K Number
K241199

Validate with FDA (Live)

Date Cleared
2025-01-23

(268 days)

Product Code
Regulation Number
870.3680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The rotatable connector is intended for use as part of a temporary cardiac pacing system. It connects a cardiovascular stimulating instrument to a cardiac pacing lead. It is intended for use by trained clinicians in a clinical environment.

Device Description

The Model 5944RL rotatable connector is a pair of single use, disposable surgical cables with two adaptors. The cables are designed to connect a cardiovascular stimulating instrument to a cardiac pacing lead system.

AI/ML Overview

The provided document is a 510(k) Premarket Notification Submission for a medical device called "Rotatable Connector (5944RL)". This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed clinical study results to prove meeting specific acceptance criteria in the way a diagnostic AI/ML device would.

Therefore, the document does not contain the information requested regarding acceptance criteria and a study that proves the device meets them in the context of an AI/ML or diagnostic device.

Specifically, the document states:

  • "The subject of this premarket submission did not require clinical studies to support substantial equivalence." This means no clinical trials were performed to establish the device's performance against specific clinical acceptance criteria.
  • The comparison is primarily focused on technical characteristics, intended use, and safety standards against a predicate device.

Given this, I cannot extract the following information from the provided text:

  1. A table of acceptance criteria and the reported device performance: No such clinical performance metrics or acceptance criteria are listed.
  2. Sample sizes used for the test set and the data provenance: No test set or data provenance from a clinical study are mentioned.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as there's no clinical test set for ground truth establishment.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is a physical medical device (a connector), not an AI/ML diagnostic.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
  8. The sample size for the training set: Not applicable.
  9. How the ground truth for the training set was established: Not applicable.

The document primarily focuses on non-clinical tests (e.g., biocompatibility, sterilization, electrical safety) to demonstrate substantial equivalence to a previously cleared predicate device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 23, 2025

Shenzhen Launch Electrical Co. Ltd Christina Zheng Assistant of General Manager Bldg F, Zhonggangxing Industrial Estate Zhangge Community Guanian Street, Longhua New District Shenzhen. 518110 China

Re: K241199

Trade/Device Name: Rotatable Connector (5944RL) Regulation Number: 21 CFR 870.3680 Regulation Name: Cardiovascular Permanent Or Temporary Pacemaker Electrode Regulatory Class: Class II Product Code: LDF Dated: December 23, 2024 Received: December 23, 2024

Dear Christina Zheng:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

Page

2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Image /page/2/Picture/3 description: The image shows a digital signature. The signature includes the name "Sara M. Royce -S" on the left side and "Digitally signed by Sara M. Royce -S" on the right side. The date and time of the signature are also included, with the date being "2025.01.23" and the time being "15:22:58 -05'00'".

Sara Royce Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K241199

Device Name

Rotatable Connector (5944RL)

Indications for Use (Describe)

The rotatable connector is intended for use as part of a temporary cardiac pacing system. It connects a cardiovascular stimulating instrument to a cardiac pacing lead. It is intended for use by trained clinicians in a clinical environment.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for UNHmed. The logo is in two colors, red and blue. The letters "UNH" are in red, and the letters "med" are in blue.

510(k) Summary

K241199

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date Prepared:Jan 21st, 2025
Manufacturer:Shenzhen Launch Electrical Co. LtdBuilding F, Zhonggangxing Industrial EstateZhangge Community, Guanian Street,Longhua New District, Shenzhen 518110,China
Contact Person:Christina Zheng

Assistant of General Manager Shenzhen Launch Electrical Co. Ltd Tel: +86-18825224607 email: christina.zheng@lnhmed.com

ldentification of the Device:

Proprietary/Trade Name:5944RL Rotatable Connector
Classification Name:Electrode, Pacemaker, Temporary
Regulation Name:Cardiovascular permanent or temporary pacemaker electrode
Regulatory Number:21 CFR 870.3680
Product Code:LDF
Device Class:Class II
Review Panel:Cardiovascular

Primary Predicate Device:

Proprietary/Trade Name:MEDTRONIC 5833SL
Classification Name:Electrode, Pacemaker, Temporary
Regulation Name:Cardiovascular permanent or temporarypacemaker electrode
Regulatory Number:21 CFR 870.3680
Product Code:LDF
Device Class:Class II
Review Panel:Cardiovascular

{5}------------------------------------------------

UNHmed

510(k) Premarket Notification Submission

Submitter/510(k) Holder:MEDTRONIC VASCULAR 7000 CENTRALAVE. N.E. MINNEAPOLIS, MN 55432
Clearance:K923407 (cleared Oct 19, 1992)

Device Description:

The Model 5944RL rotatable connector is a pair of single use, disposable surgical cables with two adaptors. The cables are designed to connect a cardiovascular stimulating instrument to a cardiac pacing lead system.

