K Number
K150951

Validate with FDA (Live)

Date Cleared
2015-06-25

(77 days)

Product Code
Regulation Number
874.4680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The aeris Balloon Dilation Catheter is intended for use in adult and pediations to dilate strictures of the airway.

Device Description

The aeris Balloon Dilation Catheter is comprised of a sinqle lumen catheter with a high pressure balloon near the distal tip. A stylet is provided to facilitate advancement of the balloon dilation catheter to the desired location. The stylet must be removed before inflation of the high pressure balloon. A luer lock at the proximal end is used for placement of the stylet and injecting sterile water into the balloon. Two radiopaque markers, located on the catheter, inside the balloon, can be used to confirm balloon placement under Fluoroscopy.

The aeris Balloon Dilation Catheter includes inflation balloons of diameters 5, 7, 8, 9, 10, 12, 14, and 16 mm with complementing characteristics as shown in Table 1.

AI/ML Overview

The provided text describes the acceptance criteria and the results of various bench tests for the aeris Balloon Dilation Catheter as part of a 510(k) submission for substantial equivalence. It does not include information about a study involving human readers, AI, or specific details on studies for ground truth establishment for a test set. This is a technical device submission, not an AI/algorithm-based diagnostic device.

Here's the breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The document provides multiple tables summarizing the bench testing results for different sizes of the aeris Balloon Dilation Catheter and a comparison with predicate devices. For conciseness, a combined table of the types of tests and their general acceptance criteria and performance will be presented, referencing the detailed tables for specific values.

Test CategoryStudy EndpointRequirementAcceptance RequirementReported Device Performance (aeris Balloon Dilation Catheter)
Functional Performance
Balloon Working LengthBalloon pressure at 1 ATMBalloon working length within tolerance5x30mm: 30 ± 1 mm9x30mm: 30 ± 0.6 mm16x40mm: 40 ± 1 mmPass
Inflation TimeBalloon pressure at recommended use pressure (17 ATM for 5/9mm, 10 ATM for 16mm)Balloon able to inflate< 10 secondsPass
Deflation TimeBalloon Pressure at 0 ATMBalloon able to deflate≤ 15 secondsPass
Catheter FatigueBalloon inflated 10 times to recommended use pressure10 cycles to recommended use pressureNo evidence of balloon leakagePass
Maximum Inflation PressureBalloon rupturesBalloon failure pressure > recommended use pressure5x30mm & 9x30mm: Balloon failure pressure > 17 ATM16x40mm: Balloon failure pressure > 10 ATMPass
Balloon Hub MeasurementBalloon pressure at recommended use pressure (17 ATM for 5/9mm, 10 ATM for 16mm)Measure diameter of hub and working length5x30mm & 9x30mm: Hub diameter minus working length diameter < 1 mm16x40mm: Hub diameter minus working length diameter < 2 mmPass
Balloon DiameterBalloon pressure at recommended use pressure (17 ATM for 5/9mm, 10 ATM for 16mm)Balloon diameter within tolerance5x30mm: 5 ± 0.4 mm9x30mm: 9 ± 0.4 mm16x40mm: 16 ± 0.6 mmPass
InsertabilityDistal end inserted through a 10° bend in a tube with specific diameterDistal end inserted through 10° bend in tubeNo visual signs of kinking or damage done to catheter or styletPass
RemovabilityDistal end withdrawn through a tube with specific diameterDistal end withdrawn through tubeNo visual signs of damagePass
Proximal Luer FittingTestedPer ISO 594-1 & 594-2Meet acceptance criteria of ISO 594-1 & 594-2Pass
TorqueableCatheter rotated 2 timesCatheter turned at least 2 timesNo visual signs of damagePass
Luer/Catheter Shaft ConnectionConnection failureBond tensile force per ISO 105555x30mm & 9x30mm: ≥ 10 N16x40mm: ≥ 15 NPass
Biocompatibility
L929 MEM Elution CytotoxicityTest article must have a reactivity grade of mild or less (≤ 2)There was no biological reactivity (Grade 0) of the cells exposed to the test article extract.Pass
Guinea Pig Maximization SensitizationTest article must have a reaction sensitivity grade of no visible change (Grade 0)There was no sensitization response (Grade 0) in the test articles.Pass
Acute Systemic ToxicityAnimals treated with test article extract must not show signs indicative of toxicity.None of the test article extract treated animals observed with clinical signs consistent with toxicity.Pass
Intracutaneous IrritationTissue reaction to the test article to control dermal observation score is ≤ 1.Differences in mean test and control scores of extract dermal operations were less than 1.Pass
Comparison to Predicate Devices
RadiopacityX-ray performedThe balloon will have marker bands, one under each balloon neck, that are visible under fluoroscopyEquivalent visibility of the aeris marker bands to the predicate devices (specifically CRE Pulmonary)Equivalent

