K Number
K960070
Manufacturer
Date Cleared
1996-04-02

(88 days)

Product Code
Regulation Number
870.1200
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This catheter is intended for balloon atrioseptostomy. Used for the palliation of several congenital heart defects: transposition of the great arteries, total anomalous pulmonary venous drainage without pulmonary obstruction, tricuspid atresia, mitral stenosis, mitral atresia, and pulmonary atresia with intact ventricular septum.

Device Description

The NuMED Septostomy Catheter is a new balloon catheter designed for the neonate with congenital heart disease requiring septostomy. It is a dual lumen catheter, 50cm in length with a 13.5mm ± 0.5mm non-compliant balloon on the distal end. The catheter also features an end hole that will accommodate an 0.018" guidewire. The inflated geometry of the balloon at the rated inflation volume of 2cc is a 13.5mm sphere. There is an imaging band under the balloon for balloon positioning in the left atrium. The catheter tip is angled at 35° to facilitate passage through the interarterial opening in the left atrium. To inflate the balloon to it's maximum diameter, 2cc of diluted contrast media is pushed into the balloon extension after purging. Catheters marked every 5cm (up to 50cm) and are supplied with a one way stopcock for balloon sealing.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the studies that demonstrate the NuMED Atrioseptostomy Catheter meets them, based on the provided text:

Acceptance Criteria and Device Performance

TestAcceptance CriteriaReported Device Performance
1. Balloon Volume vs. DiameterBalloon diameter must be within ± 10% of the rated diameter (13.5mm) at 2cc (i.e., 12.15mm to 14.85mm).Mean diameter at 2.0cc inflation volume was 13.49mm (range 13.0 - 13.9mm), which falls within the 12.15mm to 14.85mm acceptance range.
2. Bond IntegrityMinimum of 3 lb. at all test points.Distal hub to 'Y', Balloon hub to extension, Balloon extension to 'Y' connector, Catheter body to 'Y' connector, Tip to balloon: All > 10.0 lbs. Proximal balloon bond: 8.0 lbs (range 7.0 - 9.0 lbs, std dev. = 1 lb.). Tip to catheter: 8.0 lbs (range 7.0-9.0 lbs, std dev. = 1 lb.). All exceed minimum 3 lbs.
3. Guidewire CompatibilityMinimal resistance while pushing catheter over guidewire. Force may not be great enough to bend guidewire.In no case was the guidewire kinked or was the force in excess of that necessary to insert a guidewire into a similar catheter. All samples were acceptable.
4. Maximum Luminal Injection PressureMust withstand 500 psi.All samples passed 750 psi injection. Average burst pressure was 905.5 psi (range 875 - 950 psi). A limit of 600 psi will be added to the injection criteria.
5. Multiple Inflation TestingBalloon diameter must still be within ± 10% after 40 inflation/deflation cycles.All samples were tested 50 times with no failures or leaks. One sample was inflated 500 times without failure. The diameter, at maximum volume of 2cc, was "virtually unchanged" from the first to the fiftieth inflation. (Statistical analysis provided showed a mean change of 0.15, range 0.1-0.3, std dev 0.085).
6. Injection RateDemonstrate that this device can be used for fluid injection.Mean injection rate of 4cc/sec (range 3.6 to 4.4 cc/sec).
8. Balloon DeflatabilityDeflation achieved in less than 20 seconds.Deflation was achieved by applying negative pressure to the balloon using a 3 cc syringe. Deflation times were 5 - 10 seconds to achieve 50% deflation and all balloons were fully deflated within 15 seconds.
9. Balloon Non-Compliance During UseMinimal deformation, as observed visually.When pulled through a deer heart septum, no deformation was visually observed. This was also demonstrated in a submitted videotape of animal testing.
10. Balloon Protector TestingAbility to form balloon to accept a 6F introducer.All catheters (10 samples) fitted with balloon protectors were admitted through the recommended 6F introducer without incident. No resistance, kinking, or detrimental effects were noted.
11. Introducer CompatibilityCatheter must be admitted through a 6F Introducer.All catheters (10 samples) were admitted through the recommended 6F introducer without incident. No resistance, kinking, or detrimental effects were noted. Catheters showed no signs of damage after inflation and microscopic examination.

