AtriAmp, AtriAmp - Pacing Cable Medtronic Connector, AtriAmp - Pacing Cable Oscor Connector

K200674 · Atrility Medical, LLC · DSA · Jul 2, 2020 · Cardiovascular

Device Facts

Record IDK200674
Device NameAtriAmp, AtriAmp - Pacing Cable Medtronic Connector, AtriAmp - Pacing Cable Oscor Connector
ApplicantAtrility Medical, LLC
Product CodeDSA · Cardiovascular
Decision DateJul 2, 2020
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 870.2900
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The AtriAmp supports pacing by acting as an extension cable between an external pacemaker and cardiac pacing leads. The AtriAmp also supports monitoring of intra-cardiac signals to one of the unipolar precordial chest leads (leads V1-V6) of a Type CF, defibrillation-proof ECG monitor.

Device Story

AtriAmp is a sterile, single-use passive interface cable; acts as a connection hub between an external pacemaker, patient-contacting epicardial pacing lead wires, and a cardiac monitor. Device enables simultaneous real-time, continuous atrial electrogram monitoring and cardiac pacing. Used in clinical settings; operated by healthcare professionals. Input: electrical signals from cardiac pacing leads. Output: transmitted signals to ECG monitor and pacing current to heart. Benefits patient by allowing continuous monitoring of intra-cardiac signals while maintaining pacing capability.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety (ANSI/AAMI ES60601-1, IEC 60601-2-31), noise performance (ANSI/AAMI EC53:2013), biocompatibility (ISO 10993-5, ISO 10993-10), and sterile barrier integrity.

Technological Characteristics

Passive interface cable; sterile, single-use. Sterilization: e-beam (SAL 10^-6). Biocompatibility: Cytotoxicity, Irritation, Sensitization (ISO 10993). Standards: ANSI/AAMI EC53:2013, ANSI/AAMI ES60601-1, IEC 60601-2-31. Shelf life: 2 years (ASTM F88, ASTM F2096, ASTM 4169-16).

Indications for Use

Indicated for patients requiring temporary cardiac pacing and monitoring of intra-cardiac signals via an external pacemaker and a Type CF, defibrillation-proof ECG monitor.

