K Number
K992849
Date Cleared
1999-09-23

(30 days)

Product Code
Regulation Number
864.5220
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QBC STAR™ Centrifugal Hematology System provides a diagnostic hematology profile on venous or capillary blood: hematocrit, hemoglobin, mean corpuscular hemoglobin concentration (MCHC), platelet count, white blood cell count, granulocyte count (% and number) and lymphocyte/monocyte count (% and number).

Device Description

The QBC STAR™ Centrifugal Hematology System is a self-contained, whole blood automated hematology system. Testing can be performed on venous or capillary samples. All tests are performed in the QBC STAR™ Blood Collection Tube. The system is powered by a universal voltage internal power supply that plugs directly into an AC power source. The instrument is factory preset and does not require user calibration. The methodology of the QBC STAR centrifugal hematology system is based on electro-optical linear measurements of the discrete layers of packed blood cells in a microhematocrit-type tube. The cell layering results from density gradients formed during high-speed centrifugation of the blood. Nine primary hematology values, including, hematocrit, hemoglobin, mean corpuscular hemoglobin concentration (MCHC), white blood cell count, platelet count, granulocyte count (% and number) and lymphocyte/monocyte count (% and number) are derived. QBC hematology tests utilize precision-bore glass tubes pre-coated with potassium oxalate, acridine orange fluorochrome stain and an agglutinating agent. During high-speed centrifugation of the blood-filled tube, the cells form in packed layers, according to their density, around the float, which has descended into the red blood cells. The buffy coat is automatically scanned and fluoresence and absorbance readings are made to identify the expanded layers of differentiated cells. Volumes of these packed cell layers are then computed to obtain quantitative values for the listed parameters. On the QBC STAR system, the hematocrit, white blood cell counts and the platelet count are direct measurements of the cell lavers. The hemoglobin measurement is directly related to the density of the red blood cells and based on the depth of penetration of the float into the red blood cell layer. Mean Corpuscular Hemoglobin Concentration (MCHC) is a function of hemoglobin and hematocrit and is electronically calculated according to the standard equation (Hab/Hct x 100).

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the QBC STAR™ Centrifugal Hematology System, formatted as requested:

Acceptance Criteria and Study for QBC STAR™ Centrifugal Hematology System

1. Table of Acceptance Criteria and Reported Device Performance

The provided document doesn't explicitly state quantitative "acceptance criteria" in a typical numerical format (e.g., "accuracy must be >95%"). Instead, the acceptance criteria are implicitly defined by demonstrating equivalent performance to the predicate device, the QBC™ AUTOREAD™ Plus Hematology System, for all measured parameters. The reported device performance is therefore the demonstrated equivalence.

Acceptance Criteria CategorySpecific Criteria (Implicit for Equivalence)Reported Device Performance
Overall PerformanceEquivalent performance to predicate device across all hematology parameters.Validation through accuracy testing of venous and capillary samples showed equivalent performance to the previously cleared product.
Technical EquivalenceModifications do not present new issues of safety and effectiveness.Modifications were assessed through risk analysis. Verification tests were performed.
Software IntegritySystem software appropriately rejects or accepts processed samples.Assessed.
Optical Component FunctionEquivalence between new optical components and those of the previously cleared product.Comparison of scans from multiple blood and control samples.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document mentions "multiple blood and control samples" and "venous and capillary samples."
  • Data Provenance: Not explicitly stated. Given the context of a 510(k) submission to the FDA, it is highly likely the studies were conducted in the USA, but this is not confirmed. The document does not specify if the data was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. This type of device (hematology analyzer) typically uses established laboratory reference methods or predicate devices for ground truth comparison, rather than human experts interpreting results. The predicate device's performance serves as the benchmark for equivalence.

4. Adjudication Method for the Test Set

Not applicable. As described above, human expert adjudication is not typically used for establishing ground truth for automated hematology analyzers. The device's measurements are directly compared against established reference methods or the predicate device's results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study evaluates human reader performance, typically in imaging diagnostics, and is not relevant for an automated hematology analyzer.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the study demonstrates the standalone performance of the QBC STAR™ Centrifugal Hematology System. This device is an automated hematology analyzer, meaning it operates without human intervention in the interpretation or direct measurement of results from a sample. Its performance is evaluated independently of a human operator's judgment apart from proper sample collection and loading.

