K Number
K981815
Date Cleared
1998-07-30

(69 days)

Product Code
Regulation Number
864.5425
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HEPtrac Electronic Quality Control, manufactured by Medtronic Blood Management, is an interactive mechanical and software controlled verification cartridge for use with instruments that utilize the Hepcon Hemostasis Management cartridges. The purpose of the HEPtrac is to provide a means for performing quality control of the instruments in the clinical testing environment.

Device Description

The HEPtrac is an electronic control that has been developed to assist users of the Hepcon Hemostasis Management (HMS) Instrument (K894317) and associated cartridges and controls (K894317), to perform control testing as required under 42CFR493.1253. Sec. ( c ) "for all automated coagulation testing systems, the laboratory must include two levels control each eight hours of operation and each time a change in reagents occurs". The HEPtrac, electronic quality control is designed to minimize the use of liquid based controls by providing the operator a means to determine the operating condition of the instrument.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the HEPtrac Electronic Quality Control device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly define quantitative acceptance criteria in the way one might expect for a modern AI/ML device (e.g., target specificity, sensitivity, or AUC). Instead, the performance is demonstrated by its ability to perform specific checks on the HMS instrument. The criteria are implicitly tied to the proper functioning of these checks.

Acceptance Criteria (Implicit)Reported Device Performance
Ability to check Flag release force"Performance testing was done to verify the HEPtrac's ability to check the condition of the HMS instrument for the following functions: 1. Flag release force" (The text indicates this test was successfully performed and verified the device's ability to check this function, implying it meets the implicit criterion).
Ability to check Reagent delivery"Performance testing was done to verify the HEPtrac's ability to check the condition of the HMS instrument for the following functions: 2. Reagent delivery" (The text indicates this test was successfully performed and verified the device's ability to check this function, implying it meets the implicit criterion).
Ability to check Flag lift wire height"Performance testing was done to verify the HEPtrac's ability to check the condition of the HMS instrument for the following functions: 3. Flag lift wire height" (The text indicates this test was successfully performed and verified the device's ability to check this function, implying it meets the implicit criterion).
Ability to check Flag sensors"Performance testing was done to verify the HEPtrac's ability to check the condition of the HMS instrument for the following functions: 4. Flag sensors" (The text indicates this test was successfully performed and verified the device's ability to check this function, implying it meets the implicit criterion).
Substantial equivalence to the predicate device (ACTtrac) in:The HEPtrac "is very similar to the ACTtrac electronic control for the ACT instrument in the following areas; intended use, operation, and function." The key difference is the number of channels (2 for ACTtrac vs. 6 for HEPtrac) and the addition of a "Flag release test" in HEPtrac. This implies that the device's performance is acceptable because it maintains the core functionalities of the predicate while enhancing capacity.
- Control Test TypeBoth are "Electronic" (Table 1.1)
- Reagent delivery checkBoth have "yes" (Table 1.1)
- Flag Sensor checkBoth have "yes" (Table 1.1)
- Clot detectionBoth have "yes" (Table 1.1)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify a numerical sample size for the test set. It broadly states, "Laboratory testing was done to confirm the devices performance in relation to the predicate device."
  • Data Provenance: The testing was "Laboratory testing," implying it was conducted in a controlled environment. There is no information about the country of origin of the data or whether it was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. Given the nature of the device (an electronic quality control for an instrument), the "ground truth" would likely be defined by engineering specifications and the instrument's known functional parameters rather than expert medical interpretation.

4. Adjudication Method for the Test Set

This information is not provided and is not applicable given the device's function. The testing described focuses on functional verification against engineering parameters rather than subjective interpretations requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This device is an electronic quality control for a medical instrument, not an AI/ML diagnostic or assistive tool for human readers. Therefore, this type of study is not relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

The device itself is an "electronic control," which operates without human "interpretation" of results in the traditional sense of a diagnostic algorithm. Its function is to verify the performance of the main instrument. The performance testing "to verify the HEPtrac's ability to check the condition of the HMS instrument" can be considered a standalone performance evaluation of the control device in its intended function. There is no "human-in-the-loop" aspect to the HEPtrac's function itself; it's a verification tool.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth used for this device's performance evaluation would be the engineering specifications and known functional parameters of the Hepcon Hemostasis Management (HMS) Instrument. The HEPtrac is designed to test if the HMS instrument is operating within its defined parameters for flag release force, reagent delivery, flag lift wire height, and flag sensors.

