 510(k) premarket notification letter and summary for a device called "Quick-20m," which is a physiological signal amplifier (EEG headset). While it discusses the device's technical characteristics and claims substantial equivalence to a predicate device (g.Nautilus PRO), it explicitly states the following:

  • "Clinical studies were not required to demonstrate that this device is as effective and at least as safe as the predicate device." (page 10)
  • "Bench testing on the Quick-20m includes Intertek certifications to 80601-2-26. in addition to internal company design verification testing that successfully demonstrated compliance to user inputs, and in accordance with 21 CFR 820.40. Clinical testing on the Quick-20m was not applicable." (page 10)

This submission relies on technical comparisons to a predicate device and bench testing for safety standards rather than a clinical study or performance data against specific acceptance criteria for a diagnostic/interpretive AI.

Therefore, I cannot extract:

  1. A table of acceptance criteria and reported device performance: No such criteria or performance data are presented.
  2. Sample size for the test set and data provenance: No test set is described.
  3. Number of experts and their qualifications for ground truth: No expert ground truth establishment is described as clinical studies were not performed.
  4. Adjudication method for the test set: Not applicable as there's no test set.
  5. MRMC comparative effectiveness study or effect size: No clinical study involving human readers or AI assistance is mentioned.
  6. Standalone (algorithm only) performance: The device is a hardware amplifier; performance here refers to its ability to acquire and transmit EEG, not an AI algorithm's diagnostic performance.
  7. Type of ground truth: Not applicable as no clinical ground truth was established for performance evaluation.
  8. Sample size for training set: Not applicable; this is a hardware device, not an AI model requiring a training set in the conventional sense.
  9. How ground truth for training set was established: Not applicable.

The document focuses on establishing substantial equivalence based on technical specifications and safety standards, not on a performance study against specific diagnostic or interpretive endpoints.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

CGX, LLC % Prithul Bom Regulatory Technology Services, LLC 1000 Westgate Drive, Suite #510k Saint Paul, Minnesota 55114

Re: K203331

Trade/Device Name: Quick-20m Regulation Number: 21 CFR 882.1835 Regulation Name: Physiological signal amplifier Regulatory Class: Class II Product Code: GWL, GXY Dated: May 4, 2021 Received: May 5, 2021

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203331

Device Name Quick-20m

Indications for Use (Describe)

The Quick-20m is intended to be used to acquire the electroencephalogram (EEG) and transmit it wirelessly to a computer.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo with the letters "CGX" inside a circular shape. The circle is divided vertically into two halves, with the left half in a gradient of light to dark teal and the right half in a solid blue color. The letters "CGX" are white and positioned in the center of the circle, with each letter taking up roughly one-third of the space.

510(k) Summary

Submitter:CGX, LLC
Address:8445 Camino Santa Fe, Suite 104, San Diego, CA 92121
Phone number:858-864-9400
Contact person:Spencer Linton
Phone number:(858) 864-9400
Email:Spencer@CGXSystems.com
Date prepared:June 23, 2020
Trade name:Quick-20m
Common name:Physiological Signal Amplifier
Product Code, Primary:GWL, GXY
Regulation:21 CFR 882.1835

Substantial equivalence claimed to: g.tec medical engineering GmbH g.Nautilus PRO, K171669, a Class II device.

Device Description:

The Quick-20m is a wireless, battery-operated 10-20 montage EEG headset utilizing dry sensor technology. The headset provides an integrated approach to the wireless acquisition of EEG signals. A seated patient is free to exhibit natural movements while real-time data is collected. The Quick-20m includes advanced amplification and shielding to reject ambient electrical noise.

The headset obtains high-quality EEG with minimal scalp preparation. Patented mechanisms and a range of replaceable dry sensors align to various head shapes and sizes, maintaining sensor positions in a standard 10-20 layout. EEG channels are digitized with 24 bits of resolution at 500 Hz. The Quick-20m is suitable for general-purpose EEG.

