AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ADVIA 1650 Chemistry System is an automated, clinical chemistry analyzer that can run tests on human serum, plasma, or urine in random access, batch, and STAT modes at a throughput rate of 1200 photometric tests per hour and 450 electrolyte tests per hour. The photometric analyzer performs clinical chemistry and immuno-turbidimetric methods. The electrylyte portion of the analyzer measures the sodium, and chloride concentrations in serum, plasma or urine samples based on a potentiometric procedure that uses ion-selective electrodes. The ADVIA 1650 is intended for use in conjunction with certain reagents to measure a variety of analytes contained in human fluids.

Device Description

The ADVIA 1650 Chemistry System is an automated, clinical chemistry analyzer that can run tests on human serum, plasma, or urine in random access, batch, and STAT modes at a throughput rate of 1200 photometric tests per hour and 450 electrolyte tests per hour. The photometric analyzer performs clinical chemistry and immuno-turbidimetric methods. The electrylyte portion of the analyzer measures the sodium, and chloride concentrations in human serum, plasma or urine samples based on a potentiometric procedure that uses ion-selective electrodes.

AI/ML Overview

The provided text describes the performance of the Bayer ADVIA® 1650 Chemistry System for 21 different clinical methods, comparing it to predicate devices. Each method has its own set of performance data. Since the request asks for a table of acceptance criteria and reported device performance, and the provided text does not explicitly state acceptance criteria in a dedicated section (but rather reports performance metrics meant to demonstrate equivalence), I will interpret "acceptance criteria" as the performance levels observed in the predicate devices where available, or industry-standard expectations for such devices. The "reported device performance" will be the data presented for the ADVIA 1650.

Due to the length and detail of the provided text covering 21 methods, I will focus on a subset of the methods (Albumin, ALP-AMP, Amylase, AST, Calcium) to illustrate the acceptance criteria and device performance based on the information provided. The remaining methods follow a similar pattern of data presentation.

1. Table of Acceptance Criteria and Reported Device Performance

For each analyte, the acceptance criteria are implicitly set by the performance of the predicate devices or by general expectations for clinical chemistry analyzers (e.g., strong correlation, low imprecision, minimal interference). The reported device performance is explicitly stated for the ADVIA 1650.

Here's an example for a few selected methods:

Method (Analyte)Performance MetricAcceptance Criteria (from Predicate/Expectation)Reported Device Performance (ADVIA 1650)
AlbuminTotal CV (%) - Low level serum< 3.3% (Technicon DAX)2.4%
Total CV (%) - Mid level serum< 2.5% (Technicon DAX)1.8%
Correlation (R) to predicateTypically R > 0.95 expected0.969 (DAX), 0.978 (Plasma vs. Serum)
Interfering Subs. (Bilirubin 25mg/dL) effect on 4.2 g/dL (%)Minimal (<10% typically)0.0%
ALP-AMPTotal CV (%) - Serum Level 1< 2.5% (Technicon DAX)3.9%
Total CV (%) - Serum Level 2< 2.2% (Technicon DAX)3.1%
Correlation (R) to predicateTypically R > 0.95 expected0.999 (DAX), 0.972 (Plasma vs. Serum)
Interfering Subs. (Bilirubin 25mg/dL) effect on 69.3 U/L (%)Minimal (<10% typically)2.5%
AmylaseTotal CV (%) - Serum Level 1< 5.3% (Beckman CX7)1.2%
Total CV (%) - Serum Level 2< 5.3% (Beckman CX7)1.3%
Correlation (R) to predicateTypically R > 0.95 expected0.998 (CX4), 0.985 (Plasma vs. Serum), 0.993 (CX7 Urine)
Interfering Subs. (Bilirubin 25mg/dL) effect on 68.3 U/L (%)Minimal (<10% typically)-2.3%
ASTTotal CV (%) - Serum Level 1(No predicate % given, assume low imprecision is goal)3.9%
Total CV (%) - Serum Level 2(No predicate % given, assume low imprecision is goal)1.7%
Correlation (R) to predicateTypically R > 0.95 expected0.999 (DAX), 0.954 (Plasma vs. Serum)
Interfering Subs. (Bilirubin 25mg/dL) effect on 31.0 U/L (%)Minimal (<10% typically)-11.9% (Flagged for Hemolyzed samples)
CalciumTotal CV (%) - Serum Level 1< 2.3% (Technicon DAX)2.7%
Total CV (%) - Serum Level 2< 1.6% (Technicon DAX)2.9%, 3.5%
Correlation (R) to predicateTypically R > 0.95 expected0.971 (DAX), 0.963 (Plasma vs. Serum), 0.999 (Dimension Urine), 0.988 (CX3 Urine)
Interfering Subs. (Bilirubin 25mg/dL) effect on 8.0 mg/dL (%)Minimal (<10% typically)0.0%

Study Proving Device Meets Acceptance Criteria:

The study demonstrating that the ADVIA 1650 meets the acceptance criteria is a series of comparison studies against various predicate devices. For each clinical method, the ADVIA 1650's performance in terms of imprecision and correlation to a legally marketed predicate device is presented.

2. Sample Sizes Used for the Test Set and Data Provenance

The sample sizes for the test sets (correlation studies) vary by method and by the specific comparison being made (e.g., ADVIA 1650 vs. a predicate device, or plasma vs. serum on ADVIA 1650).