8. The Sample Size for the Training Set

This information is not provided and is not applicable in the context of this device. The HEPtrac is described as an "electronic control" and a "mechanical and software controlled verification cartridge." It does not appear to utilize machine learning or AI that would require a "training set" in the conventional sense. Its function is based on programmed logic and mechanical interactions to test an existing instrument.

9. How the Ground Truth for the Training Set was Established

As there is no indication of a "training set" for AI/ML, this question is not applicable. The device's "knowledge" of what constitutes correct instrument operation is built into its design and programming, based on the specifications of the HMS instrument it is designed to test.

{0}------------------------------------------------

K98181815

JUL 30 1998

510(k) SUMMARY

Submitted by:Medtronic Blood Management18501 E. Plaza DriveParker, Colorado 80134
Manufacturing Facility:Medtronic Blood Management18501 E. Plaza DriveParker, Colorado 80134
Submitted Device:Trade Name:Common Name:HEPtrac Electronic Quality ControlElectronic Quality Control

Device Classification: Class II

21 CFR § 864.5425 "A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays." The HEPtrac is a Electronic Control for such a device.

  • Product Description: The HEPtrac is an electronic control that has been developed to assist users of the Hepcon Hemostasis Management (HMS) Instrument (K894317) and associated cartridges and controls (K894317), to perform control testing as required under 42CFR493.1253. Sec. ( c ) "for all automated coagulation testing systems, the laboratory must include two levels control each eight hours of operation and each time a change in reagents occurs". The HEPtrac, electronic quality control is designed to minimize the use of liquid based controls by providing the operator a means to determine the operating condition of the instrument.
  • Intended Use: The HEPtrac Electronic Quality Control, manufactured by Medtronic Blood Management, is an interactive mechanical and software controlled verification cartridge for use with instruments that utilize the Hepcon Hemostasis Management cartridges. The purpose of the HEPtrac is to provide a means for performing quality control of the instruments in the clinical testing environment.

{1}------------------------------------------------

Performance Standards:

Per section 514 of the Food, Drug and Cosmetic Act, there are no specific performance standards for this device.

Statement of substantial equivilance-

The HEPtrac electonic control for the HMS instrument is very similar to the ACTtrac electronic control for the ACT instrument in the following areas; intended use, operation, and function. The HEPtrac and the ACTtrac are both used to asses the operating condition of automated coagulation timers. The HEPtrac has been designed to function with a six channel instrument rather than the predicate device that was design to function in a two channel device.

A summary of the essential features between the ACTtrac (predicate device) and HEPtrac is contained in Table 1.1

Table 1.1 Comparison between the ACTtrac and HEPtrac Electronic Quality Control tests

Predicate device
ACTtracHEPtrac
Control Test TypeElectronicElectronic
Number of channels26
Reagent delivery checkyesyes
Flag Sensor checkyesyes
Clot detectionyesyes
Flag release testnoyes

From Table 1.1 it can be seen that the two types of devices share the same basic features for testing automated coagulation timers. The primary difference between these two devices lies in the model number of coagulation timer it is designed to test.

Testing:

Laboratory testing was done to confirm the devices performance in relation to the predicate device.

Performance testing was done to verify the HEPtrac's ability to check the condition of the HMS instrument for the following functions:

    1. Flag release force
    1. Reagent delivery
    1. Flag lift wire height
    1. Flag sensors

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features the department's name arranged in a circular pattern around a stylized eagle emblem. The eagle is depicted with three human profiles incorporated into its design, symbolizing the department's focus on people's health and well-being.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUL 30 1998

Karl Steineck . Regulatory Affairs Specialist Medtronic Blood Management 18501 E. Plaza Drive Parker, Colorado 80134-9061

Re : K981815 HEPtrac Requlatory Class: II Product Code: JPA Dated: May 20, 1998 Received: May 22, 1998

Dear Mr. Steineck:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If vour device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ಗ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Aitman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INTENDED USE STATEMENT

510(k) Number (if known): K981815

Device Name: HEPtrac Electronic Quality Control

Indications For Use:

The HEPtrac Electronic Quality Control, manufactured by Medtronic Blood Management, is an interactive mechanical and software controlled verification cartridge for use with instruments that utilize the Hepcon Hemostasis Management cartridges. The purpose of the HEPtrac is to provide a means for performing quality control of the instruments in the clinical testing environment.

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number. K981815
Prescriptive Use ✓
OR
Over-the-Counter Use _

§ 864.5425 Multipurpose system for in vitro coagulation studies.

(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.