(84 days)
The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Gloves were tested for use with Chemotherapy drugs, Fentanyl Citrate, Simulated Gastric acid and Xylazine in Fentanyl Citrate in accordance with ASTM D6978-05 Standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.
Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate is a Class I patient examination glove and specialty chemotherapy gloves, that made from synthetic nitrile latex. They are pink color, non-sterile, powder-free, fingertip textured, ambidextrous with beaded cuff, and single use only, and come in different sizes- XS, S, M, L, XL and XXL.
The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been been tested for biocompatibility and permeability to chemotherapy drugs, Fentanyl Citrate, Gastric Acid and Xylazine in Fentanyl Citrate.
This document describes the acceptance criteria and the study that proves the device meets the acceptance criteria for a medical device submitted for FDA 510(k) clearance.
The device in question is Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate.
Acceptance Criteria and Device Performance Study
The acceptance criteria for this device are primarily based on established industry standards for medical examination gloves and specific permeation resistance tests for various chemical substances. The study conducted to prove the device meets these criteria is a series of non-clinical performance tests.
1. Table of Acceptance Criteria and Reported Device Performance
| Test Performed | Methodology | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Freedom From Holes | ASTM D6319-19, ASTM D5151-19 | Meet requirement inspection level G-1, AQL 2.5 (ISO2859-1) | Pass |
| Dimension - Length | ASTM D6319-19 | Minimum 220mm for size XS-S Minimum 230mm for size M-XXL | Pass |
| Dimension - Width | ASTM D6319-19 | XS: 70±10mm S: 80±10mm M: 95±10mm L: 110±10mm XL: 120±10mm XXL: 130±10mm | Pass |
| Dimension - Thickness | ASTM D6319-19 | Finger: 0.05mm (min) Palm: 0.05mm (min) | Pass |
| Physical Properties | ASTM D6319-19, ASTM D412-16 | Before aging: Tensile Strength (Min 14 Mpa) Elongation (Min 500%) After aging: Tensile Strength (Min 14 Mpa) Elongation (Min 400%) | Pass (All criteria met) |
| Powder Residue | ASTM D6319-19, ASTM D6124-06 | Not more than 2 mg per glove | Pass |
| Skin Irritation (Biocompatibility) | ISO10993-23:2021 | Under the conditions of the study, not a primary skin irritant | Pass (Not a primary skin irritant) |
| Skin Sensitization (Biocompatibility) | ISO10993-10:2021 | Under the conditions of the study, not a contact sensitizer | Pass (Not a contact sensitizer) |
| Acute Systemic Toxicity (Biocompatibility) | ISO 10993-11:2017 | Under the conditions of the study, no signs of acute systemic toxicity were observed | Pass (No signs of acute systemic toxicity were observed) |
| Chemotherapy Drugs Permeation | ASTM D6978-05 (2023) | Minimum Breakthrough Detection Time (minutes) Carboplatin: >240 Carmustine: Varies (15.3 min reported) Cisplatin: >240 Cyclophosphamide: >240 Cytarabine HCl: >240 Dacarbazine: >240 Daunorubicin HCl: >240 Doxorubicin HCl: >240 Etoposide: >240 5-Fluorouracil: >240 Gemcitabine HCl: >240 Ifosfamide: >240 Irinotecan HCl: >240 Mechlorethamine HCl: >240 Melphalan HCl: >240 Methotrexate: >240 Mitomycin-C: >240 Mitoxantrone HCl: >240 Pacilitaxel: >240 Thiotepa: Varies (28.2 min reported) | Carboplatin: >240 min. Carmustine: 15.3 min. Cisplatin: >240 min. Cyclophosphamide: >240 min. Cytarabine HCl: >240 min. Dacarbazine: >240 min. Daunorubicin HCl: >240 min. Doxorubicin HCl: >240 min. Etoposide: >240 min. 5-Fluorouracil: >240 min. Gemcitabine HCl: >240 min. Ifosfamide: >240 min. Irinotecan HCl: >240 min. Mechlorethamine HCl: >240 min. Melphalan HCl: >240 min. Methotrexate: >240 min. Mitomycin-C: >240 min. Mitoxantrone HCl: >240 min. Pacilitaxel: >240 min. Thiotepa: 28.2 min. Note: The document explicitly states that Carmustine and Thiotepa have "extremely low permeation times" and are accompanied by a "Warning: Do not use with Carmustine and Thiotepa," indicating that the stated permeation times for these two drugs are not >240 minutes and are accepted for the specified usage. |
| Fentanyl, Gastric Acid, Xylazine Permeation | ASTM D6978-05 (2023) | Minimum Breakthrough Detection Time (minutes) Fentanyl Citrate Injection: >240 Simulated Gastric Acid: >240 Fentanyl Citrate:Xylazine HCl (50:50): >240 | Fentanyl Citrate Injection: >240 min. Simulated Gastric Acid: >240 min. Fentanyl Citrate (50mcg/ml): Xylazine HCl (100mg/ml), 50:50: >240 min. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample sizes (number of gloves) used for each individual test. However, it indicates that the tests were conducted according to established ASTM and ISO standards, which typically specify appropriate sample sizes for such tests to ensure statistical validity.
