CRE PULMONARY BALLOON DILATATION CATHETER, MODELS K00550300, M00550310, M00550320, M00550330, M00550340, M00550350

K023337 · Boston Scientific Corp · KTI · Nov 18, 2002 · Ear, Nose, Throat

Device Facts

Record IDK023337
Device NameCRE PULMONARY BALLOON DILATATION CATHETER, MODELS K00550300, M00550310, M00550320, M00550330, M00550340, M00550350
ApplicantBoston Scientific Corp
Product CodeKTI · Ear, Nose, Throat
Decision DateNov 18, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4680
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CRE™ Pulmonary Balloon Dilatation Catheter is intended to be used endoscopically to dilate strictures of the airway tree.

Device Story

CRE Pulmonary Balloon Dilatation Catheter; endoscopic device for airway stricture dilation. Consists of inflatable balloon on catheter shaft with inflation and guidewire lumens. Used via bronchoscope by physicians in clinical settings. Device inserted into airway; balloon inflated to mechanically dilate strictures. Benefits patient by restoring airway patency. No electronic or software components.

Clinical Evidence

No clinical data provided; substantial equivalence based on descriptive and technological characteristics.

Technological Characteristics

Balloon dilatation catheter; consists of inflatable balloon, catheter shaft, and proximal hub. Mechanical device; no energy source, software, or electronic components.

Indications for Use

Indicated for endoscopic dilation of airway tree strictures in patients requiring airway intervention.

Regulatory Classification

Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K023337 ## NOV 1 8 2002 # SECTION 14 510(K) SUMMARY #### FOI RELEASABLE Pursuant to §513(I)(3)(A) of the Food, Drug, and Cosmetic Act, Boston Scientific Corporation is required to submit with this Premarket Notification "…adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." Boston Scientific Corporation chooses to submit a summary of information respecting safety and effectiveness. | Date: | October 4, 2002 | | | |-------------------------------------------------|---------------------------------------------|--------|------------| | Common/Usual Names: | Balloon Dilatation Catheter | | | | Trade/Proprietary Names: | CRETM Pulmonary Balloon Dilatation Catheter | | | | Classification Name &<br>Device Classification: | Class II | | | | | Name | Number | 21CFR Ref. | | | Bronchoscope & Acc. | 77 KTI | 874.4680 | | Device Panel/Branch: | Ear, Nose and Throat (ENT) | | | | Owner/Operator: | Boston Scientific Corporation | | | | | One Boston Scientific Place | | | | | Natick, MA 01760 | | | | Contact Person: | Paige Sweeney | | | | | Regulatory Affairs Specialist | | | {1}------------------------------------------------ #### Description of Devices The CRE™ Pulmonary Balloon Dilatation Catheter is used to access the airway tree via a bronchoscope for the purpose of dilating strictures. It consists of an inflatable balloon on a catheter shaft with lumens for inflation and guidewire. #### Indications for Use The CRE™ Pulmonary Balloon Dilatation Catheter is intended to be used endoscopically to dilate strictures of the airway tree. #### Descriptive and Technological Characteristics of Proposed and Predicate Devices Boston Scientific Corporation believes that the CRE™ Pulmonary Balloon Dilatation Catheter is substantially equivalent to the currently marketed CRE™ Balloon Dilatation Catheter. The major components of these devices are the balloon, catheter shaft, and proximal hub. A thorough comparison of the descriptive characteristics between the proposed devices and the predicate devices show equivalence. #### Conclusion Boston Scientific Corporation has demonstrated that the CRE™ Pulmonary Balloon Dilatation Catheter is substantially equivalent to the currently marketed CRE™ Balloon Dilatation Catheter. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus. The logo is black and white. #### Public Health Service The page you requested could not be found. Every effort has been made to ensure the accuracy of the information on this website; however, errors and omissions are possible. We advise you to contact us directly if you have any questions or concerns. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # NOV 1 8 2002 Boston Scientific, Corp. c/o Paige Sweeney Regulatory Affairs Specialist Microvasive Endoscopy One Boston Scientific Place Natick, MA 01760 Re: K023337 Trade/Device Name: CRE™ Pulmonary Balloon Dilation Catheter Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: KTI Dated: October 4, 2002 Received: October 7. 2002 Dear Ms. Sweeney: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that I·DA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Paige Sweeney This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address . http://www.fda.gov/cdrlv/dsma/dsmamain.html Sincerely yours, A. Ralph Rosenthal A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ #### SECTION 3 INDICATIONS FOR USE K023337 510(k) Number: To Be Determined Device Name: CRETM Pulmonary Balloon Dilatation Catheter - The CRE™ Pulmonary Balloon Dilatation Catheter is intended Indication for Use: to be used endoscopically to dilate strictures of the airway tree. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Over-The-Counter Use _ Prescription Use _ OR (Per 21CFR 801.1091) (Optional Format 1-2-96) Karen H. Bohrer (Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises Ko 23337 510(k) Number _ > 510(k) Premarket Notification for CRE™ Pulmonary Balloon Dilatation Catheter Proprietary and Confidential Information of Boston Scientific Corporation 5
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...