K Number
K973819
Date Cleared
1998-03-10

(154 days)

Product Code
Regulation Number
864.7320
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VIDAS D-Dimer (DD) Assay is for the quantitative detection of fibrin degradation products (FbDP) in human plasma. It is intended to aid in the diagnosis of deep venous thrombosis and pulmonary embolism disease.

The VIDAS D-Dimer (DD) assay is intended for use with a VIDAS (Vitek ImmunoDiagnostic Assay System) instrument as an automated quantitative enzyme-linked fluorescent immunoassay (ELFA) for the determination of fibrin degradation products (FbDP) in plasma (trisodium citrate). The VIDAS D-Dimer assay is intended for use as an aid in the diagnosis of deep venous thrombosis and pulmonary embolism disease.

Device Description

The VIDAS D-Dimer assay is an enzyme-linked fluorescent immunoassay (ELFA) performed on an automated VIDAS instrument. All assay steps and assay temperature are controlled by the instrument.

A pipette tip-like disposable device known as the Solid Phase Receptacle (SPR), serves as a solid phase for the assay as well as a pipetting device. Reagents for the assay are located in the sealed VIDAS DD Reagent Strips.

The VIDAS D-Dimer kit contains 60 SPRs, 60 Reagent Strips, 2 Bottles of Calibrator, 2 Bottles each level of Positive Control (three levels) and 1 bottle of Diluent. The Kit contains a sufficient number of SPR's and Strips to perform 60 Tests.

The SPR is coated with mouse anti-FbDP antibodies. The Strip contains the reagents necessary to perform the assay, as well as a sample well for placement of the specimen. Each DD test requires one DD Reagent Strip and one DD SPR.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the VIDAS D-Dimer (DD) Assay, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance Criteria CategorySpecific Criteria/MetricReported Device Performance
CorrelationComparison with predicate device (American Bioproducts Asserachrom D-Di Kit)Line equation: y = 1.06x - 50.4. Correlation coefficient: 0.90.
Sensitivity (Analytical)Lowest measurable level of FbDP (limit of detection) at 95% probability45 ng/ml. The assay is designed to measure FbDP between 45 ng/ml and 1000 ng/ml.
SpecificityCross-reactivity with fibrinogen and its degradation productsTwo monoclonal antibodies used in combination recognized D-Dimer and D-Dimer containing fibrin derivatives, and did not cross-react with fibrinogen or its degradation products. (Individual antibodies showed weak cross-reaction with fibrinogen degradation products, but this was overcome in combination).
Interfering SubstancesInterference from hemolyzed, icteric, or lipemic specimensNo assay interference was demonstrated when testing spiked hemolyzed, icteric, or lipemic specimens. (However, use of such samples is not recommended).
Precision/ReproducibilityIntra-assay precision (Coefficients of Variation - CV)Ranged from 3.8% to 5.8% over the reportable range of the assay.
Inter-assay reproducibility (CV over 8 weeks)Did not exceed 7.6%.
Inter-assay, inter-instrument reproducibility (CV for 4 samples across 8 runs on 8 instruments)Did not exceed 4.9%.
Calibrator ValidityMaster curve validity for the kit's shelf lifeThe calibrator in the kit ensures the master curve stored by the VIDAS instrument is valid. The body of data supports the use of a single calibrator for this purpose.

Study Details

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document does not explicitly state the sample size used for the test set. It mentions "body of data" for calibrator validity and "four different serum samples in 8 runs on 8 different instruments" for inter-instrument reproducibility, but a comprehensive sample size for all performance metrics is not provided.
    • Data Provenance: Not explicitly stated. The document describes laboratory performance characteristics, which would typically involve controlled laboratory settings. There is no information regarding country of origin or whether the data was retrospective or prospective.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. The document describes analytical performance characteristics of an immunoassay, not a diagnostic interpretation that would typically require expert ground truth establishment. The "ground truth" here is the actual concentration of FbDP, as measured by reference methods or spiked samples.
  3. Adjudication method for the test set:

    • Not applicable/Not mentioned. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation or clinical endpoint assessments, which are not the focus of this analytical performance study.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is an analytical performance study for an automated immunoassay. It does not involve human readers interpreting results, nor does it involve AI assistance for human interpretation.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes. This study exclusively details the performance of the VIDAS D-Dimer assay performed on an automated VIDAS instrument. The device itself is an automated system providing quantitative results without human interpretation of the primary measurement.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for the analytical performance studies appears to be based on:
      • Reference measurements/Predicate device comparison: For correlation, the predicate device (American Bioproducts Asserachrom D-Di Kit) served as a reference.
      • Known concentrations: For sensitivity, specificity, and interfering substances, components were likely tested at known, purified, or spiked concentrations (e.g., purified fibrinogen degradation products, spiked hemolyzed specimens).
      • Statistical analysis of repeated measurements: For precision and reproducibility.
    • It is not based on expert consensus, pathology, or outcomes data, as these are typically relevant for clinical diagnostic studies, not analytical assay validation.
  7. The sample size for the training set:

    • Not applicable/Not mentioned. This is an immunoassay, not a machine learning algorithm that typically requires a distinct "training set." The assay development and calibration would involve its own internal validation, but it's not described in terms of a "training set" in the context of AI/ML.
  8. How the ground truth for the training set was established:

    • Not applicable/Not mentioned. As this is not an AI/ML algorithm requiring a training set with established ground truth in the conventional sense, this information is not relevant to the provided text.