Study Details:

Bench Testing (Laboratory Testing)

  1. Sample sizes used for the test set and the data provenance:

    • Balloon Volume vs. Diameter: 10 full-length, sterile catheters. Data provenance is laboratory bench testing, presumably conducted in the USA (where NuMED is located).
    • Bond Integrity: 10 full-length, sterile catheters. Data provenance is laboratory bench testing.
    • Guidewire Compatibility: 20 full-length, sterile catheters. Data provenance is laboratory bench testing.
    • Maximum Luminal Injection Pressure: 10 sterilized, full-length test catheters. Data provenance is laboratory bench testing.
    • Multiple Inflation Testing: 10 full-length catheters. Data provenance is laboratory bench testing.
    • Injection Rate: 10 full-length sterilized catheters. Data provenance is laboratory bench testing.
    • Balloon Deflatability: 5 full-length, sterile catheters. Data provenance is laboratory bench testing.
    • Balloon Non-Compliance During Use: Not explicitly stated, but at least two catheters (NuMED and Baxter Miller). Data provenance is R&D lab testing using a deer heart.
    • Balloon Protector Testing: 10 Atrioseptostomy catheters. Data provenance is laboratory bench testing.
    • Introducer Compatibility: 10 catheters. Data provenance is laboratory bench testing.

    All bench testing appears to be retrospective as the results are being reported after the tests were conducted.

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Bench Testing: No external experts were used to establish ground truth for the measurements in bench testing. The results are objective measurements from equipment (micrometers, pull testers, pressure gauges).
    • Balloon Deflatability: Dr. Ziyad Hijazi, a medical professional, provided input on the maximum allowable deflation time (20 seconds) and also evaluated deflation during animal testing. His qualifications are stated as "New England Medical Center Pediatric Cardiology."
    • Balloon Non-Compliance During Use: Visual observation by the R&D lab staff, and confirmed by Dr. Hijazi via a videotape of animal testing (G940143).
  3. Adjudication method for the test set:

    • Not applicable for the objective measurements performed in bench testing. The results are direct measurements.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a medical device (catheter), not an AI diagnostic tool.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a medical device, not an algorithm.
  6. The type of ground truth used:

    • Bench Testing: Objective physical measurements using calibrated equipment (e.g., micrometer, Chatillon pull tester, pressure gauge).
    • Balloon Deflatability & Non-Compliance: A combination of objective timing (stopwatch) and expert clinical observation/assessment by Dr. Ziyad Hijazi.
  7. The sample size for the training set:

    • Not applicable. This is a medical device, not a machine learning model. The concept of a "training set" doesn't directly apply here. The device was designed and iterated based on engineering principles and prior device experience (referenced as approved materials from IDE G940143, K931009, and G890030) rather than iterative training on a dataset.
  8. How the ground truth for the training set was established:

    • Not applicable (as above). The closest analogy might be the established safety and performance profiles of the predicate devices and the materials approved under previous IDEs, which serve as foundational knowledge for the design.

Animal Testing

  1. Sample sizes used for the test set and the data provenance:

    • Animal Testing: 5 piglets. Data provenance is prospective animal testing performed at the cath lab, as reported by Dr. Ziyad Hijazi.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Animal Testing: Dr. Ziyad Hijazi (New England Medical Center Pediatric Cardiology) performed the testing and provided the summary. He is the expert establishing the direct observations and conclusions from the animal procedures and autopsies.
  3. Adjudication method for the test set:

    • Not applicable. The observations were direct clinical procedures and subsequent autopsy findings, reported by the performing expert.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable.
  6. The type of ground truth used:

    • Animal Testing: Direct observation of the procedure by the performing cardiologist (Dr. Hijazi), and post-mortem autopsy findings to assess septal defect size and any complications. Videotapes and slides were also used as evidence.
  7. The sample size for the training set:

    • Not applicable. The animal testing represents validation of the finished device design.
  8. How the ground truth for the training set was established:

    • Not applicable.

