K Number
K962573
Date Cleared
1996-07-22

(21 days)

Product Code
Regulation Number
862.1660
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Control for Thyroid Stimulating Hormone assays.

Device Description

Horse serum albumin control with recombinant Thyroid Stimulating Hormone for use in TSH assays.

AI/ML Overview

Here's an analysis of the provided text, focusing on the requested acceptance criteria and study information:

This submission (K962573 for the Elecsys® PreciControl TSH Assay) is a 510(k) Pre-market Notification for a quality control material, not a diagnostic device that performs measurements. Therefore, the "acceptance criteria" and "device performance" are related to its suitability as a control, such as its stability and ability to be assigned a value, rather than diagnostic accuracy metrics like sensitivity or specificity.

1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of a quality control material, the "acceptance criteria" relate to its manufacturing, stability, and value assignment, which are implicitly demonstrated through comparison to the predicate device and standardized methods. The document doesn't explicitly list "acceptance criteria" in the traditional sense of performance targets (e.g., minimum sensitivity). Instead, it describes the characteristics of the control material and confirms they are adequate for its intended use.

Acceptance Criteria CategoryReported Device Performance (Elecsys® PreciControl TSH)
Intended UseTo be a control for Thyroid Stimulating Hormone assays.
FormulationHorse serum albumin control with recombinant Thyroid Stimulating Hormone. (Same as predicate: Horse serum matrix, recombinant TSH)
Target Concentration0.18 μU/mL (Different from predicate, which had 0 and 1.4 μU/mL, but appropriate for a control material)
StandardizationStandardized against WHO-reference material (#80/558). (Same as predicate)
Storage Conditions2 - 8°C. (Same as predicate)
ReconstitutionNone required, liquid. (Same as predicate)
Open Vial Stability4 weeks at 2 - 8°C. (Same as predicate)
Application Test SystemElecsys® 2010. (Same as predicate)
ManufacturerBoehringer Mannheim Germany. (Same as predicate)
Substantial EquivalenceSubstantially equivalent to Elecsys® TSH CalSet (K961491). Intended use changed to quality control material.

2. Sample Size Used for the Test Set and Data Provenance

The document does not describe a "test set" in the context of diagnostic performance. For a quality control material, testing would involve verifying its properties and stability. The specific sample sizes for these internal manufacturing and validation tests are not provided in the 510(k) summary. Given the date (1996), such detailed information for a control material was less frequently included in the public 510(k) summary than it might be today. There is no information on data provenance (country of origin, retrospective/prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

Not applicable. As this is a quality control material, there is no "ground truth" established by experts in the diagnostic sense (e.g., radiologists interpreting images). The "ground truth" for its value assignment comes from standardization against the WHO-reference material.

4. Adjudication Method for the Test Set

Not applicable. There is no diagnostic "test set" requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

Not applicable. This is not a diagnostic device, and therefore, no MRMC study involving human readers' performance with or without AI assistance would be relevant or conducted.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a quality control material, not an algorithm or a diagnostic device, so standalone performance in that context is not relevant.

7. The Type of Ground Truth Used

The "ground truth" for the Elecsys® PreciControl TSH is its assigned value, which is established by standardization against the World Health Organization (WHO) reference material #80/558. This is a recognized international standard for TSH assays.

8. The Sample Size for the Training Set

Not applicable. As a quality control material, there is no "training set" in the context of an algorithm or diagnostic model. Development would involve manufacturing and validation processes, but not machine learning training.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for a quality control material.

{0}------------------------------------------------

K962573

JUL 22 1996

Image /page/0/Picture/2 description: The image shows the logo for Boehringer Diagnostics. The logo consists of the word "boehringer" inside of a circle, which is inside of a black square. To the right of the square is the word "Diagnostics" in a serif font.

510(k) Summary

According to the requirements of 21 CFR 807.92, the following information Introduction provides sufficient detail to understand the basis for a determination of substantial equivalence. Boehringer Mannheim Corporation 1. Submitter 2400 Bisso Lane n a me, Concord, CA 94524-4117 address, (510) 674-0667 contact Contact Person: Patricia Klimley Date Prepared: June 26, 1996 2. Proprietary name: Elecsys® PreciControl TSH Assay Device name Common name: Control material Classification name: Enzyme controls (Assayed and Unassayed) 3. The Boehringer Mannheim Elecsys® PreciControl TSH is substantially Predicate equivalent to the Elecsys® TSH CalSet (K961491). The intended use has device been changed to a quality control material.

Continued on next page

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Boehringer Diagnostics. The logo consists of the words "mannheim" and "boehringer" stacked on top of each other inside of a black square with a white circle. To the right of the logo is the word "Diagnostics" in a serif font.

510(k) Summary, Continued

ﺍﻟﻤﺴﺎﻋﺪ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ

4.DeviceDescriptionHorse serum albumin control with recombinant Thyroid Stimulating Hormonefor use in TSH assays.
5.Intended useControl for Thyroid Stimulating Hormone assays.
6.Comparisonto predicatedeviceThe Boehringer Mannheim Elecsys® PreciControl TSH is substantiallyequivalent to other products in commercial distribution intended for similaruse. Most notably it is substantially equivalent to the Elecsys® TSH CalSet(K961491).
The following table compares the Elecsys® PreciControl TSH with thepredicate device, Elecsys® TSH CalSet.
Similarities:• Same formulation.
•Same matrix: Horse serum matrix
•Same analyte: recombinant Thyroid Stimulating Hormone
•Standardization: World Health Organization (WHO-reference material #80/558)
•Storage conditions: 2 - 8°C
•Application test system: Elecsys® 2010
•Reconstitution: none required, liquid
•Open vial stability: 4 weeks at 2 - 8°C
•Manufacturer: Boehringer Mannheim Germany

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains the logo for Boehringer Diagnostics. The logo consists of the word "mannheim" written vertically on the left side of a black square. Inside the square is a white circle with the word "boehringer" written inside. To the right of the logo is the word "Diagnostics" in a serif font.

510(k) Summary, Continued

Comparison
to predicate
device cont.

Differences:

FeatureElecsys® TSH CalSetElecsys ® PreciControl TSH
Targetconcentration0 and 1.4 μU/mL0.18 μU/mL
Intended UseCalibration of the ElecsysTSH AssayMonitoring assays for TSH

Performance Characteristics:

FeatureElecsys® TSH CalSetElecsys ® PreciControl TSH
Open VialStability4 weeks at 2 - 8°C4 weeks at 2 - 8°C
ValueAssignmentstandardized against WHO-reference material (#80/558)standardized against WHO-reference material (#80/558)

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.