K242635 · Alton (Shanghai) Medical Instruments Co., Ltd. · FDS · May 30, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K242635
Device Name
Endoscopic Distal Attachment (AF-D series)
Applicant
Alton (Shanghai) Medical Instruments Co., Ltd.
Product Code
FDS · Gastroenterology, Urology
Decision Date
May 30, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
Endoscopic Distal Attachment is intended to be used in combination with compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection.
Device Story
Endoscopic Distal Attachment (AF-D series) is a single-use, medical-grade silicone rubber accessory designed to attach to the distal end of compatible endoscopes. The device consists of an attaching portion, a distal portion, and a drain portion (on some models). It functions by maintaining an appropriate endoscopic field of view during procedures like mucosal resection. The drain portion prevents fluid accumulation on the endoscope surface. The device is used in clinical settings by physicians (gastroenterologists) during endoscopic procedures. It is provided sterile via Ethylene Oxide sterilization. The device is available in various models with different surface designs (flat or inclined), with or without drainage holes, and with or without scales to accommodate different endoscope specifications. It does not involve electronic components, software, or automated analysis.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-5, -10, -11), sterilization validation (ISO 11135, ISO 11737-2, ISO 10993-7), shelf-life/packaging validation (ASTM F1980, ISO 11607, ASTM F1929, ASTM D3078, ASTM F88, ASTM D4169), and mechanical performance testing (ISO 8600-1, ISO 8600-4).
Indicated for use in combination with compatible endoscopes to maintain the field of view during endoscopic procedures, such as mucosal resection, in patients undergoing endoscopic examination.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
FUJIFILM Hood Models DH-28GR, DH-29CR and DH-30CR (K162749)
Related Devices
K162749 — FUJIFILM Hood Models DH-28GR, DH-29CR and DH-30CR · Fujifilm Medical Systems U.S.A, Inc. · Jun 15, 2017
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 30, 2025
Alton (Shanghai) Medical Instruments Co. Ltd
Song Wei
Project Engineer
No.24 Building Jinshao Rd.1688.Baoshan District
Shanghai, Shanghai 200949
China
Re: K242635
Trade/Device Name: Endoscopic Distal Attachment (AF-D series)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: FDS, FDF
Dated: April 30, 2025
Received: April 30, 2025
Dear Song Wei:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242635 - Song Wei
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K242635 - Song Wei
Page 3
Sincerely,
Shanil P. Haugen -S
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242635 | |
| Device Name Endoscopic Distal Attachment (AF-D series) | |
| Indications for Use (Describe) Endoscopic Distal Attachment is intended to be used in combination with compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K242635
Page 1 of 8
# 510(k) Summary
## I. SUBMITTER
Name: Alton (Shanghai) Medical Instruments Co. Ltd.
Address: No.24 Building, JinShao Rd. 1688, Baoshan District, 200949 Shanghai, P. R. China
Name of contact person: Vivian Li
Position: Director of Quality Department
Tel: +86 21 56771811
Fax: +86 21 66307823
Mail: vivian@alton.com.cn
Date prepared: 2025-04-28
## II. Identification of Subjective Device
Device trade name: Endoscopic Distal Attachment (AF-D series)
Regulation Name: Endoscope and accessories
Regulation number: 21 CFR 876.1500
Regulation class: 2
Review Panel: Gastroenterology/Urology
Product Code Descriptions: Gastroscope And Accessories, Flexible/Rigid;
Colonoscope And Accessories, Flexible/Rigid
Product code: FDS, FDF
## III. Identification of Predicate device
Predicate Submission Number: K162749
Trade/Device Name: FUJIFILM Hood Models DH-28GR, DH-29CR and DH-30CR
Regulation Name: Endoscope and accessories
Regulation Number: 21 CFR 876.1500
Regulatory Class: 2
Review Panel: Gastroenterology/Urology
Product Code Descriptions: Gastroscope And Accessories, Flexible/Rigid;
Colonoscope And Accessories, Flexible/Rigid
Product Code: FDS, FDF
{5}
K242635
Page 2 of 8
## IV. Device description
The Endoscopic Distal Attachment is an additional device, made of TPU designed to attach to the distal end of the endoscope. The device is composed of three main portions: an attaching portion, a distal portion, and a drain portion. The attaching portion is a wider diameter opening which is used to connect the attachment to a compatible endoscope; the distal portion is the ending portion of the attachment which tapers into narrower diameter opening. The hole, which is only available for some models, forms drain portion which prevent the fluids lodging on the surface of the endoscope.
The Endoscopic Distal Attachment can be used in combination with compatible endoscopes to maintain an appropriate endoscopic field of view.
The Endoscopic Distal Attachment has six models, depending on different surface design (flat or incline), with or without hole and with scale or without scale. The models include models of flat face without hole, flat face with hole, inclined face without hole, inclined face with hole, flat face without hole with scale and flat face.
