K Number
K211894

Validate with FDA (Live)

Device Name
Trachealator
Manufacturer
Date Cleared
2021-11-24

(156 days)

Product Code
Regulation Number
874.4680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Trachealator is indicated for use in dilatation techniques to open or expand stenosis in the airway.

Device Description

The D35 Airway Dilation Catheter (marketing name: Trachealator) is a sterile single-use device used during airway dilatation procedures. The proposed device consists of a series of parallel balloons on a catheter which, when deployed, provides an outward radial force while also creating an inter balloon space or passage for airflow. Thus, airway occlusion is avoided. The D35 Airway Dilation Catheter is also packaged with a double-ended PTFE-coated stainlesssteel guidewire. The guidewire consists of a two-tip configuration with a length of 1500 mm. One end has a straight tip configuration, while the other end consists of a J-tip with a 6 mm diameter. The catheter is also used with an inflation device that is not provided with the catheter. The D35 Airway Dilation Catheter is provided in nine models. The only differences between the models are the balloon length and diameter.

AI/ML Overview

The provided text describes the 510(k) summary for the Trachealator (D35 Airway Dilation Catheter). It details the device, its intended use, comparison to predicate devices, and performance data. However, it explicitly states that clinical data and animal testing are "Not applicable" and "Clinical studies are not necessary to establish the substantial equivalence of this device."

Therefore, the document does not contain information about acceptance criteria and a study that proves the device meets those criteria in the context of clinical performance or human expert evaluation. The performance data provided consists solely of bench testing and biocompatibility testing to demonstrate substantial equivalence to predicate devices, rather than establishing clinical efficacy against specific acceptance criteria.

Below is a summary of the requested information based on the provided text, primarily focusing on the engineering and safety aspects for which data exists:


1. Table of Acceptance Criteria and Reported Device Performance

Since no clinical acceptance criteria or corresponding clinical performance data are presented, this table will focus on the bench testing that was conducted to demonstrate device specifications and substantial equivalence. The document describes several tests performed but does not explicitly list "acceptance criteria" values for each, rather it states that "the D35 Airway Dilation Catheters performs as intended and that the catheter's meets their device specifications."

Performance Test CategorySpecific TestReported Device Performance / OutcomeComments on Acceptance Criteria (as implied)
BiocompatibilityCytotoxicityPassedMet ISO 10993-5 standards
SensitizationPassedMet ISO 10993-10 standards
Irritation/Intracutaneous ReactivityPassedMet ISO 10993-10 standards
Acute Systemic ToxicityPassedMet ISO 10993-11 standards
PyrogenicityPassedMet relevant standards
SterilizationEO Sterilization Validation (Half cycle / over-kill)ValidatedSterility Assurance Level (SAL) 10⁻⁶ (ISO 11135:2014)
Ethylene Oxide Sterilization ResidualsVerifiedMet ISO 10993-7:2008
Shelf LifeShelf-life testingSupports a 3-year shelf lifeN/A (Supports claimed shelf life)
Packaging Seal StrengthTestedN/A
Packaging Dye LeakTestedN/A
Bench TestingVisual inspection and dimensional verificationPerformedMet device specifications
Force to remove balloon protectorPerformedMet device specifications
Crossing diameterPerformedMet device specifications
Simulated usePerformedPerformed as intended
Friction on the guide wirePerformedMet device specifications
Visibility/radiopacityPerformedMet device specifications
Compliance and nominal lengthPerformedMet device specifications
Inflation/deflation timesPerformed (2-17 seconds)Within acceptable range, improved vs predicate
Rated fatigue pressurePerformedMet device specifications
Rated burst pressurePerformed (12 atm)Within acceptable range (within predicate/reference range)
Bond strengthPerformedMet device specifications
Catheter flexibilityPerformedMet device specifications
Catheter kinkPerformedMet device specifications
Torque strengthPerformedMet device specifications
Air leakPerformedMet device specifications
Liquid leakPerformedMet device specifications
Cluster crushingPerformedMet device specifications
Radial strengthPerformedMet device specifications
Dilation forcePerformedMet device specifications
Tip perforationPerformedMet device specifications
High tensile load balloon deflationPerformedMet device specifications
Inter balloon area verificationPerformedMet device specifications

(Note: "Met device specifications" is inferred as the acceptance criteria for each bench test, as specific numerical criteria are not detailed in the provided text.)

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: Not specified for any of the bench tests.
  • Data Provenance: Not explicitly stated, but the submission is from DISA Medinotec, located in Johannesburg, South Africa. The testing was presumably conducted by or for the manufacturer. The data is retrospective in the context of this 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. The document explicitly states that "Clinical Data" is "Not applicable." The ground truth for bench and biocompatibility tests is established by laboratory measurements against defined standards and internal device specifications, not by human experts adjudicating clinical cases.