2. Sample Size Used for the Test Set and the Data Provenance

The document details bench testing for various physical attributes of the device and biocompatibility testing. It does not describe a clinical "test set" in the context of diagnostic AI models (i.e., human patient data with ground truth labels).

  • Bench Testing: The sample sizes for each specific bench test (e.g., number of catheters tested for inflation time, fatigue, etc.) are not explicitly stated in the provided summary tables. The results are presented as "Pass" for each test.
  • Biocompatibility Testing: The number of biological samples or animals used for tests like L929 MEM Elution Cytotoxicity, Guinea Pig Maximization Sensitization, Acute Systemic Toxicity, and Intracutaneous Irritation are not specified.
  • Data Provenance: Not applicable in the context of clinical patient data. The tests are laboratory-based bench tests and preclinical biocompatibility assessments.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This question is not applicable to the provided document. The submission is for a medical device (balloon dilation catheter) and involves physical and biological performance testing, not a diagnostic AI device requiring expert-established ground truth from clinical images or data. Bench test acceptance criteria are based on engineering specifications and regulatory standards (e.g., ISO standards).

4. Adjudication Method for the Test Set

Not applicable. There is no "test set" involving expert review or adjudication in the context of AI/diagnostic performance described in this document.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device for dilating strictures of the airway, not an AI or imaging diagnostic device that would involve human readers or AI assistance in interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This document pertains to a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the performance tests is established by:

  • Engineering Specifications: Defined tolerances for physical dimensions (e.g., balloon working length, diameter), time limits (e.g., inflation/deflation time), force requirements (e.g., bond tensile force), and operational limits (e.g., maximum inflation pressure).
  • Regulatory Standards: Adherence to ISO standards (e.g., ISO 594-1 & 594-2 for luer fittings, ISO 10555 for bond tensile force).
  • Visual Inspection: "No visual signs of kinking or damage" for insertability/removability, "No evidence of balloon leakage" for fatigue, "No visual signs of damage" for torqueable.
  • Biocompatibility Assessment: Biological responses meeting defined safety thresholds (e.g., reactivity grade of mild or less, no sensitization response, no signs of toxicity, differences in scores less than 1).
  • Equivalence to Predicate Devices: For radiopacity, the visibility of the aeris marker bands needed to be "Equivalent" to that of the predicate devices.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/algorithm-based device.

9. How the Ground Truth for the Training Set was Established

Not applicable. There is no training set for an AI/algorithm described in this document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines extending from their faces. The profiles are arranged in a way that suggests unity and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 25, 2015

Bryan Medical, Inc. c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street, NW Buffalo, MN 55313

Re: K150951

Trade/Device Name: Aeris Balloon Dilation Catheter Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: KTI Dated: June 9, 2015 Received: June 10, 2015

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Mark Job

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Eric A. Mann -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use Statement

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

Device Name

aeris Balloon Dilation Catheter

Indications for Use (Describe)

The aeris Balloon Dilation Catheter is intended for use in adult and pediations to dilate strictures of the airway.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Services (200) 483-6740

