K Number
K974080
Device Name
CRITIKON SOFT BLOOD PRESSURE CUFF
Date Cleared
1998-05-18

(201 days)

Product Code
Regulation Number
870.1120
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The CRITIKON Soft Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The device is non-sterile and semidisposable (may be single-patient use or optional limited reuse). It is available in neonatal, pediatric and adult sizes. The device is not designed, sold, or intended for use except as indicated.
Device Description
The device comprises tubing attached to a soft inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a noninvasive blood pressure measurement system. The device is available in the following configurations: - five neonatal and eight pediatric and adult sizes t - single and dual cuff tubing, connectors, adapters and bulb/valve . assemblies available for use with a variety of manual and automated sphygmomanometers
More Information

No
The description focuses on the mechanical components and intended use of a blood pressure cuff, with no mention of AI or ML.

No
This device is an accessory (blood pressure cuff) used for measurement, not for treatment of a disease or condition. It is a diagnostic tool, not a therapeutic one.

No
The CRITIKON Soft Blood Pressure Cuff is an accessory used for noninvasive blood pressure measurement; it does not diagnose a disease or condition, but rather assists in collecting a physiological parameter.

No

The device description clearly outlines physical components like tubing, a sleeve, an inflatable bladder, and hook and loop closure, indicating it is a hardware device.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to be an accessory for noninvasive blood pressure measurement systems. This is a physiological measurement taken directly from the patient's body, not a test performed on a sample of biological material in vitro (outside the body).
  • Device Description: The description details a physical cuff that wraps around a limb and connects to a blood pressure system. There is no mention of analyzing biological samples.
  • Lack of IVD Characteristics: The document does not mention any of the typical characteristics of an IVD, such as:
    • Analysis of biological samples (blood, urine, tissue, etc.)
    • Reagents or assays
    • Detection of specific analytes or markers
    • Diagnosis or screening based on laboratory test results

The device is clearly designed for a direct physiological measurement, which falls outside the scope of In Vitro Diagnostics.

N/A

Intended Use / Indications for Use

The CRITIKON Soft Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The device is non-sterile and semidisposable (may be single-patient use or optional limited reuse). It is available in neonatal, pediatric and adult sizes. The device is not designed, sold, or intended for use except as indicated.

Product codes

DXQ

Device Description

The device comprises tubing attached to a soft inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a noninvasive blood pressure measurement system.
The device is available in the following configurations:

  • five neonatal and eight pediatric and adult sizes t
  • single and dual cuff tubing, connectors, adapters and bulb/valve . assemblies available for use with a variety of manual and automated sphygmomanometers

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

patient's limb

Indicated Patient Age Range

neonatal, pediatric and adult

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the CRITIKON Soft Blood Pressure Cuff: Biocompatibility Validation of reprocessing Repeated inflations

Key Metrics

Not Found

Predicate Device(s)

K790810, K883977, K911213, K921997, K940214

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.1120 Blood pressure cuff.

(a)
Identification. A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.(b)
Classification. Class II (performance standards).

0

1974080

MAY 1 8 1998

510(k) Summary
-----------------------
DateOctober 27, 1997
ContactDarlene T. Korab
Sr. Project Manager
Regulatory Affairs
Johnson & Johnson Medical, Inc.
4110 George Road
Tampa, FL 33634
Telephone: 813-887-2663
Telefax: 813-887-2263
E Mail: dkorab@crtus.jnj.com
Device NameCRITIKON* Soft Blood Pressure Cuff
The subject of this premarket notification is a modification of the currently
marketed Johnson & Johnson Medical, Inc. CRITIKON Soft Blood Pressure
Cuff (formerly DISPOSA-CUF* Soft Blood Pressure Cuff).
Common NameBlood Pressure Cuff
ClassificationThe classification name, 21 Code of Federal Regulations (CFR) Part and
Paragraph number, product code and classification of the CRITIKON Soft
Blood Pressure Cuff follow. The tier categorization based on the list (January
27, 1994) distributed by the Office of Device Evaluation is also included.
Classification Name21 CFR SectionProduct CodeClassTier
Blood Pressure Cuff870.1120DXQII2

Continued on next page

:

.

(12.05

1

510(k) Summary, Continued

Predicate Devices

The CRITIKON Soft Blood Pressure Cuff is substantially equivalent to the CRITIKON (formerly DISPOSA-CUF) Soft Blood Pressure Cuff currently marketed by Johnson & Johnson Medical, Inc. The original DISPOSA-CUF Blood Pressure Cuff received market clearance on May 25, 1979 via Crest Medical Equipment 510(k) K790810. The device is virtually the same as the CRITIKON or DISPOSA-CUF Soft Blood Pressure Cuff . The CRITIKON Soft Blood Pressure Cuff is also substantially equivalent to the CUFF-ABLE* Blood Pressure Cuff currently marketed by Vital Signs, Inc. [Biomedical Dynamics 510(k) K911213] and other soft blood pressure cuffs.