4. Adjudication method for the test set

  • Not applicable. No expert adjudication of clinical outcomes or diagnostic interpretations was performed as no clinical studies were deemed necessary.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a physical medical device (catheter) and does not involve AI or human readers for diagnostic interpretation.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not applicable. This device is a physical medical device, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • For the non-clinical performance evaluations (biocompatibility, sterilization, bench testing), the "ground truth" implicitly refers to adherence to established international standards (e.g., ISO) and the manufacturer's internal device specifications.

8. The sample size for the training set

  • Not applicable. This device is a physical medical device; there is no "training set" in the context of an algorithm.

9. How the ground truth for the training set was established

  • Not applicable. There is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 9, 2021

DISA Medinotec % Matthew Krueger Senior Consultant Biologics Consulting Group, Inc. 1555 King Street, Suite 300 Alexandria, Virginia 22314

Re: K211894

Trade/Device Name: Trachealator Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: KTI

Dear Matthew Krueger:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated 11/24/2021. Specifically, FDA is updating this SE Letter for a typo in the Sponsor name, from "DISA Medintec" to "DISA Medinotec" as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Shu-Chen Peng, Ph.D., OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices, 301-796-6481, shu-chen.peng(@fda.hhs.gov.

Sincerely,

Shuchen Peng -S

Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 24, 2021

DISA Medintec % Matthew Krueger Senior Consultant Biologics Consulting Group, Inc. 1555 King Street. Suite 300 Alexandria, Virginia 22314

Re: K211894

Trade/Device Name: Trachealator Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: KTI Dated: October 27, 2021 Received: October 28, 2021

Dear Matthew Krueger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{2}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continung-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shuchen Peng -S

Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

In accordance with 21 CFR 807.87(h) and 21 CFR 807.92 the 510(k) Summary for the D35 Airway Dilation Catheter (Trachealator) is provided below.

1. SUBMITTER

Applicant:DISA MedinotecStand 171, Northlands Business Park,Bush Telegraph Street, Northriding,Johannesburg, 2169, South Africa+27 (0) 87 330 2301
Contact:Michelle NieuwoudtQuality & Regulatory Affairs ManagerDISA Medinotec+27 (0) 87 330 2301michelle@disals.co.za
Submission Correspondent:Matthew KruegerSenior ConsultantBiologics Consulting Group, Inc.1555 King Street, Suite 300Alexandria, VA 22314571-777-9548mkrueger@biologicsconsulting.com
Date Prepared:October 26, 2021

2. DEVICE

Device Trade Name:Trachealator
Device Common Name:D35 Airway Dilation Catheter
Classification Name21CFR 874.4680, Bronchoscope (flexible or rigid) and
accessories
Regulatory Class:Class II
Product Code:KTI

3. PREDICATE DEVICE

K170759 – CRE Pulmonary Balloon Dilatation Catheter (Boston Predicate Device: Scientific) K110218 - Inspira AIR Balloon Dilation System (Acclarent, Inc.) Reference Device:

{4}------------------------------------------------

DEVICE DESCRIPTION 4.

The D35 Airway Dilation Catheter (marketing name: Trachealator) is a sterile single-use device used during airway dilatation procedures.

The proposed device consists of a series of parallel balloons on a catheter which, when deployed, provides an outward radial force while also creating an inter balloon space or passage for airflow. Thus, airway occlusion is avoided.

The D35 Airway Dilation Catheter is also packaged with a double-ended PTFE-coated stainlesssteel guidewire. The guidewire consists of a two-tip configuration with a length of 1500 mm. One end has a straight tip configuration, while the other end consists of a J-tip with a 6 mm diameter. The catheter is also used with an inflation device that is not provided with the catheter.

The D35 Airway Dilation Catheter is provided in nine models. The only differences between the models are the balloon length and diameter. Refer to Table 1 for the specifications for each of the 9 models.

ProductReferenceNumberBalloon Length(mm)BalloonDiameter (mm)NominalPressure (atm)MaximumPressure (atm)
TRD1-06030306.0612
TRD1-07030307.0612
TRD1-08030308.0612
TRD1-09030309.0612
TRD1-100303010.0612
TRD1-120404012.0612
TRD1-145404014.5612
TRD1-160404016.0612
TRD1-180404018.0612

Table 1: Catheter Models and Balloon Specifications

5. INTENDED USE/INDICATIONS FOR USE

The device is indicated for use in dilatation techniques to open or expand stenosis in the airway.

{5}------------------------------------------------

SUBSTANTIAL EQUIVALENCE 6.