{3}------------------------------------------------

510(k) Summary

Bryan Medical, aeris Balloon Dilation Catheter

The assigned 510(k) number is: K150951 510(k) Owner: Andrew Georgilis Bryan Medical, Inc. 3906 Oak St. Cincinnati, OH 45227 Phone: (513) 272-1600 Fax: (513) 272-1610 Contact Person: Andy Black Medical Murray, Inc. 400 North Rand Road North Barrington, IL 60010 Phone: (847) 620-7990 Fax: (847) 620-7995 January 30, 2015 Date Prepared: Device Name and Classification Classification Name: Bronchoscope (Flexible or Rigid) and Accessories Common Name: Balloon Dilation Catheter

Name: aeris Balloon Dilation Catheter

Device Classification: Class II

Number: 21 CFR Ref. § 874.4680

Product Code: KTI

Review Panel: Ear, Nose, & Throat

Predicate Device: Acclarent Inspira AIR Balloon Dilation System (K090660)

Boston Scientific CRE Pulmonary Balloon Dilation Catheter (K023337)

{4}------------------------------------------------

Device Description

The aeris Balloon Dilation Catheter is comprised of a sinqle lumen catheter with a high pressure balloon near the distal tip. A stylet is provided to facilitate advancement of the balloon dilation catheter to the desired location. The stylet must be removed before inflation of the high pressure balloon. A luer lock at the proximal end is used for placement of the stylet and injecting sterile water into the balloon. Two radiopaque markers, located on the catheter, inside the balloon, can be used to confirm balloon placement under Fluoroscopy.

The aeris Balloon Dilation Catheter includes inflation balloons of diameters 5, 7, 8, 9, 10, 12, 14, and 16 mm with complementing characteristics as shown in Table 1.

aeris BalloonDilationCatheter SizeProductNumberBalloonØ (mm)BalloonWorkingLength (mm)RecommendedUse Pressure(ATM)CatheterOD (mm)CatheterLength(cm)
5 x 30 mmKG0530530171.5555
7 x 30 mmKG0730730171.5555
8 x 30 mmKG0830830171.5555
9 x 30 mmKG0930930171.5555
10 x 30 mmKG10301030172.3455
12 x 40 mmKG12401240102.3455
14 x 40 mmKG14401440102.3455
16 x 40 mmKG16401640102.3455

Table 1: aeris Balloon Dilation Catheter sizes and corresponding characteristics.

The aeris Balloon Dilation Catheter is provided in sterile packaging and with a protective sheath covering the balloon. After removing the catheter from the packaging and protective sheath, it is wiped down with a gauze pad soaked in sterile water. The catheter should then be gently advanced into the airway, to the site of the stricture under endoscopic visualization. The balloon portion of the device is then centered across the restriction and the stylet is removed while holding the catheter securely in place. The balloon dilation catheter luer is attached to the Inflation Device and the balloon is inflated to desired pressure with sterile water. Monitor the balloon via endoscopy during inflation assessing the diameter, shape, and position of the balloon ensuring that the proximal end of the balloon remains proximal to the stricture.

The balloon can be completely deflated while maintaining an endoscopic view of the balloon using a vacuum. Once visual confirmation of the fully deflated balloon is achieved, the balloon may be removed from the airway.

Indications for Use

The aeris Balloon Dilation Catheter is intended to dilate strictures of the airway.

{5}------------------------------------------------

Substantial Equivalence Claim

Based on comparison of device features, materials, intended use, and performance, the Bryan Medical, aeris Balloon Dilation Catheter is substantially equivalent to the commercially available predicate device Acclarent Inspira AIR approved by the FDA under 510k number K090660 and Boston Scientific CRE approved by the FDA under 510k number K023337. Table 2 shows a substantial equivalence comparison summary of the aeris Balloon Dilation Catheter compared to the predicate devices.