The data provenance is through non-clinical testing performed in accordance with recognized international and national standards (ASTM, ISO). This implies the data were prospectively generated for the purpose of this submission. The country of origin of the data is not specified, but the use of international standards suggests a globally accepted testing methodology.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable (N/A) as the device is a physical product (medical glove) and the performance is evaluated through objective, standardized laboratory tests (e.g., measuring permeation time, tensile strength, dimensions) rather than expert interpretation of images or clinical data. There is no "ground truth" established by human experts in the context of diagnostic accuracy for this type of device.
4. Adjudication Method for the Test Set
This information is not applicable (N/A) for the same reason as above. Adjudication methods like 2+1 or 3+1 are relevant for studies involving human interpretation (e.g., radiology reads) where discrepancies need to be resolved. The tests performed for this device are objective measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable (N/A). This document pertains to the clearance of a physical medical device (gloves), not an AI algorithm or a diagnostic imaging device that would involve human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
This information is not applicable (N/A). As mentioned, this is a physical medical device, not a software algorithm.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is established by objective, quantitative measurements obtained through standardized laboratory testing methodologies (ASTM and ISO standards). For example:
- Permeation: Measured breakthrough time of specific chemicals through the glove material.
- Physical Properties: Measured tensile strength, elongation, length, width, and thickness.
- Biocompatibility: Observed biological responses (e.g., irritation, sensitization, systemic toxicity) in controlled in-vitro or in-vivo (animal) tests as per ISO standards.
This is not based on expert consensus, pathology, or outcomes data in the clinical sense, but rather on direct physical and chemical testing.
8. The Sample Size for the Training Set
This information is not applicable (N/A). This device is a manufactured product that undergoes performance and biocompatibility testing. There is no "training set" in the context of machine learning. The manufacturing process is designed and validated to consistently produce gloves meeting the specified criteria.
9. How the Ground Truth for the Training Set was Established
This information is not applicable (N/A) as there is no training set for this device. The "ground truth" for the device's design and manufacturing is derived from the established standards (ASTM, ISO) that define the performance characteristics expected of such gloves. Quality control during manufacturing ensures conformity to these standards.
FDA 510(k) Clearance Letter - K251716
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.00
August 27, 2025
Basic Medical Technology Inc.
John Zhao
General Manager
5300 Concours Street
Ontario, California 91764
Re: K251716
Trade/Device Name: Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate
Regulation Number: 21 CFR 880.6250
Regulation Name: Non-Powdered Patient Examination Glove
Regulatory Class: Class I, reserved
Product Code: LZC, LZA, OPJ, QDO
Dated: June 4, 2025
Received: June 4, 2025
Dear John Zhao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K251716 - John Zhao
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K251716 - John Zhao
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ALLAN GUAN -S
For Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K251716
Device Name: Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate
Indications for Use (Describe)
The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Gloves were tested for use with Chemotherapy drugs, Fentanyl Citrate, Simulated Gastric acid and Xylazine in Fentanyl Citrate in accordance with ASTM D6978-05 Standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.
Chemotherapy Drugs - Minimum Breakthrough Detection Time
| Drug | Concentration | Time |
|---|---|---|
| Carboplatin | 10 mg/ml | >240 min. |
| Carmustine | 3.3 mg/ml | 15.3min. |
| Cisplatin | 1 mg/ml | >240 min. |
| Cyclophosphamide | 20 mg/ml | >240 min. |
| Cytarabine HCl | 100 mg/ml | >240 min. |
| Dacarbazine | 10 mg/ml | >240 min. |
| Daunorubicin HCl | 5 mg/ml | >240 min. |
| Doxorubicin HCl | 2 mg/ml | >240 min. |
| Etoposide | 20 mg/ml | >240 min. |
| 5-Fluorouracil | 50 mg/ml | >240 min. |
| Gemcitabine HCl | 38 mg/ml | >240 min. |
| Ifosfamide | 50 mg/ml | >240 min. |
| Irinotecan HCl | 20 mg/ml | >240 min. |
| Mechlorethamine HC1 | 1 mg/ml | >240 min. |
| Melphalan HCl | 5 mg/ml | >240 min. |
| Methotrexate | 25 mg/ml | >240 min. |
| Mitomycin-C | 0.5 mg/ml | >240 min. |
| Mitoxantrone HC1 | 2 mg/ml | >240 min. |
| Pacilitaxel | 6 mg/ml | >240 min. |
| Thiotepa | 10 mg/ml | 28.2 min. |
Fentanyl, Gastric Acid and Xylazine - Minimum Breakthrough Detection Time
| Substance | Time |
|---|---|
| Fentanyl Citrate Injection, 50mcg/ml | >240 min. |
| Simulated Gastric Acid | >240 min. |
| Fentanyl Citrate (50mcg/ml): Xylazine HCl(100mg/ml), 50:50 | >240 min. |
*Please note that the following drugs have extremely low permeation times:
Carmustine: 15.3 minutes, Thiotepa: 28.2 minutes
Warning: Do not use with Carmustine and Thiotepa.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D) ☒ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/23) Page 1 of 2
Page 5
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K251716
Device Name: Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate
Indications for Use (Describe)
The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Gloves were tested for use with Chemotherapy drugs, Fentanyl Citrate, Simulated Gastric acid and Xylazine in Fentanyl Citrate in accordance with ASTM D6978-05 Standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.