{0}------------------------------------------------

1973819

MAR 1 0 1998

I.1. 510(k) SUMMARY

Submitter:Terry McGovernbioMérieux Vitek, Inc.1022 Hingham St.Rockland, MA 02370(617) 871-4442
Date:September , 1997
Device:VIDAS D DIMER (DD Assay)
ClassificationName:21CFR 864.7320. Fibrinogen and fibrin split products, antigen,antiserum, control (81DAP)
Common orUsual Name:Enzyme-linked Fluorescent Immunoassay (ELFA), D-Dimer (DD)
Predicate Device:American Bioproducts Asserachrom D-Di Kit (K862156)
Intended Use:The VIDAS D-Dimer (DD) Assay is for the quantitative detection offibrin degradation products (FbDP) in human plasma. It is intended toaid in the diagnosis of deep venous thrombosis and pulmonaryembolism disease

Device Description:

The VIDAS D-Dimer assay is an enzyme-linked fluorescent immunoassay (ELFA) performed on an automated VIDAS instrument. All assay steps and assay temperature are controlled by the instrument.

A pipette tip-like disposable device known as the Solid Phase Receptacle (SPR), serves as a solid phase for the assay as well as a pipetting device. Reagents for the assay are located in the sealed VIDAS DD Reagent Strips.

The VIDAS D-Dimer kit contains 60 SPRs, 60 Reagent Strips, 2 Bottles of Calibrator, 2 Bottles each level of Positive Control (three levels) and 1 bottle of Diluent. The Kit contains a sufficient number of SPR's and Strips to perform 60 Tests.

The SPR is coated with mouse anti-FbDP antibodies. The Strip contains the reagents necessary to perform the assay, as well as a sample well for placement of the specimen. Each DD test requires one DD Reagent Strip and one DD SPR.

{1}------------------------------------------------

510(k) SUMMARY (CONT.) SYNOPSIS OF PERFORMANCE TESTING

    1. Correlation: Comparison of the VIDAS D-Dimer assay with the American Bioproducts Asserachrom DDi kit yields a line with the equation y = 1.06x -50.4 and a correlation coefficient of 0.90.
    1. The calibrator in the kit ensures that the master curve stored by the VIDAS instrument is valid for the shelf life of that kit. The body of data supports the use of a single calibrator for this purpose.
    1. Sensitivity (analytical): The VIDAS D-Dimer assay is designed to measure fibrin degradation products (FbDP) between 45 ng/ml and 1000 ng/ml. The lowest measurable level of FbDP (limit of detection) that can be distinguished from zero with 95 % probability is 45 ng/ml.
    1. Specificity: The specificity of the two monoclonal antibodies used in the VIDAS D-Dimer assay was demonstrated by testing the antibodies against different concentrations of purified fibrinogen degradation products (X, Y and D) and fibrin degradation products (D-Dimer). Each antibody was tested separately and showed a weak cross reaction with fibrinogen degradation products. Testing the antibodies in combination demonstrated that the antibodies recognized D-Dimer and D-Dimer containing fibrin derivatives. However, they do not cross react with fibrinogen or its degradation products.
    1. Interfering Substances: No assay interference was demonstrated when testing the VIDAS D-Dimer assay using spiked hemolyzed, icteric or lipemic specimens. Although interference linked to the presence of hemoglobin, bilirubin or lipids has not been observed at the concentrations tested, using hemolyzed, icteric or lipemic samples is not recommended when using the VIDAS D-Dimer assay.
    1. Precision/Reproducibility:
    • a. Intra-assay precision studies showed coefficients of variation ranging from 3.8 % to 5.8 % over the reportable range of the assay.
    • Inter-assay reproducibility over an eight-week time period yields coefficients of ﻓﻠ variation that do not exceed 7.6 %.
    • c. Inter-assay, inter-instrument reproducibility for four different serum samples in 8 runs on 8 different instruments yields coefficients of variation that do not exceed 4.9 %.

When the VIDAS D-Dimer assay is used as instructed in the package insert, the above statements are true. The package insert should always be consulted along with the Operator's Manual to ensure that the assay is being performed properly. For additional information, references are listed in the package insert.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a single snake entwined around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAR 1 0 998

Terry McGovern · Manager, Quality Assurance/Requlatory Affairs bioMerieux Vitek, Inc. 1022 Hingham Street Rockland, Massachusetts 02370

Re : K973819 VIDAS D-Dimer (DD) Assay Requlatory Class: II Product Code: DAP . . . . . . Dated: January 15, 1998 Received: January 20, 1998

Dear Mr. McGovern:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostie devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Litman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):K973819
------------------------------------

Device Name: VIDAS D-Dimer Assay (DD)

Indications for Use:

The VIDAS D-Dimer (DD) assay is intended for use with a VIDAS (Vitek ImmunoDiagnostic Assay System) instrument as an automated quantitative enzyme-linked fluorescent immunoassay (ELFA) for the determination of fibrin degradation products (FbDP) in plasma (trisodium citrate). The VIDAS D-Dimer assay is intended for use as an aid in the diagnosis of deep venous thrombosis and pulmonary embolism disease.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) NumberK184973819

Concurrence of CDRH, Office of Device Evaluation (ODE) ...............................................................................................................

Prescription Use

or

Over-the-Counter Use__________________________________________________________________________________________________________________________________________________________

§ 864.7320 Fibrinogen/fibrin degradation products assay.

(a)
Identification. A fibrinogen/fibrin degradation products assay is a device used to detect and measure fibrinogen degradation products and fibrin degradation products (protein fragments produced by the enzymatic action of plasmin on fibrinogen and fibrin) as an aid in detecting the presence and degree of intravascular coagulation and fibrinolysis (the dissolution of the fibrin in a blood clot) and in monitoring therapy for disseminated intravascular coagulation (nonlocalized clotting in the blood vessels).(b)
Classification. Class II (performance standards).