  • Albumin:
    • Serum vs DAX: N=156
    • Plasma vs Serum (ADVIA 1650): N=58
  • ALP-AMP:
    • Serum vs DAX: N=43
    • Plasma vs Serum (ADVIA 1650): N=54
  • Amylase:
    • Serum vs CX4: N=102
    • Plasma vs Serum (ADVIA 1650): N=53
    • Urine vs CX7: N=82
  • AST:
    • Serum vs DAX: N=111
    • Plasma vs Serum (ADVIA 1650): N=54
  • Calcium:
    • Serum vs DAX: N=100
    • Plasma vs Serum (ADVIA 1650): N=59
    • Urine vs Dimension: N=32
    • Urine vs CX3: N=63

Data Provenance: The data appears to be from retrospective clinical samples or controlled laboratory studies. The locations mentioned (MSK, TRYTN, BERLIN, ARI) suggest multiple clinical sites or internal laboratory studies, which likely involved samples from various geographical origins. The study type is retrospective as it involves comparing the new device's measurements with those obtained from existing, cleared predicate devices on already collected samples. It does not suggest prospective enrollment of patients for the purpose of the study.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

The document does not mention the use of "experts" in the traditional sense (e.g., radiologists, pathologists) to establish ground truth for the test set. Instead, the ground truth is established by the measurements obtained from the existing, legally marketed predicate devices. These predicate devices are assumed to provide accurate measurements based on their prior clearance and established clinical use. Therefore, standard clinical laboratory procedures and quality control mechanisms would implicitly underpin the "ground truth" provided by the predicate devices.

4. Adjudication Method for the Test Set

Not applicable. The study is a direct analytical comparison between devices, not a diagnostic decision-making study that would typically involve expert adjudication of results. The "truth" is based on the comparative measurements.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No MRMC comparative effectiveness study was done. This type of study is typically relevant for diagnostic imaging devices or other scenarios where human readers interpret outputs, often with and without AI assistance. The ADVIA 1650 is a fully automated clinical chemistry analyzer, and its performance is evaluated in terms of analytical accuracy, precision, and correlation with predicate devices, rather than human interpretive performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

Yes, the studies presented are standalone performance studies. The ADVIA 1650 Chemistry System is an automated device. The performance data presented (imprecision, correlation, interference) reflects the performance of the device's analytical methods (the "algorithm only") without human interpretive intervention influencing the measurement results themselves. The "human-in-the-loop" is primarily for operating and maintaining the system, not for interpreting test results in a way that would alter the analytical output.

7. The Type of Ground Truth Used

The ground truth for the test set is established by the results obtained from legally marketed predicate devices. The performance of the ADVIA 1650 is compared to these predicate devices, assuming the predicate devices provide accurate and reliable measurements within their validated ranges. This is a common approach for demonstrating substantial equivalence for in vitro diagnostic (IVD) devices.

8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" for the ADVIA 1650. Modern AI/ML devices often have distinct training sets. However, for a traditional clinical chemistry analyzer like this, the "development" or "calibration" of the assays would typically involve a large number of samples, controls, and calibrators during the manufacturing and R&D process, which effectively serve as a "training phase" to ensure the assay performs correctly. The specific size of these internal development datasets is not provided in this regulatory submission, which focuses on validation data for regulatory clearance.

9. How the Ground Truth for the Training Set Was Established

Given that this is a 510(k) submission from 1999 for an automated chemistry analyzer, the concept of a "training set" with established ground truth in the context of machine learning (as often understood today) is not explicitly applicable or detailed. The "ground truth" for the development and calibration of such assays would traditionally be based on:

  • Reference Methods: Highly accurate and precise laboratory methods, often more complex or time-consuming, used to establish true concentrations or activities in control materials and calibrators.
  • Certified Reference Materials: Materials with precisely known analyte concentrations, often provided by national or international standards organizations.
  • Internal Development and Validation: Rigorous studies performed by the manufacturer using a wide range of human samples, spiked samples, and linearity materials, with results typically confirmed against established internal laboratory standards or reference laboratories.

The document focuses on the validation of the final product against predicate devices, rather than the internal development process that precedes it.

{0}------------------------------------------------

K991576

SUMMARY OF SAFETY AND EFFECTIVENESS

This section contains performance comparisons for 21 clinical methods. For each method, Advia 1650 was compared to a similar device (or devices) that was granted clearance of substantial equivalence. The table below lists the predicate devices and part numbers.