Chemotherapy Drugs - Minimum Breakthrough Detection Time
| Drug | Concentration | Time |
|---|---|---|
| Carboplatin | 10 mg/ml | >240 min. |
| Carmustine | 3.3 mg/ml | 15.3min. |
| Cisplatin | 1 mg/ml | >240 min. |
| Cyclophosphamide | 20 mg/ml | >240 min. |
| Cytarabine HCl | 100 mg/ml | >240 min. |
| Dacarbazine | 10 mg/ml | >240 min. |
| Daunorubicin HCl | 5 mg/ml | >240 min. |
| Doxorubicin HCl | 2 mg/ml | >240 min. |
| Etoposide | 20 mg/ml | >240 min. |
| 5-Fluorouracil | 50 mg/ml | >240 min. |
| Gemcitabine HCl | 38 mg/ml | >240 min. |
| Ifosfamide | 50 mg/ml | >240 min. |
| Irinotecan HCl | 20 mg/ml | >240 min. |
| Mechlorethamine HCl | 1 mg/ml | >240 min. |
| Melphalan HCl | 5 mg/ml | >240 min. |
| Methotrexate | 25 mg/ml | >240 min. |
| Mitomycin-C | 0.5 mg/ml | >240 min. |
| Mitoxantrone HCl | 2 mg/ml | >240 min. |
| Pacilitaxel | 6 mg/ml | >240 min. |
| Thiotepa | 10 mg/ml | 28.2 min. |
Fentanyl, Gastric Acid and Xylazine - Minimum Breakthrough Detection Time
| Substance | Time |
|---|---|
| Fentanyl Citrate Injection, 50mcg/ml | >240 min. |
| Simulated Gastric Acid | >240 min. |
| Fentanyl Citrate (50mcg/ml): Xylazine HCl(100mg/ml), 50:50 | >240 min. |
*Please note that the following drugs have extremely low permeation times:
Carmustine: 15.3 minutes, Thiotepa: 28.2 minutes
Warning: Do not use with Carmustine and Thiotepa.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D) ☒ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23) Page 2 of 2
Page 6
Basic Medical Technology Inc.
510(K) Summary
1. Submission Information
Date Prepared: August 27, 2025
Submission Format: Traditional 510(k)
510(K)#: K251716
2. Submitter Information
Applicant Name: Basic Medical Technology Inc.
Address: 5300 Concours Ontario, CA 91764
Contact Person: John Zhao
Tel: (909) 980-1678
Email: hotmailtojohn@yahoo.com
3. Device Information
Trade Name: Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate
Common Name: Nitrile Powder-Free Exam Gloves
Classification Name: Non-Powdered Patient Examination Glove
Regulation: 21 CFR 880.6250
Product Code: LZA, LZC, QDO, OPJ
Classification Panel: General Hospital
Device Class: Class I
4. Predicate Device Information
Primary Predicate Device:
Applicant: Pt. Maja Agung Latexindo
510(k) #: K212408
Trade Name: Powder Free Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs, Simulated Gastric Acid, and Fentanyl Citrate)
Reference Device:
Applicant: Basic Medical Technology Inc.
510(k) #: K250630
Page 7
Trade Name: Nitrile Powder-Free Exam Gloves with Hyaluronic Acid, and tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate
5. Device Description:
Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate is a Class I patient examination glove and specialty chemotherapy gloves, that made from synthetic nitrile latex. They are pink color, non-sterile, powder-free, fingertip textured, ambidextrous with beaded cuff, and single use only, and come in different sizes- XS, S, M, L, XL and XXL.
The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs, Fentanyl Citrate, Gastric Acid and Xylazine in Fentanyl Citrate.
