(95 days)
Immunoassay for the in vitro quantitative determination of Thyroid Stimulating Hormone in human serum and plasma.
Sandwich principle: Total duration of assay: 18 minutes (37°C).
• 1st Incubation (9 minutes): Sample (50µL), a biotinylated monoclonal TSH-specific antibody (60 µL), and a monoclonal TSH-specific antibody labeled with a ruthenium complex (50 µL) react to form a sandwich complex.
• 2nd Incubation (9 minutes): After addition of streptavidin-coated microparticles (40 µL) the complex becomes bound to the solid phase via interaction of biotin and streptavidin.
• The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell.
• Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame).
• Results are determined via a calibration curve which is instrument-specifically generated by a 2-point calibration curve and a master curve provided via the reagent bar code.
This document describes the Elecsys® TSH Assay, an electrochemiluminescence assay for the quantitative determination of Thyroid Stimulating Hormone (TSH) in human serum and plasma. The K961491 submission claims substantial equivalence to the Enzymun-Test® TSH (K915195).
Here's an analysis of the acceptance criteria and study information provided:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implied by the comparison to the predicate device, Enzymun-Test® TSH, and general scientific principles for immunoassay performance. The document presents performance characteristics for both the Elecsys® TSH and the predicate device, demonstrating that the new device meets or exceeds the predicate's performance in key areas.
| Feature | Acceptance Criteria (Implied by Predicate Performance / General Standards for Immunoassay) | Reported Elecsys® TSH Performance |
|---|---|---|
| Precision | Typically, %CV for immunoassays at various TSH levels should be within acceptable ranges (e.g., single-digit %CV for mid-to-high concentrations, possibly higher for very low concentrations). Comparing to the predicate, Elecsys® TSH should show comparable or better precision. | Within-Run %CV:- Sample Control 1 (0.914 µU/ml): 2.08%- Control 2 (2.451 µU/ml): 1.88%- Control 3 (10.670 µU/ml): 1.47%Total %CV:- Sample Control 1 (0.914 µU/ml): 3.28%- Control 2 (2.451 µU/ml): 2.20%- Control 3 (10.670 µU/ml): 1.76% |
| Sensitivity | Should have a lower detection limit and functional sensitivity suitable for clinical applications, ideally comparable to or better than the predicate's 0.03 µU/ml. | Functional Sensitivity: 0.01 µU/mlLower Detection Limit: 0.005 µU/mL |
| Assay Range / Linearity | Reportable range should cover clinically relevant TSH levels, and values within this range should be linear (e.g., within ±10% deviation from the linear line). Predicate range: 0.03 µU/ml - 40.00 µU/ml. | Reportable Range: 0.01 µU/ml - 100 µU/ml(within ±10% deviation from the linear line) |
| Method Comparison (Correlation with Predicate) | Should demonstrate good correlation with the predicate device (e.g., r > 0.95, slope near 1, intercept near 0). | Vs Enzymun-Test® TSH:- Least Squares: y = 1.09x + 0.14, r = 0.991, SEE = 0.798 (N=132)- Passing/Bablok: y = 1.12X - 0.05, r = 0.991, SEE = 0.798 (N=132)Vs Nichols 3rd Generation:- Least Squares: y = 1.02x - 0.24, r = 0.985, SEE = 1.12 (N=142) |
| Interfering Substances | Should show no significant interference from common biological substances at clinically relevant concentrations. | No interference at:- Bilirubin 25 mg/dL- Hemoglobin 1 g/dL- Lipemia 1500 mg/dL- Biotin 30 ng/mL- RF 339 IU/mL |
| Specificity (Cross-Reactivity) | Should demonstrate minimal or no cross-reactivity with other related hormones to avoid false positives/negatives. | Conc % Cross-Reactivity:- HCG 200: 0- LH 400: 0.038- FSH 400: 0.008- HGH 400: 0.00004 |
2. Sample Size Used for the Test Set and Data Provenance
- Precision: For "Within-Run" and "Total" precision,
N = 60was used for each of the three control levels for the Elecsys® TSH. (The predicate used N=120 for each of its three levels.) - Method Comparison:
N = 132samples were used for comparison between Elecsys® TSH and Enzymun-Test® TSH.N = 142samples were used for comparison between Elecsys® TSH and Nichols 3rd Generation.
- Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. Given the context of a 510(k) summary for a diagnostic assay, it is highly probable that these studies involved prospective collection of human serum and plasma samples according to standard clinical laboratory practices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
For an in vitro diagnostic (IVD) immunoassay like the Elecsys® TSH, the "ground truth" is typically established by the reference method (the predicate device, in this case, Enzymun-Test® TSH, and an additional comparator, Nichols 3rd Generation TSH assay) or by expert consensus on the clinical state of the patient using standard clinical diagnostic criteria for thyroid function, rather than by human interpretation of images by experts. Therefore, the concept of "experts" as in radiologists, for example, is not applicable in this context. The "ground truth" values for TSH concentration are derived directly from the laboratory measurements by the reference assays.