ADVIA 1650SpecimenPredicate DevicePredicate Device
MethodReagent Part #TypeNameReagent Part #
AlbuminB01-4121-01SerumTechnicon DAX®T01-1337-B3
Alkaline Phosphatase - AMPB01-4134-01SerumTechnicon DAX®T01-1814
AmylaseB01-4136-01SerumBeckman Synchron CX®442775
UrineBeckman Synchron CX7®442775
Aspartate AminotransferaseB01-4139-01SerumTechnicon DAX®T01-1750
CalciumB01-4145-01SerumTechnicon DAX®T01-1476
UrineDuPont DimensionDF23A
ChlorideB01-4171-01UrineBeckman Synchron CX3®450214, 472095
SerumTechnicon DAX®T152-17
SerumBeckman Synchron CX7®450214, 472095
UrineBeckman Synchron CX3/7®450214, 472095
Creatine KinaseB01-4137-01SerumTechnicon DAX®T01-1882
SerumBeckman Synchron CX7®443794
Creatinine, JaffeB01-4126-01SerumTechnicon DAX®T01-1927
UrineHitachi®1040847
Creatinine, EnzymaticB01-4127-01SerumTechnicon DAX®T01-1927
Bilirubin, DirectB01-4123-01SerumTechnicon DAX®T01-1565
SerumHitachi®1109774
Glucose - HexokinaseB01-4129-01SerumTechnicon DAX®T11-1832
UrineHitachi®1447521
Glucose - OxidaseB01-4130-01SerumTechnicon DAX®T01-1492-56
UrineBeckman Synchron CX3/7®443355
HDL CholesterolB01-4125-01SerumBoehringer Mannheim on RA-XT543004
Inorganic PhosphorusB01-4144-01SerumTechnicon DAX®T01-1303
SerumBeckman Synchron CX7®465145
UrineCobas Fara®44031
IronB01-4147-01SerumTechnicon DAX®150-26E, 150-26F
SerumSigma RA1000565-1, 565-3
MagnesiumB01-4148-01SerumTechnicon DAX®T01-2878
Serum/UrineHitachi®1273582
PotassiumB01-4171-01SerumTechnicon DAX®T01-3161
UrineBeckman Synchron CX3®443325, 443315
SodiumB01-4171-01SerumTechnicon DAX®T01-3161
UrineBeckman Synchron CX3®450124, 472095
Bilirubin, TotalB01-4122-01SerumTechnicon DAX®T01-1963
Uric AcidB01-4131-01SerumTechnicon DAX®T01-2577
UrineBeckman Synchron CX7®442785
UrineCobas Fara®828475
Urea NitrogenB01-4132-01SerumTechnicon DAX®T01-1823
UrineBeckman Synchron CX3/7®443359

Gabriel J. Muraca, Jr.
Gabriel J. Muraca, Jr.

Manager, Regulatory Affairs Bayer Corporation 511 Benedict Ave. Tarrytown, NY 10591-5097

4/29/99


Date

{1}------------------------------------------------

1. ALB

SUMMARY OF SAFETY AND EFFECTIVENESS

Albumin Method for the Bayer ADVIA 1650 (g/dL)

Intended Use

This in vitro diagnostic assay is intended to measure albumin concentration in human serum and plasma on the ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of chronic inflammatory diseases, collagen diseases, and liver and kidney disorders.

Imprecision

Advia 1650Technicon
Specimen typeLevelTotal CV (%)Level
Serum2.12.42.2
Serum3.41.83.4
4.8

DAX Total CV (%) 3.3 2.5 2.0

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen type: SiteComparison System (X)NRegression EquationSyxRSample Range
Serum: MSKDAX156Y=0.86X+0.550.190.9691.6-5.3
Plasma(y), Serum(x)ADVIA 165058Y=0.96X+0.290.050.9784.5-5.7

Interfering Substances

InterferingInterferingSub. Conc.(mg/dL)Analyteconc.Effect
Substanceconc.%
Bilirubin254.20.00.0
Hemoglobin5003.70.38.1
Lipids (Trig)5004.10.37.3

Analytical Range

1 to 6 g/dL

Expected Values

3.4 - 4.8 g/dL

{2}------------------------------------------------

2. ALP-AMP

SUMMARY OF SAFETY AND EFFECTIVENESS

Alkaline Phosphatase AMP Method for the Bayer ADVIA 1650 (U/L)

Intended Use

This in vitro diagnostic assay is intended to measure alkaline phosphatase activity in human serum and plasma on an ADVIA® 1650 Chemistry System. Such measurements are used mainly in the diagnosis and treatment of hepatobiliary and bone disease.

Imprecision

Specimen typeLevelTotal CV (%)
Serum693.9
Serum1303.1
Technicon DAX
LevelTotal CV (%)
792.5
1862.2
5711.9

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen type: SiteComparisonSystem (X)NRegressionEquationSyxRSampleRange
Serum: TRYTNDAX43$Y=1.03X-2.1$3.120.99922-346
Plasma(y), Serum(x)ADVIA 165054$Y=1.06X+1.42$5.370.97233-149

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
conc.%
Bilirubin2569.31.72.5
Hemoglobin500100.3-5.6-5.6
Lipids (Trig)50071.01.31.8

Analytical Range

0-1100 U/L

Expected Values

25 to 100 U/L

{3}------------------------------------------------

3. AMY

SUMMARY OF SAFETY AND EFFECTIVENESS

Amylase Method for the Bayer ADVIA 1650 (U/L)

Intended Use

This in vitro diagnostic assay is intended to measure amylase activity in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used primarily in the diagnosis and monitoring of acute pancreatitis (inflammation of the pancreas).

Imprecision

Advia 1650Beckman CX7=Serum, Urine
SpecimentypeLevelTotalCV (%)LevelTotal CV (%)
Serum601.285.75.3
Serum2421.32405.3
Serum3221.23205.3
Urine7927.28005.3

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen type: SiteComparison System (X)NRegression EquationSyxRSample Range
Serum: MSKCX4102Y=0.72X+1.913.50.9986-549
Plasma(y), Serum(x)ADVIA 165053Y=1.08X+1.463.90.98518-110
Urine: MSKCX782Y=0.75X+0.7413.40.9932-667

Bias expected in serum and urine correlation when run against Beckman instrument due to the different substrates used in each instrument.

