(151 days)
GradiLeiden V is a simple functional clotting test system intended for screening of resistance to Activated Protein C in plasma from individuals with the Factor V (Leiden) defect. It can also be performed on plasma from patients on stabilized oral anticoagulant or heparin therapy.
The GradiLeiden V Test is a lyophilized paired reagent containing 5 vials of whole diluted Agkistrodon contortrix venom and 5 vials of phospholipid rich Russell's Viper Venom time reagent.
Here's a breakdown of the acceptance criteria and study information for the GradiLeiden V Test, based on the provided 510(k) summary:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Sensitivity | 100% |
| Specificity | 98.8% |
| Within Run Precision for FVL+ plasma | 1.1% |
| Within Run Precision for normal plasma | 1.9% |
| Within Run Precision at cutoff | 0.9% |
| Total Precision for FVL+ plasma | 8.9% |
| Total Precision for borderline normal plasma | 1.6% |
| Total Precision for normal plasma | 5.6% |
| Identifies Factor V Leiden status correctly | 163/164 individuals |
| Identifies Oral Anticoagulated plasmas | 35/36 |
| Identifies heparinized plasmas | 21/21 |
| Identifies Lupus Anticoagulant positive plasmas | 12/12 |
Study Details
-
Sample size used for the test set and the data provenance:
- Test Set Size: 164 individuals (This number includes 35 oral anticoagulated plasmas, 21 heparinized plasmas, and 12 Lupus Anticoagulant positive plasmas, alongside other samples).
- Data Provenance: Not explicitly stated (e.g., country of origin). The submission is from Australia, suggesting the data may originate from there or an international collaboration. It describes the comparison as "GradiLeiden V was compared against the predicate device in a series of clotting assays," implying a prospective approach for the comparison study, but the source of the patient samples (retrospective/prospective collection) is not specified.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified. The ground truth was established by "DNA analysis," which implies a laboratory-based, molecular diagnostic method rather than expert interpretation of a diagnostic image or clinical presentation.
-
Adjudication method for the test set:
- Not applicable as the ground truth was based on DNA analysis, not expert consensus requiring adjudication.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI-based diagnostic device; it's a laboratory clotting test. The comparison was between two laboratory tests (GradiLeiden V and Coatest APC Resistance V) with ground truth established by DNA analysis.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone performance was done for the GradiLeiden V device. Its performance was evaluated against DNA analysis as the ground truth.
-
The type of ground truth used:
- DNA analysis (specifically mentioned: "all results confirmed by DNA analysis"). This is a highly accurate, molecular diagnostic method for determining Factor V Leiden status.
-
The sample size for the training set:
- Not explicitly stated. The document focuses on the test set used for validation. The device's cut-off of 1.57 was "obtained by ROC analysis," which implies a dataset was used to determine this threshold, but whether this constitutes a distinct "training set" in a machine learning sense, or if it was part of the overall validation process, is not detailed.
-
How the ground truth for the training set was established:
- Assuming the "training set" (or data used for ROC analysis) was drawn from similar sources as the test set, the ground truth would have been established by DNA analysis.
{0}------------------------------------------------
$$\llcorner \approx \sharp \sharp \sharp \sharp \flat \omega$$
510(k) Summary
GradiLeiden V Test
The submitter's name, address, telephone number, a contact person, and the date the (a)(1) summary was prepared;
Submitter's Name: Gradipore Ltd
Submitter's Address: Lot 16 Riverside Corporate Park 35 - 105 Delhi Rd, North Ryde 2113 Australia
Submitter's Telephone: 1800 762 2620
Submitter's Contact: Rhonda Pilgrim Regulatory Affairs Manager
Date 510(k) Summary Prepared: 21 July, 1999
The name of the device, including the trade or proprietary name if applicable, the common (a)(2) or usual name, and the classification name, if known;
Trade or Proprietary Name: GradiLeiden V
Common or Usual Name: APC Resistance Test
Classification Name: Hematology
An identification of the legally marketed device to which the submitter claims substantial (a)(3) equivalence;
Device Equivalent to: Coatest APC Resistance V
A description of the device. (a)(4)
Device Description: The GradiLeiden V Test is a lyophilized paired reagent containing 5 vials of whole diluted Agkistrodon contortrix venom and 5 vials of phospholipid rich Russell's Viper Venom time reagent.
A statement of the intended use of the device. (a)(5)
Device Intended Use:
Device Intended Use. OrialDoron - 15 a smillion with the Factor V (Leiden) defect. It can also be performed on plasma from patients on stabilized oral anticoagulant or heparin therapy.
Image /page/0/Picture/22 description: The image contains the words "Gradipore ORIGINAL (if in red)". The text is in a sans-serif font and is black on a white background. The text is slightly blurry and pixelated.
{1}------------------------------------------------
GRADILEIDEN V 510(k) Submission July, 1999
A summary of the technological characteristics of the new device in comparison to (a)(6) those of the predicate device.
GradiLeiden V is a clotting test based on lyophilised Russell's Viper Venom, which activates clotting at Factor X in the common pathway, while the Coatest APC Resistance V test is a clotting test based on an APTT system. The Factor V Leiden mutation affects Factor V in the common pathway, so both clotting tests are equally sensitive to the defect.
GradiLeiden V uses a lyophilised snake venom to activate the patient's own protein C while the Coatest APC Resistance V test uses lyophilised purified activated protein C.
Both tests can be automated in 2 stage clotting test systems with similar activation and acquisition times.
