K Number
K983178
Device Name
COAMATIC HEPARIN
Date Cleared
1998-11-03

(53 days)

Product Code
Regulation Number
864.7525
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Coamatic® Heparin is an in vitro diagnostic test for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human citrated plasma using automated and microplate methods. The amount of UFH or LMWH is determined from the anti-FXa activity expressed by the [AT*Heparin] complex formed in plasma.

Device Description

Coamatic® Heparin is an in vitro diagnostic test for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human citrated plasma using automated and microplate methods.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Coamatic® Heparin device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as numerical targets in the provided document. Instead, the study aims to demonstrate "substantial equivalence in its performance" to the predicate device (IL Test™ Heparin). This is primarily assessed through correlation studies, where a slope close to 1, an intercept close to 0, and a high correlation coefficient (r) indicate substantial equivalence. Precision is also evaluated using %CV.

Therefore, the acceptance criteria are inferred from what constitutes "substantially equivalent performance" in such comparative studies.

Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
Comparative Studies (vs. Predicate IL Test™ Heparin)
SlopeClose to 1.00 (e.g., 0.90 to 1.10)Range: 0.91 to 1.04
InterceptClose to 0.00 (e.g., -0.10 to 0.10)Range: -0.01 to 0.06
Correlation (r)High (e.g., ≥ 0.95)Range: 0.97 to 0.99
Precision (Microplate Method)
Within Run %CV (UFH)Low (e.g., < 5%)2.8%, 3.4%
Between Run %CV (UFH)Low (e.g., < 5%)1.2%, 1.5%
Total %CV (UFH)Low (e.g., < 5%)2.8%, 3.7%
Within Run %CV (LMWH)Low (e.g., < 5%)3.6%, 2.4%
Between Run %CV (LMWH)Low (e.g., < 5%)2.8%, 2.3%
Total %CV (LMWH)Low (e.g., < 5%)4.4%, 3.2%

Note: The exact numerical acceptance criteria (e.g., specific ranges for slope, intercept, and r) are not explicitly stated but are common expectations for demonstrating substantial equivalence for such devices. The reported performance consistently falls within generally accepted ranges for good correlation and precision in clinical assays.

2. Sample Sizes Used for the Test Set and Data Provenance

The "test set" in this context refers to the samples used in the comparative studies against the predicate device and the precision studies.

  • Comparative Studies (against IL Test™ Heparin):

    • Patient Samples (UFH and LMWH treated):
      • Cobas Mira: 87 samples
      • Microplate: 70 samples
      • ACL 300: 62 samples
      • ACL Futura: 113 samples
      • MLA Electra: 80 samples
    • Normal Plasmas and Pooled Plasmas Spiked with UFH:
      • Sysmex 6000: 30 samples
      • AMAX: 30 samples
      • Hitachi 911: 30 samples
      • Hitachi 917: 30 samples
  • Precision Data (Microplate Method):

    • The sample size for the precision studies is not explicitly stated as a number of individual patient samples. The %CVs are reported at specific mean concentrations (0.7 IU/mL UFH, 0.4 IU/mL UFH, 0.7 IU/mL LMWH, 0.4 IU/mL LMWH), implying replicate measurements were performed to determine within-run, between-run, and total precision. The "n" in the comparative table refers to the number of samples compared across methods, not necessarily the number of replicates for precision.
  • Data Provenance: The document does not specify the country of origin of the data. It states samples were from "patients treated with UFH and LMWH" and "normal plasmas and pooled plasmas spiked with UHF heparin," implying these were likely clinical or laboratory samples. The study appears to be retrospective in the sense that existing samples (from patients or prepared in a lab) were used, rather than prospectively enrolling and following patients solely for this study.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This type of device (in vitro diagnostic for quantitative determination of substances) does not typically involve human expert "ground truth" establishment in the same way an image analysis AI might.

  • Number of Experts: Not applicable.
  • Qualifications of Experts: Not applicable.

For quantitative assays like this, the "ground truth" is established by a reference method or a predicate device. In this study, the predicate device, IL Test™ Heparin, run on an ACL 300, served as the de facto reference standard against which the new Coamatic® Heparin device was compared. The performance of the predicate device itself would have been validated against an even more fundamental reference method or through extensive clinical use.

4. Adjudication Method for the Test Set

Not applicable. This is a quantitative laboratory assay for which adjudication by experts is not a standard practice. The "result" is the numerical concentration reported by the instrument. Discrepancies would be resolved through re-testing, calibration checks, or investigation of assay interference, not through expert consensus.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done. These studies are typically relevant for diagnostic imaging or pathology, where human readers interpret and classify cases. This device is an in vitro diagnostic assay that produces a quantitative numerical output, not an image or case requiring human interpretation in the same way.

6. Standalone (Algorithm Only) Performance

Yes, the studies presented in the "Summary of Performance Data" are essentially standalone performance studies for the Coamatic® Heparin device. The device is a laboratory assay kit which, when run on various automated analyzers (e.g., Cobas Mira, Microplate, ACL 300, ACL Futura, MLA Electra, Sysmex 6000, AMAX, Hitachi 911, Hitachi 917), produces quantitative results. The performance metrics (slope, intercept, correlation against the predicate, and precision) represent the device's inherent capability to measure heparin directly. There is no human-in-the-loop component being evaluated in these specific performance summaries.

7. Type of Ground Truth Used

The ground truth used for the comparative studies was the measurements obtained from the predicate device, IL Test™ Heparin, particularly when run on an ACL 300. This approach is standard for demonstrating substantial equivalence for new in vitro diagnostic devices against already-cleared predicate devices. For precision, the ground truth is statistical variation around a known concentration of analyte.