7. The Type of Ground Truth Used

The ground truth was established by:

  • Predicate Device Performance: The QBC™ AUTOREAD™ Plus Hematology System's performance served as the primary benchmark for demonstrating substantial equivalence.
  • Established Laboratory Reference Methods: Although not explicitly detailed, "accuracy testing" usually implies comparison to well-established and validated laboratory methods for hematology parameters.

8. The Sample Size for the Training Set

Not explicitly stated. The document focuses on verification testing and comparison to a predicate device, rather than detailing a separate "training set" in the context of machine learning. The device's algorithms are pre-programmed and based on physical principles of blood separation and optical measurement, not on machine learning models that require a distinct training set in the modern sense.

9. How the Ground Truth for the Training Set Was Established

Not applicable in the typical machine learning sense. The "ground truth" for developing the underlying algorithms and measurement principles of the QBC STAR™ (and its predicate) would have been established through:

  • Known physical properties of blood cells: Density, size, and fluorescent properties.
  • Extensive laboratory testing and empirical data: To refine the algorithms that convert optical measurements to cell counts and other hematology parameters.
  • Clinical correlation: Ensured the device's measurements correlated with patient health status and clinical needs.

The document indicates that existing algorithms from the predicate device were likely re-used or adapted ("A series of numerical algorithms convert the volume of measured cell material in each layer to an equivalent cell count," identical description for both devices in Table 1). Therefore, the "training" (development and refinement) of these algorithms would have occurred during the development of the predicate device and potentially minor adjustments during the development of the QBC STAR system.

{0}------------------------------------------------

SEP 2 3 1999

510(k) SUMMARY

  • BECTON DICKINSON MICROBIOLOGY SYSTEMS SUBMITTED BY: 7 LOVETON CIRCLE SPARKS, MD 21152
  • CONTACT: Monica E. Giguere Requlatory Affairs Associate
  • (410) 316-4287 TELEPHONE:
  • PREPARED: August 20, 1999
  • QBC STAR™ Centrifugal Hematology System DEVICE NAME:

PREDICATE DEVICE:

QBC™ AUTOREAD™ Plus Hematology System

The QBC STAR™ Centrifugal Hematology System INTENDED USE: provides a diagnostic hematology profile on venous or capillary blood: hematocrit, hemoglobin, mean corpuscular hemoglobin concentration (MCHC), platelet count, white blood cell count, granulocyte count (% and number) and lymphocyte/monocyte count (% and number).

{1}------------------------------------------------

DEVICE DESCRIPTION:

The QBC STAR™ Centrifugal Hematology System is a self-contained, whole blood automated hematology system. Testing can be performed on venous or capillary samples. All tests are performed in the QBC STAR™ Blood Collection Tube. The system is powered by a universal voltage internal power supply that plugs directly into an AC power source. The instrument is factory preset and does not require user calibration.

The methodology of the QBC STAR centrifugal hematology system is based on electro-optical linear measurements of the discrete layers of packed blood cells in a microhematocrit-type tube. The cell layering results from density gradients formed during high-speed centrifugation of the blood. Nine primary hematology values, including, hematocrit, hemoglobin, mean corpuscular hemoglobin concentration (MCHC), white blood cell count, platelet count, granulocyte count (% and number) and lymphocyte/monocyte count (% and number) are derived.

QBC hematology tests utilize precision-bore glass tubes pre-coated with potassium oxalate, acridine orange fluorochrome stain and an agglutinating agent. During high-speed centrifugation of the blood-filled tube, the cells form in packed layers, according to their density, around the float, which has descended into the red blood cells. The buffy coat is automatically scanned and fluoresence and absorbance readings are made to identify the expanded layers of differentiated cells. Volumes of these packed cell layers are then computed to obtain quantitative values for the listed parameters.

On the QBC STAR system, the hematocrit, white blood cell counts and the platelet count are direct measurements of the cell lavers. The hemoglobin measurement is directly related to the density of the red blood cells and based on the depth of penetration of the float into the red blood cell layer. Mean Corpuscular Hemoglobin Concentration (MCHC) is a function of hemoglobin and hematocrit and is electronically calculated according to the standard equation (Hab/Hct x 100).

{2}------------------------------------------------

DEVICE COMPARISON:

The QBC STAR Centrifugal Hematology System is a modification of the QBC AUTOREAD Plus System and is substantially equivalent to it. Table 1 summarizes the similarities and differences between the QBC STAR system and the modified/predicate device.

