K Number
K982482

Validate with FDA (Live)

Device Name
HANDMASTER NMS1
Date Cleared
1998-09-30

(76 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HANDMASTER NMS1 is intended to be used to exercise the lower arm and hand by activating the muscles thereof with electrical stimulation. As a powered muscle stimulator, the HANDMASTER NMS1 is intended to be used for the following indications: maintenance or increase of range of motion, reduction of muscle spasm, prevention or retardation of disuse atrophy, muscle reeducation, and increasing local blood circulation.

Device Description

The HANDMASTER NMS1 is a portable, one-channel electrical neuromuscular stimulator for personal use in exercising the upper extremity. The stimulator serves five surface electrodes held on to the upper limb by a splint. The control unit housing the stimulator may be worn using the shoulder strap provided, or it may be placed on any stable surface. The splint is worn on the hand and forearm. The splint is connected to the control unit by a light cable. A single channel of constant-voltage symmetrical biphasic Russian waveform stimulation is delivered to the muscles through five surface electrodes. Microprocessor-controlled switching of the stimulation between these five electrodes allows the muscles to be activated in combinations either cyclically or continuously. The stimulation is ramped up at the beginning and down at the end of each cycle. The electrode locations allow the HANDMASTER NMS1 to give finger and thumb extension and flexion. The user can select from five stimulation programs by pressing the mode button on the control unit. The stimulation programs are supplied as microprocessor firmware. They comprise either cyclic or continuous activation of the finger and thumb extensors and flexors. The user can increase or decrease the stimulation intensity in ten discrete levels by pressing on buttons labeled "+" or "-" on the control unit. This alters the duration of the stimulation pulse. During the initial system set-up, the clinician opens a clinical panel within the control unit. Adjustments are provided for limiting the maximum current to the extensor muscles and to the flexor muscles, along with a global timing factor which increases or decreases the duration of the stimulation cycles, effectively speeding or slowing the cyclic hand motion. The user starts or stops the stimulation program by pressing a "trigger" button. If required, the user may also stop all stimulation immediately.by switching OFF the device. The HANDMASTER NMS1 splint is used to hold the wrist joint at a comfortable extension angle (20°), and also to hold the electrodes on the forearm and hand segments. It is constructed from fiber-reinforced plastic with soft polyurethane cushion sections to distribute stress over bony regions. The electrodes are made from metal foil coated with carbon-impregnated polymer. Replaceable water-soaked cloth pads are arranged over the electrodes to provide a conductive interface with the skin. A sponge-capped bottle is provided to facilitate wetting of the electrode pads. Rechargeable nickel-cadmium batteries power the device.

AI/ML Overview

The provided document, K982482 for the HANDMASTER NMS1, describes a 510(k) premarket notification for a modified powered muscle stimulator. The purpose of this submission is to demonstrate substantial equivalence to a predicate device, not necessarily to establish new safety and effectiveness criteria through a comprehensive clinical study with acceptance criteria in the typical sense.

Here's an analysis of the provided information relative to your requested structure:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance Criteria (Implicit from Substantial Equivalence Claim)Reported Device Performance
Intended UseSame as predicate device (K952273)"The modified HANDMASTER NMS1 and its predicate device have the same intended use, which is to exercise the limb by activating the muscles with electrical stimulation."
Indications for UseSame as predicate device"The modification to the original HANDMASTER NMS1 will not alter the cleared device's intended use or indications for use."
Technological CharacteristicsSimilar to predicate device, with one specific change (carrier wave frequency) not impacting safety/effectiveness.The modified device has a higher carrier wave frequency (11,000 Hz) compared to the predicate (5,000 Hz). The submission states, "the stimulation frequency, not the carrier wave frequency, determines the effect and comfort of the stimulation."
Safety and EffectivenessEquivalent to predicate device. The change should not impact safety or effectiveness."Importantly, the modification to the HANDMASTER NMS1 could not impact safety or effectiveness because it is well recognized in the medical literature that the stimulation frequency, not the carrier wave frequency, determines the effect and comfort of the stimulation."
Performance DataOscilloscope Tracing Specification tests completed per Draft FDA Guidance."Oscilloscope Tracing Specification tests have been carried out on the modified HANDMASTER NMS1 per Section 1 of the Draft FDA Guidance entitled "Technological Reporting for Powered Muscle Stimulators". This testing measured the stimulator output for the modified HANDMASTER NMS1 at various intensity levels, output loads and operating modes." (Specific quantitative results are not provided in this summary.)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: This document does not describe a clinical test set with human subjects or a defined "sample size" in the conventional sense for performance evaluation in a clinical study. The performance data mentioned (Oscilloscope Tracing Specification tests) are technical tests performed on the device itself, likely on a limited number of units for engineering verification.
  • Data Provenance: Not applicable in the context of a clinical test set. The technical testing was performed on the modified HANDMASTER NMS1 device. The document does not specify the country of origin of the device's manufacture for these specific tests, but the submitter is based in Israel. The data would be considered prospective in relation to the submission, as it was generated to support the 510(k).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not applicable. The device performance was assessed via technical tests (Oscilloscope Tracing Specification tests), not against a "ground truth" derived from expert clinical assessment. The "ground truth" here is the expected electrical output characteristics described in the FDA guidance and device specifications.