Table 2. Comparison table of aeris Balloon Dilation Catheter characteristics to predicate devices
AttributeAcclarent Inspira AIRBoston Scientific CREBryan Medical aeris
A. Intended Use
A.1 Indications forUse StatementThe Airway BalloonCatheter is an instrumentintended to dilate stricturesof the airway tree.The CRE Pulmonary BalloonDilatation Catheter is intendedto be used endoscopically todilate strictures of the airwaytree.The aeris Balloon DilationCatheter is intended to dilatestrictures of the airway.
A.2 LabelingEquivalent in content (labelincluded in Section N)Equivalent in content (labelincluded in Section N)Equivalent content (labelincluded in Section N)
A.3 TargetPopulationAll patients meetingintended use and labelingrequirements.All patients meeting intended useand labeling requirements.All patients meeting intendeduse and labelingrequirements.
A.4 Product code& 21 CFRregulation #KTI, 874.4680KTI, 874.4680KTI, 874.4680
A.5Contraindications• Balloon dilation iscontraindicated in anypatient whose degree ofrespiratory failure wouldnot allow the patient totolerate the manipulationrequired to accomplishballoon dilation.• Balloon dilation iscontraindicated in thepresence of:o significant activebleeding from the siteof the proposeddilationo and/or presence of aknown perforation atthe site of proposeddilationo and/or presence of aknown fistula betweenthe tracheobronchialtree and esophagus,mediastinum orpleural space.• Balloon dilation iscontraindicated in any patientwhose general medicalcondition and degree ofrespiratory failure would notallow the patient to toleratebronchoscopy (rigid orflexible) and/or themanipulation required toaccomplish balloon dilation.• Balloon dilation iscontraindicated in thepresence of:significant active bleedingfrom the site of the proposeddilatation,and/or presence of a knownperforation at the site ofproposed dilatation,and/or presence of a knownfistula between thetracheobronchial tree andesophagus, mediastinum orpleural space unless thedilatation was beingperformed in preparation forthe placement of a stent totreat the perforation or fistula.• Balloon dilation of theairway is contraindicated inany patient whose degreeof respiratory failure wouldnot allow the patient totolerate the manipulationrequired to accomplishballoon dilation.• Balloon dilation iscontraindicated in thepresence of:o significant activebleeding from the site ofthe proposed dilationo and/or presence of aknown perforation at thesite of proposed dilationo and/or presence of aknown fistula betweenthe tracheobronchial treeand esophagus,mediastinum or pleuralspace
Table 2 cont.
AttributeAcclarent Inspira AIRBoston Scientific CREBryan Medical aeris
B. Technology
B.1 Balloon Shapeat < 2 ATM / Max.Inflation PressureStraight body and conicalends / straight body andspherical endsStraight body and conicalends / straight body andspherical endsDog-bone body and conicalends / straight body withspherical ends
B.2 Used withStyletOptionalYes (Guidewire)Optional
B.3 RadiopaqueMarkersNoYesYes
B.4 PatientContactingMaterialsUnknownUnknownNylon balloon, Pebax catheter
B.5 BalloonDiameters5 mm7 mm8.5 mm10 mm12 mm14 mm16 mm8 – 9 – 10 mm10 – 11 – 12 mm12 – 13.5 – 15 mm15 – 16.5 – 18 mm18 – 19 – 20 mm5 mm7 mm8 mm9 mm10 mm12 mm14 mm16 mm
B.6 BalloonWorking Lengths(mm)24 for diameters 5, 7, & 8.540 for diameters 10, 12, 14,& 1630 for diameters 8 - 1255 for diameters 12 – 18 permanufacturer's website30 for diameters 5 - 1040 for diameters 12 - 16
B.7 CatheterOutside Diameter(mm)2.12.51.55 for diameters 5 - 92.34 for diameters 10 - 16
B.8 CatheterLength (cm)457555
B.9 Deflation Time(Seconds)≤ 15 & 25 per clearanceletter for K110218Unknown≤ 15
B.10 MaximumInflation Pressure8 – 16 ATM6 - 9 ATM10 & 17 ATM
B.11 FlexibleYesYesYes
B.12 Shaft DesignCoaxial lumenCoaxial lumenSingle lumen
B.13 Single-UseYesYesYes
B.14 SterilizationEthylene OxideEthylene OxideEthylene Oxide
C. Performance Specifications
C.1 Inflation TimeEquivalentEquivalent≤ 10 seconds
C.2 InsertabilityEquivalentEquivalentThrough equal ID tubing with 10°bend around 4.5" radius
C.3 RemovabilityEquivalentEquivalentThrough airway ≤ 60% ofballoon diameter

{6}------------------------------------------------

{7}------------------------------------------------

Table 2 cont.