Regulatory Classification

Identification

A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other. July 1, 2020 Atrility Medical, LLC % Gary Syring Principal Consultant Quality & Regulatory Associates, LLC 800 Levanger Lane Stoughton, Wisconsin 53589 Re: K200674 Trade/Device Name: AtriAmp - Pacing Cable Medtronic Connector, AtriAmp - Pacing Cable Oscor Connector Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer And Electrode Cable (Including Connector) Regulatory Class: Class II Product Code: DSA Dated: May 15, 2020 Received: May 15, 2020 Dear Gary Syring: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Hetal Odobasic Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200674 Device Name AtriAmp Indications for Use (Describe) The AtriAmp supports pacing by acting as an extension cable between an external pacemaker and cardiac pacing leads. The AtriAmp also supports monitoring of intra-cardiac signals to one of the unipolar precordial chest leads (leads V1-V6) of a Type CF, defibrillation-proof ECG monitor. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Atrility Medical. The word "Atrility" is in red, with the "A" being larger than the other letters. Below "Atrility" is the word "MEDICAL" in gray, smaller font. The logo is simple and modern. #### 510(k) Summary This 510(k) summary was prepared to provide an understanding of the determination of substantial equivalence in accordance with the requirements 21 CFR 807.92. | Submitters Name: | Atrility Medical, LLC<br>313 Price Place, Suite #13<br>Madison, WI 53705 | |------------------------|-----------------------------------------------------------------------------------------------------------------------| | Contact Person: | Matthew Knoespel, Regulatory Engineer<br>Atrility Medical, LLC | | Contact Phone: | Phone: 920-619-8737 | | Date Summary Prepared: | June 22, 2020 | | Device Trade Name: | AtriAmp | | Common Name: | Passive interface cable to external pacemaker, epicardial pacing lead wire and<br>cardiac monitor | | Classification Name: | 21 CFR 870.290 Patient transducer and electrode cable (including connector)<br>Product Code: DSA, Class II | | Predicate Devices: | K080621, Fastlock Pacing Wire Extension (Primary Predicate)<br>K070926, Electrode/Extension Cables, Model series ATAR | #### Device Description The AtriAmp acts as a connection hub for these three systems (external pacemaker, patient contacting external pacemaker lead wires and cardiac monitor) to provide a real-time, continuous atrial electrogram to the cardiac monitor while simultaneously allowing pacing of the atriAmp is a sterile, single patient use device. #### Intended Use of the Device The AtriAmp is a single-use device that is intended to connect an electrode/lead from a patient to a Type CF, defibrillation-proof ECG Monitor and/or an external pacemaker. #### Summary of Technological Characteristics The following table provides a side-by-side comparison of the predicate devices applied to support this pre-market notification. | Technical Characteristics Comparison | | | | | |--------------------------------------|--------------------------------------------|-------------------------------------------------------------------------------|---------------------------------------------------------------------------|------------------------------------| | Feature | AtriAmp and<br>Accessories<br>Under Review | Fastlock Pacing<br>Wire Extension<br>Cable<br>(Primary Predicate:<br>K080621) | Electrode/Extension<br>Cables, Model series<br>ATAR<br>Predicate: K070926 | Equivalence<br>Comments | | Product Code,<br>Classification | DSA,<br>21 CFR 870.2900,<br>Class II | DSA,<br>21 CFR 870.2900,<br>Class II | IKD,<br>21 CFR 890.1175<br>Class II | Identical to primary<br>predicate. | {4}------------------------------------------------ | Technical Characteristics Comparison | | | | | |---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Feature | AtriAmp and<br>Accessories<br>Under Review | Fastlock Pacing<br>Wire Extension<br>Cable<br>(Primary Predicate:<br>K080621) | Electrode/Extension<br>Cables, Model series<br>ATAR<br>Predicate: K070926 | Equivalence<br>Comments | | Indications for<br>Use | The AtriAmp supports<br>pacing by acting as an<br>extension cable between<br>an external pacemaker<br>and cardiac pacing<br>leads. The AtriAmp also<br>supports monitoring of<br>intra-cardiac signals by<br>transmitting signals to<br>one of the unipolar<br>precordial chest leads<br>(leads V1-V6) of a Type<br>CF, defibrillation-<br>proof ECG monitor. | Fastlock pacing wire<br>extension cable with<br>connectors is indicated<br>for use as an electrical<br>extension cable used to<br>transmit signal from, or<br>power or excitation<br>signal to patient-<br>connected electrodes. | Electrode/extension<br>cable is intended to<br>connect an<br>electrode/lead from a<br>patient or another<br>cable to a diagnostic<br>machine or an<br>external pacemaker. | Equivalent<br>The same as the<br>predicate devices, the<br>AtriAmp interfaces<br>between an external<br>pacemaker and patient<br>applied intra-cardiac<br>pacing lead wires.<br>Equivalent to the<br>secondary predicate, the<br>AtriAmp interfaces the<br>signal from the<br>epicardial lead wires to<br>a, defibrillation-proof<br>ECG Monitor, one of<br>the leads (V1 to V6) of<br>an ECG monitor. | | Duration of<br>use | Anticipated to be less<br>than 14 days | < 7 Days | < 7 Days | Similar | | Single use,<br>disposable | Yes | Yes | Yes | Same | | Provided sterile | Yes | Yes | Yes | Same | | Sterilization<br>method | e-beam | ethylene oxide | ethylene oxide | Similar | | Sterility<br>Assurance Level<br>(SAL) | $ ≤ 10^{-6} $ | Unknown | Unknown | Acceptable for the<br>intended use | | Sterile barrier<br>package | Flexible Double Pouch | Tyvek Flexible Barrier | Tyvek Flexible Barrier | Similar | | Biocompatibility<br>evaluation | Cytotoxicity, Irritation,<br>Sensitization | Unknown | Unknown | Similar | # Non-clinical Performance Tests To establish the technical equivalency of the AtriAmp with accessories, evaluations were conducted to confirm compliance with performance requirements, including: | Test | Test Method Summary | Result | |-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | ANSI/AAMI<br>EC53:2013<br>compliance | The AtriAmp complies with applicable requirements of EC53:2013<br>and performs the same as the predicate for Subclause 5.3.2 Cable<br>and leadwire noise. | Pass | | ANSI/AAMI ES<br>60601-1 and IEC<br>60601-2-31<br>compliance | The AtriAmp with accessory external pacemaker interface cables<br>complies with applicable requirements of ANSI/AAMI ES60601-<br>1:2005 & A1:2012 and IEC 60601-2-31:2011. | Pass | {5}------------------------------------------------ | Test | Test Method Summary | Result | |--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | ES60601-1<br>Subclause 8.5.2.3 | In support of 21 CFR Part 898 compliance, the AtriAmp complies<br>with ES60601-1 Subclause 8.5.2.3 Patient leads or patient cables. | Pass | | Sterility | Sterilization of the subject device was validated according to ISO<br>11137-1:2006(R)2015 & A1:2013 & A2:2018, demonstrating the<br>device is provided with a minimum sterility assumed level (SAL) of<br>$10^{-6}$ . | Pass | | Sterile barrier<br>integrity | The sterile barrier package is a flexible pouch. The methods applied<br>to evaluate the sterile barrier package integrity included post<br>terminal sterilization simulated distribution, seal peel and bubble<br>emission testing. | Pass | | Biocompatibility | The ISO 10993-5:2009 and ISO 10993-10:2020 test methods were<br>applied, with passing results: for Cytotoxicity, Sensitization,<br>Irritation. | Pass | | Shelf life<br>two (2) years | Confirmation of device functional performance and sterile barrier<br>pouch integrity (seal peel per ASTM F88 and bubble leak per ASTM<br>F2096) with accelerated aging and simulated distribution per ASTM<br>4169-16 and bubble leak (ASTM F2096). | Pass | | Duration of use | Operation use was evaluated to verify use for up to 14 days. | Pass | Clinical data are not needed. ## Conclusion The AtriAmp with accessories meet performance requirements equivalent to the predicate device. The intended use and technology of the AtriAmp with accessories is the same as the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...