(13 days)
Syntron's QuikPac II One Step Cocaine assay is a rapid, qualitative, competitive binding immunoassay for the determination of Cocaine in urine at the cutoff level of 300 ng/ml (NIDA screening cutoff is 300 ng/ml). The test provides only preliminary data which should be confirmed by other methods such as gas chromatography/mass spectrophotometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated6. Syntron's QuikPac II One Step Cocaine Test is not intended to monitor drug levels, but only to screen urines for the presence of Cocaine and its metabolites.
Syntron's QuikPac II One Step Cocaine Test consists of a chromatographic absorbent device in which the drug or drug metabolites in the sample compete with a drug conjugate immobilized on a porous membrane support for the limited antibody sites. As the test sample flows through the absorbent device, the labeled antibody-dye conjugate binds to the free drug in the specimen forming an antibody:antigen complex. This complex competes with immobilized antigen conjugate in the positive reaction zone and will not produce a magenta color band when the drug is above the detection level of 300 ng/ml. Unbound dye conjugate binds to the reagent in the control zone, producing a magenta color band, demonstrating that the reagents and device are functioning correctly.
Here's a breakdown of the acceptance criteria and study information for the QuikPac II One Step Cocaine Test based on the provided text:
Acceptance Criteria and Device Performance
| Metric | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Relative Sensitivity (Positive Samples) | High agreement with a commercial EIA test and GC/MS for positive samples. | In-house: 1.000 (100%) |
| Clinical Trial: 99.89% | ||
| Relative Specificity (Negative Samples) | High agreement with a commercial EIA test and GC/MS for negative samples. | In-house: 1.000 (100%) |
| Clinical Trial: 100% | ||
| Accuracy | High overall accuracy compared to a commercial EIA test and GC/MS. | In-house: 100% |
| Clinical Trial: 99.56% | ||
| Detection Level | Should detect cocaine at a cutoff of 300 ng/ml. | Confirmed by testing (implied by design and correlation to NIDA cutoff). |
Note on Acceptance Criteria: The document primarily reports the device performance and then states that it "yielded" those results or that "the combined data yielded" them. The exact explicit acceptance criteria are not formally listed, but they are implied by the high performance metrics themselves, suggesting the device successfully met the sponsor's internal standards and what would be expected for a device of this type.
Study Information
1. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: 318 samples (for the clinical trial).
- Data Provenance: The document does not explicitly state the country of origin. It indicates it was an "in-house testing" and a "clinical trial." Given Syntron Bioresearch, Inc. is located in Carlsbad, California, it's reasonable to infer the "in-house" testing was done in the USA and the clinical trial likely also involved US-based samples, though this is not explicitly stated. The nature of the samples for the clinical trial is implied to be urine samples from individuals, presumably collected prospectively or retrospectively and then anonymized. The text does not specify if the clinical trial data was retrospective or prospective.
2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- The document does not mention the use of experts to establish the ground truth for the test set in the traditional sense of human review.
- Instead, the ground truth was primarily established through laboratory methods: "All positive samples by either screening method were confirmed by GC/MS."
3. Adjudication Method for the Test Set:
- The primary adjudication method involved GC/MS confirmation for all positive samples from either the QuikPac II or the commercial EIA test.
- For discrepant samples, the text notes an analysis of "2 discrepant samples" where "Two samples were positive by both GC/MS and QuikPac II, but negative for Cocaine by Emit II." This suggests further investigation and comparison with the definitive GC/MS result to understand the nature of discrepancies among the screening methods.
4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
-
No, an MRMC comparative effectiveness study was not done. This device is a rapid, qualitative, competitive binding immunoassay, which does not involve human "readers" interpreting results in the same way an imaging device or diagnostic tool typically would. The "reading" is a simple visual interpretation of color bands, and the focus is on the agreement with laboratory reference methods.
- Effect Size of Human Readers: Not applicable since an MRCM study was not performed.
5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, the performance reported is essentially a "standalone" or "device-only" performance when compared to reference methods. While a human visually interprets the color bands, the device's inherent sensitivity, specificity, and accuracy are being evaluated against the gold standard (GC/MS) and a predicate device (commercial EIA). There's no "human-in-the-loop" AI component to evaluate.
6. The Type of Ground Truth Used:
- Gas Chromatography/Mass Spectrometry (GC/MS): This was the definitive ground truth for confirming the presence or absence of cocaine and its metabolites, especially for positive samples and discrepant results.
- Commercial EIA Test for Cocaine: This served as a comparative screening method against which the QuikPac II device's performance was initially measured. While not the absolute ground truth, it was used as a benchmark, with GC/MS as the ultimate arbiter for confirmation.