8. Sample Size for the Training Set

The document does not mention a "training set" in the context of machine learning or algorithm development. This device is a biochemical assay kit, not an AI/ML algorithm that requires training data in that sense. The methods are established based on known chemical reactions and principles.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no mention of a machine learning "training set" for this device. The principles of the assay are based on established scientific understanding of Factor Xa inactivation using a synthetic chromogenic substrate.

{0}------------------------------------------------

NOV 3 1998

Section 3 Coamatic® Heparin - 510(k) SUMMARY (Summary of Safety and Effectiveness)

Submitted by:

Carol Marble Regulatory Affairs Manager Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02421 Phone: 781-861-4467 Fax: 781-861-4464

Contact Person:

Carol Marble Phone: 781-861-4467 / Fax: 781-861-4464

Summary Prepared:

September 10, 1998

Name of the device:

Coamatic® Heparin

Classification name(s):

864.7525Heparin AssayClass II
81KFFAssay, Heparin

Identification of predicate device(s):

IL Test™ Heparin K980242

Description of the device/intended use(s):

Coamatic® Heparin is an in vitro diagnostic test for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human citrated plasma using automated and microplate methods.

Statement of How the Technological Characteristics of the Device Compare to the Predicate Device:

The new Coamatic® Heparin is based on a synthetic chromogenic substrate and Factor Xa inactivation, as is the predicate device: IL Test™ Heparin, and is substantially equivalent in its performance, intended use and safety and effectiveness.

Summary of Performance Data:

The results from comparative studies of Coamatic® Heparin on different methods vs. IL Test™ Heparin on an ACL 300 using samples from patients treated with UFH and LMWH are shown below:

ApplicationnSlopeInterceptr
Cobas Mira871.04-0.010.98
Microplate701.000.020.97
ACL 300621.020.010.98
ACL Futura1130.970.010.97
MLA Electra801.010.000.97
Section 3Coamatic® Heparin 510(k)Page 1 of 2

{1}------------------------------------------------

Summary of Performance Data (Continued):

The results from additional comparative studies of Coamatic® Heparin on other methods vs. IL Test™ Heparin on an ACL 300 using normal plasmas and pooled plasmas spiked with UHF heparin are shown below:

ApplicationnSlopeInterceptr
Sysmex 6000300.910.060.99
AMAX300.970.030.99
Hitachi 911300.980.010.99
Hitachi 917301.000.020.98

Precision data summarized below was obtained with the microplate method using unfractionated heparin (UFH) and low molecular weight heparin (LMWH):

Microplate Method:

Within RunBetween RunTotal
Mean Concentration%CV%CV%CV
0.7 IU/mL UFH2.81.22.8
0.4 IU/mL UFH3.41.53.7
0.7 IU/mL LMWH3.62.84.4
0.4 IU/mL LMWH2.42.33.2

NOTE: Instrument-specific precision results are available in the application sheets.

{2}------------------------------------------------

September 8, 1998

Image /page/2/Picture/1 description: The image shows the logo for Instrumentation Laboratory. The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is a stylized cube-like structure, and the text reads "Instrumentation Laboratory" in a bold, sans-serif font.

PREMARKET NOTIFICATION TRUTHFUL AND ACCURATE STATEMENT [AS REQUIRED BY 21 CFR 807.87(j)]

I certify that, in my capacity at Instrumentation Laboratory Company, I believe to the best of my knowledge, that all data and information submitted in the premarket notification are truthful and accurate and that no material fact has been omitted.

William Wood
Signature

Signature

William Wood Name

IL Director of Regulatory Affairs/Quality Assurance Title 9 98 8

Date

K983178


Premarket Notification 510(k) Number

Coamatic® Heparin

Coamatic® Heparin 510(k)

Instrumentation Laboratory Company

Page 5 of 7 A CH-Werfen Company

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The text is in all caps and is evenly spaced around the circle.

Public Health Service

3 1998 NOV

Ms. Carol Marble Requlatory Affairs Manager Instrumentation Laboratory Company 101 Hartwell Avenue Lexington, Massachusetts 02173-3190

Re: K983178 Trade Name: Coamatic® Heparin Regulatory Class: II Product Code: KFF Dated: September 10, 1998 Received: September 11, 1998

Dear Ms. Marble:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affectinq your device can be found in the Code of Federal Requlations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

{4}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Dutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radioloqical Health

Enclosure

راني

{5}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Coamatic® Heparin

Indications for Use:

Coamatic® Heparin is an in vitro diagnostic test for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human citrated plasma using automated and microplate methods. The amount of UFH or LMWH is determined from the anti-FXa activity expressed by the [AT*Heparin] complex formed in plasma.

Heparin is the most frequently used antithrombotic drug. The biological activity of this sulphated glycosaminoglycan resides in its ability to accelerate (up to 2000-fold) the inhibitory effect of antithrombin on coagulation proteases.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

E. Make
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K983778

Prescription Use /
(Per 21 CFR 801.019)
OR Over-The-Counter Use

§ 864.7525 Heparin assay.

(a)
Identification. A heparin assay is a device used to determine the level of the anticoagulant heparin in the patient's circulation. These assays are quantitative clotting time procedures using the effect of heparin on activated coagulation factor X (Stuart factor) or procedures based on the neutralization of heparin by protamine sulfate (a protein that neutralizes heparin).(b)
Classification. Class II (performance standards).