AttributeAcclarent Inspira AIRBoston Scientific CREBryan Medical aeris
C.4 Proximal LuerFittingYesYesYes
C.5 TorqueableEquivalentEquivalentMin. 2 turns
C.6 Bond ForceUnknownUnknownMin. 10 N for diameters 5 - 9Min. 15 N for diameters 10 - 16

The intended use of the aeris Balloon Dilation Catheter is substantially equivalent to the predicate devices in intended use, contraindications, and content of labeling. All devices meet the description set forth by the FDA requlation number 21 CFR Ref. § 874.4680.

The technology incorporated into the aeris Balloon Dilation Catheter is substantially equivalent to the technology for the predicate devices. The major components of these devices include a distally located inflation balloon, a flexible catheter shaft, and proximal luer to which an inflation device may be connected. The catheter shaft can be stiffened with a stylet (guidewire) running axially in an interior lumen. Attached to the catheter shaft under each end of the inflation balloon are radiopaque marker bands similar to the Boston Scientific CRE predicate device. All materials of the aeris Balloon Dilation Catheter were tested for biocompatibility for their intended use.

The inflation balloon sizes of the aeris Balloon Dilation Catheter are equivalent to those of the predicate devices. The predicate device's catalog of sizes includes a 5 mm through 18 mm diameter balloon at inflation pressure with intermediate balloon sizes approximately every 2 millimeters.

The aeris Balloon Dilation Catheter differs from the predicate devices in the design of the balloon. The aeris Balloon Dilation Catheter has hubs proximally and distally located on the balloon which begin to take shape upon inflation creating a dog bone balloon shape. As the pressure approaches the inflation pressure, each balloon end begins to take a spherical end shape. The predicate devices' balloon ends have a conical radius shape at pressures approximately equal to or less than two atmospheres. Above two atmospheres, a spherical end shape, similar to the aeris Balloon Dilation Catheter device takes shape.

The aeris Balloon Dilation Catheter is a sinqle lumen catheter design opposed to a dual lumen design of the predicates. The predicate devices have one lumen for the stylet (quidewire) to run through and a separate inflation lumen. The single aeris Balloon Dilation Catheter lumen is able to serve both of these functions without creating any additional risk.

The performance of the aeris Balloon Dilation Catheter was compared to the predicate devices in the insertability and removability tests. A sample device of each the Acclarent Inspira AIR and Boston Scientific CRE were tested for equivalency. Each predicate device passed the tests, showing equivalency between the aeris Balloon Dilation Catheter and each of the predicates were

{8}------------------------------------------------

applicable. The aeris Balloon Dilation Catheter also shows equivalency to performance specifications of the predicate devices in the proximal luer fitting design. Equivalency of the remaining performance specifications for both predicate devices is unknown at this time due to the specifications being unspecified by the manufactures. The aeris Balloon Dilation Catheter performance specification values confirm there are no increased risks or safety concerns introduced to the patient.

Summary of Testing

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs at least as safely and effectively as the legally marketed predicate devices identified in this submission.