4. Adjudication Method for the Test Set

  • This is not applicable as there was no expert-adjudicated clinical test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission is for a powered muscle stimulator, and clinical trials of this nature are generally not required for 510(k) submissions unless substantial equivalence cannot be demonstrated through other means (e.g., technical performance, material safety, and reference to existing literature).

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

  • This concept is not directly applicable. The HANDMASTER NMS1 is a physical device that delivers electrical stimulation. Its "performance" refers to its electrical output characteristics. There isn't an "algorithm only" component that would be evaluated in a standalone manner distinct from the physical device's operation. The microprocessor firmware controls the stimulation programs, but its performance is inherently tied to the device's physical output.

7. The Type of Ground Truth Used

  • For the technical "Performance Data" mentioned, the "ground truth" or reference standard would be the specifications outlined in Section 1 of the Draft FDA Guidance entitled "Technological Reporting for Powered Muscle Stimulators" and the internal design specifications of the device. This is a technical (engineering) ground truth, not a clinical one (like pathology or expert consensus).

8. The Sample Size for the Training Set

  • Not applicable. This is a medical device, not an AI/ML algorithm that is "trained" on a dataset in the typical sense. The microprocessor firmware contains pre-programmed stimulation programs, but these are designed based on established physiological principles and prior device iterations, not a "training set."

9. How the Ground Truth for the Training Set was Established

  • Not applicable, as there is no "training set" for this type of device in the context of the provided documentation. The "ground truth" for the device's operational characteristics (e.g., waveform shape, frequency, intensity levels) is based on engineering design, medical literature regarding neuromuscular electrical stimulation, and regulatory guidance for powered muscle stimulators.

{0}------------------------------------------------

K982/482

SEP 3 0 1998

510(K) SUMMARY

N.E.S.S. NEUROMUSCULAR ELECTRICAL STIMULATION SYSTEMS LTD.'S MODIFIED HANDMASTER NMS1 POWERED MUSCLE STIMULATOR

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Submitter:

N.E.S.S. Neuromuscular Electrical Stimulation Systems Ltd. 19 Ha-Haroshet Street Keidar Center Suite 207 P.O. Box 2500 Industrial Zone Ra'anana, 43465 ISRAEL

Phone: 011-972-7485738 Facsimile: 011-972-7485740

Contact Persons:

Giora Arbel N.E.S.S. Neuromuscular Electrical Stimulation Systems Ltd. 19 Ha-Haroshet Street Keidar Center Suite 207 P.O. Box 2500 Industrial Zone Ra'anana, 43465 ISRAEL

011-972-7485738 Phone: Facsimile: 011-972-7485740

Jonathan S. Kahan, Esq. Hogan & Hartson, L.L.P. 555 Thirteenth Street, N.W. Washington, D.C. 20004-1109

Phone: 202-637-5794 Facsimile: 202-637-5910

{1}------------------------------------------------

July 16, 1998 Date Prepared:

Name of Device and Name/Address of Sponsor

The HANDMASTER NMS1

N.E.S.S. Neuromuscular Electrical Stimulation Systems Ltd. 19 Ha-Haroshet Street Keidar Center Suite 207 P.O. Box 2500 Industrial Zone Ra'anana, 43465 ISRAEL

Phone: 011-972-7485738 Facsimile: 011-972-7485740

Common or Usual Name

External Neuromuscular Stimulator

Classification Name

Powered Muscle Stimulator

Predicate Devices

N.E.S.S. Neuromuscular Electrical Stimulation Systems HANDMASTER NMS1 (K952273).

Intended Use

The HANDMASTER NMS1 is intended to be used to exercise the lower arm and hand by activating the muscles thereof with electrical stimulation. As a powered muscle stimulator, the HANDMASTER NMS1 is intended to be used for the following indications: maintenance or increase of range of motion, reduction of

{2}------------------------------------------------

muscle spasm, prevention or retardation of disuse atrophy, muscle reeducation, and increasing local blood circulation.

Technological Characteristics and Substantial Equivalence

The HANDMASTER NMS1 is a portable, one-channel electrical neuromuscular stimulator for personal use in exercising the upper extremity. The stimulator serves five surface electrodes held on to the upper limb by a splint. The control unit housing the stimulator may be worn using the shoulder strap provided, or it may be placed on any stable surface. The splint is worn on the hand and forearm. The splint is connected to the control unit by a light cable.

The HANDMASTER NMS1 is used for exercising the hand, conditioning selected muscles of the forearm and hand. It is intended for use by patients suffering from upper limb paralysis due to injury or disease of the central nervous system such as cervical spinal cord injuries or stroke.