Comparison of Indications

Predicate Device Indications for Use

K170759 - Boston Scientific CRE Pulmonary Balloon Dilatation Catheter

The CRE Pulmonary Balloon Dilatation Catheter is intended to be used endoscopically to dilate strictures of the airway tree.

Subject Device Indications for Use

The Trachealator is indicated for use in dilatation techniques to open or expand stenosis in the airway.

The intended use of the Trachealator (internal device name: D35 Airway Dilation Catheter) is the same as that of the predicate Boston Scientific CRE Pulmonary Balloon Dilation Catheter. The proposed Indications for Use statement is similar to the predicate IFU but uses slightly different wording. Despite the difference in wording, both Indications for Use statements convey the same clinical use of the device.

Technological Comparisons

The table below compares the key technological feature of the subject device to the predicate device (CRE Pulmonary Balloon Dilatation Catheter-Boston Scientific [K170759]) and the reference device (Inspira AIR Balloon Dilation System-Acclarent, Inc. [K110218]). The reference device is used because the Inspira device helps to better define the "box" of device attributes that are cleared to demonstrate that the Trachealator's key parameters fall within the range of cleared devices.

Proposed DevicePredicate DeviceReference DeviceComments onSimilarities/Differences
510(k) NumberTBDK170759K110218
ApplicantDISA MedinotecBoston ScientificAcclarent
Device NameD35 Airway DilationCatheter(Trachealator)CRE Pulmonary BalloonDilation CatheterInspira AIR Balloon System
ClassificationRegulation21 CFR 874.468021 CFR 874.468021 CFR 874.4680Identical
Product CodeKTIKTIKTIIdentical
Indications forUseIndicated for use indilatation techniquesto open or expandstenosis in theairway.Intended to be usedendoscopically to dilatestrictures of the airwaytree.Intended to dilatestrictures of the airwaytree.Equivalent
Table 2:Technological Comparison
------------------------------------

{6}------------------------------------------------

Proposed DevicePredicate DeviceReference DeviceComments onSimilarities/Differences
Guidewire/StyletlumenYesYesYesIdentical
Packaged withGuidewire orStyletGuidewireNoneStyletEquivalent
Guidewire/StyletDiameterGuidewire0.018" (6 - 12 mm)0.035" (14.5 -18 mm)Guidewire≤ 0.035" (0.89 mm)Stylet0.84 mm (tip)*EquivalentThis difference in diameter does notsignificantly alter performance, as0.018" is also a standard guidewiresize used in hospitals, and the wire issupplied as an accessory with theTrachealator.
Strain reliefYesYesYesIdentical
Usable CatheterLength (mm)650 ± 10 mm1100 mm450 mmEquivalentThe Trachealator has a usable catheterlength similar to other devices on themarket.
Outer LumenCatheterDiameter (mm)3.0 ± 0.05 (6 - 10mm clusters)3.5 ± 0.05 (12 - 18mm clusters)2.5 mm2.15 mm (14x40)2.7 mm (16x40)EquivalentThis small difference in diameter doesnot affect its performance.
Inflated BalloonCluster/ BalloonShapeSix/eight cylindricalballoons, arranged inparallel in a circularconfiguration, andbonded together attheir contact pointsStraight body withspherical endsStraight body withconical endEquivalentThe Trachealator balloon is made of 6or 8 smaller cylindrical balloonsarranged in a circular configurationdesigned to create a center channel,which the other devices do not have.All three balloons dilate an airway in aradial manner; therefore, theirfunction is the same.
Effective BalloonLength (mm)(BalloonDiameter, (mm))27 (6 - 10 mmdiameter clusters)37 (12 – 18 mmdiameter clusters)30 (8 - 12)55 (12 - 18)24 (5, 7, 8.5)40 (10,12,14,16)EquivalentSmall differences in balloon lengthsexist, but the subject device lengthsare within the range of lengths for thecleared predicate and referencedevices.
Capable BalloonDiameter (mm)Diameters rangingfrom 6 to 18 mm8-9-1010-11-1212-13.5-1515-16.5-1818-19-205, 7, 8.5,10, 12,14,16, 18EquivalentThe Trachaelator sizes (diameters) arewithin the range of other cleareddevices.
Proposed DevicePredicate DeviceReference DeviceComments onSimilarities/Differences
Markers1 x Radiopaque1 x Visual marker2 x RadiopaqueNo markersEquivalentThe predicate and subject device bothprovide markers to assist invisualization of balloon.
Nominalpressure:atm (balloon ODmm)6 atm(6 - 18)3 atmOnly MaximumPressure GivenEquivalentPressure at which nominal diameter isachieved differs. Higher pressuresprovide a greater ability to dilate alesion, but the principle is similar.
Rated BurstPressure /MaximumPressure(Balloon ODmm)12 atm(6 - 18)Balloon Length 55 mm:6 atm (18-19-20)7 atm (15-16.5-18)8 atm (12-13.5-15)Balloon Length 30 mm:8 atm (10-11-12;12-13.5-15)9 atm (8-9-10)Balloon Length 24mm:16 atm (5 - 7)12 atm (8.5)Balloon Length 40mm:12 atm (10)10 atm (12 - 14)8 atm 16)EquivalentThe subject device has a higher burstpressure rating than the predicatedevice. However, the maximum burstpressure rating is within the burstpressure rating of the reference device.
BalloonComplianceSemi-compliantSemi-CompliantNon-compliantIdentical
MaximumDeflation Time(seconds)2-1710 - 30≤ 15 & 25EquivalentThe Trachealator provides animproved deflation time compared tothe predicate device, which is animprovement.
CapableBronchoscopeDiameterMinimum workingchannel of 5 mm2.8 mmThe balloon dilationsystem can be usedside by side with abronchoscope.Compatibility for usewithin the workingchannel of abronchoscope has notbeen established.EquivalentThe functionality of the subject deviceis similar to the predicate device anddoes not raise any new questions ofsafety and effectiveness.
Soft TipYesYesNo*Equivalent
Shaft DesignCoaxial LumenCoaxial LumenCoaxial LumenIdentical