## V. Indication for use
Endoscopic Distal Attachment is intended to be used in combination with compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection.
| Attribute | Subject device | Predicative device (K220884) | Discussion/Conclusion |
| --- | --- | --- | --- |
| Manufacturer | Alton (Shanghai) Medical Instruments Co. Ltd | Fujifilm Medical Systems USA, Inc. | / |
| Trade name | Endoscopic Distal Attachment | FUJIFILM Hood | / |
| Regulation name | Endoscope and accessories | Endoscope and accessories | Same |
| Regulatory Class | II | II | Same |
| Product code | FDS, FDF | FDS, FDF | Same |
| Clinical characteristics | | | |
| Indications for use | Endoscopic Distal Attachment is intended to be used in combination with compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection. | The FUJIFILM Hood Models DH-28GR, DH-29CR and DH-30CR are intended to be used in combination with compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection. | Same |
| Compatible endoscopes | Olympus® series: GIF-H190; GIF-Q260; | Fujifilm's endoscopes: EG-590WR, EG-580RD, | Different, see below |
{6}
K242635
Page 3 of 8
| Attribute | Subject device | Predicative device (K220884) | Discussion/Conclusion |
| --- | --- | --- | --- |
| | PCF-PQ260L; PCF-PQ260I; SIF-Q260; GIF-H260; GIF-H290Z; GIF-Q260J; CF-Q260DL; CF-Q260DI; CF-Q260AL; CF-Q260AI; CF-H290; CF-HQ290ZL/I; CF-HQ290L/I; CF-H290L/I; CF-HQ190L/I; CF-H190L/I; CF-HQ290ZL/I; CF-HQ290L/I; CF-HQ190L/I; CF-H190L/I; CF-FH260AZI
Fujifilm®: EG-760R; EG-720R; EG-600WR; EG-L600WR7; EG-L580RD7; EG-L580RD; EG-580RD; EC-450RD5/M; EC580RD/M; EC-740T/M, L; EG-L600ZW; EG-600ZW; EG-760Z; EG-760CT; EG-590ZW; EG-590ZW2; EG-L590ZW; EI-740D/S; EC-760Z-V/M, L; EC-760S; EC-720R/M, I, L; EC-L590ZW; EC-590ZW3/M; EC-590ZW/M; EC-600ZW/M; EC-590WM
Pentax®: EG27-i10; EG29-i10; EG29-i10N; EG29-i20c; EC34-i10M; EC34-i10L; EC34-i10F; EC38-i10L; EC38-i10F/F2; EC38-i10M/M2; EC38-i20c; | EC-580RD/M, EG-580RD/L, EG-600ZW, EG-530CT, EG-590ZW, EC-590WM, C-590ZW/M, EC-590ZW/L, EC-530WM3, EC-530WI3, EC-530WL3, EC-530DL, ES-530WE, EC-590WM4, EC-590WI4, EC-590WL4, EC-590ZW3/M, EC-590ZW3/L, EC-600WM, EC-600WI, EC-600WL, EC-600WL v2, EC-600HL, EC-600ZW/M, EC-600ZW/L | discussion |
| General technological characteristics | | | |
| Device composition and Principle of Operation | The Endoscopic Distal Attachment is an additional device, made of medical grade silicone rubber designed to attach to the distal end of the endoscope. The device is composed of three main portions: an attaching portion, a distal portion, and a drain portion. The attaching | The FUJIFILM Hood Models DH-28GR, DH-29CR and DH-30CR are comprised of three main sections: an attaching portion, a distal portion, and a drain portion. The attaching portion is a wider diameter opening which is used to connect the hood to an applicable endoscope; a distal portion is the | Same |
{7}
K242635
Page 4 of 8
| Attribute | Subject device | Predicative device (K220884) | Discussion/Conclusion |
| --- | --- | --- | --- |
| | portion is a wider diameter opening which is used to connect the attachment to a compatible endoscope; the distal portion is the ending portion of the attachment which tapers into narrower diameter opening. The hole, which is only available for some models, forms drain portion which prevent the fluids lodging on the surface of the endoscope. | ending portion of the hood which tapers into narrower diameter opening, the drain slits on the distal portion form drain portion which prevent the fluids lodging on the surface of the endoscope. | |
| Outer diameter | 11.7mm; 12.1mm; 12.2mm; 12.6mm; 12.9mm; 13.7mm; 14mm; 14.4mm; 14.8mm; 16.4mm | DH-28GR: 11.8mm; DH-29CR: 13.0mm; DH-30CR: 14.8mm | Different, see below discussion |
| Maximum diameter of attaching endoscope | 13.7mm~18.4mm | DH-28GR: 15.5mm DH-29CR: 16.5mm DH-30CR: 18.4mm | Different, see below discussion |
| Total length | 14mm | 17.0mm | Different, see below discussion |
| Distance from the tip | 4mm | 7.0mm | Different, see below discussion |
| Diameter of attaching portion | 9.2mm; 9.8mm; 9.9mm; 10.5mm; 10.8mm; 11.5mm; 11.8mm; 12.2mm; 12.8mm; 13.2mm; 14.8mm | DH-28GR: 10.4 - 11.3 mm; DH-29CR: 11.6 - 12.3 mm; DH-30CR: 13.4 - 14.2 mm | Different, see below discussion |
| Size and number the drain | Models without hole: No holes Models with hole: Circular hole, φ 3mm | Square hole, 5.0mm×1.25mm | Different, see below discussion |
| Sterilization | Ethylene Oxide sterilization | Ethylene Oxide sterilization | Same |
| Single use/reuse | For single use | For single use | Same |
| Shelf life | 3 years | Not known | Different, see below discussion |
{8}
K242635
Page 5 of 8
**Discussion on differences between the subject device and the predicate device** The principle of operation and intended use of the subject device AF-D series are identical to that of the predicate device FUJIFILM Hood Model DH-28GR, DH-29CR, DH-30CR (K162749). The modifications done to the subject device include changes in dimensions to make it be applicable to different specifications of endoscopes from different manufacturers. Comparison of performance tests have been conducted on the subject device and the predicate device to validate the mechanical performance of the subject device is equivalent to the predicate device. In addition, tests to address safety against tissue damage, connection reliability and resistance against endoscope damage in combination assembly with the compatible endoscopes have been conducted on the subject device, the favorable test results demonstrated the subject device's safety on such issue concerns. Such differences do not alter the intended use or fundamental technology of the subject devices neither affects their safety and effectiveness.
