K Number
K190030

Validate with FDA (Live)

Date Cleared
2019-03-05

(57 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Embrella™ Endoscopic Distal Attachment is intended to the distal end of the endoscope to facilitate endoscopic therapy and maintain an appropriate endoscopic field of view.

Device Description

The proposed device Embrella™ Endoscopic Distal Attachment is an additional device, made of medical grade silicone rubber designed to attach to the distal end of the endoscope. Its shape is umbrella shaped, with six flexible wings that easily to fold backwards and forwards. It can be used to facilitate endoscopic therapy and maintain an appropriate endoscopic field of view.

Embrella™ Endoscopic Distal Attachment has two models, one is EM-S-01, which is compatible with endoscopy distal diameter 12.8-14.5mm; the other is EM-S-02, which is compatible with endoscopy distal diameter 11.5-13.0mm.

AI/ML Overview

The provided text is a 510(k) Summary for the Embrella™ Endoscopic Distal Attachment. It states that the device is substantially equivalent to a predicate device (EndoRing™ K133359) and lists various tests performed. However, it does not contain specific acceptance criteria or detailed results of a study that proves the device meets specific performance metric acceptance criteria in the way one would expect for a detailed AI/ML device performance study.

Instead, the document focuses on demonstrating substantial equivalence through:

  • Comparison of technological characteristics (materials, design, configuration, indications for use, sterilization, etc.).
  • Performance Bench Testing: "Dimension; Operating Performance; Matching Performance; Tension Performance;"
  • Animal Testing: "Endoscopy View Testing."

The document explicitly states: "No clinical study is included in this submission." This means there is no multi-reader multi-case (MRMC) study, standalone algorithm performance study, or human-in-the-loop study with detailed metrics like sensitivity, specificity, or AUC, as might be expected for an AI/ML diagnostic device.

Therefore, many of the requested points cannot be directly answered from the provided text. I will address the points based on the information available and note where information is not present.


Summary of Acceptance Criteria and Study Performance for Embrella™ Endoscopic Distal Attachment

Based on the provided 510(k) summary, the acceptance criteria and study proving device performance are primarily centered around demonstrating substantial equivalence to a predicate device (EndoRing™ K133359) through bench testing and animal testing, rather than a clinical trial with statistical metrics, given that this is a mechanical accessory device, not an AI/ML diagnostic.

1. Table of Acceptance Criteria and Reported Device Performance

The document broadly states that bench tests and animal tests were performed to demonstrate equivalence to the predicate device. Specific quantitative acceptance criteria (e.g., "Tension performance must be > X Newtons") and exact numerical results are not provided in this summary. The acceptance criterion implied is "equivalent to the predicate device."

Test CategoryAcceptance Criteria (Implied)Reported Device Performance (Implied)
Dimension PerformanceDimensions meet specified requirements and are equivalent to predicate."The testing performed demonstrated that the proposed device and predicate device are equivalent."
Operating PerformanceDevice operates as intended and equivalent to predicate."The testing performed demonstrated that the proposed device and predicate device are equivalent."
Matching PerformanceDevice properly matches with endoscope and is equivalent to predicate."The testing performed demonstrated that the proposed device and predicate device are equivalent."
Tension PerformanceDevice withstands specified tension and is equivalent to predicate."The testing performed demonstrated that the proposed device and predicate device are equivalent."
Endoscopy View TestingMaintains appropriate endoscopic field of view, equivalent to predicate."The testing performed demonstrated that the proposed device and predicate device are equivalent." (Performed on animals.)

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: The document does not specify exact sample sizes for the bench tests or the "Endoscopy View Testing" on animals. Details like the number of devices tested or the number of animals used are not provided.
  • Data Provenance: Not explicitly stated, but the submission is from a Chinese company (Micro-Tech (Nanjing) Co., Ltd.) to the FDA. The tests would presumably have been conducted in a controlled environment to support regulatory submission. The document implies these were retrospective tests conducted specifically for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • No information is provided regarding the number or qualifications of experts involved in establishing "ground truth" for the performance tests. Given these are bench and animal tests for a mechanical device, formal expert adjudication in the sense of image interpretation (like for AI/ML) is not mentioned.

4. Adjudication method for the test set:

  • Not applicable/Not described. The tests appear to be physical/mechanical performance validations rather than subjective interpretation requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs. without AI assistance:

  • No MRMC study was done. The document explicitly states: "No clinical study is included in this submission." Therefore, there is no information on human reader improvement with AI assistance (as this is not an AI device).

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is a mechanical endoscope accessory, not an algorithm or AI system.

7. The type of ground truth used:

  • For the bench tests (Dimension, Operating, Matching, Tension), the "ground truth" would be engineering specifications and measurements.
  • For the "Endoscopy View Testing" (animal testing), the "ground truth" would be direct observation and assessment of the endoscopic field of view during the procedure. The "truth" is established by the observed physical performance of the device itself.

8. The sample size for the training set:

  • Not applicable. This device is a mechanical accessory, not an AI/ML system that requires training data.

9. How the ground truth for the training set was established:

  • Not applicable. As above, no training set for an AI/ML model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 5. 2019

Micro-Tech (Nanjing) Co., Ltd. Becky Li Quality and Regulatory Affairs Director NO. 10 Gaoke Third Road Nanjing, Jiangsu 210032 China

Re: K190030

Trade/Device Name: Embrella™ Endoscopic Distal Attachment Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FED Dated: December 27, 2018 Received: January 7, 2019

Dear Becky Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark J. Antonino -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K190030

Device Name

Embrella™ Endoscopic Distal Attachment

Type of Use (Select one or both, as applicable)

Indications for Use (Describe)

Embrella™ Endoscopic Distal Attachment is intended to the distal end of the endoscope to facilitate endoscopic therapy and maintain an appropriate endoscopic field of view.