TestStudy EndpointRequirementAcceptanceRequirementResult
BalloonWorking LengthBalloon pressureat 1 ATMBalloon workinglength withintolerance30 ± 1 mmPass
Inflation TimeBalloon pressureat 17 ATMBalloon able toinflate< 10 secondsPass
Deflation TimeBalloon Pressureat 0 ATMBalloon able todeflate≤ 15 secondsPass
CatheterFatigueBalloon inflated10 times to 17ATM10 cycles torecommended usepressureNo evidence ofballoon leakagePass
MaximumInflationPressureBalloon rupturesBalloon failurepressure >recommended usepressureBalloon failurepressure > 17 ATMPass
Balloon HubMeasurementBalloon pressureat 17 ATMMeasure diameter ofhub and workinglengthHub diameter minusworking lengthdiameter < 1 mmPass
BalloonDiameterBalloon pressureat 17 ATMBalloon diameterwithin tolerance5 ± .4 mmPass
InsertabilityBalloon locatedDistal end insertedthrough 10° bend intube with 5 mmdiameterNo visual signs ofkinking or damagedone to catheter orstyletPass
RemovabilityCatheterretractedDistal end withdrawnthrough 3 mm tubeNo visual signs ofdamagePass
Proximal LuerFittingTestedPer ISO 594-1 &594-2Meet acceptancecriteria of ISO 594-1 &594-2Pass
TorqueableCatheter rotated2 timesCatheter turned atleast 2 timesNo visual signs ofdamagePass
Luer/CatheterShaftConnectionConnectionfailureBond tensile forceper ISO 10555≥ 10 NPass
Table 3. Summary of baseline bench testing for 5x30mm aeris Balloon Dilation Catheter

{9}------------------------------------------------

TestStudy EndpointRequirementAcceptanceRequirementResult
BalloonWorking LengthBalloon pressureat 1 ATMBalloon workinglength withintolerance30 ± .6 mmPass
Inflation TimeBalloon pressureat 17 ATMBalloon able toinflate< 10 secondsPass
Deflation TimeBalloon Pressureat 0 ATMBalloon able todeflate≤ 15 secondsPass
CatheterFatiqueBalloon inflated10 times to 17ATM10 cycles torecommended usepressureNo evidence ofballoon leakagePass
MaximumInflationPressureBalloon rupturesBalloon failurepressure >recommended usepressureBalloon failurepressure > 17 ATMPass
Balloon HubMeasurementBalloon pressureat 17 ATMMeasure diameter ofhub and workinglengthHub diameter minusworking lengthdiameter < 1 mmPass
BalloonDiameterBalloon pressureat 17 ATMBalloon diameterwithin tolerance9 ± .4 mmPass
InsertabilityBalloon locatedDistal end insertedthrough 10° bend intube with 9 mmdiameterNo visual signs ofkinking or damagedone to catheter orstyletPass
RemovabilityCatheterretractedDistal end withdrawnthrough 5 mm tubeNo visual signs ofdamagePass
Proximal LuerFittingTestedPer ISO 594-1 &594-2Meet acceptancecriteria of ISO 594-1 &594-2Pass
TorqueableCatheter rotated2 timesCatheter turned atleast 2 timesNo visual signs ofdamagePass
Luer/CatheterShaftConnectionConnectionfailureBond tensile forceper ISO 10555> 10 NPass
Table 4. Summary of baseline bench testing for 9x30mm aeris Balloon Dilation Catheter

Table 5. Summary of baseline bench testing for 16x40mm aeris Balloon Dilation Catheter

TestStudy EndpointRequirementAcceptance RequirementResult
BalloonWorking LengthBalloon pressure at 1 ATMBalloon working length within tolerance$40 ± 1 mm$Pass
Inflation TimeBalloon pressure at 10 ATMBalloon able to inflate< 10 secondsPass
Deflation TimeBalloon Pressure at 0 ATMBalloon able to deflate$≤ 15 seconds$Pass
CatheterFatigueBalloon inflated 10 times to 10 ATM10 cycles to recommended use pressureNo evidence of balloon leakagePass

{10}------------------------------------------------

Table 5 cont.
TestStudy EndpointRequirementAcceptanceRequirementResult
MaximumInflationPressureBalloon rupturesBalloon failurepressure >recommended usepressureBalloon failurepressure > 10 ATMPass
Balloon HubMeasurementBalloon pressureat 10 ATMMeasure diameter ofhub and workinglengthHub diameter minusworking lengthdiameter < 2 mmPass
BalloonDiameterBalloon pressureat 10 ATMBalloon diameterwithin tolerance$16 \pm .6$ mmPass
InsertabilityBalloon locatedDistal end insertedthrough 10° bend intube with 16 mmdiameterNo visual signs ofkinking or damagedone to catheter orstyletPass
RemovabilityCatheterretractedDistal end withdrawnthrough 9.6 mm tubeNo visual signs ofdamagePass
Proximal LuerFittingTestedPer ISO 594-1 &594-2Meet acceptancecriteria of ISO 594-1 &594-2Pass
TorqueableCatheter rotated2 timesCatheter turned atleast 2 timesNo visual signs ofdamagePass
Luer/CatheterShaftConnectionConnectionfailureBond tensile forceper ISO 10555≥ 15 NPass

