K Number
K955175
Date Cleared
1996-04-05

(144 days)

Product Code
Regulation Number
870.1330
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACS ANCHORTM Exchange Device is used within a guiding catheter to facilitate the exchange of an interventional device while maintaining the position of a guide wire within the vasculature.

Device Description

The ACS ANCHORTM Exchange Device consists of a proximal control handle and a superelastic core wire within a single lumen hypotube approximately 96 cm. long. The control handle provides the userinterface to control the movement of the hypotube and the core wire. During an exchange procedure, the distal end of the core wire is wrapped around the guide wire to lock its position in the vasculature. The ACS ANCHOR™ Exchange Device is compatible with 0.010" through 0.018" diameter non-hydrophilic-coated guide wires and associated interventional devices (except the ROTOBLATOR® Rotational Angioplasty System) when at least 0.015" free space is available within the guiding catheter.

AI/ML Overview

The provided text describes a medical device, the ACS ANCHOR™ Exchange Device, and its comparison to a predicate device. However, it does not include detailed acceptance criteria or a study designed to prove the device meets specific performance criteria in the way a clinical trial for an AI diagnostic device would.

Here's an analysis based on the provided text, addressing your questions where possible:

1. Table of Acceptance Criteria and Reported Device Performance:

The document describes tests performed rather than explicit numerical acceptance criteria. The acceptance is broadly stated as "acceptable for its intended use" or "substantially equivalent."

Acceptance Criteria (Inferred)Reported Device Performance
Biocompatibility: Device must be biocompatible for intended application.Met ISO 10993-1:1992(E) and showed it is biocompatible.
Tensile Strength: Device must withstand tensile forces.Passed tensile tests; results demonstrated acceptability.
Torsional Strength: Device must withstand torsional forces.Passed torsional strength tests; results demonstrated acceptability.
Locking Force: Device must provide sufficient locking force to hold a guidewire.Passed locking force tests; showed substantial equivalence to predicate device.
Fatigue (Cycling): Device must withstand repeated use/stress.Passed fatigue (cycling) tests; results demonstrated acceptability.
In Vivo Performance: Device must perform as intended in a living system with various interventional devices.In vivo testing conducted to substantiate use; implied successful performance.
Sterility: Device must achieve a sterility assurance level (SAL).Sterilization validation achieved SAL (10d).

2. Sample Size for the Test Set and Data Provenance:

The document does not specify exact sample sizes for the "bench" or "in vivo" testing.

  • Bench Testing: No specific numbers provided.
  • In vivo Testing: "In vivo testing was conducted to substantiate the use..." No specific sample size or details on data provenance (e.g., animal model, human cadaver, specific country of origin) are given.
  • Retrospective/Prospective: Not applicable or specified. This device is a mechanical tool, not an AI diagnostic, so these terms don't directly apply in the same way.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

Not applicable. This is a medical device for facilitating procedures, not an AI diagnostic tool requiring expert interpretation of images or data to establish a "ground truth" in the diagnostic sense. The "ground truth" here is functional performance.

4. Adjudication Method for the Test Set:

Not applicable. See point 3.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

Not applicable. An MRMC study is relevant for evaluating the impact of an AI diagnostic tool on human readers' performance. This device is a procedural accessory, not a diagnostic AI.

6. Standalone (Algorithm Only) Performance:

Not applicable. This is a physical medical device, not an algorithm.

7. Type of Ground Truth Used:

The "ground truth" for this device's performance is primarily engineering specifications and functional performance metrics.

  • For biocompatibility: Adherence to ISO standards.
  • For mechanical tests (tensile, torsional, locking, fatigue): Measurement against design specifications and comparison to a predicate device.
  • For in vivo testing: Successful demonstration of intended function within a living system (though details are sparse).
  • For sterility: Achievement of a validated sterility assurance level.

8. Sample Size for the Training Set:

Not applicable. This is a physical device, not an AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable. See point 8.

