K Number
K142564
Date Cleared
2014-12-23

(103 days)

Product Code
Regulation Number
876.5540
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Use for temporary cannulation non-implantable, less than 30 days) to vascular access for extra corporeal blood treatment. The device is intended for single use only and has an anti-stick feature integrated as part of the Needle Set which aids in prevention of needle-stick injuries

Device Description

SysLoc® MINI (V4) is intended as non-implanted blood access device, which consists of flexible tube and needle with integrated sharp safety features as described in 21 CFR 876.5540. SysLoc® MINI (V4) comes with a rotational feature and the needle is retracted with the wing sheath after deliberate release of secured external lock, and final locking is assured by an audible 'click' sound when the hub/tube is pulled rearwards.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device (JMS SysLoc® MINI A.V. Fistula Needle Set (V4) and JMS SysLoc® MINI Apheresis Needle Set (V4)). This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets acceptance criteria through a traditional clinical study with performance metrics in terms of sensitivity, specificity, or similar.

Therefore, the requested information elements related to diagnostic performance (e.g., sample size for test set, number of experts, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to this type of regulatory submission and the information provided. The "acceptance criteria" here refer to testing to ensure the device performs as intended and is safe, similar to the predicate, not diagnostic accuracy.

Here's a breakdown of the available information based on your request, with an emphasis on what is not present in this type of document:

1. Table of acceptance criteria and the reported device performance

The document does not specify "acceptance criteria" in terms of diagnostic performance metrics. Instead, it refers to "Performance Specification" and "Bench testing" to demonstrate that the new device (V4) performs as intended and is safe and effective, similar to the predicate (V3).

Acceptance Criteria (Implicit from Predicate V3)Reported Device Performance (JMS SysLoc® MINI (V4))
Require very minimum force when retraction and needle is locked after use.Remains the same as (V3) (Implied: performs with minimum retraction force).
Sheath will not be punctured by needle when properly used.Remains the same as (V3) (Implied: sheath integrity is maintained).
Able to be operated by one hand and two-handed.Remains the same as (V3) (Implied: operable by one or two hands).
Anti-stick feature integrated as part of the Needle Set which aids in prevention of needle-stick injuries. (Technical Characteristic of Anti-stick: Retract needle into Wing Sheath after releasing external lock. Needle slides into the sheath and locks in position to prevent needle stick)Remains the same as (V3) (Implied: anti-stick mechanism functions identically).

2. Sample size used for the test set and the data provenance

  • Not Applicable. This document describes bench testing to demonstrate substantial equivalence, not a clinical study to evaluate diagnostic performance. Therefore, there is no "test set" in the context of diagnostic data. The number of devices tested in bench testing is not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. Ground truth for diagnostic accuracy is not relevant here.

4. Adjudication method for the test set

  • Not Applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is a medical device (fistula needle set), not an AI-powered diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not Applicable. This is not an algorithm.

7. The type of ground truth used

  • Not Applicable. Ground truth for diagnostic accuracy is not relevant here. The "ground truth" for this device would be its physical and functional specifications meeting design requirements and demonstrating safety and effectiveness through mechanical and material testing, as well as biocompatibility, which are confirmed to be similar to the predicate.

8. The sample size for the training set

  • Not Applicable. This is not a machine learning or AI device.

9. How the ground truth for the training set was established

  • Not Applicable.

Summary of focus (based on the provided document):

This 510(k) submission primarily focuses on demonstrating that the modified device (V4) is substantially equivalent to a legally marketed predicate device (V3). This is achieved by showing that the V4 device has:

  • The same intended use as the V3.
  • Similar technological characteristics (e.g., anti-stick feature, physical specifications, performance specifications) to the V3.
  • Passed bench testing to verify that it performs as intended as a safe and effective medical device.
  • Satisfied biocompatibility and sterilization requirements comparable to the V3.

The "study" that proves the device meets (implicit) acceptance criteria is the bench testing and other evaluations (biocompatibility, sterilization) which concluded that the "SysLoc® MINI (V4) device is performing as intended to be a safe and effective medical device" and is "substantially equivalent" to the predicate. Specific details about the methodology and sample sizes of this bench testing are not provided in this summary document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, resembling a bird-like shape.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 23, 2014

JMS North America Corporation Sho Hosoki Regulatory 22320 Foothill Blvd., Suite 350 Hayward, CA 94541

Re: K142564

Trade/Device Name: JMS SysLoc® MINI A.V. Fistula Needle Set (V4) JMS SysLoc® MINI Apheresis Needle Set (V4) Regulation Number: 21 CFR§ 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: FIE Dated: November 20, 2014 Received: December 1, 2014

Dear Sho Hosoki,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Image /page/1/Picture/6 description: The image shows the name "Herbert P. Lerner -S" in a large, sans-serif font. The text is black and appears to be centered on a white background. The letters are evenly spaced and the text is easy to read.