{0}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS DATA

January 2, 1996

K960070

APR - 2 1996

Submitted By: NuMED, Inc., Main St., Hopkinton, NY 12940 (Ph) 315-328-4491

Contact Person: Susan D. Jones

Device Name: NuMED Atrioseptostomy Catheter; Class II

Predicate Devices: Baxter Miller Balloon Atrioseptostomy catheter

Device Description: The NuMED Septostomy Catheter is a new balloon catheter designed for the neonate with congenital heart disease requiring septostomy. It is a dual lumen catheter, 50cm in length with a 13.5mm ± 0.5mm non-compliant balloon on the distal end. The catheter also features an end hole that will accommodate an 0.018" guidewire. The inflated geometry of the balloon at the rated inflation volume of 2cc is a 13.5mm sphere. There is an imaging band under the balloon for balloon positioning in the left atrium. The catheter tip is angled at 35° to facilitate passage through the interarterial opening in the left atrium. To inflate the balloon to it's maximum diameter, 2cc of diluted contrast media is pushed into the balloon extension after purging. Catheters marked every 5cm (up to 50cm) and are supplied with a one way stopcock for balloon sealing.

The non-compliant balloon and dual lumen design of the NuMED Septostomy cathers from other marketed catheters:

  • · The balloon will not deform during the procedure like a latex balloon. The result is a similar diameter opening produced using a smaller diameter balloon.
  • · The non-compliant balloon is heat bonded to the catheter shaft, thus reducing the risk of balloon detachment.
  • · The balloon design facilitates the use of a 5F introducer sheath. This will minimize blood loss in the neonates.
  • · The catheter end hole can be utilized for guidewire, contrast injection, or blood analysis.

Materials and colorants used in this device were approved under IDE G940143 and are the same as those used in previously approved NuMED products: PTA K931009 and PTV G890030.

Biocompatibility Testing

The materials used in the NuMED Atrioseptostomy Balloon Catheter are the same as those used in our PTA (K931009) and PTV Catheters (IDE #G890030) which have been tested for biocompatibility in compliance with the Tripartite Biocompatibility Guidance for Medical Devices.

Test results indicate that all materials demonstrate the biocompatibility of the NuMED catheter and are on file at NuMED, Inc.

{1}------------------------------------------------

Laboratory (Bench) Testing All bench testing was performed in accordance with GMP's and the results are as follows.

Balloon Volume vs. Diameter 1 .

Acceptance Criteria: Balloon diameter must be within ± 10% of the rated diameter at the rated volume of 2cc. (i.e., 12.15mm to 14.85mm)

Justification: ±10% demonstrates the noncompliance of the balloon, and is still well below the maximum inflated diameter of 19.0mm of the predicate device .

Equipment: 3cc syringe, micrometer (CC-0078), water bath @ 98°F ± 5°F, water @ 68°F ± 10°F

Methodology:

Ten full length, sterile catheters with a distally mounted balloon were tested at body temperature - room temperature fluid was injected into the balloon, which was submerged in a body temperature bath. The balloons were inflated to each volume increment and measured with a micrometer (CC-0078).

Results:

InflationVolumeSample1 Dia.Sample2 Dia.Sample3 Dia.Sample4 Dia.Sample5 Dia.Sample6 Dia.Sample7 Dia.Sample8 Dia.Sample9 Dia.Sample10 Dia.
0.5cc7.87.97.88.18.08.28.18.08.17.9
1.0cc10.010.210.110.710.510.810.610.710.610.5
1.5cc11.811.911.812.212.012.112.112.112.012.0
2.0cc13.013.213.113.713.513.613.913.813.613.5
2.5cc13.513.613.613.913.713.913.913.913.713.6
3.0cc******************************

*** Burst

Volume (cc)Mean Diameter (mm)RangeStandard Deviation
0.57.997.8 - 8.20.137
1.010.4710.0 - 10.80.275
1.512.0011.8 - 12.20.133
2.013.4913.0 - 13.90.300
2.513.7313.5 - 13.90.157

2. Bond Integrity

Acceptance Criteria: Minimum of 3 lb. at all test points

Justification: The predicate device latex balloon bond failed at a value of 2.7 Ibs. in laboratory testing, using the same test methodology.

