K Number
K973168
Manufacturer
Date Cleared
1997-11-21

(88 days)

Product Code
Regulation Number
892.1750
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SmartView Option for HiSpeed CT/i is indicated for head and whole body x-ray computed tomography applications. This option is a modifications which can be added to the existing family of HiSpeed CT/i Systems. The option provides almost "real time" reconstruction (SmartView) with a display latency of less than one second. An in-room monitor provides an image display with 6 frame/sec display of up to ninety, one second rotation, scans. Controls are provided for the control of the table and gantry by a physician or their assistant during procedures that require imaging.

Device Description

The SmartView Option for HiSpeed CT/i is an x-ray computed tomography scanner based on the HiSpeed CT/i platform consisting of a gantry, patient table, console, computer and associated accessories.

AI/ML Overview

The provided document is a 510(k) summary for the GE Medical Systems SmartView Option for HiSpeed CT/i. It focuses on demonstrating substantial equivalence to predicate devices, rather than a detailed study proving performance against defined acceptance criteria. Therefore, most of the requested information cannot be extracted from this document, as it describes a clearance process based on equivalence for a modification to an existing device, not a performance study for a novel device.

However, based on the limited information, here's what can be gathered:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Display latencyLess than one second
Image display frame rate6 frames/sec for up to ninety, one-second rotation, scans
SafetyNo new potential safety risks compared to predicate devices; performs as well as or better than currently marketed devices regarding safety. Adherence to UL and IEC standards.

2. Sample size used for the test set and the data provenance

Not applicable. The document does not describe a clinical study with a test set of images or patients. It focuses on demonstrating safety and effectiveness through substantial equivalence and compliance with existing standards for a device modification.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. No test set requiring expert ground truthing is described in this document.

4. Adjudication method for the test set

Not applicable. No test set adjudication is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a modification to a CT scanner that provides "real-time" reconstruction and display for procedural guidance. It is not an AI-assisted diagnostic tool that would typically undergo an MRMC study.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. The SmartView Option is an integrated component of a CT system designed to provide real-time image feedback to a physician during procedures. It's not a standalone algorithm in the typical sense.

7. The type of ground truth used

Not applicable as there is no described performance study for a diagnostic task. The "ground truth" for the device's functionality would be its ability to display images within the specified latency and frame rate, which appears to be based on engineering specifications and testing rather than a clinical ground truth.

8. The sample size for the training set

Not applicable. This document describes a hardware/software modification to an existing CT system, not an AI model requiring a training set in the conventional sense.

9. How the ground truth for the training set was established

Not applicable. As above, no training set for an AI model is described.

{0}------------------------------------------------

GE Medical Systems

General Electric Company P.O. Box 414, Milwaukee, WI 53201

K973168

2 | 1991

Summary of Safety & Effectiveness

This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h)

Larry A. Kroger, Ph.D. Submitter. Regulatory Programs Manager Who may be contacted by telephone at 414-544-3894 or by FAX at 414-544-3863. Summary prepared 21 August 1997

Product Identification Name: SmartView Option for HiSpeed CT/i

Manufacturer. General Electric Medical Systems 16800 W. Rverson Road New Berlin, WI 53151

Same as Manufacturer Distributor:

Marketed Devices:

The SmartView Option for HiSpeed CT/i is of a type and substantially equivalent to currently marketed Computed Tomography X-ray Systems that comply with the same or equivalent standards and have the same intended uses.

Device Description:

The SmartView Option for HiSpeed CT/i is an x-ray computed tomography scanner based on the HiSpeed CT/i platform consisting of a gantry, patient table, console, computer and associated accessories.

Materials: Materials and construction are equivalent to the HiSpeed CT/i and are compliant with UL 187, IEC 601-1, and 21 CFR Subchapter J.

Design: The design is essentially the same as the HiSpeed CT/i (K940606) and CT/i with Performix Tube Option (K964746) the difference being that the SmartView option includes an inroom monitor, a hand control, and a footswitch for initiating x-rays. The hand control allows for gantry tilt and table positioning as well as control of viewing of the scans displayed in the room.

Energy Source and Exposure Levels: The energy source and exposure levels are the same as those previously submitted for HiSpeed CT/i and HiSpeed CT/i with Performix Tube Option.

Principals of Operation: The same as HiSpeed CT/i and HiSpeed CT/i with Performix Tube Option.

{1}------------------------------------------------

Summary (cont.) SmartView Option

Indications for Use:

The SmartView Option for HiSpeed CT/i is indicated for head and whole body x-ray computed tomography applications. This option is a modifications which can be added to the existing family of HiSpeed CT/i Systems. The option provides almost "real time" reconstruction (SmartView) with a display latency of less than one second. An in-room monitor provides an image display with 6 frame/sec display of up to ninety, one second rotation, scans. Controls are provided for the control of the table and gantry by a physician or their assistant during procedures that require imaging.

Comparison with Predicate:

It is the opinion of GE Medical Systems that the SmartView Option for HiSpeed CT/i is of a comparable type and substantially equivalent to currently marketed head and whole body x-ray computed tomography systems with respect to design, material composition, energy source, and radiation characteristics.

Adverse Effects on Health:

Potential electrical, mechanical and radiation hazards are identified in a hazard analysis and controlled by:

  • System evaluation to insure performance to specifications and Federal Regulations. .
  • Adherence to Industry and International Standards. (UL and IEC) .

Conclusions:

Use of the SmartView Option for HiSpeed CT/i does not result in any new potential safety risks and performs as well as or better than devices currently on the market.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 2 1 1997

Larry A. Kroger, Ph.D. Regulatory Affairs Program Manager GE Medical Systems, Inc. P.O. Box 414 Milwaukee, WI 53201

Re:

K973168

SmartView Option for HiSpeed CT/i Dated: August 25, 1997 Received: August 25, 1997 Regulatory class: II 21 CFR 892.1750/Procode: 90 JAK

Dear Dr. Kroger:

We have reviewed your Section 510(k) notification of intent to market the device and we have determined the device is substantially equivalent (for the including) to devices marketed in interstate in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Fremarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (2) CFR Part 801 and additionally 809.10 for in viro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours.

h.Liau Yu

Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K973168 510(k) Number (if known):

Device Name: SmartView Option for HiSpeed CT/i

Indications For Use:

The SmartView Option for HiSpeed CT/i is indicated for head and whole body x-ray computed tomography applications.

The SmartView Option is a "real time" reconstruction with a display latency less than one second. An in-room monitor provides an image display with 6 frame/sec display of ninety, one second rotation, scans. Controls are provided for the control of the table and gantry by a physician or their assistant during procedures that require imaging.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use __ (Per 21 CFR 801-109)

Over-The Counter Use_ OR

Daniel A. Seymore

(Division Sign-Off) Division of Reproductive, bdominal. ENT and Radiological Devic 510(k) Number

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.