K Number
K972609
Manufacturer
Date Cleared
1997-08-06

(26 days)

Product Code
Regulation Number
862.3450
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VARIANT Express Glycohemoglobin Program is intended for the separation and area percent determination of glycohemoglobin (GHb) in whole blood using boronate affinity high performance liquid chromatography (HPLC).

Device Description

The Bio-Rad VARIANT Express Glycohemoglobin Program is designed for use on the Bio-Rad VARIANT Express Instrument. The analytical system, consisting of instrument, software and reagent kit, provides an assay for the separation and percent determination of glycohemoglobin in whole blood. The VARIANT Express Glycohemoglobin Program is for use on the fully automated VARIANT Express analyzer. The analytical system of instrument and reagent kit provides a means of measuring a mixture of hemoglobins, each formed by the non-enzymatic attachment of circulating blood glucose to a receptive amino acid site on the hemoglobin molecule (HbAo). Attachment of glucose to hemoglobin is achieved in a two step process: formation of an unstable aldimine (Schiff base) intermediate followed by production of a stable ketoamine linkage. Formation of the Schiff base intermediate is a reversible reaction and is influenced by short-term blood glucose fluctuations. In contrast, conversion of the Schiff base intermediate to the "ketoamine" products (Total GHb) is much slower and irreversible. The percentage of total glycated hemoglobin in whole blood is dependent upon the level of sustained blood glucose and is indicative of mean blood glucose over the lifetime of red blood cells (~120 days). Thus, monitoring of total GHb levels has become a routine test for the long-term management of patients with diabetes mellitus.

AI/ML Overview

The Bio-Rad VARIANT™ Express Glycohemoglobin Program is an in-vitro diagnostic device intended for the separation and area percent determination of glycohemoglobin (GHb) in whole blood using boronate affinity high performance liquid chromatography (HPLC).

Here's an analysis of its acceptance criteria and the study that demonstrates its performance:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Predicate Device Performance)Reported Device Performance (Bio-Rad VARIANT Express Glycohemoglobin Program)
PrecisionMust meet requirements of the National Glycohemoglobin Standardization Program (NGSP) and be comparable to the predicate (Bio-Rad VARIANT Total GHb Program).Within-run % CV: Low: 3.54%, Medium: 2.90%, High: 2.30% Between-run % CV: Low: 3.97%, Medium: 4.47%, High: 3.55% Total Precision: Low: 5.03%, Medium: 4.70%, High: 3.67% The method meets NGSP precision requirements.
AccuracyHigh correlation with the predicate device (Bio-Rad VARIANT Total GHb Program).Correlation coefficient ("r") = 0.9895 when compared to the Bio-Rad VARIANT Total GHb Program.
Intended UseSame as predicate: Separation and area percent determination of glycohemoglobin (GHb) in whole blood using boronate affinity high performance liquid chromatography (HPLC).Same as predicate: Separation and area percent determination of glycohemoglobin (GHb) in whole blood using boronate affinity high performance liquid chromatography (HPLC).
Underlying TechnologySame as predicate: Boronate affinity gel chromatography, direct dilution of whole blood for sample preparation, visible detection at the same wavelength.Same as predicate: Boronate affinity gel chromatography, direct dilution of whole blood in aqueous medium for sample prep, visible detection at the same wavelength.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size:
    • Precision Study: "low, medium and high patient" samples were used. The exact number of individual patient samples for each level or total number of runs is not specified, but it was sufficient to satisfy NCCLS EP5-T2 protocol.
    • Accuracy/Correlation Study: Not explicitly stated, but the study "followed NCCLS Document EP9-T," which would typically involve a sufficient number of samples across the measuring range to establish a robust correlation.
  • Data Provenance: Not explicitly stated in the provided text. It's likely from an internal Bio-Rad study, but details regarding country of origin or whether the data was retrospective or prospective are not provided. Given it's a 510(k) submission for a diagnostic device, the studies are typically prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as this device is a diagnostic assay, not an AI or imaging device that relies on human expert interpretation for ground truth. The "ground truth" for this device's performance is established by comparison to a legally marketed predicate device (Bio-Rad VARIANT Total GHb Program) using established laboratory protocols and measurements.