GradiLeiden V does not require dilution of plasma samples as the test is insensitive to heparin, Lupus Anticoagulants and to factor deficiencies caused by oral anticoagulants. Coatest APC Resistance V test requires dilution of patient samples in Factor V depleted plasma in order to eliminate sensitivity to heparin, Lupus Anticoagulants and oral anticoagulants.
A brief discussion of the nonclinical tests submitted, referenced, or relied on in the (b)1) premarket notification submission for a determination of substantial equivalency.
GradiLeiden V and Coatest APC Resistance V test both have the same intended use and can be used with the same patient groups. Both tests are automatable two stage clotting tests, wherein a blank test and a test in which either APC (Coatest APC Resistance V) or Venom activator (GradiLeiden V) are compared and a ratio obtained. The normal range for this ratio in both tests is reported as approximately 2-3. While both tests can be used for patients on oral anticoagulant or heparin therapy and for Lupus Anticoagulant positive patients. GradiLeiden V does not require plasma dilution while the Coatest APC Resistance V does. Within run precision for both tests is less than 5%.
A brief discussion of the clinical tests submitted, referenced, or relied on in the premarket (b)(2) notification submission for a determination of substantial equivalency.
GradiLeiden V was compared against the predicate device in a series of clotting assays, with all results confirmed by DNA analysis. Using a cut-off of 1.57 obtained by ROC analysis, Gradil.eiden V correctly identified the Factor V Leiden status of 163/164 individuals (82 negative). These included 35/36 Oral Anticoagulated plasmas, 21/21 heparinised plasmas and 12/12 Lupus Anticoagulant positive plasmas. GradiLeiden V was tested using the Stago STA, Dade Behring BCS, Organon Teknika MDA180 and manual tilt tube technique. The test was found to have 100% sensitivity and 98.8% specificity at a cut-off of 1.57 determined by ROC analysis.
Total precision estimates for FVL+, borderline normal, and normal plasma are 8.9%, 1.6%, and 5.6%, respectively. Within run precision estimates for FVL+ and normal plasma are 1.1% and 1.9%. Within run precision at the cutoff is 0.9%
(b)3) The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).
Gradipore Ltd considers GradiLeiden V to be substantially equivalent to the Chromogenix AB Coatest APC Resistance V Test in terms of intended use, method comparison and overall performance characteristics.
Image /page/1/Picture/13 description: The image shows the words "Gradipore ORIGINAL (if in red)" in a bold, sans-serif font. The words are stacked on top of each other, with "Gradipore" on the top line, "ORIGINAL" on the second line, and "(if in red)" on the third line. The text is black and the background is white. The text appears to be part of a label or document.
{2}------------------------------------------------
GRADILEIDEN V 510(k) Submission July, 1999
17.0 QUALITY ASSURANCE/QUALITY CONTROL
The following is Gradipore Ltd's recommended quality control for the use of GradiLeiden V Test in clinical laboratories:
Each laboratory should establish its own normal range and abnormal range for each lot of GradiLeiden V Test reagent.
The use of control plasma is recommended for monitoring coagulation assays following established laboratory quality control procedures. NCCLS recommends controls be assayed at the initiation of testing, at least once each shift, or with each group of assays. In high volume laboratories, controls should be tested with at least every 40 samples, as per NCCLS tentative guidelines H28-T, 1992.
If control values are out of range, do not report patient results. Determine which part of the instrument/reagent/control system is not functioning properly and correct it. After corrective measures are implemented and documented following good laboratory practice, retest the controls. If they are within range, patient samples can be tested and reported.
Image /page/2/Picture/6 description: The image shows the words "Gradipore ORIGINAL (if in red)" stacked on top of each other. The word "Gradipore" is at the top, followed by the word "ORIGINAL" in a slightly larger font. The phrase "(if in red)" is at the bottom in a smaller font. The text appears to be slightly distressed or textured.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 2 1 1999
Ms. Rhonda Pilgrim M.SC. Regulatory Affairs Manager Gradipore Ltd. Haemostasis Lot 16 Riverside Corporate Park 35 – 105 Delhi Road North Ryde 2113 Australia
Re: K992456 Trade Name: GradiLeiden V Test Regulatory Class: II Product Code: GGW Dated: November 10, 1999 Received: November 17, 1999
Dear Ms. Pilgrim:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
GRADILEIDEN V 510(k) Sudmissium July, 1999
18.0 INTENDED USE STATEMENT
510(k) Number (If known): K992456
Device Name: GradiLeiden V Test
Indications For Use:
GradiLeiden V is a simple functional clotting test system intended for sceening of resistance to Gradil eiden V is a smillional cloiding cloridize with the Factor V (Leiden) defect. It can also be
Activated Protein C in plasma from individuals with the Factor V (Leidene Activated Protein C in plasma from marrialians was more anticoagulant or heparin therapy.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
| Division of Clinical Laboratory Devices | |
| 510(k) Number | K952456 |
| Prescription Use (Per 21 CFR 801.109) | OR | Over-The-Counter Use ______ |
|---|---|---|
| (Optional Format 1-2-96) |
Image /page/5/Picture/8 description: The image shows the words "Gradipore ORIGINAL (if in red)" stacked on top of each other. The words are in a bold, sans-serif font. The words are slightly distorted, as if they were printed on a textured surface. The words are all in black ink.
§ 864.7925 Partial thromboplastin time tests.
(a)
Identification. A partial thromboplastin time test is a device used for primary screening for coagulation abnormalities, for evaluation of the effect of therapy on procoagulant disorders, and as an assay for coagulation factor deficiencies of the intrinsic coagulation pathway.(b)
Classification. Class II (performance standards).