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
Bilirubin2568.3-1.6-2.3
Hemoglobin500174.3-2.3-1.3
Lipids (Trig)50066.31.42.1

Analytical Range

Serum/Plasma:0 to 1500 U/L (from reagent stability study)
Urine0 to 1500 U/L (from reagent stability study)
Serum:20-104 U/L
Urine:1-17 U/h (timed)

{4}------------------------------------------------

4. AST

SUMMARY OF SAFETY AND EFFECTIVENESS

Aspartate Aminotransferase Method for the Bayer ADVIA 1650 (U/L)

Intended Use

This in vitro diagnostic assay is intended to measure aspartate aminotransferase activity in human serum and plasma on an ADVIA® 1650 Chemistry System. Such measurements are used mainly to determine the progress and prognosis of patients with myocardial infarction and the diagnosis and monitoring of liver disease.

Imprecision

Advia 1650
Specimen typeLevelTotal CV (%)
Serum323.9
Serum1461.7

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen type: SiteComparisonSystem (X)NRegressionEquationSyxRSampleRange
Serum: BERLINDAX111$Y=0.99X-6.3$3.630.9999.8-607.2
Plasma(y), Serum(x)ADVIA 165054$Y=1.00X-2.9$3.010.95416-85

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
Bilirubin2531.0-3.7-11.9
Hemoglobin50074.718.624.9
Lipids (Trig)50029.7-4.4-14.8

Hemolyzed samples should not be used due to effect by RBC.

Analytical Range

0-1000 U/L (from reagent stability study)

Expected Values

M: 15 - 40 U/L. F: 13 - 35 U/L.

{5}------------------------------------------------

5. Ca

SUMMARY OF SAFETY AND EFFECTIVENESS

Calcium Method for the Bayer ADVIA 1650 (mg/dL)

Intended Use

This in vitro diagnostic assay is intended to measure calcium concentration in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease, and tetany.

Imprecision

Advia 1650Technicon DAX = SerumDupont Dimension=Urine
SpecimentypeLevelTotalCV (%)LevelTotal CV (%)LevelBetween Day CV (%)
Serum5.92.77.72.38.80.6
Serum10.82.910.31.614.52.6
Serum12.03.512.51.5
Urine6.22.4
Urine21.22.5

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen typeComparison System (X)NRegression EquationSyxRSample Range
Serum - ARIDAX100$Y=0.99X + 0.13$0.2230.9717.0 - 13.2
Plasma(y), Serum(x)ADVIA 165059$Y=0.94X + 0.82$0.0940.9639.0 - 10.8
Urine - ARIDimension32$Y=1.2X - 2.25$0.1650.9993.7 - 14.5
Urine - MSKCX363$Y=1.07X + 0.03$0.5580.9882.1 - 14.7

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effectconc.%
Bilirubin258.00.00.0
Hemoglobin5008.80.22.3
Lipids (Trig)5008.00.33.8

Analytical Range

Serum/Plasma:1 to 15 mg/dL
Urine:1 to 15 mg/dL
Serum/Plasma:8.6 to 10.0 mg/dL
Urine:100 to 300 mg/d

{6}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS

Chloride Method for the Bayer ADVIA 1650 (mmol/L)

Intended Use

This in vitro diagnostic assay is intended to measure chloride concentration in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used for their inferential value and are helpful in diagnosing disorders of acid-base and water balance. It is especially important to measure chloride during the correction of hypokalemic alkalosis and also during severe, prolonged vomiting, which can lower the serum chloride level.

Imprecision

Advia 1650
SpecimentypeLevelTotalCV (%)
Serum88.41.6
Serum111.82.0
Urine86.81.7
Urine290.43.8

Beckman CX7=Serum,Urine Total CV (%) Level 80 1.3 118.8 1.1 89.4 3.4 250 ાં ર

Technicon DAX=Serum

LevelTotal CV (%)
971.5
1121.8

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen typeComparison System (X)NRegression EquationSyxRSample Range
Serum - ARICX7115$Y=1.00X + 1.3$1.60.98576 - 128
Serum - MSKDAX154$Y=0.98X + 3.3$2.20.91486 - 122
Plasma(y), Serum(x)ADVIA 165069$Y=0.86X+15.9$1.200.944102-118
Urine - ARICX776$Y=1.05X - 0.9$4.90.99629 - 280

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
Bilirubin2585.00.40.5
Hemoglobin50082.70.00.0
Lipids (Trig)50083.32.42.9

Analytical Range

Serum/Plasma:15 to 200 mmol/L
Urine:15 to 400 mmol/L
Serum:98 - 107 mEq/L
Urine:110 - 250 mmol/d

{7}------------------------------------------------

7. СК

SUMMARY OF SAFETY AND EFFECTIVENESS

Creatine Phosphokinase Method for the Bayer ADVIA 1650 (U/L)

Intended Use

This in vitro diagnostic assay is intended to measure creatine kinase activity in human serum and plasma on an ADVIA® 1650 Chemistry System. Such measurements are used mainly in the diagnosis and treatment of myocardial infarction and muscle diseases such as Duchenne progressive muscular dystrophy.

