K Number
K961579
Date Cleared
1996-07-10

(77 days)

Product Code
Regulation Number
864.5400
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The STA Compact instrument is a fully automated laboratory analyzer designed to perform tests which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.

Device Description

The STA Compact instrument is a fully automated laboratory analyzer designed to perform tests which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy. Once samples and test reagents are loaded onto the instrument, sample handling. reagent delivery, analysis and reporting of test results are performed automatically, making the STA Compact instrument a true walkaway system. A central processing unit controls instrument functions, including management of patient results, quality control, scheduling of instrument maintenance and workload organization. The instrument utilizes Diagnostica Stago (France) reagents, and is open to adaptation of other reagents currently available from third-party manufacturers. Barcoding of test reagents, calibrators and controls facilitates their use on the system and makes reagent management simple. Manual entry of reagent information enables the use of non-barcoded reagents. The instrument performs multiple test methodologies in random access (as selected by the user); these include clot-based tests and photometric assays at predetermined wavelengths.

AI/ML Overview

The provided text describes a medical device, the STA Compact® Automated Multi-Parametric Analyzer, and asserts its substantial equivalence to a previously cleared device. However, it does not contain any information regarding specific acceptance criteria, study data, or performance metrics in the format requested.

Therefore, I cannot populate the table or answer the specific questions about sample size, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set details from the provided text.

The text focuses on:

  • Device Function: An automated laboratory analyzer for coagulation tests.
  • Automation: "True walkaway system" with automatic sample handling, reagent delivery, analysis, and reporting.
  • Control: Central processing unit manages results, QC, maintenance, and workload.
  • Reagents: Uses Diagnostica Stago (France) reagents, open to third-party reagents, barcoding for management.
  • Test Methodologies: Clot-based and photometric assays in random access.
  • Carryover: Very low (0.05%).
  • Substantial Equivalence: Claimed to be substantially equivalent to the full-size STA® analyzer (K942117).

To provide the requested information, specific study reports detailing performance metrics and validation methodologies would be required.

{0}------------------------------------------------

Page 53 of 53

AMERICAN BIOPRODUCTS COMPANY Pre-Market 510(k) Notification STA Compact® Automated Multi-Parametric Analyzer

K961579

11 1 0 1996

IX. SAFETY AND EFFECTIVENESS SUMMARY

The STA Compact instrument is a fully automated laboratory analyzer designed to perform tests which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy. Once samples and test reagents are loaded onto the instrument, sample handling. reagent delivery, analysis and reporting of test results are performed automatically, making the STA Compact instrument a true walkaway system. A central processing unit controls instrument functions, including management of patient results, quality control, scheduling of instrument maintenance and workload organization.

The instrument utilizes Diagnostica Stago (France) reagents, and is open to adaptation of other reagents currently available from third-party manufacturers. Barcoding of test reagents, calibrators and controls facilitates their use on the system and makes reagent management simple. Manual entry of reagent information enables the use of non-barcoded reagents.

The instrument performs multiple test methodologies in random access (as selected by the user); these include clot-based tests and photometric assays at predetermined wavelengths.

Sample-to-sample carryover has been demonstrated to be very low, being of the order of 5/100th of 1% (i.e., 0.05%).

The STA Compact instrument is substantially equivalent to the commercially available full-size STA® analyzer that received FDA 510(k) clearance under K942117.

§ 864.5400 Coagulation instrument.

(a)
Identification. A coagulation instrument is an automated or semiautomated device used to determine the onset of clot formation for in vitro coagulation studies.(b)
Classification. Class II (special controls). A fibrometer or coagulation timer intended for use with a coagulation instrument is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.