FEATUREQBC STAR SystemQBC™ AUTOREAD™ Plus
Intended UseThe QBC STAR CentrifugalHematology System providesa diagnostic hematologyprofile on venous or capillaryblood: Hematocrit,Hemoglobin, MeanCorpuscular HemoglobinConcentration (MCHC),Platelet Count, White bloodcell count, Granulocyte (%and number) andLymphocyte/Monocyte (% andnumber).The QBC AUTOREAD PlusSystem provides a diagnostichematology profile on venousor capillary blood: Hematocrit,Hemoglobin, MeanCorpuscular HemoglobinConcentration (MCHC),Platelet Count, White bloodcell count, Granulocyte (%and number) andLymphocyte/Monocyte (% andnumber).
MethodologyElectro-optical linearmeasurements of the discretelayers of blood cells in amicrohematocrit-type tube.Electro-optical linearmeasurements of the discretelayers of blood cells in amicrohematocrit-type tube.
Sample TubeUses the QBC™ AccuTubeenclosed in a plastic sleeve.(a.k.a. QBC STAR™ BloodCollection Tube)Uses the QBC™ AccuTube. Itis not enclosed in a plasticsleeve.
FloatInjection molded polystyreneplasticInjection molded polystyreneplastic
Float InsertionInserted by user duringcapping operation.Inserted by user as a distinctstep.
Fill Volume$70 \mu L \pm 5 \mu L$$70 \mu L \pm 5 \mu L$
Closed-end stopperPolyethylene plug with smallair vent hole that does notrequire specific plug seatingstep by the user. Theextruded plug has afluorescent dye additive so thebottom of the red blood cellscan be locatedSantoprene plug with molded-in vents that must be seatedby the user. The injection-molded plug also uses afluorescent dye additive so thebottom of the red blood cellscan be located.
FEATUREQBC STAR SystemQBC™ AUTOREAD™ Plus
Instrument ModulesCentrifuge and analyzer are in one housing.Centrifuge and analyzer are separate modules.
CapacitySpins and analyzes one tube at a time.Can spin 20 tubes at once.Analyzes one tube at a time.
Centrifuge CyclePacking cycle is preceded by a mixing cycle where the rotor spins at 5000 RPM until the float reaches the bottom of the tube. The rotor then spins at 11,000 RPM for 4.75 minutes to separate and pack the blood into cell layers.There is no mixing cycle.The centrifuge provided with the system spins at 12,000 RPM for 5 minutes to separate and pack the blood into cell layers.
Optical Measurement SubsystemOptical measurement system is based on capturing the intensity profile (image) of the tube in the centrifuge rotor as it rotates past the detector assembly.Optical measurement system is based on moving the tube linearly through a fixed optical system to build an intensity profile of the tube.
Fluorescent MeasurementElectronic scan of tube:• Region of tube illuminated by blue flash lamp• Array of photodetectors• Electronic sequencing of arrayMechanical scan of tube:• Point on tube illuminated by blue incandescent lamp.• Single detector• Mechanical scan of tube.
Cell Layer DeterminationThe intensity profile obtained, using 2 wavelengths of emitted light (red and green), is used to determine the beginning and the end of each cell layer.The intensity profile is measured at 45-degree intervals around the tube so that any variations in the length of the cell layers may be averaged.The intensity profile obtained, using 2 wavelengths of emitted light (red and green), is used to determine the beginning and the end of each cell layer.The intensity profile is measured at 45-degree intervals around the tube so that any variations in the length of the cell layers may be averaged.
Transmission MeasurementElectronic scan of tube:• Region of tube illuminated by six 630nm LEDs.• Array of photodetectors.• Electronic sequencing of arrayMechanical scan of tube:• Point on tube illuminated by a single 610nm LED.• Single detector• Mechanical scan of tube.
Blue Excitation Light SourceXenon linear arc lamp with elliptical reflector illuminating a region of tube through a band pass filter with an average %T of 70% over a wavelength of 390nm to 490nm.Tungsten filament lamp focused into a spot on the tube through a single aspherical lens and a band pass filter with an average %T of 45% over a wavelength of 410nm to 485nm.
Red Emission FilterPlasma coated interference filter. The average %T is 80% over a wavelength of 610nm to 670nm.A 610nm longpass red glass filter combined with infra-red glass blocking filter and blue blocking glass filter. The average %T is 10%.
FEATUREQBC STAR SystemQBC™ AUTOREAD™ Plus
Green emission filterPlasma coated interferencefilter. The average %T = 70%over a wavelength of 520nmto 570nm.A 530 green transmitting glassfilter combined with glassinfra-red blocking and blueblocking filters. The average%T is 20% over a wavelengthof 530nm to 590nm.
Locating Bottom of FloatA transmittance scan is usedto locate the float bottom, 8red fluorescent scans, 8 greenfluorescent scans are taken ofthe float region of the tube.A transmittance scan is usedto locate the float bottom, 8red fluorescent scans, 8 greenfluorescent scans are taken ofthe float region of the tube.
Identification of the Bottom ofRBCsThe instrument scans thebottom of the sampleassembly in fluorescencemode to pick up thefluorescent signal from theclosed end stopper.The instrument scans thebottom of the sampleassembly in fluorescencemode to pick up thefluorescent signal from theclosed end stopper.
Fill Volume DeterminationThe instrument scans theplasma area of the sample influorescence mode (blue lightexcitation). The meniscusprovides a large signaltransition that identifies thetop of the sample column.The instrument scans theplasma area of the sample influorescence mode (blue lightexcitation). The meniscusprovides a large signaltransition that identifies thetop of the sample column.
Identification of buffy coatlayers.A sequence of processing theRed and Green float scansidentifies the location of thethree layers of buffy coat cells.This sequence uses patternsof similarities as well asdissimilarities in emissioncolor.A sequence of processing theRed and Green float scansidentifies the location of thethree layers of buffy coat cells.This sequence uses patternsof similarities as well asdissimilarities in emissioncolor.
Verification of processedsample integrityThe data scans are evaluatedto determine if the layers haveformed properly, that thecolors of the buffy coat havethe proper signal intensity andthat the patterns of thesecolors follow a known set ofrules.The data scans are evaluatedto determine if the layers haveformed properly, that thecolors of the buffy coat havethe proper signal intensity andthat the patterns of thesecolors follow a known set ofrules.
AlgorithmsA series of numericalalgorithms convert the volumeof measured cell material ineach layer to an equivalentcell count.A series of numericalalgorithms convert the volumeof measured cell material ineach layer to an equivalentcell count.
Calibration CheckUses internal calibrationtargets to verify properfunction of the centrifuge andoptical subsystems for eachtest run. The results fromeach run are compared withresults stored at the time ofmanufacturing.Uses an external calibrationrod that the operator must runon a daily basis. The resultsare compared againstacceptance ranges by the userto verify proper systemoperation.