Equipment: Chatillon pull tester (CC-0013, 0-10 lbs.), circular weight, test fixture Methodology:

Ten full length, sterile catheters with a distally mounted balloon were tested. The specified bonds were pulled at room temperature with a Chatillon pull tester.

The balloon bond testing performed simulates an actual use condition. To test the proximal balloon bond and balloon operation, the balloon was used to lift a circular weight until failure. To test the tip bond, catheters were gripped at the balloon and pulled until inner tubing failure. This tip was then gripped and the balloon was pulled until bond failure occurred.

{2}------------------------------------------------

The methodology for balloon bond testing was as follows:

BALLOON BOND

  • [a] Inflate catheter with 2.0cc of fluid.
  • [b] Insert extensions through the weight and slide weight into place over vertically held balloon.
  • [c] Holding the catheter by the shaft, lift the weight off the counter (see Fig. below)

Image /page/2/Figure/6 description: The image shows a diagram of a catheter shaft with a septostomy balloon. A 2lb "doughnut" weight with a 6mm opening is attached to the catheter shaft above the balloon. The diagram also indicates the direction of lift with an arrow labeled "LIFT". The catheter shaft is positioned vertically, with the balloon at the bottom and the weight in the middle.

  • [d] Repeat this testing by adding weight rings until the balloon fails when lifting.
    The additional testing was conducted on these same samples after balloon failure. The balloon was cut to reveal the balloon to catheter tip bond. The balloon was then pulled to 10 lbs. to insure that it would not come off. Since the grip points for this test were the catheter shaft and the broken balloon this test also formulated the tip to catheter bond test. In all cases these bonds exceeded 10 lbs. (see Fig below).

Image /page/2/Figure/9 description: This image shows a diagram of a balloon with labels pointing to different parts of the balloon. The labels include 'GRIP POINT', 'PULL', 'TIP AREA', 'TIP TO BALLOON BOND', and 'BALLOON WITH PROXIMAL BOND REMOVED'. The diagram shows the balloon being pulled from the grip point.

Results:

Test SiteAverageRange
Distal hub to 'Y'> 10.0 lbs.> 10.0 lbs.
Balloon hub to extension> 10.0 lbs.> 10.0 lbs.
Balloon extension to 'Y'connector> 10.0 lbs.> 10.0 lbs.
Catheter body to 'Y' connector> 10.0 lbs.> 10.0 lbs.
Proximal balloon bond8.0 lbs7.0 - 9.0 lbs (std dev. = 1 lb.)
Tip to catheter8.0 lbs7.0-9.0 lbs. (std dev. = 1 lb.)
Tip to balloon> 10.0 lbs.> 10.0 lbs.

{3}------------------------------------------------

3. Guidewire Compatibility

Acceptance Criteria: minimal resistance while pushing catheter over guidewire. Force may not be great enough to bend guidewire.

Justification: This resistance is equal to or less than the resistance of 0.021" guidewires in balloon dilatation catheters of comparable size (i.e., same recommended guidewire size).

Equipment: Micrometer CC-0078; 0.021" guidewires

Methodology: Guidewires of a 0.021" labeled diameter were measured to ensure compliance of ± 0.001". These guidewires were then inserted through 20 full length sterile catheters. The guidewires were gripped at 2" from the guidewire hub and pushed through the catheter tip.

Results:

In no case was the guidewire kinked or was the force in excess of that necessary to insert a guidewire into a similar catheter. NuMED catheters are 100% inspected with the appropriate pin gauge for QC acceptance prior to release for packaging.

ModelnAverageStd. Deviation99.9%ConfidenceRange
2100-0120NANANAAll acceptable

Maximum Luminal Injection Pressure 4.

Acceptance Criteria: Must withstand 500 psi

Tustification: Predicate device does not allow for pressure iniection, this value is equivalent to the currently marketed Berman Angiographic catheter, of the same French size.