4. Adjudication Method for the Test Set
Not applicable for this type of IVD device. The "ground truth" is the quantitative measurement from the reference methods, not subjective interpretation requiring adjudication among human experts.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an immunoassay for TSH, not an imaging device requiring human readers or AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This device is a standalone algorithm/instrument system. The Elecsys® TSH Assay on the Elecsys® 2010 instrument performs the measurement and provides a quantitative TSH value without human interpretation in the loop impacting the result generation directly. The reported "Performance Characteristics" (precision, sensitivity, linearity, method comparison, interference, specificity) represent the standalone performance of the assay and instrument system.
7. The Type of Ground Truth Used
The ground truth used for establishing the performance of the Elecsys® TSH Assay was:
- Reference Method Comparison: TSH values obtained from samples analyzed by the predicate device (Enzymun-Test® TSH) and another established method (Nichols 3rd Generation TSH). These assays provide quantitative measurements that serve as the "true" or reference values for TSH concentration.
- Known Concentrations/Spiked Samples: For studies like linearity, sensitivity, and interfering substances, samples with known TSH concentrations or spiked with interfering substances would be used as the ground truth.
- Clinical Samples: Samples from patients with various TSH levels (e.g., hyperthyroid, euthyroid, hypothyroid) would be used for method comparison studies to ensure broad clinical utility.
8. The Sample Size for the Training Set
The document does not explicitly mention a "training set" in the context of machine learning. For an immunoassay, the "training" analogous to adjusting an algorithm would be the development and optimization of the assay reagents, protocols, and calibration procedures. The calibration curve for the Elecsys® TSH is generated by a 2-point calibration curve and a master curve provided via the reagent bar code. This essentially "calibrates" the system, allowing it to accurately quantitate unknown samples. The details of the samples used to establish the master curve are not provided but would typically involve highly characterized standard materials.
9. How the Ground Truth for the Training Set was Established
As discussed above, the concept of a training set as understood in AI/ML is not directly applicable. However, the "ground truth" for the development and calibration of the assay would be established through:
- WHO Standardization: The assay standardization is stated to be via the World Health Organization (WHO), indicating that the assay's measurements are traceable to international reference standards for TSH. These standards are rigorously characterized and assigned specific TSH values by expert international committees.
- Internal Validation and Reference Materials: Boehringer Mannheim would have used a series of internal reference materials and standards, carefully characterized for their TSH concentrations, during the assay development and for establishing the master curve. These internal standards would be assigned values based on comparisons to WHO international standards and/or highly accurate internal reference methods.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Boehringer Diagnostics. The logo consists of the word "Diagnostics" in a serif font, next to a black square containing the words "boehringer mannheim" in white, with the word "boehringer" inside of a white circle. The logo is simple and professional.
JUL 22 1996 510(k) Summary
According to the requirements of 21 CFR 807.92, the following information Introduction provides sufficient detail to understand the basis for a determination of substantial equivalence.
- Submitter name, address, contact
Boehringer Mannheim Corporation 2400 Bisso Lane Concord, CA 94524-4117 (510) 674-0667
Contact Person: Betsy Soares-Maddox
Date Prepared: April 4, 1996
- Device name Proprietary name: Elecsys® TSH Assay
Common name: Electrochemiluminescence assay for the thyroid stimulating hormone.
Classification name: Radioimmunoassay-thyroid stimulating hormone
- Predicate device
We claim substantial equivalence tc the Enzymun-Test® TSH (K915195).
Continued on next page
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Boehringer Diagnostics. The logo consists of a black square with a white circle inside. The word "boehringer" is written inside the circle, and the word "mannheim" is written vertically along the left side of the square. To the right of the logo, the word "Diagnostics" is written in a serif font.