Principle of Operation

The Quick-20m headset is a wireless physiological signal amplifier that integrates cutaneous electrodes in a 10-20 montage. The headset captures EEG from the electrodes with 24 bits of resolution at 500 Hz, amplifies the signal and sends the data to a computer for further analysis by a healthcare professional.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a logo with the letters 'CGX' inside a circle that is split vertically into two halves. The left half of the circle is a light teal color, while the right half is a darker blue. The letters 'CGX' are white and are positioned in the center of the circle, with 'C' on the left half and 'GX' on the right half. The overall design is simple and modern.

The environment of use for this device is in a home or healthcare facility setting, such as a doctor's office. The device is not sterile.

The CGX Quick-20m works in the same manner as the predicate device, the g.tec medical engineering GmbH g.Nautilus PRO (K171669), with no safety concerns.

Intended Use(s)

The CGX Quick-20m is intended to be used to acquire the electroencephalogram (EEG) and transmit it wirelessly to a computer.

Technological Characteristics

The CGX Quick-20m is substantially equivalent to the predicate device, the g.tec medical engineering GmbH g.Nautilus PRO for wireless transmission of EEG to a computer. It utilizes dry sensors on 20 monopolar channels to acquire EEG data via a headset worn on the head and transmits this data, via 24-bit Delta-Sigma A/D conversion at a sampling rate of 500 Hz/channel, wirelessly to a computer. The Quick-20m is battery powered with exactly the same CMRR, noise and power-on LED indicators as the predicate. The CGX Quick-20m meets all the applicable safety standards as the predicate, as listed in the Substantial Equivalence Table below with the same system components, patient connections/inputs and firmware/driver software as the predicate.

ItemCGX Quick-20mThis Submissiong.tec medicalengineering GmbHg.Nautilus PROSubstantialEquivalenceComments
Intended UseThe Quick-20m isintended to be used toacquire theelectroencephalogram(EEG) and transmit itwirelessly to acomputerThe g.Nautilus PRO isintended to be used toacquire theelectroencephalogram(EEG) and transmit itwirelessly to acomputerSame as predicate

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a logo with the letters "CGX" inside of a circle. The circle is divided vertically into two halves. The left half is a light teal color, and the right half is a darker blue color. The letters "CGX" are white and are placed in the center of the circle, overlapping both halves.