Indications for Use:

The rotatable connector is intended for use as part of a temporary cardiac pacing system. It connects a cardiovascular stimulating instrument to a cardiac pacing lead. It is intended for use by trained clinicians in a clinical environment.

Comparison with Predicate Device:

The 5944RL Rotatable Connector and its predicate device the 5833SL Disposable surgical cables (K923407), have the same intended use, and similar physical characteristics.

DescriptionSubject DevicePredicate Device
5944RLRotatable Connector5833SLDisposable surgicalcables
510K numberK241199K923407
ManufacturerLaunchMedtronic Vascular
Product code,ClassificationLDF21 CFR 870.3680Class IILDF21 CFR 870.3680Class II
Indications for useThe rotatable connectoris intended for use aspart of a temporarycardiac pacing system.It connects acardiovascularstimulating instrumentto a cardiac pacinglead. It is intended foruse by trained cliniciansin a clinicalenvironment.The Medtronic Model5833SL disposablesurgical cables areintended for use aspart of a temporarycardiac pacing system.They connect a singlechamber or dualchambercardiovascularstimulating instrumentto cardiac pacing leadsor heart wires
DescriptionSubject DevicePredicate Device
5944RLRotatable Connector5833SLDisposable surgicalcables
Intended UserThe Model 5944RLpatient cable isintended for use bytrained clinicians.It is intended for use bytrained clinicians.
Intended UseEnvironmentclinical environmentclinical environment
Duration of useLimited : ≤24HLimited : ≤24H
Single use, disposableYesYes
Provided sterileYesYes
Sterilization methodEORadiation
Sterility AssuranceLevel (SAL)≤ 10-6≤ 10-6
Sterile barrier packageTyvek Flexible BarrierTyvek Flexible Barrier
ProductpictureImage: [red cable with alligator clips]Image: [white cable with red and black connectors]
DesignPrinciples ofoperationThe device is used aspart of a temporarycardiac pacing systembased on the principleof transmitting electricalsignals between thecardiovascularstimulating instrumentand the cardiac pacingleads. The deviceallows real-timecontinuous andaccurate pacing andsensing electricalconnection during leadimplantation. After theThe Medtronic Model5833SL cable is singleuse, disposablesurgical cables. Thecables are designed toconnect acardiovascularstimulating instrumentto cardiac pacing leadsystem.
DescriptionSubject DevicePredicate Device
5944RLRotatable Connector5833SLDisposable surgicalcables
implantation procedureis finished, the device isdisconnected anddiscarded.
FunctionThe Model 5944RLrotatable connector is apair of single use,disposable surgicalcables with twoadaptors. The cablesare designed to connecta cardiovascularstimulating instrumentto a cardiac pacing leadsystem.The Medtronic Model5833SL cables aresingle use, disposablesurgical cable. Thecable is designed toconnect acardiovascularstimulating instrumentto a cardiac pacinglead system.
ConstructionThis device consists ofa pair of connectioncables, which includesa cathode cable (blackwire) and an anodecable (red wire) and apair of un-shrouded pinadaptors.The anode cable has analligator clip at the distalend, and a 2 mmshroud pin at theproximal end.The cathode cable hasa rotatable leadconnector at the distalend of the cable, and a2 mm shroud pin at theproximal end.The distal end of therespective cablesThe proximal end ofeach cable consists ofa connector with 2recessed pins thatplugs into acardiovascularstimulating instrument.The distal end of eachcable consists of 2alligator clips withcolor-coded boots thatattach to the connectorpins ofcardiac leads. The redclip is positive (+). Theblack clip is negative(-).Plug is Type CFconnection plug, whichis connected to Type
DescriptionSubject DevicePredicate Device
5944RLRotatable Connector5833SLDisposable surgicalcables
clip and rotatableconnector that attach tothe electrode contactsof the cardiac pacinglead or a unipolarreference. The alligatorclip with the red wire ispositive (+). Therotatable connector withblack wire is negative(-).Plug type is shroud pinor safety connectorpin, which isconnected to type of pinreceptacle.
Method ofclampintroductionImage: 5944RL Rotatable ConnectorImage: 5833SL Disposable surgical cables
Male to femaleconnector,Alligator clipsMale to femaleconnector,Alligator clips
PerformanceKey performance-Cable segment: 22AWG-Resistance ≤5Ω-Insulationresistance >1GΩ-Dielectric withstandingvoltage: capable of1500 volts A.C. for oneKey performance-Cable segment: 22AWG-Resistance ≤5Ω-Insulationresistance >1MΩ-Dielectric withstandingvoltage: capable of1500 volts A.C. for one
DescriptionSubject DevicePredicate Device
5944RLRotatable Connector5833SLDisposable surgicalcables
Patient ContactMaterialsCytotoxicity, Irritation,Sensitization and meetISO 10993-1 and FDAguidanceCytotoxicity, Irritation,Sensitization and meetISO 10993-1 and FDAguidance
Electrical Safety andElectromagneticcompatibilityANSI/AAMI ES60601-1IEC60601-1-2ANSI/AAMI ES60601-1IEC60601-1-2
LabelingConforms to 21 CFRPart 801Conforms to 21 CFRPart 801