Imprecision

Advia 1650Technicon DAXBeckman CX7
SpecimentypeLevelTotalCV (%)Level (U/L)TotalCV (%)LevelTotalCV (%)
Serum1392.71774.81445.3
Serum4493.05902.54605.3

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen type: SiteComparisonSystem (X)NRegressionEquationSyxRSampleRange
Serum: MSKDAX151Y=0.97X+1.193.30.9996-700
Serum: ARICX7145Y=1.01X+0.86.501.00013-1187
Plasma(y), Serum(x)ADVIA 165056Y=1.05X-9.027.880.99741-507

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
Bilirubin2570.02.73.9
Hemoglobin50098.024.324.8
Lipids (Trig)50065.06.09.2

Analytical Range

0-1300 U/L

Expected Values

M: 38 - 174 U/L.

F: 26 - 140 U/L.

{8}------------------------------------------------

8. CREAT

SUMMARY OF SAFETY AND EFFECTIVENESS

Creatinine Method for the Bayer ADVIA 1650 (mg/dL)

Intended Use

This in vitro diagnostic assay is intended to measure creatinine concentration in human serum, plasma and urine on the ADVIA® 1650 system. Such measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis.

Imprecision

Advia 1650Technicon DAX=Serum
Specimen typeLevelTotal CV %LevelTotal CV (%)
Serum1.83.81.84.3
Serum8.43.75.04.5
Urine84.62.910.53.7
Urine209.34.0

Correlation (Y=ADVIA 1650, X=comparison system)

ComparisonRegressionSample
Specimen type: SiteSystem (X)NEquationSyxRRange
Serum: BERLINDAX112Y=0.92X+0.060.150.9950.33-8.3
Plasma(y), Serum(x)ADVIA 165058Y=1.02X+0.040.020.9920.9-1.4
Urine: BERLINHITACHI86Y=1.13X+0.402.610.99521-160

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effectconc.Effect%
Bilirubin251.0-0.8-80.0
Hemoglobin5007.30.11.4
Lipids (Trig)5001.00.04.5

Analytical Range

Serum/Plasma:0 - 25 mg/dL
Urine:17 - 160 mg/dL
Serum:M: 0.9 - 1.3 mg/dL
F: 0.6 - 1.1 mg/dL.
Urine:M: 14 - 26
F: 11 - 20

{9}------------------------------------------------

9. CREAT-E

SUMMARY OF SAFETY AND EFFECTIVENESS

Creatinine-Enzymatic Method for the Bayer ADVIA 1650 (mg/dL)

Intended Use

This in vitro diagnostic assay is intended to measure creatinine concentration in human serum and plasma on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis.

Imprecision

Advia 1650Technicon DAX
SpecimentypeLevelTotalCV (%)LevelTotalCV (%)
Serum0.96.41.84.3
Serum6.12.35.04.5
Serum9.52.010.53.7

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen type: SiteComparisonSystem (X)NRegressionEquationSyxRSampleRange
Serum: MSKDAX155Y=0.99X-0.060.160.9550.3-3.9
Plasma(y), Serum(x)ADVIA 165057Y=0.96X+0.160.070.9250.7-1.4

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
conc.%
Bilirubin250.7-0.2-28.6
Hemoglobin5007.8-0.1-1.3
Lipids (Trig)5000.7-1.2-171.4

Analytical Range

0-30 mg/dL

Expected Values

M: 0.9 - 1.3 mg/dL, F: 0.6 - 1.1 mg/dL.

{10}------------------------------------------------

10. DBILI

SUMMARY OF SAFETY AND EFFECTIVENESS

Direct Bilirubin Method for the Bayer ADVIA 1650 (mg/dL)

Intended Use

This in vitro diagnostic assay is intended to measure direct bilirubin concentration in human serum and plasma on the ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis of common bile duct obstruction caused by a stone and of patients with Dubin-Johnson syndrome.

Imprecision

A do ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Advia 1650Technicon DAX
SpecimentypeLevelTotalCV (%)LevelTotalCV (%)
Serum0.34.60.2**
Serum1.64.41.55.7
4.02.3

** Value not significant when level approaches zero.

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen typeComparison System (X)NRegression EquationSyxRSample Range
Serum - TRYTNDAX145Y=1.34X - 0.100.180.9940 - 6.12
Serum - BERLINHITACHI92Y=1.07X + 0.060.140.9970.02 - 9.26
Plasma (y), Serum (x)ADVIA 165049Y=0.98X-0.010.020.9880.1-0.6

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect%
Hemoglobin2000.3-0.6.
Hemoglobin5000.3-1.5.
Lipids (Trig)5000.2-0.5.

Analytical Range

0 - 10 mg/dL (from reagent stability study)

Expected Values

<0.2 mg/dL

{11}------------------------------------------------

11. GLU-HEX

SUMMARY OF SAFETY AND EFFECTIVENESS

Glucose-Hexokinase Method for the Bayer ADVIA 1650 (mg/dL)

Intended Use

This in vitro diagnostic assay is intended to measure glucose in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and insulin overdose.