**VII. Summary of substantial equivalence discussion**
Based on the above comparison table as well as discussion on differences, the subject device and the predicate device have similar design features and performance specifications. There are no differences on device composition and component material used on both devices. Although there are some differences on size specification parameters on the subject device and predicate device, such differences will not affect the effectiveness and safety of the subject device as different size design is dependent on the different size of the applicable endoscope. In addition, a performance comparison testing between the subject device and the predicate device is implemented and all performances including mechanical property and chemical property are verified to confirm the performance of the subject device is substantially equivalent to the predicate device. These technological differences between the subject device and the predicate device do not affect the safety and effectiveness of the subject device when used as labeled.
**VIII. Summary of Non-clinical Data**
Non-clinical testing for Endoscopic Distal Attachment was conducted to verify that the subject device met all design specifications, demonstrated safety and essential
{9}
K242635
Page 6 of 8
performance based on current applicable standards, and to demonstrate substantial equivalence to the predicate. The following tests were performed.
> Biocompatibility
The biocompatibility evaluation for the Endoscopic Distal Attachment was conducted in accordance with the FDA guidance “Use of International Standard ISO 10993-1, ‘Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process’” September 4, 2020, and International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,” as recognized by FDA.
- MTT Cytotoxicity Test: ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity
- Skin Sensitization Test, Skin Irritation Test: ISO 10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for irritation and sensitization
- Acute Systemic Toxicity Test and Pyrogen Test: ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
> Sterilization Validation
The EO sterilization of the Endoscopic Distal Attachment has been validated according to the following applicable standards:
- ISO11135:2014+A1:2018 Sterilization of medical device- validation and routine control of ethylene oxide sterilization
- ISO 11737-2:2019 Sterilization of Medical Device-Microbiological methods-part 2: Tests of sterility performed in the validation of a sterilization process
- ISO 10993-7:2008+A1: 2019 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
> Shelf Life and Sterile Barrier System
Shelf Life and Sterile Barrier System of the Endoscopic Distal Attachment has been validated according to the following applicable standards:
- ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier
{10}
K242635
Page 7 of 8
Systems for Medical Devices
- ISO11607-1:2019 Packaging for terminally sterilized Medical Device Part 1: Requirement for materials, sterile barrier systems and packaging systems
- ISO11607-2:2019 Packaging for terminally sterilized Medical Device Part 2: Validation Requirement for forming, sealing and assembly process
- ASTM F 1929-15: Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- ASTM D 3078-02(2013): Standard Test Method for Determination of Leaks in Flexible Packaging by Bubble Emission
- DIN 58593-6: 2016 Sterilization – Sterile Supply – Part 6: Microbial Barrier Testing of Packaging Materials for Medical Devices Which Are to Be Sterilized
- ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials
- ASTM D4169-22 Standard practice for performance testing of shipping containers and systems (DC-13, Level II)
> Performance Data – Bench
The performance tests were implemented on both the subject device and the predicate device to demonstrate substantial equivalence according to the product specification as below listed:
- Appearance
- Sizes
- Endoscope Compatibility
- Operational performance according to ISO 8600-1: 2015 and ISO 8600-4: 2023
- Safety against tissue damage
- Connection reliability and Resistance against endoscope damage in combination assembly with the compatible endoscope
> Performance Data – Animal
N/A, no animal studies are available for the subject device.
{11}
K242635
Page 8 of 8
## IX. Summary of Clinical Data
N/A, no clinical studies are available for the subject device.
## X. Conclusions
In conclusion, the technological characteristics, features, specifications, materials, principle of operation, and intended use of the subject device substantially equivalent to the predicate devices quoted above. The differences between the subjective device and its predicate devices do not introduce a new intended use and do not raise new issues of safety and effectiveness. Verification and Validation testing demonstrated that no adverse effects have been introduced by these differences and that the device performs as intended. From the results of nonclinical testing described, it can be concluded that the subject device is substantially equivalent to the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.