4. Adjudication Method for the Test Set

This information is not applicable for the same reason as point 3. Adjudication methods like 2+1 or 3+1 are used when human expert opinions need to be reconciled for a ground truth label, typically in situations where qualitative assessment or image interpretation is involved.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

This information is not applicable. The Bio-Rad VARIANT Express Glycohemoglobin Program is an automated laboratory assay for measuring glycohemoglobin. It does not involve human readers interpreting images or data that would be assisted by AI.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the studies conducted (precision and accuracy/correlation) were for the standalone performance of the Bio-Rad VARIANT Express Glycohemoglobin Program as an analytical system (instrument, software, and reagent kit). The device itself is "fully automated" and provides quantitative results directly.

7. The Type of Ground Truth Used

The "ground truth" for the performance evaluation was established by:

  • Predicate Device Comparison: For accuracy, the Bio-Rad VARIANT Express Glycohemoglobin Program's results were correlated with those obtained from the legally marketed Bio-Rad VARIANT Total GHb Program (K941616). This predicate device served as the reference standard for accuracy.
  • Established Analytical Protocols: For precision, the studies followed NCCLS Evaluation protocol, Vol. 12, No 4, EP5-T2. These protocols define how to measure and verify precision in laboratory assays.
  • National Standardization Program: The device's precision performance was evaluated against the requirements of the National Glycohemoglobin Standardization Program (NGSP).

8. The Sample Size for the Training Set

This information is not applicable in the context of this device's development as described. The Bio-Rad VARIANT Express Glycohemoglobin Program is an analytical system based on established chemical and physical principles (affinity chromatography and HPLC), not a machine learning or AI model that requires a distinct "training set" in the computational sense. The development of such systems involves extensive analytical validation, calibration, and optimization rather than machine learning training.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reasons as point 8.

{0}------------------------------------------------

K 712609

Image /page/0/Picture/1 description: The image shows the logo for BIO-RAD. The logo is white text on a black, rounded rectangle. The text is in all caps and bolded.

Bio-Rad Laboratories Diagnostics Group 4000 Alfred Nobel Dr. Hercules, CA 94547-1803 Telephone: 510 724-7000 Fax: 510 741-5824

AUG -6 in

SUMMARY OF SAFETY AND EFFECTIVENESS

Submitter:Bio-Rad Laboratories, Inc.Clinical Systems Division4000 Alfred Nobel DriveHercules, California 94547Phone 1-510-741-6015FAX 1-510-741-5824
Contact Person:John W. NelsonManager, Regulatory Affairs
Date Prepared:June 30, 1997
Product Trade Name:Bio-Rad VARIANT™ Express GlycohemoglobinProgram
Common Name:VARIANT Express
Classification Name:Assay, Glycosylated Hemoglobin, 81LCP
Predicate Devices1. VARIANT Total GHb Program (K941616)2. Pierce Glycotest II GlyHb Assay Kit (K 895661

The Bio-Rad VARIANT Express Glycohemoglobin Program is designed for use on the Bio-Rad VARIANT Express Instrument. The analytical system, consisting of instrument, software and reagent kit, provides an assay for the separation and percent determination of glycohemoglobin in whole blood.

The VARIANT Express Glycohemoglobin Program is for use on the fully automated VARIANT Express analyzer. The analytical system of instrument and reagent kit provides a means of measuring a mixture of hemoglobins, each formed by the non-enzymatic attachment of circulating blood glucose to a receptive amino acid site on the hemoglobin molecule (HbAo). Attachment of glucose to hemoglobin is achieved in a two step process: formation of an unstable aldimine (Schiff base) intermediate followed by production of a stable ketoamine linkage. Formation of the Schiff base intermediate is a reversible reaction and is influenced by short-term blood glucose fluctuations. In contrast, conversion of the Schiff base intermediate to the "ketoamine" products (Total GHb) is much slower and irreversible. The percentage of total glycated hemoglobin in whole blood is dependent upon the level of sustained blood glucose and is indicative of mean blood glucose over the lifetime of red blood cells (~120 days). Thus, monitoring of total GHb levels has become a routine test for the long-term management of patients with diabetes mellitus.

{1}------------------------------------------------

To establish substantial equivalence to an existing device, and thus establish the safety and effectiveness of the Bio-Rad VARIANT Express Glycohemoglobin Program , the Bio-Rad VARIANT Express Glycohemoglobin Program has been compared to the Bio-Rad VARIANT Total GHb Program (K941616). A review of the intended use of each system shows them to be the same in that they both measure total qlycosylated hemoglobin in a sample of whole blood using boronate affinity high performance liquid chromatography. The intended use of both the Bio-Rad VARIANT Express Glycohemoglobin Program and the VARIANT Total GHb Program is stated as:.... for the separation and area percent determination of glycohemoglobin (GHb) in whole blood using boronate affinity high performance liquid chromatography (HPLC) (see Appendix C and Appendix L).