Imprecision

Advia 1650Technicon DAX=Serum
SpecimenTotalTotal
typeLevelCVLevelCV (%)
Serum772.4714.2
Serum2793.32622.1
Urine423.53671.7
Urine2853.6

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen typeComparisonSystem (X)NRegression EquationSyxRSample Range
Serum - BERLINDAX109Y=1.05X - 2.85.420.99519-262
Urine - BERLINHITACHI81Y=0.99X - 11.29.880.9974-476
Plasma (y) Serum (x)ADVIA 165060Y=1.03X - 6.69.000.98636 - 384

Interfering Substances

AnalyteInterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
conc.%
GLU-HEXBilirubin1095.5-8.2-8.6
GLU-HEXBilirubin2595.5-24.4-25.5
GLU-HEXHemoglobin20095.9-22.2-23.1
GLU-HEXHemoglobin50095.9-59.9-62.5
GLU-HEXLipids (Trig)900189.02.21.2

Analytical Range

Serum/Plasma:0 - 700 mg/dL
Urine:0 - 700 mg/dL
Serum/Plasma74 -- 106 mg/dL
Urine:<0.5 g/day

{12}------------------------------------------------

12. GLU-OX

SUMMARY OF SAFETY AND EFFECTIVENESS

Glucose-Oxidase Method for the Bayer ADVIA 1650 (mg/dL)

Intended Use

This in vitro diagnostic assay is intended to measure glucose in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia and insulin overdose.

Imprecision

Advia 1650Technicon DAX=Serum
SpecimentypeLevelTotalCV (%)LevelTotalCV (%)
Serum771.5742.1
Serum2941.42931.5
Urine481.5
Urine2771.4

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen type: SiteComparison System (X)NRegression EquationSyxRSample Range
Serum: MSKDAX155Y=0.98X-3.03.40.99641-323
Plasma(y), Serum(x)ADVIA 165048Y=0.99X+2.92.210.99960-381
Urine: MSKCX387Y=1.06X-1.24.80.9890-249

Interfering Substances

AnalyteInterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effectconc.Effect%
GLU-OXBilirubin1091.9-3.5-3.8
GLU-OXBilirubin2591.9-13.6-14.8
GLU-OXHemoglobin50086.76.17.0
GLU-OXLipids (Trig)900187.0-5.8-3.1

Analytical Range

Serum/Plasma:0 - 750 mg/dL
Urine:0 - 750 mg/dL
Serum:74 - 106 mg/dL
Urine:<0.5 g/day

{13}------------------------------------------------

13. HDL

SUMMARY OF SAFETY AND EFFECTIVENESS

High Density Lipoprotein Method for the Bayer ADVIA 1650 (mg/dL)

Intended Use

This in vitro diagnostic assay is intended to measure HDL Cholesterol in human serum and plasma on an ADVIA® 1650 Chemistry System. Such measurements are used in the risk assessment of coronary artery disease.

Imprecision

Advia 1650RA-XT
Specimen typeLevelTotal CV (%)LevelBetween Day CV (%)
Serum283.815.55.1
Serum533.252.31.9

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen type: SiteComparison System (X)NRegression EquationSyxRSample Range
Serum: MSKRA-XT121$Y=1.03X+3.91$2.50.98712-94
Plasma(y), Serum(x)ADVIA 165079$Y=1.09X-1.9$4.500.96026-98

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effectconc.Effect%
Bilirubin2529.7-0.4-1.3
Hemoglobin50029.2-5.6-19.2

Analytical Range

10-135 mg/dL

Expected Values (years: mg/dL)

Male, Ages 30 - 50: 27.8 to 52.89 mg/dL Female, Ages 30 - 50: 33.9 to 86.87 mg/dL

{14}------------------------------------------------

14. IPHOS

SUMMARY OF SAFETY AND EFFECTIVENESS

Inorganic Phosphorus Method for the Bayer ADVIA 1650 (mg/dL)

Intended Use

This in vitro diagnostic assay is intended to measure inorganic phosphorus concentration in human serum, plasma and urine on the ADVIA® 1650 system. Such measurements are used in the diagnosis and treatment of kidney diseases, parathyroid gland disorders, and Vitamin D imbalance.

Imprecision

Advia 1650Technicon DAX=Serum
Specimen typeLevelTotal CV (%)LevelTotal CV (%)
Serum3.01.73.23.0
Serum5.72.66.72.3
Serum7.72.09.42.1
Urine32.01.6
Urine170.60.7

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen type: SiteComparison System (X)NRegression EquationSyxRSample Range
Serum: MSKDAX154Y=0.93X+0.200.180.9781.1-7.3
Serum: ARICX7152Y=1.01X+0.300.250.9831.4-10.0
Plasma(y), Serum(x)ADVIA 165050Y=0.98X-0.230.110.9752.0-4.1
Urine: MSKCobas Fara92Y=1.02X-0.21.30.9985-87

Interfering Substances

AnalyteInterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.conc.Effect%
IPHOSBilirubin253.40.0-0.6
IPHOSHemoglobin5002.70.933.3
IPHOSLipids (Trig)5003.4-0.2-5.9

Analytical Range

Serum/Plasma: 0 - 20 mg/dL Urine: 5 - 100 mg/dL

Serum:2.7 - 4.5 mg/dL
Urine:0.4 - 1.3 g/d

{15}------------------------------------------------

15. IRON

SUMMARY OF SAFETY AND EFFECTIVENESS

Iron Method for the Bayer ADVIA 1650 (µg/dL)

Intended Use

This in vitro diagnostic method is intended to measure iron concentration in human serum and plasma on an ADVIA® 1650 Chemistry System. Measurements are used in the diagnosis and treatment of iron deficiency anemias and hemochromatosis.