Equipment: pressure gauge CC-0115, 0-4000 psi, pressure generating apparatus

Methodology:

Ten sterilized, full length test catheters with balloon attached were used. The catheter tip was plugged (using a stainless steel pin) on each catheter to give a static pressure measurement. Catheters were pressurized to 750 PS1, a value in excess of that normally used in the cath lab, and then pressurized to failure. The following values are the data acquired from this testing. A limit of 600 psi will be added to the injection criteria.

Results:

Sample #750 psi InjectionBurst Pressure
1Pass875 psi
2Pass950 psi
3Pass950 psi
4Pass900 psi
5Pass950 psi
6Pass880 psi
7Pass890 psi
8Pass890 psi
9Pass890 psi
10Pass880 psi
SUMMARYAll Samples Passed 750 psi905.5 psi Average Burst
MeanRangeStandard Deviation99% Confidence
905.5 psi875 - 950 psi31.486741.7 lb.

9

{4}------------------------------------------------

  • Multiple Inflation Testing 5.
    Acceptance Criteria: The balloon diameter must still be within ± 10% after 40 inflation/deflation cycles.

Justification: Industry standard for multiple inflations is 40 cycles.

Equipment: micrometer CC-0078, 3cc syringe, water bath @ 98°F ± 5°F, water @ 68°F ± 10°F

Methodology:

Ten full length catheters were sterilized and inflated with the recommended volume. The test was conducted at body temperature. The test sample was inflated, the O.D. measured, then deflated. All samples were inflated 50 times with no failures or leaks. One sample was inflated 500 times without failure. The diameter, at maximum volume of 2cc, was virtually unchanged from the first to the fiftieth inflation, since the non-compliant balloon was not stressed during inflation. This explains why the one sample inflated 500 times did not fail.

Results:

The raw data for the multiple inflation testing is as follows: All samples were tested 50 times with #10 tested 500 times. Individual diameter measurements were not taken on all 500 inflations of the sample. The following is the statistical analysis of balloon diameter change after 50 inflations. The raw data is enclosed within this section.

Mean: 0.15Carolina Company of ChildrenTHE LEASE CHANAL CHANAL CHARACT CHARACT COLLEGIANand and the contract of the control of the commend of the many of the many of the may be any of the comments of the may------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Range: 0.1 - 0.3with the programment of the desires of the desires of the desiresStandard Deviation: 0.085And Analy of Children Commend College of Children Comments of ChildrenAnd Antiques and Children Comments of Children Comments of Children Comments of Children Comments of Children Comments of Children Comments of Children Comments of Children C
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    1. Injection Rate
      Acceptance Criteria: Demonstrate that this device can be used for fluid injection.

Justification: The predicate device does not have a through-lumen and cannot be used for this purpose.

Equipment: NuMED hydraulic pressure injector: Pressure Gauge CC-0115, 0-4000 psi: water @ 68°F ± 10°F

Methodology:

The iniection rate test was conducted on 10 full length sterilized catheters. The purged catheters were attached to the NuMED hydraulic pressure injector and the injection pressure was set at 300 psi. The injection machine consists of a hydraulic cylinder with a pressure gauge mounted at the fluid outlet. The hydraulic arm exerts a pressure on a metal syringe to expel fluid. This machine was attached directly to the distal port of the catheter and the iniection pressure read 300 psi. The catheters were then subjected to a 5 second pressure injection. The injection fluid was then measured and a cc/sec value was calculated. The injection media used was room temperature water.

{5}------------------------------------------------

Results:

Sample #Qty Fluid InjectedCalculated Rate (cc/sec)
121cc4.2
219cc3.8
321cc4.2
420cc4.0
519cc3.8
618cc3.6
720cc4.0
818cc3.6
922cc4.4
1022cc4.4
MeanRangeStandard Deviation99% Confidence
4cc/sec3.6 to 4.40.2982.45 cc/sec
    1. Luminal Frequency Response
      These catheters are not recommended for use in measuring luminal frequency.
    1. Balloon Deflatability
      Acceptance Criteria: Deflatation achieved in less than 20 seconds.

Justification: As per Dr. Ziyad Hijazi, the maximum allowable deflation time was set at 20 seconds.

