K Number
K992711
Date Cleared
1999-12-06

(117 days)

Product Code
Regulation Number
864.7925
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sigma Diagnostics ALEXIN™ HS is a device used for primary screening for coagulation abnormalities, for evaluation of the effect of therapy on procoagulant disorders, and as an assay for coagulation factor deficiencies of the intrinsic coagulation pathway.

Device Description

Sigma Diagnostics ALEXIN 14 HS contains purified rabbit brain and soy phospholipids with an ellagic acid activator for the determination of the activated partial thromboplastin time and related coagulation procedures in citrated plasma.

AI/ML Overview

Sigma Diagnostics ALEXIN™ HS is a device used for primary screening for coagulation abnormalities, for evaluation of the effect of therapy on procoagulant disorders, and as an assay for coagulation factor deficiencies of the intrinsic coagulation pathway. The study aimed to demonstrate its substantial equivalence to Dade Actin-FS.

1. Table of acceptance criteria and reported device performance:

The document doesn't explicitly state "acceptance criteria" in a numerical sense, but rather demonstrates substantial equivalence through correlation and regression with a predicate device, and provides precision data. For the purpose of this analysis, we will infer the acceptance criteria from the reported performance, implying that the achieved correlation and precision were considered acceptable for substantial equivalence.

MetricAcceptance Criteria (Inferred from Predicate Equivalence)Reported Device Performance (ALEXIN™ HS)
Correlation with Dade Actin-FS (Optical Mode)High correlation (e.g., > 0.95 or similar to predicate)0.97
Correlation with Dade Actin-FS (Mechanical Mode)High correlation (e.g., > 0.95 or similar to predicate)0.97
Correlation on KC 4A analyzerHigh correlation (e.g., > 0.95 or similar to predicate)0.96
Regression Equation (Optical Mode)Regression line close to y = x (i.e., slope ≈ 1, intercept ≈ 0)y = 0.81 x + 0.31
Regression Equation (Mechanical Mode)Regression line close to y = x (i.e., slope ≈ 1, intercept ≈ 0)y = 0.84 x + 0.28
Regression Equation (KC 4A analyzer)Regression line close to y = x (i.e., slope ≈ 1, intercept ≈ 0)y = 0.94 x + 0.08
Within-Run Precision (Control I CV%)Low variability (e.g., typically < 5%)2.00 %
Within-Run Precision (Control II CV%)Low variability (e.g., typically < 5%)1.68 %
Within-Run Precision (Control III CV%)Low variability (e.g., typically < 5%)1.92 %
Total Precision (Control I CV%)Low variability (e.g., typically < 10%)3.49 %
Total Precision (Control II CV%)Low variability (e.g., typically < 10%)2.74 %
Total Precision (Control III CV%)Low variability (e.g., typically < 10%)2.85 %

2. Sample size used for the test set and the data provenance:

  • Sample Size for Equivalence Study: 153 random frozen patient samples were used for comparison between ALEXIN™ HS and Dade Actin-FS in both optical and mechanical modes. An additional 97 random frozen samples were used for correlation on the KC 4A analyzer.
  • Data Provenance: The data provenance (e.g., country of origin, specific clinics) is not specified. The samples were "random frozen patient samples," implying they were clinical samples, but whether they were retrospective or prospective is not explicitly stated. Given the context of a 510(k) submission for substantial equivalence, it's common for such data to be retrospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable to this study. The study uses a predicate device (Dade Actin-FS) as the reference for comparison, not an independent "ground truth" established by experts.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not applicable to this study. The study involves direct comparison of numerical results from two analytical devices, not subjective assessments requiring expert adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable to this study. This device is an in-vitro diagnostic (IVD) reagent for automated laboratory testing (Activated Partial Thromboplastin Time, APTT), not an AI-assisted diagnostic imaging or interpretation tool that would involve human "readers" or an MRMC study.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The device itself is an "algorithm only" in the sense that it's a reagent used in an automated analyzer to produce a numerical result. The performance described is the standalone performance of the reagent in conjunction with the analytical instruments (optical, mechanical, KC 4A) in comparison to a predicate reagent. There is no human interpretation or intervention in the generation of the APTT value itself.

7. The type of ground truth used:

The "ground truth" in this study is the results obtained from the predicate device, Dade Actin-FS. The study aims to demonstrate substantial equivalence to this legally marketed device, rather than establishing biological or clinical truth via pathology, outcomes data, or expert consensus.

8. The sample size for the training set:

This information is not applicable to this study. This device is a chemical reagent, not an AI or machine learning model that requires a "training set" in the conventional sense. The "development" of the reagent would involve chemical formulation and optimization, not data-driven model training.

9. How the ground truth for the training set was established:

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

DEC - 6 1999 SUMMARY OF SAFETY AND EFFECTIVENESS ALEXINTM HS (Catalog No. A 0969/A 1094)

Sigma Diagnostics ALEXIN 14 HS contains purified rabbit brain and soy phospholipids with an ellagic acid activator for the determination of the activated partial thromboplastin time and related coagulation procedures in citrated plasma.