Table 6. Summary of transit & 1 year accelerated aging bench testing for 5x30mm aeris Balloon Dilation Catheter

TestStudy EndpointRequirementAcceptanceRequirementResult
BalloonWorking LengthBalloon pressureat 1 ATMBalloon workinglength withintolerance30 ± 1 mmPass
Inflation TimeBalloon pressureat 17 ATMBalloon able toinflate< 10 secondsPass
Deflation TimeBalloon Pressureat 0 ATMBalloon able todeflate≤ 15 secondsPass
CatheterFatiqueBalloon inflated10 times to 17ATM10 cycles torecommended usepressureNo evidence ofballoon leakagePass
MaximumInflationPressureBalloon rupturesBalloon failurepressure >recommended usepressureBalloon failurepressure > 17 ATMPass
Balloon HubMeasurementBalloon pressureat 17 ATMMeasure diameter ofhub and workinglengthHub diameter minusworking lengthdiameter < 1 mmPass

{11}------------------------------------------------

TestStudy EndpointRequirementAcceptance RequirementResult
BalloonDiameterBalloon pressureat 17 ATMBalloon diameterwithin tolerance$5 \pm .4$ mmPass
InsertabilityBalloon locatedDistal end insertedthrough 10° bend intube with 5 mmdiameterNo visual signs ofkinking or damagedone to catheter orstyletPass
RemovabilityCatheterretractedDistal end withdrawnthrough 3 mm tubeNo visual signs ofdamagePass
Proximal LuerFittingTestedPer ISO 594-1 &594-2Meet acceptancecriteria of ISO 594-1 &594-2Pass
TorqueableCatheter rotated2 timesCatheter turned atleast 2 timesNo visual signs ofdamagePass
Luer/CatheterShaftConnectionConnectionfailureBond tensile forceper ISO 10555$\geq 10$ NPass

Table 7. Summary of transit & 1 year accelerated aging bench testing for 9x30mm aeris Balloon Dilation Catheter

TestStudy EndpointRequirementAcceptanceRequirementResult
BalloonWorking LengthBalloon pressureat 1 ATMBalloon workinglength withintolerance$30 \pm .6$ mmPas
Inflation TimeBalloon pressureat 17 ATMBalloon able toinflate< 10 secondsPass
Deflation TimeBalloon Pressureat 0 ATMBalloon able todeflate≤ 15 secondsPass
CatheterFatigueBalloon inflated10 times to 17ATM10 cycles torecommended usepressureNo evidence ofballoon leakagePass
MaximumInflationPressureBalloon rupturesBalloon failurepressure >recommended usepressureBalloon failurepressure > 17 ATMPass
Balloon HubMeasurementBalloon pressureat 17 ATMMeasure diameter ofhub and workinglengthHub diameter minusworking lengthdiameter < 1 mmPass
BalloonDiameterBalloon pressureat 17 ATMBalloon diameterwithin tolerance$9 \pm .4$ mmPass
InsertabilityBalloon locatedDistal end insertedthrough 10° bend intube with 9 mmdiameterNo visual signs ofkinking or damagedone to catheter orstyletPass
RemovabilityCatheterretractedDistal end withdrawnthrough 5 mm tubeNo visual signs ofdamagePass

{12}------------------------------------------------

Table 7 cont.
TestStudy EndpointRequirementAcceptanceRequirementResult
Proximal LuerFittingTestedPer ISO 594-1 &594-2Meet acceptancecriteria of ISO 594-1 &594-2Pass
TorqueableCatheter rotated2 timesCatheter turned atleast 2 timesNo visual signs ofdamagePass
Luer/CatheterShaftConnectionConnectionfailureBond tensile forceper ISO 10555≥ 10 NPass