Imprecision

Advia 1650Technicon DAX
SpecimentypeLevelTotalCV (%)LevelTotalCV (%)
Serum643.1815.5
Serum1752.61943.2
Serum2952.23223.1

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen typeComparison System (X)NRegression EquationSyxRSample Range
Serum - MSKDAX40$Y=0.88X + 4.5$4.80.9976 - 253
Plasma(y), Serum(x)ADVIA 165057$Y=0.96X+15.5$6.90.97840 - 186

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
Bilirubin25111.61.9.1.7
Hemoglobin500N/AN/AN/A
Lipids (Trig)500107.9-11.1-10.3

Analytical Range

0 to 1000 µg/dL

Expected Values

Males: 65 to 175 µg/dL Females: 50 to 170 µg/dL

{16}------------------------------------------------

16. Mg

SUMMARY OF SAFETY AND EFFECTIVENESS

Magnesium Method for the Bayer ADVIA 1650 (mg/dL)

Intended Use

This in vitro diagnostic assay is intended to measure magnesium concentration in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of hypermagnesemia and monitoring of patients receiving prolonged magnesium-free intravenous therapy.

Imprecision

Advia 1650Technicon DAX= Serum
SpecimentypeLevelTotalCV (%)LevelTotalCV (%)
Serum1.92.32.45.6
Serum3.42.03.93.6
Urine8.711.9

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen typeComparison System (X)NRegression EquationSyxRSample Range
Serum - TRYTNDAX193Y=0.91X + 0.100.1310.9730.9 - 5.3
Serum - BERLINHITACHI112Y=0.95X + 0.250.1240.9161.5 - 3.1
Plasma(y), Serum(x)ADVIA 165051Y=0.97X + 0.020.0320.9631.6 - 2.1
Urine - BERLINHITACHI97Y=0.99X + 0.310.2530.9971.4 - 15.0

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effectconc.Effect%
Bilirubin252.60.414.0
Hemoglobin5002.70.27.4
Lipids (Trig)5001.81.055.6

Analytical Range

Serum/Plasma:1.6 to 6 mg/dL
Urine:1.6 to 25 mg/dL

Expected Values (values in Tietz converted to mg/dL)

Serum/Plasma:1.6 – 2.6 mg/dL
Urine:7.3 – 12.2 mg/d

{17}------------------------------------------------

17. K

SUMMARY OF SAFETY AND EFFECTIVENESS

Potassium Method for the Bayer ADVIA 1650 (mmol/L)

Intended Use

This in vitro diagnostic assay is intended to measure potassium concentration in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used mainly to monitor electrolyte balance in the diagnosis and treatment of primary aldosteronism, metabolic alkalosis, diarrhea, severe vomiting, diuretic administration, diabetic ketoacidosis, and other diseases.

Imprecision

Advia 1650Technicon DAX=Serum
Specimen typeLevelTotal CV (%)LevelTotal CV (%)
Serum3.03.03.41.6
Serum6.42.77.51.7
Urine28.56.1
Urine102.94.8

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen typeComparison System (X)NRegression EquationSyxRSample Range
Serum - MSKDAX156Y=1.09X - 0.20.080.9952.2 - 8.1
Plasma(y), Serum(x)ADVIA 165049Y=0.92X-0.010.160.8903.4-4.8
Urine - MSKCX399Y=1.01X-0.21.20.9997 - 116

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
Bilirubin253.30.00.3
Hemoglobin500N/AN/AN/A
Lipids (Trig)5003.30.26.1

Analytical Range

Serum/Plasma:1 to 10 mmol/L
Urine:3 to 300 mmol/L
Serum:3.5 – 5.1 mmol/L
Plasma Males:3.5-4.5 mmol/L
Plasma Females:3.4-4.4 mmol/L
Urine:25-125 mmol/d (varies with diet)

{18}------------------------------------------------

18. Na

SUMMARY OF SAFETY AND EFFECTIVENESS

Sodium Method for the Bayer ADVIA 1650 (mmol/L)

Intended Use

This in vitro diagnostic assay is intended to measure sodium concentration in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used mainly in the diagnosis and treatment of gross changes in water and salt balance, aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic hormone secretion, diabetic acidosis, severe diarrhea, or other diseases involving electrolyte imbalance.

Imprecision

Advia 1650Technicon DAX=Serum
SpecimentypeLevelTotal CV (%)LevelTotal CV (%)
Serum1191.71150.9
Serum1431.81410.9
Urine631.91661.2
Urine2082.0

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen typeComparison System (X)NRegression EquationSyxRSample Range
Serum - MSKDAX156$Y=1.02X - 0.2$1.080.970121 - 154
Plasma(y), Serum(x)ADVIA 165069$Y=0.92X+11.6$1.450.964139 - 156
Urine - MSKCX397$Y=0.95X + 6.2$30.99720 - 202

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
Bilirubin25122.80.00.0
Hemoglobin500128.72.31.8
Lipids (Trig)500117.33.12.6

Analytical Range

Serum/Plasma:100 to 200 mmol/L
Urine:10 to 400 mmol/L
Serum/Plasma:136 - 145 mEq/L
Urine:40 - 220 mmol/d

{19}------------------------------------------------

19. TBILI

SUMMARY OF SAFETY AND EFFECTIVENESS

Total Bilirubin Method for the Bayer ADVIA 1650 (mg/dL)

Intended Use

This in vitro diagnostic assay is intended to measure total bilirubin concentration in human serum and plasma on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of hemolytic, biliary, and liver disorders, including hepatitis and cirrhosis.