Table 1: Device Comparison - QBC STAR™ Centrifugal Hematology System vs. QBC™ AUTOREAD™ Plus System

' The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without pre-market approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts

{3}------------------------------------------------

{4}------------------------------------------------

{5}------------------------------------------------

The modifications included in the QBC STAR system were assessed through risk analysis and verification tests were performed as a result of this risk analysis. The impact to the system of modifications to the sample collection tube was tested through simulated testing. The ability of the system software to appropriately reject or accept the processed samples was assessed.

Comparison of scans from multiple blood and control samples were made to determine equivalence between the new optical components and those used on the previously cleared product. Validation through accuracy testing of venous and capillary samples showed equivalent performance to the previously cleared product.

Although the QBC STAR Centrifugal Hematology System has modifications from the previously cleared QBC AUTOREAD Plus System these modifications do not present new issues of safety and effectiveness.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

SEP 2 3 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Monica E. Giguere Regulatory Affairs Associate Becton Dickinson & Company 7 Loveton Circle Sparks, Maryland 21152

K992849 Re:

Trade Name: Becton Dickinson & Company, QBC™ STAR Centrifugal Hematology System Regulatory Class: II Product Code: GKZ Dated: August 20, 1999

Received: August 24, 1999

Dear Ms. Giguere:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{7}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Page 1 of 1

K992849 510(k) Number (if known)

QBC STARTM Centrifugal Hematology System Device Name:

Indications For Use:

The QBC STAR Centrifingal Hematology System provides a diagnostic hematology profile on venous or capillary blood: hematocrit, hemoglobin, MCHC (Mean Corpuscular Hemoglobin Concentration, platelet count, white blood cell count, granulocyte count (% and number) and lymphocyte/monocyte count (% and number).

Pete E. Maker

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K992849

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluations (ODE)

Prescription

(Optional Format 3-10-98)

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”