The activated partial thromboplastin time (APTT) is a general screening procedure for the detection of coagulation abnormalities in the intrinsic pathway. This assay was developed from the plasma recalcification time (PRT) by the addition of a surface activator to a suspension of rabbit brain cephalin. The activator eliminates the effect of variable glass surfaces on coagulation assays.

The APTT is sensitive to deficiencies or abnormalities of factors VIII, IX, XI, XII, X, V and II, prekallikrein, high molecular weight kininogen (HMWK), fibrinogen and to inhibitors of blood coagulation such as lupus anticoagulants and fibrin/fibrinogen degradation products. The presence of non-specific inhibitors, such as lupus-anticoagulants may prolong the APTT. However, this effect is variable and related to the APTT reagent employed. The APTT assay has also been widely used to monitor the effectiveness of standard heparin therapy, where the clotting time is prolonged in proportion to the level of unfractionated heparin.56 In summary, the APTT is a clinically relevant screening test for the diagnosis of coagulant disorders and therapeutic monitoring of unfractionated heparin.

The safety and effectiveness of Sigma Diagnostics ALEXIN HS has been demonstrated by showing its substantial equivalence to Dade Actin-FS. One hundred fifty-three random frozen patient samples were assayed in both the optical and mechanical modes with the described ALEXIN HS reagent (y) and with Dade Actin-FS (x). Comparison of the results yielded a correlation coefficient of 0.97 and regression equation of v = 0.81 x + 0.31 using the optical mode and a correlation coefficient of 0.97 and regression equation of y = 0.84 x + 0.28 using the mechanical mode. When 97 random frozen samples were correlated on the KC 4A analyzer . correlation coefficient was 0.96 and the regression equation was y = 0.94 x + 0.08. All regression results were assessed using logarithmically transformed data.

SampleControl IControl IIControl III
Mean, seconds26.546.265.0
Within-Run
1 SD, seconds0.530.771.25
CV %2.001.681.92
Total
1 SD, seconds0.921.271.85
CV%3.492.742.85

Precision was determined on the KC 4A according to the NCCLS publication EPS-A using plasma controls at three different APTT levels. The following data were obtained:

REFERENCES

    1. Manual of Hemostasis and Thrombosis, 3rd ed. AR Thompson, LA Harker, Editors, Davis Company, Philadelphia, 1983
    1. Human Blood Coagulation, Haemostasis and Thrombosis, 3rd ed. R Biggs, CR Rizza, Editors, Blackwell Scientific Publications London, 1984
    1. Surdsmo JS, Fair DS: Relationship among the complement, kinin, coagulation and fibrinolytic systems in the inflammatory reaction. Clin Physiol Biochem 1:225, 1983.

{1}------------------------------------------------

  • Forastiero RR, Cerrato GS, Carreras LO: Evaluation of Recently Described Tests for Detection of 4. the Lupus Anticoagulant. Throm. Haem. 72(5):728,1994.
    1. Basu D, Gallus MB, Hirsh J, Cade J: A prospective study of the value of monitoring heparin treatment with the activated partial thromboplastin time. N Engl J Med 287:324, 1972
  • Deykin D: Regulation of heparin therapy. N Engl J Med 287:355, 1972 6.
  • National Committee for Clinical Laboratory Standards. Evaluation of Precision Performance of 7. Clinical Chemistry Devices: Approved Guideline. NCCLS Publication EP5-A, Villanova, PA, 1999

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three wavy lines, which are meant to represent the human form.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC - 6 1999

William R. Gilbert, Ph.D. Manager, Scientific Affairs Sigma Diagnostics, Inc. 545 South Ewing Avenue St. Louis, Missouri 63103

Re: K992711 Trade Name: Sigma Diagnostics ALEXIN™ HS Regulatory Class: II Product Code: GFO Dated: October 14, 1999 Received: October 15, 1999

Dear Dr. Gilbert:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that. through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Kg9271 510(k) Number (if known): ___

Device Name: Sigma Diagnostics ALEXIN™ HS

Indications For Use:

Prescription Use

(Per 21 CFR 801.109)

Sigma Diagnostics ALEXIN™ HS is a device used for primary screening for coagulation abnormalities, for evaluation of the effect of therapy on procoagulant disorders, and as an assay for coagulation factor deficiencies of the intrinsic coagulation pathway.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)/
Division of Clinical Laboratory Devices
510(k) Number K992711

Prescription Use √ OR Over-The-Counter Use

OR

§ 864.7925 Partial thromboplastin time tests.

(a)
Identification. A partial thromboplastin time test is a device used for primary screening for coagulation abnormalities, for evaluation of the effect of therapy on procoagulant disorders, and as an assay for coagulation factor deficiencies of the intrinsic coagulation pathway.(b)
Classification. Class II (performance standards).