(58 days)
JMS CAVEO A.V. Fistula Needle Set is intended for temporary cannulation to vascular access for extracorporeal blood treatment for hemodialysis. This device is intended for single use only. The anti-needlestick safety feature aids in prevention of needle stick injuries when removing and discarding the needle after dialysis. The device also has an integrated safety mechanism that is designed to automatically generate a partial occlusion of the internal fluid path and trigger the hemodialysis machine to alarm and shut off if a complete dislodgement of the venous needle from the arm inadvertently occurs. In vitro testing supports that this feature triggers the hemodialysis machine to alarm and shut off.
The subject device is the JMS CAVEO A.V. Fistula Needle Set (CAVEO) with an anti-needlestick safety feature. The Caveo is predicted to protect patients from the risks associated with venous needle dislodgement (VND) based on bench testing results. It contains an integrated stainless steel torsion spring mechanism and bottom footplate that provides an open/fluid path when the AV fistula set is fully cannulated into the access site. When the venous needle becomes completely dislodged from the patient's arm, this mechanism enables the footplate to partially occlude the blood path, generating an increased venous line pressure high enough to trigger automatic alarm and halt further blood pumping of the hemodialysis machine. In vitro testing supports that this feature triggers the hemodialysis machine to alarm and shut off. Based on bench testing results, this may significantly reduce patient blood loss in the event of a complete VND. The Caveo has a pre-attached anti-stick needle guard for prevention of needlestick injury at the time of needle withdraw after completion of a hemodialysis procedure.
In vitro performance testing using dialysis machine Fresenius 2008K supports the function of the Caveo VND feature with a venous pressure limit set to 200mmHg symmetric mode, a maximum dialyzer membrane surface area of 2.5 m2, minimum blood flow rate of 200 mL/min, maximum ultrafiltration rate of 4000 mL/hour, and simulated treatment duration of 8 hours. If different machine and/or setting are used, before introducing the device, refer to Directions for Use.
This document describes the FDA 510(k) clearance for the JMS CAVEO A.V. Fistula Needle Set. This device is a physical medical device, specifically a needle set for hemodialysis, and does not involve Artificial Intelligence (AI). Therefore, many of the requested criteria related to AI/software performance, ground truth establishment, expert adjudication, MRMC studies, and training datasets are not applicable.
The document primarily focuses on bench testing (in vitro performance) and a simulated clinical usability study to demonstrate device safety and effectiveness.
Here's a breakdown based on the provided text:
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria are primarily demonstrated through various performance tests, with "Passed" as the acceptance. The document doesn't explicitly state numerical acceptance thresholds for all tests (e.g., how much "Force to Depress the Footplate" is acceptable), but it implies successful completion. For some, like needle penetration resistance and retraction lock strength, numerical criteria are provided.
Acceptance Criterion (Test) | Reported Device Performance |
---|---|
Needle Penetration Resistance | |
14G | ≤ 40g (Predicate: ≤ 40g) |
15G | ≤ 35g (Predicate: ≤ 35g) |
16G | ≤ 30g (Predicate: ≤ 30g) |
17G | ≤ 30g (Predicate: ≤ 30g) |
Needle Retention Strength | > 6.0kgf (Predicate: > 6.0kgf) |
Needle Surface (Visually) | No dented/damaged needle (Predicate: No dented/damaged needle) |
Product Leak | No air bubble should appear when subjected to air pressure 0.40 kgf/cm2 and immersed in water. (Predicate: Same) |
Needle Retraction Final Lock Strength | ≤ 2.0kgf (Predicate: ≤ 2.0kgf) |
Connector (Air tightness, Luer fit) | Passed (ISO 80369-7 compliant) |
Connection Strength (Tube to Connector/Joint, Tube to Pivot Valve Core) | > 6.0kgf (Predicate: > 3.0kgf for Tube to Connector/Joint, > 6.0kgf for Tube to Hub) |
Leakage by Pressure Decay (Female Luer Lock) | Passed |
Positive Pressure Liquid Leakage (Female Luer Lock) | Passed |
Sub-atmospheric Pressure Air Leakage (Female Luer Lock) | Passed |
Stress Cracking (Female Luer Lock) | Passed |
Resistance to Separation from Axial Load (Female Luer Lock) | Passed |
Resistance to Separation from Unscrewing (Female Luer Lock) | Passed |
Resistance to Overriding (Female Luer Lock) | Passed |
Tube to Connector Pull Test (Female Luer Lock) | Passed |
Luer Lock Cover Open Torque Test (Female Luer Lock) | Passed |
Testing Activation of the Sharps Injury Protection Feature | Passed |
Needle Pushback Strength Test | Passed |
Needle Guard Detachment Strength Test | Passed |
Appearance Check (Caveo) | Passed |
Cover Pull with Hub (Caveo) | Passed |
Air Leak Test (Caveo) | Passed |
Positive Pressure Leak Test (Caveo) | Passed |