Equipment: 3.0cc syringe, stopwatch, silicone sheeting

Methodology:

Deflation testing was performed at NuMED in a simulated use condition in the laboratory. A total of (5) full length, sterile catheters were introduced into a section of silicone sheet, then inflated with a 30% solution of Renographin and water. They were then pulled through the sheet to simulate the septostomy procedure. These catheters were then deflated using the same 3.0cc syringe used for inflation.

Results:

Balloon deflatability was also evaluated by Dr. Hijazi during animal testing performed on 2-6-95. Deflation was achieved by applying negative pressure to the balloon using a 3 cc syringe (the contrast mixture was 30%). Deflation times were 5 - 10 seconds to achieve 50% deflation and all balloons were fully deflated within 15 sec. For a faster deflation time, the physician may use a larger (10 cc) syringe. These steps have been added to the Instructions for Use.

{6}------------------------------------------------

  • Balloon Non-Compliance During Use 9.
    Acceptance Criteria: minimal deformation, as observed visually

Justification: The predicate device exhibits great deformation.

Equipment: N/A

Methodology & Results::

The catheter was used on a deer heart in the R&D lab. The balloon was slowly pulled through the septum and since this is a non-compliant material no deformation was visually observed. The Baxter Miller was then also pulled slowly through the septum of the deer heart and it was visually observed that the balloon diameter reduced in diameter. The size of the hole that was accomplished was similar in each case: 12mm hole using the 13.5mm NuMED balloon, and 1 1.5mm hole using the 18mm Baxter balloon.

Noncompliance of the balloon was also demonstrated and confirmed by Dr. Hijazi on the previously submitted videotape of the animal testing performed (G940143).

    1. Balloon Protector Testing
      Acceptance Criteria: Ability to form balloon to accept a 6F introducer

Justification: This is the label recommended introducer size for this device.

Equipment: Septostomy catheters, balloon protectors

Methodology:

The balloon protectors are made of PTFE (Teflon). It has an I.D. of 0.068" and will form and protect the balloon. Ten Atrioseptostomy catheters were fitted with balloon protectors, sent to sterilization, then opened and tested with a 6F B. Braun introducer upon return.

Results:

All catheters were admitted through the recommended 6F introducer without incident. No resistance, kinking or detrimental effects to the catheter were noted.

    1. Introducer Compatibility
      Acceptance Criteria: Catheter must be admitted through a 6F Introducer

Justification: This is the size of introducer recommended in the device labeling.

Equipment: 6F B. Braun Introducers, Sterile Atrioseptostomy Catheters Methodology:

Ten catheters will be opened and tested for admittance through the recommended introducer upon return from sterilization.

Results:

All catheters were admitted through the recommended 6F introducer without incident. No resistance, kinking or detrimental effects to the catheter were noted. The catheters were then inflated and checked for any signs of damage. None were observed.

Under microscopic examination, the actual profile of the individual components is as follows: Catheter Shaft = 0.052"; Proximal Bond Area = 0.064"; Balloon Area = 0.065". All NuMED catheters are equipped with a 0.068" max. balloon protector/profiling sleeve prior to sterilization. The attachment of this sleeve ensures that the balloon will be admitted through the introducer.

{7}------------------------------------------------

Animal Testing B.

Animal testing was performed on five piglets on February 6, 1995 using finished samples of the NuMED Atrioseptostomy Catheter manufactured as per the design submitted (13.5 mm balloon, 2 cc capacity, 0.018" to 0.021" GW). The following summary was provided by Dr. Ziyad Hijazi (New England Medical Center Pediatric Cardiology) who performed the testing.