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510k summary

Image /page/3/Picture/1 description: The image shows a blue and white logo. The logo is a stylized, three-dimensional representation of the letters 'M' and 'T'. The 'M' is in white and appears to be in front of the blue 'T', creating a layered effect. The logo has a modern and geometric design.

Tab 7

510K Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: K190030

1. Date of Preparation: 2019-03-01

2. Sponsor Identification

Micro-Tech (Nanjing) Co., Ltd.

No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone,

Nanjing, Jiangsu Province, PRC

Establishment Registration Number: 3004837686

Contact Person: Becky Li

Position: Quality and Regulatory Affairs Director

Tel: +86-25-58646378

Fax: +86-25-58744269

Email: Beckyli mt@outlook.com

3. Identification of Proposed Device

Trade Name: Embrella™ Endoscopic Distal Attachment

Common Name: Endoscopic Distal Attachment

Regulatory Information

Classification Name: Endoscope and accessories

Classification: 2

Product Code: FED

Regulation Number: 21 CFR 876.1500

Review Panel: Gastroenterology/Urology

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for MICRO-TECH (Nanjing) Co., Ltd. The logo consists of the letters MT in a stylized, three-dimensional font on the left. To the right of the letters is the company name in English and Chinese. The English name is MICRO-TECH (Nanjing) Co., Ltd., and the Chinese name is 南京微创医学科技股份有限公司.

4. Identification of Predicate Device

510(k) Number: K133359 Product Name: Endoring™ Manufacturer: EndoAid, Ltd.

5. Indications for Use

Embrella™ Endoscopic Distal Attachment is intended to be attached to the distal end of the endoscope to facilitate endoscopic therapy and maintain an appropriate endoscopic field of view.

6. Device Description

The proposed device Embrella™ Endoscopic Distal Attachment is an additional device, made of medical grade silicone rubber designed to attach to the distal end of the endoscope. Its shape is umbrella shaped, with six flexible wings that easily to fold backwards and forwards. It can be used to facilitate endoscopic therapy and maintain an appropriate endoscopic field of view.

Embrella™ Endoscopic Distal Attachment has two models, one is EM-S-01, which is compatible with endoscopy distal diameter 12.8-14.5mm; the other is EM-S-02, which is compatible with endoscopy distal diameter 11.5-13.0mm.

7. Comparison of Technological Characteristics

The Embrella™ Endoscopic Distal Attachment incorporates substantially equivalent device materials, design, configuration, packaging fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate device Endoring™.

Comparison to predicate Devices:

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a blue logo with the letters "M" and "T" in a stylized, three-dimensional design. The "M" is positioned in front of the "T", creating a sense of depth. The logo has a modern and corporate feel due to its clean lines and simple color scheme.

510k summary
ItemProposed DeviceEmbrella™ Endoscopic DistalAttachmentPredicate DeviceEndoring™(K133359)SubstantialEquivalence
Product CodeFEDFEDSame
Regulation No.21 CFR 876.150021 CFR 876.1500Same
Class22Same
Supplied inSterileYesYesSame
MaterialSilicone RubberSilicone RubberSame
ConfigurationOne short tube with 6 flexiblewingsOne short tube with 6 flexiblewingsSame
Compatibleendoscopyworkingchannel11.5-13.0mm12.8-14.5mm11.5-13.0mm12.8-14.5mmSame
Indications forUseEmbrella™ Endoscopic DistalAttachment is intended to beattached to the distal end of theendoscope to facilitate endoscopictherapy and maintain anappropriate endoscopic field ofview.The EndoRings is intended to beattached to the distal end of theendoscope to facilitate endoscopictherapy, to be used for thefollowing:*Keeping the suitable depth ofendoscope's view fieldSame
Single UseYesYesSame
PackagingSingle-use EO sterilized pouchwith one device per pouchSingle-use EO sterilized pouchwith one device per pouchSame
Shelf LifeOne yearsThree yearsDifferent

8. Performance Data

The proposed device the Embrella™ Endoscopic Distal Attachment meets the requirements of ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within A Risk Management Process", ISO 11135 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for MICRO-TECH (Nanjing) Co., Ltd. The logo consists of a blue graphic on the left and the company name in blue text on the right. The company name is written in both English and Chinese. The English name is "MICRO-TECH (Nanjing) Co., Ltd." and the Chinese name is "南京微创医学科技股份有限公司".

510k summary

Development, Validation, and Routine Control of Sterilization processes for

Medical Devices", and ISO 10993-7 "Biological evaluation of medical devices -

Part 7: ethylene oxide sterilization residuals",

The following bench tests were performed on the Embrella™ Endoscopic Distal

Attachment

  • Dimension;

  • Operating Performance

  • Matching Performance;

  • Tension Performance;

The testing performed demonstrated that the proposed device and predicate device are equivalent.

9. Animal Testing

Endoscopy View Testing were performed on the Embrella™ Endoscopic Distal Attachment.

The testing performed demonstrated that the proposed device and predicate device are equivalent,

10. Clinical Test Conclusion

No clinical study is included in this submission.

11. Substantially Equivalent (SE) Conclusion

Based on the indications for use, technological characteristics, and safety and performance testing, the Embrella™ Endoscopic Distal Attachment has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared predicate device EndoRing™ (K133359).

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.