EEG/PolygraphicChannels20 monopolarchannels. Electrodegrid is standard 10-20layoutUp to 32 monopolarchannels, differentelectrode gridsavailable providing 8,16 or 32 channels and32 or 16 channels CSPlayout on predefinedpositionsFewer maximumnumber of channelsbut equivalent insafety andeffectiveness
DC ChannelAllAllSame as predicate
Full Scale InputRange± 300 mV± 187.5 mV to± 2.25 VFixed range butequivalent in safetyand effectiveness
A/D Conversion24-Bit Delta-Sigma24 Bit Delta-SigmaSame as predicate
Sampling Rate500 Hz/channelUser selectable (250,500 Hz/channel)Similar to predicatebut equivalent insafety andeffectiveness
CMRR>90 dB at 60 Hz>90 dB at 60 HzSame as predicate
Noise<0.6 µV RMS,1-30 Hz<0.6 µV RMSSame as predicate
Power SupplyBattery, 3.0 V DC(Two 1.5 V AAA)Battery, 3.7 V DCDifferent batterychemistry butequivalent in safetyand effectiveness
Rated PowerConsumption0.3 VA0.5 VALess powerconsumption butequivalent in safetyand effectiveness
Amplifier-PCInterfaceWireless to receiver,USB to computerWireless to receiver,USB to computerSame as predicate
Other InterfacesPower on LEDPower on LEDSame as predicate
Use StandardSensors AndElectrodesElectrodes areincluded (Ag/AgCl) -Commerciallyavailable sensorsElectrodes and cap areincluded (Ag/AgCl orgolden dry electrodes,elastic cap with certainelectrode positions)Similar to predicatebut equivalent insafety andeffectiveness
Dimension20 (L) x 18 (W) x 19(H) cm80 (L) x 60 (W) x 27(H) mmLarger dimensionsbut similar in safetyand effectiveness
Weight<600 g<200 gMore weight butequivalent in safetyand effectiveness
Isolationwireless, patientisolationwireless, patientisolationSame as predicate
SystemComponentsAmplifier/Digitization/Electrodes/CapCharging Device +USB cable Receiver +USB cableAmplifier/Digitization/Electrodes/CapCharging Device +USB cable Receiver +USB cableSame as predicate
Firmware / DriverSoftwareResidentResident / CDSame as predicate
DigitalInputs/Outputs16 inputs, all patientseparated, no outputs8 inputs, all patientseparated, no outputsMore digital inputsbut equivalent insafety andeffectiveness
Stimulation UnitInput/OutputNot availableNot availableSame as predicate
Patient ConnectionAnd InputsReceiver:USB - 1 connectorDIGITAL IN - 1connectorReceiver:USB - 1 connectorDIGITAL IN - 1connectorSame as predicate
Type Of AppliedPartBFBFSame as predicate
ImpedanceMeasurementPerformed with125 HzPerformed with 10 HzImpedance checkperformed atdifferent frequencybut similar in safetyand effectiveness
Input Impedance>100 MOhm>100 MOhmSame as predicate
FiltersDC up to 131 HzDC up to 200 Hz(depending onsampling frequency)Fewer digital filtersbut equivalent insafety andeffectiveness
FrequencyResponseLinear between 0.1and 100 HzLinear between 0.1and 100 HzSame as predicate
Environment OfUseElectrophysiologicalElectrophysiologicalSame as predicate
Where UsedOn the headOn the headSame as predicate
Number OfPossible Electrodes20 Channels instandard 10-20 headmapDifferent electrodegrids availableproviding 8, 16 or 32channels and 32 or 16channel CSP layout onpredefined positionsFewer number ofpossible electrodesbut equivalent insafety andeffectiveness
Size Of CapsOne size for adults:52-62 cmVarious for infants,children (mini, midi,maxi) and adults(small, medium,large), headcircumference: 32-62cmSmaller range ofsizes but equivalentin safety andeffectiveness
Style Of CapsCap is full mechanicalheadsetFull head capDifferent style ofconstruction butsimilar in safety andeffectiveness
PerformanceRequirementsNeeds to transmitelectrophysiologicalsignals from anindividual to datacollection deviceNeeds to transmitelectrophysiologicalsignals from anindividual to datacollection deviceSame as predicate
Cap MaterialHP Multi-Jet FusionPA11 and PA12Oeko-Tex certificateOther material butequivalent in safetyand effectiveness
BiocompatibilityTesting On PatientContact MaterialsEvaluation doneEvaluation doneSame as predicate
Type Of ElectrodesActive dryActive, wet and drySame as predicatefor dry
ReprocessingCleaning of allcomponents withisopropyl based wipesCleaning withmultistage enzymaticcleaner of cap and wetelectrodesSimilarreprocessing butequivalent in safetyand effectiveness
Disinfection of capand wet electrodeswith low leveldisinfectant based onGlucoprotamin
Cleaning of dryelectrodes withcleaning wipe
Disinfection of dryelectrodes withdisinfection wipe
Safety StandardsIEC 60601-1 / AAMIANSI ES60601-1IEC 60601-1 / AAMIANSI ES60601-1Similar to predicatebut equivalent insafety andeffectiveness
IEC 60601-1-2IEC 60601-1-2
IEC 60601-2-26IEC 60601-2-26
IEC 80601-2-26
IEC 60601-2-40 (notapplicable)IEC 60601-2-40
ISO 14971ISO 14971
IEC 62304 / AAMIANSI IEC 62304IEC 62304 / AAMIANSI IEC 62304
AAMI ANSI ISO10993-1AAMI ANSI ISO10993-1
AAMI ANSI IEC62366AAMI ANSI IEC62366
IEEE 2010-2012IEEE 2010-2012