"On 2/6/95, piglet #237 weight 3.5 kg was taken to the cath lab. 5F sheath in the RFV percutaneously. A 4F wedge catheter was advanced to the LA (left atrium). An 0.018" guidewire was advanced and the septostomy catheter was exchanged for the other cather over the wire. The balloon was inflated with 1.2 cc 30% mixture of contrast and saline, then the wire was removed and a syringe with mixture of contrast and saline was attached to this end (the distal lumen end). The balloon was pulled across the septum without difficulty. Initially 1.5 cc, then 1.8 cc, without any problem. To assess the result of pulling a fully inflated balloon (2 cc) all the way down to the IVC (inferior vena cava) resulted in perforation of the IVC. Autopsy revealed the heart size to be 35 g, the IVC revealed a laceration at the junction with the right atrium. Inspection of the septal defect size that was created revealed smooth edges, circular shape and it measures 10-12 mm in diameter.

"On 2/6/95, piglet #238 weight 3.5 kg was taken to the cath lab. 5F sheath in the RFV percutaneously. The septostomy catheter was advanced to the LA via the PFO without difficulty and without the need for a guidewire. Position was confirmed with hand injection of contrast/saline mixture in the LA. The balloon was inflated to 1.2 cc, pulling through was easy, then the catheter was re advanced to the LA without difficulty, again position was confirmed by contrast injection, balloon was inflated to 1.5 cc, this pulled through also easily. repeat procedure with a balloon capacity of 1.8 cc and 2.0 cc without any problem. Autopsy revealed heart size is 36 g. Nice size defect measuring 10-11 mm. No other findings.

"On 2-6-95, piglet #235 weight 3.2 kg was taken to the cath lab. 5F sheath in the RFV percutaneously. Septostomy using 1.2, 1.5, 1.8 and 2.0 cc successfully as described above. No problems. On occasions, the balloon could not be advanced through the defect, therefore, a guidewire was used to advance into the LA. Once in LA contrast injection demonstrated the location of the catheter tip in the LA. Autopsy revealed that the heart size is 36 g. The defect size is 11 mm. No other findings.

"On 2/6/95, piglet #239 weight 2.9 kg was taken to the cath lab. 5F sheath in the RFV percutaneously. Septostomy using 1.0, 1.5 and 1.7 cc was successful. 2.0 cc was not tried secondary to the small size of the animal. Autopsy revealed that the heart is 25 g. Good defect size 10 mm.

"On 2/6/95, piglet #236 weight 2.8 kg was taken to the cath lab. 5F sheath in the RFV percutaneously. Septostomy with 1.0, 1.5, 1.7 and 2.0 cc was uneventful. Autopsy revealed heart to be 24.5 g. defect size 11 mm.

" All procedures were recorded on videotape and slides were taken for all animals."

Dr. Hijazi provided us with copies of the slides and videotape which were included with IDE G940143 for this device.

{8}------------------------------------------------

Intended Use: This catheter is intended for balloon atrioseptostomy.

Comparison Information:

MODEL:NUMEDBAXTER MILLER
IndicationsUsed for the palliation of severalcongenital heart defects: transposition ofthe great arteries, total anomalouspulmonary venous drainage withoutpulmonary obstruction, tricuspid atresia,mitral stenosis, mitral atresia, andpulmonary atresia with intact ventricularseptum.Used for the palliation of severalcongenital heart defects: transpositionof the great arteries, total anomalouspulmonary venous drainage withoutpulmonary obstruction, tricuspidatresia, mitral stenosis, mitral atresia,and pulmonary atresia with intactventricular septum.
Introducer:6FR8FR
Shaft Size:5FR5FR
Guidewire Size:0.018" to 0.021"N/A
Usable Length:50cm50cm
Balloon Capacity2.0cc4.0cc
Inflated Diameter:13.5mm19.0mm
Balloon Length:1.35cmNA
Max. Injection Pressure600psiNA
Flow Rate:4cc per secondNA
Tip Angulation35°35°
Materials:Shaft: polyamideBalloon: polyamideImage Band: PlatinumPVC ShaftLatex Balloon
Construction:Dual lumen construction with distallymounted non-compliant balloon. Distallumen open to tip.PVC catheter with distally mountedlatex balloon. Single Lumen with styletfor stiffness.

catheters are marketed for balloon atrioseptostomy. The parameters of the NuMED These catheters are comparable to those of the currently marketed catheters.

§ 870.1200 Diagnostic intravascular catheter.

(a)
Identification. An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.(b)
Classification. Class II (performance standards).