510(k) Summary, Continued
| 4.DeviceDescription | Sandwich principle: Total duration of assay: 18 minutes (37°C).• 1st Incubation (9 minutes): Sample (50µL), a biotinylated monoclonal TSH-specific antibody (60 µL), and a monoclonal TSH-specific antibody labeledwith a ruthenium complex (50 µL) react to form a sandwich complex.• 2nd Incubation (9 minutes): After addition of streptavidin-coatedmicroparticles (40 µL) the complex becomes bound to the solid phase viainteraction of biotin and streptavidin.• The reaction mixture is aspirated into the measuring cell where themicroparticles are magnetically captured onto the surface of the electrode.Unbound substances are then removed with ProCell.• Application of a voltage to the electrode then induces chemiluminescentemission which is measured by a photomultiplier (0.4 second read frame).• Results are determined via a calibration curve which is instrument-specifically generated by a 2-point calibration curve and a master curveprovided via the reagent bar code. |
|---|---|
| 5.Intended use | Immunoassay for the in vitro quantitative determination of ThyroidStimulating Hormone in human serum and plasma. |
| 6.Comparisonto predicatedevice | The Boehringer Mannheim Elecsys® TSH is substantially equivalent to otherproducts in commercial distribution intended for similar use. Most notably itis substantially equivalent to the currently marketed Enzymun-Test® TSH.The following table compares the Elecsys® TSH with the predicate device,Enzymun-Test® TSH.Similarities:• Intended Use: Immunoassay for the in vitro quantitative determination ofThyroid Stimulating Hormone• Sample type: Serum and plasma• Antibody: Mouse Monoclonal anti-TSH• Solid phase binding principle: Streptavidin/Biotin• Assay standardization: World Health Organization (WHO) |
page
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Boehringer Diagnostics. The logo consists of two parts: a square with the word "mannheim" written vertically on the left side and the word "boehringer" written inside a circle on the right side, and the word "Diagnostics" written in a cursive font on the right side of the square. The logo is black and white.
510(k) Summary, Continued
Differences:
Substantial equivalence,
cont.
| Feature | Elecsys® TSH | Enzymun-Test® TSH |
|---|---|---|
| Reaction testprinciple | Electrochemiluminescence | ELISA/1-step sandwich assayusing streptavidin technology |
| Instrumentrequired | Elecsys® 2010 | ES 300 |
| Assay Range | Reportable Range:0.01 μU/ml - 100 μU/ml | Reportable Range:0.03 μU/ml - 40.00 μU/ml |
| CalibrationStability | A calibration is recommendedevery 7 days if kit is notconsumed; 4 weeks with samereagent lot if reagent isconsumed within 7 days. | Calibration is required every 2weeks |
Performance Characteristics:
| Feature | Elecsys® TSH | Enzymun-Test® TSH |
|---|---|---|
| Precision | Modified NCCLS (µU/ml ): | Modified NCCLS (µU/ml ): |
| Level | Sample Control 1 Control 2 | Low Mid High |
| N | 60 60 60 | 120 120 120 |
| Within-Run | 0.914 2.451 10.670 | 0.09 1.23 23.13 |
| %CV | 2.08 1.88 1.47 | 20.0 2.7 2.6 |
| Total | 0.914 2.451 10.670 | 0.09 1.23 23.13 |
| %CV | 3.28 2.20 1.76 | 48.9 5.2 4.6 |
| Sensitivity | Functional Sensitivity:0.01 µU/mlLower Detection Limit0.005 µU/mL | Lower Detection Limit:0.03 µU/ml |
Continued on next page
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Boehringer Diagnostics. The logo consists of the word "mannheim" written vertically on the left side of a black square. Inside the square is a white circle with the word "boehringer" written inside. To the right of the square is the word "Diagnostics".
510(k) Summary, Continued
Substantial equivalence,
Performance Characteristics:
cont.
| Feature | Elecsys® TSH | Enzymun-Test® TSH | |||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Linearity | 0.01 µU/mL - 100 µU/mL(within ±10% deviation from the linear line) | 0.03 µU/mL - 40.00 µU/mL(within ±10% deviation from the linear line) | |||||||||||||||
| MethodComparison | Vs Enzymun-Test® TSHLeast Squares:$y = 1.09x + 0.14$$r = 0.991$$SEE = 0.798$$N = 132$Passing/Bablok$y = 1.12X - 0.05$$r = 0.991$$SEE = 0.798$N=132 | Vs Nichols 3rd GenerationLeast Squares:$y = 1.02x - 0.24$$r = 0.985$$SEE = 1.12$N= 142 | |||||||||||||||
| Interferingsubstances | No interference at:Bilirubin 25 mg/dLHemoglobin 1 g/dLLipemia 1500 mg/dLBiotin 30 ng/mLRF 339 IU/mL | No interference at:64.5 mg/dL1 g/dL1250 mg/dL30 ng/mLNo concentration reported | |||||||||||||||
| Specificity | Conc % Cross-Reactivity HCG 200 0 LH 400 0.038 FSH 400 0.008 HGH 400 0.00004 | No Cross-Reactivity detected at:200 U/mL200 mU/mL200 mU/mLNot reported |
§ 862.1690 Thyroid stimulating hormone test system.
(a)
Identification. A thyroid stimulating hormone test system is a device intended to measure thyroid stimulating hormone, also known as thyrotrophin and thyrotrophic hormone, in serum and plasma. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.(b)
Classification. Class II.