Negative Pressure Leak Test (Caveo) | Passed |
Needle Guard Retraction Final Lock Test (Caveo) | Passed |
Tube to Hub Pull Test (Caveo) | Passed |
Cannula to Hub Tensile Test (Caveo) | Passed |
Dimensional Analysis of Footplate to Pivot Valve Core (Caveo) | Passed |
TPE Front & Back Ends Internal Diameter (Y-axis) Measurements (Caveo) | Passed |
TPE Surface Roughness (Caveo) | Passed |
Cannulation at 15 and 45-Degree Angles (Caveo) | Passed |
Occlusion After Taping (Caveo) | Passed |
VND Performance (Venous Needle Dislodgement) | Passed |
Baseline Pressure Comparison (Caveo) | Passed |
Force to Depress the Footplate (Caveo) | Passed |
Mechanical Hemolysis Testing (Caveo) | Passed |
Simulated Clinical Usability Study | Successful |
Transportation Test | Passed |
Human Factors Testing | Passed |
Biocompatibility (Cytotoxicity, Sensitization, Irritation, Hemocompatibility, Pyrogenicity, Acute Systemic Toxicity, Subacute Toxicity, Genotoxicity) | Passed (for all, per ISO 10993 standards) |
2. Sample Size for Test Set and Data Provenance
- Test Set (Clinical Trial): 15 subjects (2 females, 13 males).
- Data Provenance: The document does not explicitly state the country of origin. It describes the recruitment from "the general hemodialysis population," and mentions racial/ethnic demographics, but not geographic. Given the company is "JMS North America Corporation" (Hayward, CA), it is highly probable the study was conducted in the US. The study appears to be prospective as it involves recruitment and device use to confirm safety, performance, and usability.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This question is largely not applicable as the device is a physical medical device. The "ground truth" for performance is established through bench testing (objective physical measurements) and the success of the device in a simulated clinical setting. There is no mention of human experts establishing a "ground truth" for diagnostic or AI-related interpretations.
For the simulated clinical usability study, the "ground truth" is whether the device performed as intended and was usable, as observed by clinicians/researchers during the study. The qualifications of those assessing the usability are not specified, beyond the implication that they are competent to conduct a clinical trial for hemodialysis devices.
4. Adjudication Method for the Test Set
Adjudication methods (like 2+1, 3+1) are typically used in studies involving human interpretation (e.g., radiology reads) where there might be disagreement in expert opinions needing a tie-breaker. This is not applicable here as:
- The primary "test set" involves objective performance characteristics (bench testing).
- The clinical usability study likely involved observing successful function and user feedback, not a diagnostic interpretation needing adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No. An MRMC study is relevant for evaluating the performance of AI systems or diagnostic tools where multiple human readers interpret cases, often with and without AI assistance, to see if the AI improves human performance. This device is a physical hemodialysis needle set, not an AI or diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This refers to the performance of an AI algorithm on its own. The device is a physical product. Its "standalone" performance is assessed through bench testing (e.g., VND Performance
, Mechanical Hemolysis Testing
, Needle Penetration Resistance
). These tests evaluate the device's inherent functional characteristics independent of human interaction during the critical failure modes.
7. The Type of Ground Truth Used
- Bench Testing: The ground truth is based on objective physical measurements and engineering specifications (e.g., force measurements, leak tests, dimensional analyses) and functional success/failure (e.g., did the VND feature trigger the alarm?).
- Simulated Clinical Usability Study: The ground truth is based on observed device performance during simulated use and the successful delivery of hemodialysis without impedance by the device's novel features. This is akin to outcomes data in a controlled simulated environment.
8. The Sample Size for the Training Set
Not applicable. The device is a physical product and does not involve AI or machine learning models that require a "training set."
9. How the Ground Truth for the Training Set was Established
Not applicable. As there is no training set for an AI model.
In summary, the provided document details the non-clinical and limited clinical testing of a physical medical device (hemodialysis needle set). The acceptance criteria are largely met through rigorous bench testing demonstrating physical and functional robustness, and a small simulated clinical study confirming usability and safety in a controlled environment. AI-specific criteria are not relevant to this type of device.
§ 876.5540 Blood access device and accessories.
(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.