7. The Sample Size for the Training Set:
- The document does not provide information regarding a separate "training set" size. This type of immunoassay device is developed and validated through laboratory testing and clinical trials rather than through machine learning models that require distinct training and test sets.
8. How the Ground Truth for the Training Set was Established:
- Since there's no explicitly mentioned "training set" in the context of a machine learning model, this question is not applicable. The device's development would have involved extensive R&D and optimization based on known concentrations of cocaine and its metabolites, with performance verified against laboratory standards (likely including GC/MS) throughout the design and optimization phases, which could be considered an iterative development process rather than a formal "training set."
{0}------------------------------------------------
12992990
510k Submission for QuikPac II One Step Cocaine(300) Test
Syntron Bioresearch, Inc.
Page 105 of 105 Pages
Revision A 6/05/99, Printed on 08/20/1999
15.1.Summary of Safety and Effectiveness
The sponsor, Syntron Bioresearch, Inc. (2774 Loker Ave. West, Carlsbad, California, 92008), has developed, manufactured, and tested under GMP/GLP guidelines a device for the qualitative testing of urine for the presence of Cocaine and its metabolites in a screening format.
The trade name of the device is QuikPac II One Step Cocaine Test having a designated common name of Cocaine Test System and a · classification as a Class II device per 21 CFR ¶ 862.3250. This device is intended for the medical/forensic screening of urine.
Syntron's QuikPac II One Step Cocaine Test consists of a chromatographic absorbent device in which the drug or drug metabolites in the sample compete with a drug conjugate immobilized on a porous membrane support for the limited antibody sites. As the test sample flows through the absorbent device, the labeled antibody-dye conjugate binds to the free drug in the specimen forming an antibody:antigen complex. This complex competes with immobilized antigen conjugate in the positive reaction zone and will not produce a magenta color band when the drug is above the detection level of 300 ng/ml. Unbound dye conjugate binds to the reagent in the control zone, producing a magenta color band, demonstrating that the reagents and device are functioning correctly.
In-house testing of Syntron's QuikPac II One Step Cocaine Test yielded a relative sensitivity or agreement within positive samples of 1.000 and a relative specificity or agreement within negative samples of 1.000 and an accuracy of 100% when tested against a commercial EIA test for Cocaine on samples documented to be positive by GC/MS. A clinical trial consisting of 318 samples was run and the combined data yielded a relative sensitivity of 99.89%, a relative specificity of 100% with an accuracy of 99.56% when compared to a commercial EIA test for Cocaine.
All positive samples by either screening method were confirmed by GC/MS. The results on the 2 discrepant samples clearly demonstrated similar errors by the Emit II method. Two samples were positive by both GC/MS and QuikPac II, but negative for Cocaine by Emit II. Previous testing samples demonstrated to be adulterated with "Clean Jane", [sodium dodecylsulfate (Tide)], which is supposed to prevent a positive test tested positive by both screening tests (Emit II & QuikPac II). GC/MS analysis returned a negative for the drug (cocaine) with a specific signature for the adulterant.
Additional information on this submission may be obtained by contacting Dr. Cleve W. Laird, President, Drial Consultants, Inc. at 805-522-6223(Ca) or by fax at 805-522-1526.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. In the center of the circle is an abstract image of an eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
SEP 20 1999
Syntron Bioresearch, Inc. c/o Dr James M. Barquest California Department of Health Food & Drug Branch P.O. Box 942732 (MS-357) Sacramento, California 94234
Re: K992990
Trade Name: QuikPac II One Step Cocaine Assay Regulatory Class: II Product Code: DIO Dated: August 20, 1999 Received: September 7, 1999
Dear Dr. Barquest:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours.
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
FOOD & DRUG BRANCH The States
510(k) Number (if Known): Not yet assigned
Device Name: QuikPac II One Step Cocaine assay
Indications For Use:
Syntron's QuikPac II One Step Cocaine assay is a rapid, qualitative, competitive binding immunoassay for the determination of Cocaine in urine at the cutoff level of 300 ng/ml (NIDA screening cutoff is 300 ng/ml). The test provides only preliminary data which should be confirmed by other methods such as gas chromatography/mass spectrophotometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated6. Syntron's QuikPac II One Step Cocaine Test is not intended to monitor drug levels, but only to screen urines for the presence of Cocaine and its metabolites.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANDOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Sean Cooper
Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number_k 99 2990
Prescription Use: (Per 21 CFR 801.109
or
Over The Counter Use: (Optional Format 1-2-96)
§ 862.3250 Cocaine and cocaine metabolite test system.
(a)
Identification. A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.(b)
Classification. Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).