Table 8. Summary of transit & 1 year accelerated aging bench testing for 16x40mm aeris Balloon Dilation Catheter

TestStudy EndpointRequirementAcceptanceRequirementResult
BalloonWorking LengthBalloon pressureat 1 ATMBalloon workinglength withintolerance40 ± 1 mmPass
Inflation TimeBalloon pressureat 10 ATMBalloon able toinflate< 10 secondsPass
Deflation TimeBalloon Pressureat 0 ATMBalloon able todeflate≤ 15 secondsPass
CatheterFatigueBalloon inflated10 times to 10ATM10 cycles torecommended usepressureNo evidence ofballoon leakagePass
MaximumInflationPressureBalloon rupturesBalloon failurepressure >recommended usepressureBalloon failurepressure > 10 ATMPass
Balloon HubMeasurementBalloon pressureat 10 ATMMeasure diameter ofhub and workinglengthHub diameter minusworking lengthdiameter < 2 mmPass
BalloonDiameterBalloon pressureat 10 ATMBalloon diameterwithin tolerance16 ± .6 mmPass
InsertabilityBalloon locatedpast curveCatheter insertedthrough 10° bend intube with 16 mmdiameterNo visual signs ofkinking or damagedone to catheter orstyletPass
RemovabilityCatheterretracted fromtubingDistal end withdrawnthrough 9.6 mm tubeNo visual signs ofdamagePass
Proximal LuerFittingTestedPer ISO 594-1 &594-2Meet acceptancecriteria of ISO 594-1 &594-2Pass
TorqueableCatheter turned 2timesCatheter turned atleast 2 timesNo visual signs ofdamagePass
Luer/CatheterShaftConnectionConnectionfailureBond tensile forceper ISO 10555≥ 15 NPass

{13}------------------------------------------------

Table 9. Summary of predicate devices performance testing for an 8.5x24mm Inspira AIR and 15x30mm CRE Pulmonary.

TestStudy EndpointResult
RequirementAcceptanceRequirementInspiraAIRCREPulmonary
InflationTimeBalloon pressure atmax. inflation pressureBalloon able to inflate< 10 secondsPassPass
InsertabilityBalloon located pastcurveCatheter insertedthrough 10° bend intube with balloon sizediameterNo visual signsof kinking ordamage done tocatheter or styletPassPass
RemovabilityCatheter retractedfrom tubingDistal end withdrawnthrough tube with ID60% of balloon sizeNo visual signsof damagePassPass
TorqueableCatheter turned 2timesCatheter turned atleast 2 timesNo visual signsof damagePassPass

Table 10. Summary of comparison test between 9x30mm aeris Balloon Dilation Catheter 15x30mm CRE Pulmonary.

TestStudy EndpointRequirementAcceptanceRequirementResult
Inspira AIRCREPulmonary
RadiopacityX-ray performedThe balloon willhave markerbands, one undereach balloon neck,that are visibleunder fluoroscopyEquivalentvisibility of theaeris markerbands to thepredicate devicesN/AEquivalent

Table 11. All materials, biocompatibility testing summary.

Test DescriptionAcceptance CriteriaResults
L929 MEM ElutionCytotoxicityTest article must have areactivity grade of mild or less (≤2)There was no biological reactivity(Grade 0) of the cells exposed to thetest article extract.
Guinea Pig MaximizationSensitizationTest article must have a reactionsensitivity grade of no visiblechange (Grade 0)There was no sensitization response(Grade 0) in the test articles.
Acute Systemic ToxicityAnimals treated with test articleextract must not show signsindicative of toxicity.None of the test article extract treatedanimals observed with clinical signsconsistent with toxicity.
Intracutaneous IrritationTissue reaction to the test articleto control dermal observationscore is ≤ 1.Differences in mean test and controlscores of extract dermal operationswere less than 1.

§ 874.4680 Bronchoscope (flexible or rigid) and accessories.

(a)
Identification. A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.