FASTLOCK, MODEL PC175M01, PC175MCE, PC005R01
K080621 · European Custom Manufacturing BV · DSA · Apr 29, 2008 · Cardiovascular
Device Facts
| Record ID | K080621 |
| Device Name | FASTLOCK, MODEL PC175M01, PC175MCE, PC005R01 |
| Applicant | European Custom Manufacturing BV |
| Product Code | DSA · Cardiovascular |
| Decision Date | Apr 29, 2008 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.2900 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Fastlock pacing wire extension cable device is intended to be used as an interface between an external pacemaker and pacing wires, which are attached to a patient body.
Device Story
Fastlock pacing wire extension cable serves as interface between external pacemaker and patient-attached pacing wires; transmits low-voltage electrical signals. Device consists of electrical cable with safety connector for pacemaker attachment and connectors for patient leads. Used in clinical settings to facilitate connection between external pacing equipment and patient. Output is electrical signal transmission; enables pacing therapy delivery. Benefits patient by providing secure, reliable connection for temporary pacing.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Electrical cable with safety connector for pacemaker and patient-side connectors. Passive device; no energy source. Non-reusable. Male-to-female connector interface.
Indications for Use
Indicated for use as an electrical extension cable to transmit signals from, or power or excitation signals to, patient-connected electrodes.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
Predicate Devices
- CP Medical Disposable Temporary Pacing Cable (K022075)
Related Devices
- K203556 — Medtronic Model 5492A, 5492V, 5492AL, 5492VL Patient Cables · Medtronic, Inc. · Jul 12, 2021
- K070926 — ELECTRODE/EXTENSION CABLES, MODELS ATAR · Oscor, Inc. · Jun 7, 2007
- K241199 — Rotatable Connector (5944RL) · Shenzhen Launch Electrical Co., Ltd. · Jan 23, 2025
- K022075 — DISPOSABLE TEMPORARY PACING CABLE,MODELS CPW-BV-5,CPW-LG-5,CPW-SM-5,CPW-ERA300-5,CPW-QLOC-5 · Cp Medical · Sep 19, 2002
- K152947 — Disposable Temporary Pacing Cable · Tiger Medical Products · Apr 5, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K080 621" in a handwritten style. The text appears to be a code or identifier. The characters are bold and clearly visible against the background.
FastLock Premarket Notification: Traditional 510(k)
## 510(k) Summary
APR 2 9 2008
| Submitter Name:<br>Submitter Address: | European Custom Manufacturing B.V.<br>PO Box 53, NL 5420 AB Gemert, The Netherlands |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone Number: | 31-492-371234 |
| Fax Number: | 31-492-371237 |
| Contact Person: | Rogier Eijck |
| Date Prepared: | 29 November 2007 |
| Device Trade Name: | Fastlock Pacing Wire Extension Cable |
| Common Name | ECG Cables |
| Classification Name,<br>Number &<br>Product Code: | Cable, transducer and electrode, patient (including connector)<br>870.2900<br>DSA |
| Predicate Devices: | CP Medical Disposable Temporary Pacing Cable |
| Device Description<br>and Statement of<br>Intended Use | The Fastlock pacing wire extension cable device provides simple<br>connection between an external pacemaker and pacing wires that are<br>attached to a patient body, in order to transmit low voltage electrical<br>signals between the devices and sensors. The Fastlock product is<br>limited by the Indications for Use of the connected external pacemaker.<br>Intended Use: The Fastlock pacing wire extension cable device is<br>intended to be used as an interface between an external pacemaker and<br>pacing wires, which are attached to a patient body. |
| Summary of<br>Technological<br>Characteristics | The Fastlock pacing wire extension cable consists of an electrical cable<br>with a safety connector at one end that allows electrical connection to<br>the external pacemaker and connectors at the other end that will be<br>attached to the patient's leads. |
| Conclusion | The information discussed above demonstrates that Fastlock product, as<br>effective, and performs as well as or better than the predicate devices. |
| Declarations | This summary includes only information that is also covered in the body<br>of the 510(k).<br>This summary does not contain any puffery or unsubstantiated labeling<br>claims. |
ge 1
Image /page/0/Picture/8 description: The image shows the numbers 05 and 002. The numbers are printed in a simple, sans-serif font. The numbers are arranged horizontally, with 05 on the left and 002 on the right. There is a line above the number 05.
{1}------------------------------------------------
This summary does not contain any raw data, i.e., contains only summary data.
This summary does not contain any trade secret or confidential commercial information.
This summary does not contain any patient identification information.
K080421 Page 2 of 3
Section 5.0: 510(k) Summary
{2}------------------------------------------------
| Feature | FastLock Extension<br>Cable | Disposable Temporary<br>Pacing Cable |
|------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | | K022075 |
| Manufacturer | European Custom<br>Manufacturing BV | CP Medical, Inc. |
| Classification # &<br>Product Code | 870.2900<br>DSA | 870.2900<br>DSA |
| Intended Use | interface between<br>external pacemaker and<br>pacing wires, which are<br>attached to a patient<br>body. | Interface between various<br>diagnostic and<br>physiological monitoring<br>devices (not manufactured<br>by CP Medical) and<br>disposable sensor<br>devices(not manufactured<br>by CP Medical) which are<br>attached to a patient body. |
| Mode of Action | Connects to pacemaker | Connects to pacemaker |
| Reusable | No | No |
| Method of Clamp<br>Introduction | Male to female connector | Alligator clips |
## Summary of Technical Characteristics
K080621 page 3 of 3
Section 5.0: 510(k) Summary
.
:
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 9 2008
European Custom Manufacturing BV c/o Mr. William Greenrose President Oserve America, Inc. 220 River Road Claremont, NH 03743
Re: K080621
Trade/Device Name: Fastlock Regulation Number: 21 CFR 870.2900 Regulation Name: Patient transducer and electrode cable Regulatory Class: Class II (two) Product Code: DSA Dated: December 13, 2007 Received: March 5, 2008
Dear Mr. Greenrose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Mr. William Greenrose
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Blymmmofa
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K080621
Device Name: Fastlock
Indications For Use:
Fastlock pacing wire extension cable with connectors is indicated for use as an electrical extension cable used to transmit signal from, or power or excitation signal to patient-connected electrodes.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummenauer