K Number
K993520
Manufacturer
Date Cleared
2000-05-03

(198 days)

Product Code
Regulation Number
876.5665
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

H2Only Water Purification System for Hemodialysis is intended for use in hemodialysis treatment. It is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. When used as a medical device, Federal law restricts this device to sale by or on the order of a physician.

Device Description

The H2Only Water Purification System for Hemodialysis is a complete water purification system and consists of either one (1) or two (2) reverse osmosis units (RO). Our recommendation to use an RO individually is based upon feed water quality and the demand of water to be used. Pretreatment includes activated carbon filtration media to remove organics, namely chlorine and chloramines. In order to optimize the performance of the RO system, the following items, are utilized: 1) a water break tank or back flow preventer as required; - booster pumps as required to increase the feed pressure to the treatment equipment; 2) - a water softener to remove scale-forming minerals; 3) - cartridge filters to remove suspended solids; 4) - a temperature blending valve to produce the desired water temperature. 5) In order to compensate for fluctuations in demand, an air-tight storage tank with a 0.2 micron vent filter, level control and internal spray array is utilized to store RO product water until needed. The first RO system feeds the storage tank, the second RO system is fed from the storage tank and its product water is directly fed to the distribution plumbing. What ever water is not used in the distribution loop is returned to the storage tank, thus a constant recirculation flow is maintained in the distribution plumbing loop. - when only one RO system is utilized, Ultraviolet disinfection, Deionization Exchange 6) Tanks and Ultrafilters are also incorporated into the system design. By utilizing this twin RO system concept a few components can be eliminated from the system design. These components are: Deionization backup tanks, ultraviolet sanitizer, ultrafilters and repressurization pumps. The elimination of these additional components improves system reliability and reduces overall system operational costs . Also by using the twin RO design the system may be safely operated on either RO system feeding the distribution loop, this feature provides system backup and eliminates down time. When feed water quality demands or when requested by the facility physician, optional devices can include the following components: - cartridge filtration for the removal of silt or feed water sediment; 1) - 2) automatic backwashing multimedia fitter for the removal of suspended solids; - 3) ultraviolet disinfection to reduce bacteria either in the influent water supply or on the product supply; - 4) a booster pump to increase influent water pressure sufficient to provide adequate flow and pressure to the water treatment system. Water treatment components are recommended based on the demand of water required, a feed water analysis, and the ability to produce water which meets AAMI standards. H2Only Water Systems will always recommend a water purfication system which will exceed Hemodialysis Water Quality Standards as set forth by AAMI.

AI/ML Overview

The provided text describes the H2Only Complete Water Purification System for Hemodialysis. However, it does not contain information about specific acceptance criteria, a detailed study proving performance against those criteria, or the methodology typically associated with validating AI/software-based medical devices (like sample sizes for test/training sets, ground truth establishment, expert adjudication, or MRMC studies).

Instead, the document is a 510(k) submission summary for a physical water purification system, focusing on its design, components, and how it achieves "substantial equivalence" to predicate devices. The performance is generally stated in terms of compliance with AAMI standards for water quality in hemodialysis.

Therefore, I cannot populate most of the requested fields because the information is not present in the provided text. I will indicate "Not Applicable" or "Not Provided" where the information is missing or not relevant to this type of device submission.

Here's a breakdown of the available and missing information based on your request:

1. Table of acceptance criteria and the reported device performance:

Acceptance Criteria (Stated)Reported Device Performance (Implied or Stated)
AAMI standards for dissolved inorganics"The water produced meets or exceeds AAMI standards for dissolved inorganics."
AAMI standards for chlorine/chloramines"AAMI recommends that chlorine and chloramines be removed from the treated water supply." (Implied compliance by design)
Microbial contaminants removalIntended to remove microbial contaminants; UV disinfection (optional) and ultrafilters used for bacterial reduction.
Organic and inorganic substances removalIntended to remove organic and inorganic substances; reverse osmosis, carbon filters, DI tanks (optional) are used.
Substantially equivalent to predicatesDeclared substantially equivalent to U.S. Filter, Osmonics Inc., and Better Water systems.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
Not provided. This is a physical device, and the submission does not detail a "test set" in the context of data for an AI/software device. Performance is typically validated through engineering tests and adherence to standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. No "ground truth" establishment in the context of expert review for a physical water purification system.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is a physical water purification system, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):
Not applicable in the context of AI/software validation. For this physical device, "ground truth" would be established by laboratory testing of the water quality against AAMI standards.

8. The sample size for the training set:
Not applicable. No training set for an AI/software device.

9. How the ground truth for the training set was established:
Not applicable.


Summary of the "Study" (as presented in the document):

The "study" or validation presented in this 510(k) submission is primarily based on technical specifications, design principles, and claiming substantial equivalence to legally marketed predicate devices.

  • Device Meets Acceptance Criteria: The submission implicitly states that the device meets acceptance criteria by asserting its design is engineered to produce water that "meets or exceeds AAMI standards" for hemodialysis water quality. The various components (RO units, carbon filters, softeners, UV, DI tanks, ultrafilters) are described with their specific functions that contribute to achieving these standards by removing organic/inorganic substances and microbial contaminants.
  • Proof of Performance: The proof of performance relies on the design and component selection of the H2Only system, stating that "Water treatment components are recommended based on the demand of water required, a feed water analysis, and the ability to produce water which meets AAMI standards." The system is designed to "exceed Hemodialysis Water Quality Standards as set forth by AAMI." Furthermore, the substantial equivalence claim implies that the FDA found sufficient evidence (likely from design documentation and performance claims) that this new device performs as safely and effectively as the predicate devices, which are already known to meet relevant standards.
  • Key Design Features for Performance:
    • Reverse Osmosis (RO) Units: Main component for rejecting inorganics, dissolved minerals, suspended solids, and microbiological contaminants.
    • Activated Carbon Filters: Remove chlorine, chloramines, and other organics.
    • Water Softener: Removes calcium and magnesium to prevent scaling of RO membranes.
    • Storage Tank with 0.2 micron vent filter: Prevents airborne bacteria.
    • Ultraviolet (UV) Disinfection Units (optional): For bacterial reduction.
    • Deionization Exchange Tanks (optional): For inorganic and dissolved gas removal.
    • Ultra Filters & Sub micron Post Filters: Remove suspended solids, bacteria, and inhibit pyrogens.
    • Monitoring Devices: Pressure gauges, flow meters, water quality monitors etc., ensure proper performance.

In essence, the "study" for this type of device is the comprehensive engineering design and component selection, coupled with the assertion that this design, when operated correctly, will consistently meet the established (AAMI) water quality standards for hemodialysis. The 510(k) clearance confirms the FDA's acceptance of this evidence for substantial equivalence, allowing the device to be marketed.

{0}------------------------------------------------

K993520 Page 1 of 5

MAY - 3 2000

Section 5 Summary of Safety & Effectiveness

H2Only Complete Water Purification System for Hemodialysis

submitted by Serv-A-Pure Company 1101 Columbus Ave. Bay City, MI 48708 517/892-7745 Fax 517/892-1092

Contact: Richard Herzberger September 16, 1999

Common or Classification Name: Complete Water Purification System for Hemodialysis

Proprietary Name: H2Only Complete Water Purification System for Hemodialysis

Claiming substantial equivalence to:

U.S.Filter Company, (800) 466-7873Water Treatment System for HemodialysisK980182
Osmonics Inc., (612) 933-2277Water Treatment System for HemodialysisK931595
Better Water, (615) 355-6063Water Treatment System for HemodialysisK920186

Intended Use: H2Only Water Purification System for Hemodialysis is intended for use in hemodialysis treatment. It is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. When used as a medical device, Federal law restricts this device to sale by or on the order of a physician.

Device Description

The H2Only Water Purification System for Hemodialysis is a complete water purification system and consists of either one (1) or two (2) reverse osmosis units (RO). Our recommendation to use an RO individually is based upon feed water quality and the demand of water to be used. Pretreatment includes activated carbon filtration media to remove organics, namely chlorine and chloramines. In order to optimize the performance of the RO system, the following items, are utilized:

  1. a water break tank or back flow preventer as required;

{1}------------------------------------------------

  • booster pumps as required to increase the feed pressure to the treatment equipment; 2)
  • a water softener to remove scale-forming minerals; 3)
  • cartridge filters to remove suspended solids; 4)
  • a temperature blending valve to produce the desired water temperature. 5)

In order to compensate for fluctuations in demand, an air-tight storage tank with a 0.2 micron vent filter, level control and internal spray array is utilized to store RO product water until needed. The first RO system feeds the storage tank, the second RO system is fed from the storage tank and its product water is directly fed to the distribution plumbing. What ever water is not used in the distribution loop is returned to the storage tank, thus a constant recirculation flow is maintained in the distribution plumbing loop.

  • when only one RO system is utilized, Ultraviolet disinfection, Deionization Exchange 6) Tanks and Ultrafilters are also incorporated into the system design.
    By utilizing this twin RO system concept a few components can be eliminated from the system design. These components are: Deionization backup tanks, ultraviolet sanitizer, ultrafilters and repressurization pumps. The elimination of these additional components improves system reliability and reduces overall system operational costs . Also by using the twin RO design the system may be safely operated on either RO system feeding the distribution loop, this feature provides system backup and eliminates down time.

When feed water quality demands or when requested by the facility physician, optional devices can include the following components:

  • cartridge filtration for the removal of silt or feed water sediment; 1)
    1. automatic backwashing multimedia fitter for the removal of suspended solids;
    1. ultraviolet disinfection to reduce bacteria either in the influent water supply or on the product supply;
    1. a booster pump to increase influent water pressure sufficient to provide adequate flow and pressure to the water treatment system.

Water treatment components are recommended based on the demand of water required, a feed water analysis, and the ability to produce water which meets AAMI standards. H2Only Water Systems will always recommend a water purfication system which will exceed Hemodialysis Water Quality Standards as set forth by AAMI.

{2}------------------------------------------------

Summary of Technological Characteristics

H2Only Water Purification Systems designs, recommends and assembles the components of the system. H2Only Water Systems is not the manufacturer of several of the components and is not the manufacturer of component parts which are assembled by H2Only Water Systems.

Although, H2Only Water Purification Systems recommends the water treatment system and components, the dialysis facility's physician has ultimate authority of the type of equipment installed for water purification. The recommended system consists of the following components. A complete system and the intended purposes of each components are described below. Our design is of equivalent to that of the predicate devices.

Hot and cold filtered water supplies are mixed together using a temperature blending valve to the desired 77 degrees. Sufficient water pressure must be supplied in order to operate the water treatment effectively. If sufficient pressure is not available, a booster pump is installed to supply the required flow and pressure of feed water. If silt or fine sediment are present in the feed water supply, optional filtration is used, 5.0 micron filter cartridges can be used to remove the fine sediment and silt. Next, the water flows through dual exchange carbon filters to remove the chlorine, chloramines and other organics from the water supply. The carbon filters are sized to provide adequate empty bed contact time of 3-5 minutes for chiorine and 6-10 minutes for chloramines. Water is then directed to the water softener where calcium and magnesium ions are exchanged for non-scaling sodium ions. Water is next directed to a multi cartridge 5.0 micron filter housing where sediment is removed.

The "pre-treated" water then enters into the reverse osmosis membrane where dissolved minerals are rejected from the product stream. The product or permeate water is where dissolved minerals are rejected from this water stream. The product water, permeate, then enters Into a storage tank. The storage tank is complete with a sub micron vent filter to prevent airbome bacteria from entering into the storage tank. Stored product water is then used to feed the second RO system and its product water is directly fed to the distribution plumbing. What ever water is not used in the distribution loop is returned to the storage tank, thus a constant recirculation flow is maintained in the distribution plumbing loop.

{3}------------------------------------------------

Identification of Components

Temperature Blending Valve - is used to blend hot and cold water to achieve a temperature of 77 degrees. Tempered water increases the efficiency of the reverse osmosis unit.

Break tank with air gap or back flow preventer is used to prevent a water back flow condition from occurring. Either one meets local plumbing codes.

Booster Pump - (optional) is used to increase the influent water pressure. The reverse osmosis unit, as well as the other pretreatment equipment, requires sufficient feed water pressure to operate.

Multi-media Filter - (optional) is used to remove suspended solids from the feed water supply. Excessive suspended solids can cause malfunction of pretreatment components. i.e. media plugging, moving parts binding etc.

Water Softener - is used to remove calcium and magnesium bicarbonate from the influent water supply. Calcium and magnesium bicarbonate will cause the reverse osmosis membrane to scale, thus reducing production of treated water.

Backwashing Carbon Filters - are used to remove organics, namely chlorine and chloramines, from the influent water supply. Chlorine will attack and destroy the reverse osmosis membrane, thus reducing the quality of the treated water. Also, AAMI recommends that chlorine and chloramines be removed from the treated water supply. By occasional backwashing of the carbon beds, the carbon media is not allowed to become compacted in the tank and the possibility of channeling is greatly reduced. If a carbon bed channels, the effectiveness of the media is greatly reduced.

Reverse Osmosis Unit #1 - rejects inorganics, dissolved minerals, suspended solids and microbiological contaminants from the product water stream. In order to meet AAMI standards for dissolved inorganics, a reverse osmosis unit is utilized.

Storage Tank - stores reverse osmosis product water and acts as a integral part of the dialysis water loop. The storage tank utilized in this design is a cone bottom on a stand to insure complete drainage when sanitizing. The reverse osmosis unit produces water at a steady rate which is sometimes slower than demand, the storage tank will allow for fluctuations in this demand. The storage tank is equipped with level sensors to turn on the reverse osmosis unit when water is required, and to turn off when the tank is full. A vent filter removes airborne bacteria while allowing the tank to breathe. A spray array is utilized in the sanitization mode so that the interior of the tank is continually rinsed.

{4}------------------------------------------------

Reverse Osmosis Unit #2 - Stored product water is then used to feed the second RO system and its product water is directly fed to the distribution plumbing. What ever water is not used in the distribution loop is returned to the storage tank, thus a constant recirculation flow is maintained in the distribution plumbing loop. The water produced meets or exceeds AAMI standards for dissolved inorganics.

Remote Audible / Visual Alarm - is mounted at the nurses station and activates when an alarm / alert condition occurs in the RO system located in the water treatment room. The RO system may be shut down from this location, but must be restarted from the RO controller.

Deionizer Exchange Tanks - are sometimes incorporated into the system design as a water polish and back up to the single RO system. DI is used to remove inorganic water contaminants and dissolved gases from the water supply. By using water quality monitoring devices, this method assures treated water will meet AAMI recommended standards for inorganic removal. Whenever deionization exchange tanks are used in a water purification system for hemodialysis, sub micron membrane post filtration is used to remove any suspended solids and bacteria. Also, a temperature-compensated quality-indicating device is installed on all systems utilizing deionization exchange tanks. For additional safety, a remote audible/visual alarm is installed at the nurses' station to indicate any problems that may occur in the water treatment room.

Ultraviolet (UV) Disinfection Units are normally installed when Deionization exchange tanks are utilized in the system. Ultraviolet is a safe, clean and effective means to rid the water of bacteria, mold, virus and algae with out the use of heat or chemicals. A UV intensity monitor is also used any time a UV disinfection unit is installed in a water system.

Ultra Filters & Sub micron Post Filters - are used to remove any suspended solids, "free floating" bacteria, and inhibit pyrogens from entering the water used in dialysis. Post filters must be used in conjunction with storage tanks, DI and UV.

Monitoring Devices - are used to assure the proper performance of the water treatment equipment. Pressure gauges, flow meters, temperature-compensated water quality monitors, test ports, temperature gauge, water meter, etc. are installed at specific points throughout the system to measure each component's performance.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. The text is written in all capital letters.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY - 3 2000

Mr. Richard C. Herzberger H2Only Company Serve-A-Pure Company 1101 Columbus Avenue Bay City, MI 48708

Re: K993520 H2Only Complete Water Purification System for Hemodialysis Dated: February 11, 2000 Received: February 14, 2000 Regulatory Class: Il 21 CFR §876.5665/Procode: 78 FIP

Dear Mr. Herzberger:

We have reviewed your Section 510(k) notification of intent to market the device and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalion assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predication for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".

Sincerely yours,

Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{6}------------------------------------------------

Page 1 of

510(k) Number (if known): K993520

Device Name: H2Only Complete Water Purification System for Hemodialysis

Indications For Use:

H2Only Water Purification System for Hemodialysis is intended for use in hemodialysis treatment. It is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. When used as a medical device, Federal law restricts this device to sale by or on the order of a physician.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K993520
BS

§ 876.5665 Water purification system for hemodialysis.

(a)
Identification. A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.(b)
Classification. Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.