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for UNHmed. The letters "UNH" are in red, and the letters "med" are in blue. The logo is simple and modern.

510(k) Premarket Notification Submission

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "UNHmed" in a stylized font. The letters "UNH" are in red, while the letters "med" are in blue. The letters are connected to each other, forming a single word.

510(k) Premarket Notification Submission

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for UNI-Imed. The logo is made up of the letters "UNI" in red and the letters "med" in blue. The letters are all connected and have a rounded, modern look.

510(k) Premarket Notification Submission

{9}------------------------------------------------

UNHmed

510(k) Premarket Notification Submission

Technology:

The 5944RL Rotatable Connector employs the same fundamental scientific technology as its predicated device.

Summary of Substantial Equivalence Comparison:

The subject devices have the same Indications for use, Intended User and use environment, same technological characteristics including Principles of operation, functions and performance, biocompatibility and general safety, only the sterilization method, construction design and Insulation resistance of performance different and they do not raise new questions of safety and effectiveness.

Summary of Non-clinical test:

5944RL Rotatable Connector was evaluated for biocompatibility, sterilization, usability as well as thermal, electrical, electromagnetic, and mechanical safety, and have been found to comply with applicable medical device safety standard, The 5944RL Rotatable Connector complies with voluntary standards:

    1. ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
    1. IEC 60601-1-2 Edition 4.1 2020-09 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests

{10}------------------------------------------------

UNHmed

510(k) Premarket Notification Submission

  • ISO 10993-5:2009 Biological evaluation of medical devices Part 5: റ്റ് Tests for in vitro cytotoxicity
  • ISO 10993-10:2021 Biological evaluation of medical devices Part 10: 4. Tests for skin sensitization
    1. ISO 11737-1:2018/A1:2021 Sterilization of health care products -Microbiological methods - Part 1: Determination of a population of microorganisms on products
  • ISO 11737-2 Third edition 2019-12 Sterilization of health care products-6. Microbiological methods-Part2:Tests of sterility performed in the definition, validation and maintenance of a s sterilization process
  • ASTM F88/F88M-21 Standard Test Method for Seal Strength of Flexible 7. Barrier Materials
  • ASTM F2096-11 (Reapproved 2019) Standard Test Method for Detecting 8. Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
    1. ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
    1. IEC 60601-2-31: 2020: Medical electrical equipment Part 2-31: Particular requirements for the basic safety and essential performance of external cardiac pacemakers with internal power source

Summary of Clinical Tests:

The subject of this premarket submission did not require clinical studies to support substantial equivalence.

Conclusion:

Shenzhen Launch Electrical Co. Ltd considers the 5944RL Rotatable Connector to be substantially equivalent to the predicate device.

§ 870.3680 Cardiovascular permanent or temporary pacemaker electrode.

(a)
Temporary pacemaker electrode —(1)Identification. A temporary pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to anexternal pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class II (performance standards).(b)
Permanent pacemaker electrode —(1)Identification. A permanent pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class III (premarket approval).(c)
Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before October 4, 2012, for any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976, or that has, on or before October 4, 2012, been found to be substantially equivalent to any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976. Any other pacemaker repair or replacement material device shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.