A single channel of constant-voltage symmetrical biphasic Russian waveform stimulation is delivered to the muscles through five surface electrodes. Microprocessor-controlled switching of the stimulation between these five electrodes allows the muscles to be activated in combinations either cyclically or continuously. The stimulation is ramped up at the beginning and down at the end of each cycle.

The electrode locations allow the HANDMASTER NMS1 to give finger and thumb extension and flexion. The user can select from five stimulation programs by pressing the mode button on the control unit. The active mode is displayed by a light glowing next to the labeled mode. When the device is

{3}------------------------------------------------

The stimulation programs are supplied as stimulating, the light flashes. microprocessor firmware. They comprise either cyclic or continuous activation of the finger and thumb extensors and flexors.

The user can increase or decrease the stimulation intensity in ten discrete levels by pressing on buttons labeled "+" or "-" on the control unit. This alters the duration of the stimulation pulse. The intensity is displayed as a number (0 to 9) on a seven-segment display.

During the initial system set-up, the clinician opens a clinical panel within the control unit. Adjustments are provided for limiting the maximum current to the extensor muscles and to the flexor muscles, along with a global timing factor which increases or decreases the duration of the stimulation cycles, effectively speeding or slowing the cyclic hand motion.

The user starts or stops the stimulation program by pressing a "trigger" button. If required, the user may also stop all stimulation immediately.by switching OFF the device.

The HANDMASTER NMS1 splint is used to hold the wrist joint at a comfortable extension angle (20°), and also to hold the electrodes on the forearm and hand segments. It is constructed from fiber-reinforced plastic with soft polyurethane cushion sections to distribute stress over bony regions. The electrodes are made from metal foil coated with carbon-impregnated polymer. Replaceable water-soaked cloth pads are arranged over the electrodes to provide a

4

{4}------------------------------------------------

conductive interface with the skin. A sponge-capped bottle is provided to facilitate wetting of the electrode pads.

Rechargeable nickel-cadmium batteries power the device. Battery status can be displayed both during device operation and while recharging the batteries. Both visual and audio battery-low warnings are provided. It is necessary to disconnect splint/electrodes in order to recharge the batteries, as the same socket is used for both.

The modified HANDMASTER NMS1 is substantially equivalent to the original HANDMASTER NMS1, a marketed powered muscle stimulator. The modified HANDMASTER NMS1 and its predicate device have the same intended use, which is to exercise the limb by activating the muscles with electrical stimulation. Like the modified HANDMASTER NMS1, the predicate device is capable of delivering electrical stimulation to surface electrodes positioned on the skin surface of the limb. In addition, both the modified HANDMASTER NMS1 and the predicate device are operated either in flexion/extension cyclic mode or in a continuous mode.

The modification to the original HANDMASTER NMS1 will not alter the cleared device's intended use or indications for use. The sole change to be made is that the modified device would produce a higher frequency carrier wave than the original HANDMASTER NMS1. The currently cleared HANDMASTER NMS1 has a carrier wave frequency of 5,000 Hz, and the proposed modification would increase the carrier wave frequency to 11,000 Hz. No other changes will be made to the

5

{5}------------------------------------------------

amplitude or shape of the waveform, or any other technological characteristics. Importantly, the modification to the HANDMASTER NMS1 could not impact safety or effectiveness because it is well recognized in the medical literature that the stimulation frequency, not the carrier wave frequency, determines the effect and comfort of the stimulation. Hence, according to FDA's framework for substantial equivalence, the 11,000 Hz device is substantially equivalent to the 5,000 Hz device.

Performance Data

Oscilloscope Tracing Specification tests have been carried out on the modified HANDMASTER NMS1 per Section 1 of the Draft FDA Guidance entitled "Technological Reporting for Powered Muscle Stimulators". This testing measured the stimulator output for the modified HANDMASTER NMS1 at various intensity levels, output loads and operating modes.

రు

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an eagle with three human profiles incorporated into its design. The eagle's head and neck are formed by the profiles of three people facing to the right.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 3 0 1998

Neuromuscular Electrical Stimulation System (N.E.S.S.) Ltd. c/o Mr. Jonathan S. Kahan Partner Hogan & Hartson 555 Thirteenth Street NW Columbia Square Washington, DC 20004-1109

Re: K982482 Trade Name: Handmaster NMS1 Regulatory Class: II Product Code: IPF Dated: July 16, 1998 Received: July 16, 1998

Dear Mr. Kahan:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{7}------------------------------------------------

Page 2 - Mr. Jonathan S. Kahan

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and

Radiological Health

Enclosure

{8}------------------------------------------------

Pevice Name: HANDMASTER NMS1

Prescr

Indications For Use: The HANDMASTER is intended to be used for the following indications: maintenance or increase of range of motion, reduction of muscle spasm, prevention or retardation of disuse atrophy, muscle reeducation, and increasing local blood circulation.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General Restorative Devices
510(k) NumberK982482

Prescription Use

OR
Over-The-Counter Use

:

(Optional Formst 1-2-96)

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).