Summary of what the document does provide:

The document functions as a 510(k) summary, aiming to demonstrate substantial equivalence to a predicate device, the Magnet Exchange Device. The "studies" described are focused on:

  • Biocompatibility against international standards.
  • Bench testing of mechanical properties (tensile, torsional, locking force, fatigue).
  • In vivo testing to show compatibility with various interventional devices.
  • Sterilization validation.

The "acceptance criteria" are implied by successful completion of these tests and the conclusion that the device is "acceptable for its intended use" and "substantially equivalent" to its predicate. It lacks the detailed, quantitative acceptance criteria and rigorous study design typically expected for an AI-powered diagnostic device clinical validation.

{0}------------------------------------------------

K955175

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990.

1.Trade Name:ACS ANCHOR™ Exchange Device
Common Name:Guide Wire Extension/Exchange Device
2.Device Classification:Class II

3. Performance Standards:

Performance standards have not been established under Section 514 of the Federal Food, Drug, and Cosmetic Act for guide wire extension/exchange devices.

4. Device Description:

ACS ANCHOR™ Exchange Device

The ACS ANCHORTM Exchange Device is used within a guiding catheter to facilitate the exchange of an interventional device while maintaining the position of a guide wire within the vasculature. The ACS ANCHOR™ Exchange Device consists of a proximal control handle and a superelastic core wire within a single lumen hypotube approximately 96 cm. long. The control handle provides the userinterface to control the movement of the hypotube and the core wire. During an exchange procedure, the distal end of the core wire is wrapped around the guide wire to lock its position in the vasculature. The ACS ANCHOR™ Exchange Device is compatible with 0.010" through 0.018" diameter non-hydrophilic-coated guide wires and associated interventional devices (except the ROTOBLATOR® Rotational Angioplasty System) when at least 0.015" free space is available within the guiding catheter.

న్న Summary of Substantial Equivalence:

Comparison of the new ANCHOR™ Exchange Device to the predicate device, The Magnet Exchange Device by Scimed Life Systems, Inc. indicates that the ANCHORTM Exchange Device is substantially equivalent to the predicate exchange device. Like the Magnet Exchange Device, the ACS ANCHORTM Exchange Device is used to facilitate the exchange of an interventional device without the use of a guide wire extension or exchange length wire. Both the ACS ANCHOR™ Exchange Device and the Magnet Exchange Device are designed to apply a holding force or "lock" on the guide wire to allow withdrawal or advancement of an interventional device over the guide wire while maintaining the guide wire position in the vasculature.

{1}------------------------------------------------

Bench and in yivo testing have been completed to evaluate the performance of the ACS ANCHOR™ Exchange Device in combination with various interventional devices (e.g., PTCA dilatation catheters, atherectomy catheters, stent delivery systems).

6. Testing Data:

Biocompatibility

Biocompatibility testing was performed on sterilized ACS ANCHOR™ Exchange Devices according to the International Standard - Biological Evaluation of Medical Devices (ISO 10993-1:1992(E)) and showed that the ACS ANCHOR™ Exchange Device is biocompatible and is acceptable for its intended application.

Bench and In Vivo Testing

Bench testing of the ACS ANCHOR™ Exchange Device consisted of the following tests: tensile tests, torsional strength tests, locking force test, and fatigue (cycling) test. The results of the tests demonstrated that the ACS ANCHOR™ Exchange Device is acceptable for its intended use. Comparison of the holding force between the ACS ANCHOR™ Exchange Device and the predicate device, The Magnet Exchange Device showed that the ACS ANCHOR™ Exchange Device is substantially equivalent to this predicate device.

In vivo testing was conducted to substantiate the use of the ACS ANCHOR™ Exchange Device with different interventional devices.

7. Sterilization:

The ACS ANCHOR™ Exchange Device is sterilized by the same methods and following the same parameters as those used for currently marketed ACS guide wires and accessories. Sterilization validation of the device was performed and showed that the sterility assurance level (10d) is achieved.

8. Conclusion:

Based on the intended use and results of the testing, the ACS ANCHOR™ Exchange Device is deemed substantially equivalent to the predicate device, The Magnet Exchange Device by Scimed Life Systems, Inc.

Signed:

Joy Tan


Joy Tan, Regulatory Affairs Coordinator

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.