The underlying technology of VARIANT Express Glycohemoglobin Program and the VARIANT Total GHb Program is the same. Both assay systems use boronate affinity gel chromatography as the separation mechanism; both assay systems use a direct dilution of whole blood in an aqueous medium for sample preparation; and both systems use visible detection at the same wavelength to monitor the hemoglobin.

The performance of the Bio-Rad VARIANT Express Glycohemoglobin Program was evaluated for precision, measuring range, and accuracy. The precision studies were done according to NCCLS Evaluation protocol, Vol. 12, No 4, EP5-T2, Appendix C, pp 31-39. Using this protocol, precision of the system was determined using a low. medium and high patient. The Within-run % CV for the low was 3.54%, for the Medium 2.90%, and for the High 2.30%. The Between-run % CV for the low was 3.97%. for the medium 4.47% and the high patient 3.55%. Total precision was 5.03% for the low. 4.70% for the medium patient and 3.67 for the high patient. The VARIANT Express Glycohemoglobin Program method meets the precision requirements of the National Glycohemoglobin Standardization Program (NGSP).

A correlation study, to determine accuracy of the Bio-Rad VARIANT Express Glycohemoglobin Program, was done against the Bio-Rad VARIANT Total GHb Program. The study followed NCCLS Document EP9-T. The "r" for the correlation was 0.9895.

When considering the similarities of the general characteristics of the two assays (Appendix C), the use of the same technology and the excellent correlation between the two methods, it can be concluded that the VARIANT Express Glycohemoglobin Program and the VARIANT Total GHb Program are substantially equivalent. Based on the establishment of substantial equivalence, the safety and effectiveness of the Bio-Rad VARIANT Express Glycohemoglobin Program is confirmed.

{2}------------------------------------------------

APPENDIX C

COMPARISON OF GENERAL CHARACTERISTICS

{3}------------------------------------------------

Comparison of General Characteristics of the VARIANT Express Glycohemoglobin Program to the VARIANT Total GHb Program.

ParameterVARIANT ExpressVARIANT Total GHb
Glycohemoglobin ProgramProgram
Analytes Measured/ReportedTotal GlycohemoglobinTotal Glycohemoglobin
Analyte Reported (optional)Standardized HbA1cStandardized HbA1c
Calibration2 point1 point
Basic PrincipleAffinity chromatographyAffinity chromatography
Measurement TypeQuantitative area percentQuantitative area percent
ReagentsAnalytical cartridge, pre-cartridge, elution buffers,hemolysis reagent, washreagent, calibrators, wholeblood primerAnalytical cartridge, elutionbuffers, hemolysis reagent,wash reagent, calibrators
Analysis MediumAlkaline glycylglycine elutionbuffer 1, sorbitol/sodiumcitrate buffer 2, silica packedcartridgeAlkaline glycylglycine elutionbuffer 1, sorbitol/sodiumcitrate buffer 2, silicapacked cartridge
Analysis Temperature35°C35°C
InstrumentationVARIANT Express-AutomatedHPLC system forglycohemoglobin analysis.VARIANT HemoglobinTesting System- AutomatedHPLC system forhemoglobin analysis.
Sample TypeWhole bloodWhole blood
Sample Volume5 μL5 μL
Use of ControlsTwo levels of control per runTwo levels of control per run

{4}------------------------------------------------

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

AUG - 6 1997

Mr. John W. Nelson · Manager, Regulatory Affairs Clinical Systems Division Bio-Rad Laboratories 4000 Alfred Nobel Drive Hercules, CA 94547-1803

Re : K972609 VARIANT Express Glycohemoglobin Proqram Regulatory Class: II Product Code: LCB Dated: July 11, 1997 Received: July 11, 1997

Dear Mr. Nelson:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.

{5}------------------------------------------------

Paqe 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".

Sincerely yours,
Steven Litman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Statement of Intended Use

Page 1 of 1

510(k) Number (If Known)

Bio-Rad VARIANT™ Express Glycohemoglobin Program. Device Name:

  • The VARIANT Express Glycohemoglobin Program is intended for the Indications for Use: separation and area percent determination of glycohemoglobin (GHb) in whole blood using boronate affinity high performance liquid chromatography (HPLC).

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDHR, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ Он

§ 862.3450 Gentamicin test system.

(a)
Identification. A gentamicin test system is a device intended to measure gentamicin, an antibiotic drug, in human specimens. Measurements obtained by this device are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to ensure appropriate therapy.(b)
Classification. Class II.