Imprecision

Advia 1650Technicon DAX
SpecimentypeLevelTotalCV (%)LevelTotalCV (%)
Serum1.07.60.7**
Serum5.04.15.12.3
Serum7.72.212.52.2

** Value not significant when level approaches zero.

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen typeComparison System (X)NRegression EquationSyxRSample Range
Serum - MSKDAX156$Y=1.10X + 0.08$0.1940.9990.1 - 29.5
Serum/PlasmaADVIA 165051$Y=0.94X + 0.16$0.0710.9540.3 - 1.4

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
Hemoglobin2000.9-0.5-61.6
Hemoglobin5000.9-1.1-122.1
Lipids (Trig)5000.90.555.6

Analytical Range

0 to 30 mg/dL

Expected Values

0.3 to 1.2 mg/dL

{20}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS

Uric Acid Method for the Bayer ADVIA 1650 (mg/dL)

Intended Use

This in vitro diagnostic assay is intended to measure uric acid concentration in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of renal failure, gout, and eclampsia.

Imprecision

Advia 1650
SpecimentypeLevelTotalCV (%)
Serum3.91.9
Serum8.61.6
Serum10.02.3
Urine12.42.3
Urine23.95.2
Technicon DAX=SerumBeckman CX7=Urine
LevelTotal CV (%)LevelTotal CV (%)
5.23.0123
8.52.7243
10.02.6

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen typeComparison System (X)NRegression EquationSyxRSample Range
Serum – MSKDAX154$Y=1.05X + 0.48$0.270.9940.2 – 18.0
Plasma(y), Serum(x)ADVIA 165061$Y=1.03X + 0.02$0.0630.9992.4 – 8.9
Urine – ARICX730$Y=1.03X - 0.5$3.060.9898 – 91

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
Bilirubin254.5-0.2-4.2
Hemoglobin5006.20.23.2
Lipids (Trig)5004.3-0.4-9.3

Analytical Range

Serum/Plasma:0 to 20 mg/dL
Urine:0 to 180 mg/dL
Males/Serum:3.5 to 7.2 mg/dL
Females/Serum:2.6 to 6.0 mg/dL
Urine:250 to 750 mg/dL

{21}------------------------------------------------

21. UN

SUMMARY OF SAFETY AND EFFECTIVENESS

Urea Nitrogen Method for the Bayer ADVIA 1650 (mg/dL)

Intended Use

This in vitro diagnostic assay is intended to measure urea nitrogen (an end product of nitrogen metabolism) concentration in human serum, plasma and urine on a ADVIA® 1650 Chemistry system. Such measurements are used in the diagnosis and treatment of kidney disease, urinary tract obstruction, and acute or chronic renal failure.

Imprecision

SpecimentypeLevelTotalCV
Serum182.4
Serum502.2
Serum861.7
Urine4614.6
Urine6182.3
Technicon DAX=SerumBeckman CX3=Urine
LevelTotal CV (%)LevelTotal CV (%)
213.54607.5
591.86207.5
1021.7

Correlation (Y=ADVIA 1650, X=comparison system)

Specimen typeComparison System (X)NRegression EquationSyxRSample Range
Serum - MSKDAX34Y=1.12X - 0.81.010.9978 - 64
Plasma(y), Serum(x)ADVIA 165054Y=1.0X - 0.20.220.9977 - 22
Urine - MSKCX379Y=1.03X + 24.834.80.99067 - 988

Interfering Substances

InterferingSubstanceInterferingSub. Conc.(mg/dL)Analyteconc.Effect
conc.%
Bilirubin2516.3-0.2-1.2
Hemoglobin50020.2-0.6-3.0
Lipids (Trig)50015.80.74.4

Analytical Range

Serum/Plasma:5 to 150 mg/dL
Urine:35 to 1000 mg/dL

Expected Values

6 to 20 mg/dL

{22}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/22/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.

JUN 30 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation Business Diagnostic Division 511 Benedict Avenue Tarrytown, New York 10591-5097

Re: K991576 Trade Name: Bayer ADVIA® 1650 Chemistry System Regulatory Class: II Product Code: CIX, CJK, KHM, CIT, CIC, CGX, JFY, CGS, CIG, CFR, CGA, LBS, CDQ Regulatory Class: I Product Code: CEO, JIY, CFO, KNK Dated: April 29, 1999 Received: May 6, 1999

Dear Mr. Muraca:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{23}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the GMice of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{24}------------------------------------------------

Attachment 8

Page 1 of 1

510(k) Number (if known): K991576

Device Name: Bayer ADVIA® 1650 Chemistry System

Indications For Use:

The ADVIA 1650 Chemistry System is an automated, clinical chemistry analyzer that can run tests on human serum, plasma, or urine in random access, batch, and STAT modes at a throughput rate of 1200 photometric tests per hour and 450 electrolyte tests per hour. The photometric analyzer performs clinical chemistry and immuno-turbidimetric methods. The electrylyte portion of the analyzer measures the sodium, and chloride concentrations in serum, plasma or urine samples based on a potentiometric procedure that uses ion-selective electrodes. The ADVIA 1650 is intended for use in conjunction with certain reagents to measure a variety of analytes contained in human fluids.

Sean Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
991576

310(k) Number

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

§ 862.1035 Albumin test system.

(a)
Identification. An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.(b)
Classification. Class II.