DeviceManufacturer510(k)
CRITIKON (formerly
DISPOSA-CUF) Soft
Blood Pressure CuffJohnson & Johnson
Medical, Inc.K790810
SURGI-CUF* CuffEthox Corp.K883977
CUFF-ABLE CuffVital Signs, Inc.K911213
Cuff-m* CuffLorin MedicalK921997
Soft Check* CuffStatcorp, Inc.K940214
Safe-Cuff* CuffCAS Medical SystemsUnknown

Device Description

The device comprises tubing attached to a soft inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a noninvasive blood pressure measurement system.

The device is available in the following configurations:

  • five neonatal and eight pediatric and adult sizes t
  • single and dual cuff tubing, connectors, adapters and bulb/valve . assemblies available for use with a variety of manual and automated sphygmomanometers

Continued on next page

2

510(k) Summary, Continued

(Individual of the state of the many of the same of

: 2018-08-08 11:2


| Indications | The CRITIKON Soft Blood Pressure Cuff is an accessory used in conjunction
with non-invasive blood pressure measurement systems. The device is non-
sterile and semi-disposable (may be single-patient use or optional limited
reuse). It is available in neonatal, pediatric and adult sizes. The device is not
designed, sold, or intended for use except as indicated. |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological
Characteristics | The CRITIKON Soft Blood Pressure Cuff is virtually the same as the
CRITIKON or DISPOSA-CUF Soft Blood Pressure Cuff with the exception
of a modification of the sleeve material to afford greater patient comfort. The
CRITIKON Soft Blood Pressure Cuff is also virtually the same as the CUFF-
ABLE Blood Pressure Cuff and other soft blood pressure cuffs. All the above
devices are available in range of neonatal, pediatric and adult sizes and
configurations for use with a wide variety of manual and automated
sphygmomanometers. |
| Performance | Bench and laboratory testing was conducted to demonstrate performance
(safety and effectiveness) of the CRITIKON Soft Blood Pressure Cuff:
Biocompatibility
Validation of reprocessing
Repeated inflations |
| Conclusions | In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR
Part 807, and based on the information provided in this premarket
notification, Johnson & Johnson Medical concludes that the modified device,
the CRITIKON Soft Blood Pressure Cuff, is safe, effective and substantially
equivalent to the predicate devices as described herein. |
| Other
Information | Johnson & Johnson Medical, Inc. will update and include in this summary
any other information deemed reasonably necessary by the FDA. |

3

Johnson Johnson
MEDICAL INC.

October 6, 1997

Critikon, a division of Johnson & Johnson Medical, Inc., certifies that the instructions for reprocessing for CRITIKON™ Soft Blood Pressure Cuffs as available in the Instructions for Use (P/N 714338) have been validated according to a formal protocol (052496), comprising sound scientific principles and laboratory procedure. The protocol is conceptually based on AAMI TIR No. 12-1994, "Designing, Testing, and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers." A summary of the validation procedure is attached. The protocol and subsequent laboratory report are archived in Critikon Technical File 103-1 under change control. They will be made available for inspection or supplied to FDA upon request.

John P. Clemmons Critikon R&D

4

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Public Health Service

MAY | 8 1998

Ms. · Darlene T. Korab Johnson & Johnson Medical, Inc. 4110 George Road Tampa, FL 33634

Re : K974080 CRITIKON® Soft Blood Pressure Cuff Regulatory Class: II (two) Product Code: 74 DXQ February 13, 1998 Dated: Received: February 17, 1998

Dear Ms. Korab ':

We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in In addition, FDA may publish further announcements regulatory action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

5

Page 2 - Ms. Darlene T. Korab

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Thomas A. Gelesh

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

6

INDICATIONS FOR USE

Page 1 of 1

510(k) Number (if known):

Device Name: CRITIKON* Soft Blood Pressure Cuff

Indications for Use:

The CRITIKON Soft Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The device is non-sterile and semidisposable (may be single-patient use or optional limited reuse). It is available in neonatal, pediatric and adult sizes. The device is not designed, sold, or intended for use except as indicated.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

Stepe R. Campbell
(Division Sign-Off)

Division of Cardiovascular, Respiratory, and Neurological Devices

¥ 974080 510(k) Number ________________________________________________________________________________________________________________________________________________________________

Prescription Use ﺎ، (per 21 CFR 801.109

OR

Over-The Counter Use (Optional Format 1-2-96)

iv