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows a logo with the letters "CGX" inside of a circle. The circle is split vertically into two halves. The left half is a light teal color, and the right half is a darker blue color. The letters "CGX" are white and are placed in the center of the circle, overlapping both halves.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows a logo with the letters "CGX" inside a circle. The circle is divided vertically into two halves, with the left half being a light teal color and the right half being a darker blue color. The letters "CGX" are white and are positioned in the center of the circle, spanning both halves.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows a logo with the letters "CGX" inside of a circle. The circle is divided vertically into two halves. The left half is a light teal color, and the right half is a darker blue color. The letters "CGX" are white and are placed in the center of the circle, with each letter appearing to be in its own quadrant.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows a logo with the letters "CGX" inside of a circle. The circle is divided vertically into two halves. The left half is a light teal color, and the right half is a darker blue color. The letters "CGX" are white and are placed in the center of the circle, with the "C" on the left side and the "GX" on the right side.

Substantial Equivalence Discussion

The CGX Quick-20m is substantially equivalent to the g.Nautilus PRO based on indications for use and comparison of functional and technological characteristics. Both devices are intended to provide the healthcare professional with an electroencephalogram (EEG) and transmit this data wirelessly to a computer. Both products fit on the head, utilize dry sensors and have the same amplifier/PC interface.

The minor differences between the CGX Quick-20m and the g.Nautilus PRO do not raise new safety or effectiveness concerns or questions. These differences are discussed below and in the substantial equivalence table.

The CGX Quick-20m is slightly larger than the g.Nautilus PRO (predicate), has slightly less input range (+/- 300 mV), a 500 Hz/channel sampling rate versus the user selectable sampling rate of 250 or 500 Hz/channel on the predicate, a battery chemistry of 3.0V DC versus the 3.7V DC of the predicate, and consumes slightly less power than the predicate. None of the aforementioned differences impact safety nor effectiveness. Further, the g.Nautilus PRO (predicate) offers both wet and dry sensors, while the Quick-20m offers dry sensors.

The CGX Quick-20m device was evaluated by a third party (Intertek) for all applicable safety standards and passed without anomalies.

Further, the company completed human factors and usability evaluation according to applicable versions of IEC 62366 and applicable FDA guidance, and the device demonstrated usability across sizes and positioning, resulting in quality, usable EEG data.

Specifically, the human factors usability program consisted of three components. An In-Field Human Factors test, six published academic studies, and service and repair data.

Bench testing on the Ouick-20m includes Intertek certifications to 80601-2-26. in addition to internal company design verification testing that successfully demonstrated compliance to user inputs, and in accordance with 21 CFR 820.40. Clinical testing on the Quick-20m was not applicable.

Animal Studies

This submission does not require nor include data on animal testing.

Clinical Studies

Clinical studies were not required to demonstrate that this device is as effective and at least as safe as the predicate device.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows a logo with the letters "CGX" inside of a circle. The circle is divided vertically into two halves. The left half is a light blue-green color, and the right half is a darker blue color. The letters "CGX" are white and are placed in the center of the circle, overlapping both halves.

Conclusion:

CGX has been commercializing this EEG technology for research purposes (i.e. not clinical use) since 2015 and in over tens of thousands of uses there have been no safety related customer complaints or quality concerns. This customer experience was leveraged as the platform for the development of the medical device version - the CGX Quick-20m.

A majority of the characteristics of the CGX Quick-20m are identical to those of the g.Nautilus PRO (predicate). The minor differences discussed above and in the Substantial Equivalence Table raise no safety concerns and do not impact effectiveness of the acquisition of EEG. The CGX Quick-20m performs as effectively as the predicate device for the same intended use with no safety concerns.

§ 882.1835 Physiological signal amplifier.

(a)
Identification. A physiological signal amplifier is a general purpose device used to electrically amplify signals derived from various physiological sources (e.g., the electroencephalogram).(b)
Classification. Class II (performance standards).