{7}------------------------------------------------

As shown in the table above, the fundamental design and technology of the Trachealator is equivalent or identical to the predicate CRE Pulmonary Balloon and/or the reference Inspira AIR Balloon. The minor differences between the subject device and predicate device does not raise questions of safety and effectiveness. Performance testing supports substantial equivalence, that the subject device performs as intended and that the subject device meets the device specifications.

{8}------------------------------------------------

7. PERFORMANCE DATA

Biocompatibility Testing

The balloons and the outer shaft are direct patient contacting devices classified as a surface device, in contact with tissue, with limited contact duration (≤24h) based on the intended use. Other parts of the catheter (inner lumen shaft and tip, radiopaque marker, film on the inner surface of the balloon cluster, and adhesive bonds and the guidewire provided with the device. may be positioned in the patient's airway and therefore encounter blood, mucous, or phlegm and thereby indirectly contact patient's tissues. These are an indirect patient contacting device classified as a surface device, in contact with tissue, with limited contact duration (≤24h).

Biocompatibility testing (cytotoxicity, sensitization, irritation/intracutaneous reactivity, acute systemic toxicity, and pyrogenicity) was performed in accordance with the following standards:

ISO 10993-5: 2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

ISO 10993-11:2018, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity

ISO 10993-12:2012, Biological evaluation of medical devices - Part 12: Sample preparation and reference materials

The D35 Airway Dilation Catheter passed all biocompatibility testing.

Sterilization and Shelf Life

The device is EO sterilization, validated by a half cycle / over-kill approach in accordance with ISO 11135:2014. The sterility assurance level is 106. The D35 Airway Dilation Catheters have been verified according to ISO 10993-7:2008 "Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals". Shelf-life testing was conducted to support a 3 year shelf life. Packaging was tested for seal strength and dye leak.

Electrical Safety and Electromagnetic Compatibility (EMC)

Not applicable. It contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.

Software Verification and Validation Testing

Not applicable. The device contains no software.

Bench Testing

The following performance testing supports substantial equivalence, that the D35 Airway Dilation Catheters performs as intended and that the catheter's meets their device specifications:

  • . Visual inspection and dimensional verification

{9}------------------------------------------------

  • Force to remove balloon protector ●
  • Crossing diameter ●
  • Simulated use
  • Friction on the guide wire ●
  • Visibility/radiopacity
  • Compliance and nominal length
  • Inflation/deflation times
  • Rated fatigue pressure ●
  • Rated burst pressure
  • Bond strength
  • Catheter flexibility
  • Catheter kink
  • Torque strength
  • Air leak
  • Liquid leak
  • Cluster crushing
  • Radial strength ●
  • Dilation force
  • Tip perforation ●
  • High tensile load balloon deflation ●
  • Inter balloon area verification ●

Animal Testing

Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.

Clinical Data

Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.

CONCLUSION 8.

Based on the detailed comparison to the predicate device and the performance testing, the D35 Airway Dilation Catheter can be found substantially equivalent to the predicate device.

{10}------------------------------------------------

Indications for Use

510(k) Number (if known) K211894

Device Name Trachealator

Indications for Use (Describe)

The Trachealator is indicated for use in dilatation techniques to open or expand stenosis in the airway.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 874.4680 Bronchoscope (flexible or rigid) and accessories.

(a)
Identification. A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.