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Section 04: Indications for Use

510(K) Number (if known): K142564
Device Name:JMS SysLoc® MINI A.V.Fistula Needle Set (V4)JMS SysLoc® MINI Apheresis Needle Set (V4)
Indications for Use:Use for temporary cannulation non-implantable, less than 30 days) tovascular access for extra corporeal blood treatment. The device isintended for single use only and has an anti-stick feature integrated aspart of the Needle Set which aids in prevention of needle-stick injuries

Prescription Use ______________________________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

Section 05: 510(k) Summary [807.92(c)]

Owner [807.92(a)1]:

Company Name:JMS North America Corp.
Company Address:22320 Foothill Blvd., Suite 350Hayward, CA 94541
Telephone:510-888-9090
Fax:510-888-9099
Contact Person:Sho Hosoki

Summary Preparation Date: 08/01/2014

Device of Submission [807.92(a)(2]]

Classification Name:Needle, FistulaIntravascular Catheter
Proprietary Name:JMS SysLoc® MINI A.V. Fistula Needle SetJMS SysLoc® MINI Apheresis Needle Set
Classification:Class II
Product Code:FIE, FOZ
Code of Federal Regulations:21 CFR 876.5540

Note: Same trade name is to be used for both modified and predicate device, thus for clearer differentiation, the modified device will be denoted as SysLoc® MINI (V4) and the predicate device as SysLoc® MINI (V3).

Predicate Device [807.92(a)(3)]

K NumberProductCompany
K110157JMS SysLoc® MINI A.V.Fistula Needle Set & JMSSysLoc® MINI ApheresisNeedle Set (V3)JMS North AmericaCorporation

Device Description [807.92(a)(4)]

SysLoc® MINI (V4) is intended as non-implanted blood access device, which consists of flexible tube and needle with integrated sharp safety features as described in 21 CFR 876.5540. SysLoc® MINI (V4) comes with a rotational feature and the needle is retracted with the wing sheath after deliberate release of secured external lock, and final locking is assured by an audible 'click' sound when the hub/tube is pulled rearwards.

Intended Use [807.92(a)(5)]

SysLoc® MINI (V4) is intended for temporary cannulation (non-implantable, less than 30 days) to vascular access for extracorporeal blood treatment. The device is intended for single use only and has an anti-stick feature integrated as part of the Needle Set which aids in prevention of needle-stick injuries.

{4}------------------------------------------------

Section 05: 510(k) Summary [807.92(c)]

Predicate Comparison [807.92(a)(6)] & [807.92(b)(1)]

Predicate Comparison [807.92(a)(6)] & [807.92(b)(1)]
Modified DevicePredicate
Device NameSysLoc® MINI (V4).SysLoc® MINI (V3).
510(k) #---K110157
ClassificationClass IIClass II
Intended UseRemains the same as (V3).SysLoc® MINI (V3) is intended for temporary cannulation (non-implantable, less than 30 days) to vascular access for extracorporeal blood treatment. The device is intended for single use only and has an anti-stick feature integrated as part of the Needle Set which aids in prevention of needle-stick injuries.
Technical Characteristic of Anti-stickRemains the same as (V3).Retract needle into SysLoc® MINI (V3) Wing Sheath after releasing external lock. Needle slides into the sheath and locks in position to prevent needle stick
Physical SpecificationRemains the same as (V3) except: Reduction in filter paper of PE pouchSysLoc® MINI (V3) (with safety feature) is integrated into the wing design. There is no pre-attached component to the device that may interfere the dialysis session.
MaterialSee Section 15 for material comparison
LabelingSee Section 13 for Labeling
BiocompatibilitySee Section 15 for biocompatibility
SterilizationETO sterilization
Performance SpecificationRemains the same as (V3)ETO sterilizationSysLoc® MINI (V3) device required very minimum force when retraction and needle is locked after use. SysLoc® MINI (V3) sheath will not be punctured by needle when properly used. SysLoc® MINI (V3) device is able to be operated by one hand and two-handed

Conclusion [807.92(b)(3)]

SysLoc® MINI (V4) has the same intended usage as the predicate device SysLoc® MINI (V3) K110157. Bench testing was conducted to verify that the SysLoc® MINI (V4) device is performing as intended to be a safe and effective medical device. Evaluation data and reports are enclosed within this submission document. The information provided in this submission clearly demonstrates the substantial equivalence of SysLoo® MINI (V4) to the predicate device JMS SysLoc® MINI A.V. Fistula Needle Set (V3) & JMS SysLoc® MINI Apheresis Needle Set (V3) K110157.

§ 876.5540 Blood access device and accessories.

(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.