6. Indications for Use:
The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Gloves were tested for use with Chemotherapy drugs, Fentanyl Citrate, Simulated Gastric acid and Xylazine in Fentanyl Citrate in accordance with ASTM D6978-05 Standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.
| NO. | Chemotherapy Drugs | Minimum Breakthrough Detection Time |
|---|---|---|
| 1 | Carboplatin, 10 mg/ml | >240 min. |
| 2 | Carmustine, 3.3 mg/ml | 15.3 min. |
| 3 | Cisplatin, 1 mg/ml | >240 min. |
| 4 | Cyclophosphamide, 20 mg/ml | >240 min. |
| 5 | Cytarabine HCl, 100 mg/ml | >240 min. |
| 6 | Dacarbazine, 10 mg/ml | >240 min. |
| 7 | Daunorubicin HCl, 5 mg/ml | >240 min. |
| 8 | Doxorubicin HCl, 2 mg/ml | >240 min. |
| 9 | Etoposide, 20 mg/ml | >240 min. |
| 10 | 5-Fluorouracil, 50 mg/ml | >240 min. |
| 11 | Gemcitabine HCl, 38 mg/ml | >240 min. |
| 12 | Ifosfamide, 50 mg/ml | >240 min. |
| 13 | Irinotecan HCl, 20 mg/ml | >240 min. |
| 14 | Mechlorethamine HCl, 1 mg/ml | >240 min. |
| 15 | Melphalan HCl, 5 mg/ml | >240 min. |
| 16 | Methotrexate, 25 mg/ml | >240 min. |
| 17 | Mitomycin-C, 0.5 mg/ml | >240 min. |
| 18 | Mitoxantrone HCl, 2 mg/ml | >240 min. |
| 19 | Pacilitaxel, 6 mg/ml | >240 min. |
| 20 | Thiotepa, 10 mg/ml | 28.2 min. |
| NO. | Fentanyl, Gastric Acid and Xylazine | Minimum Breakthrough Detection Time |
|---|
Page 8
| 1 | Fentanyl Citrate Injection,50mcg/ml | >240 min. |
| 2 | Gastric Acid | >240 min. |
| 3 | Fentanyl Citrate (50mcg/ml): Xylazine HCl(100mg/ml), 50:50 | >240 min. |
*Please note that the following drugs have extremely low permeation times:
Carmustine: 15.3 minutes, Thiotepa: 28.2 minutes
Warning: Do not use with Carmustine and Thiotepa.
7. Comparison of Subject Device and Predicate Device:
General Comparison Table:
| Device | Proposed Device | Predicate Device | Reference Device | Result |
|---|---|---|---|---|
| 510K # | K251716 | K212408 | K250630 | Different |
| Applicant Name | Basic Medical Technology Inc. | Pt. Maja Agung Latexindo | Basic Medical Technology Inc. | Same as K250630 |
| Product Name | Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate | Powder Free Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs, Simulated Gastric Acid, and Fentanyl Citrate) | Nitrile Powder-Free Exam Gloves with Hyaluronic Acid, and tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate | Different |
| Product Code | LZA, LZC, QDO, OPJ | LZA, LZC, QDO, OPJ | LZA, LZC, QDO, OPJ | Same |
| Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | 21 CFR 880.6250 | Same |
| Device Classification | Class I | Class I | Class I | Same |
| Indications for use | The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves were tested for use with Chemotherapy drugs, Fentanyl Citrate, Simulated Gastric acid and Xylazine in Fentanyl Citrate in accordance with ASTM D6978-05 Standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs. | Powder Free Nitrile Examination Glove is a disposable device intended for medical purpose that is worn on examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs, Simulated Gastric Acid, and Fentanyl Citrate as per ASTM D6978-05(2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. | Nitrile Powder-Free Exam Gloves with Hyaluronic Acid, and tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves were tested for use with Chemotherapy drugs, Fentanyl, Gastric acid and Xylazine in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs. | Same |
| Powder free | Yes | Yes | Yes | Same |
Page 9
| Material | Nitrile | Nitrile | Nitrile | Same |
| Color | Pink | Blue | Blue | Different |
| Size | XS, S, M, L, XL, XXL | XS, S, M, L, XL, XXL | XS, S, M, L, XL, XXL | Same |
| Sterile | Non-Sterile | Non-Sterile | Non-Sterile | Same |
| Single Use | Single use | Single use | Single use | Same |
| Design feature | Ambidextrous | Ambidextrous | Ambidextrous | Same |
| Other | N/A | N/A | Hyaluronic Acid coated on inside | Different |
| Chemo Drugs Claim | See below comparison table | See below comparison table | See below comparison table | Different |
Analysis: * The finished subject device has been tested with performance and Biocompatibility, all the test results meet the requirements, so the difference of color and other does not raise questions of safety and effectiveness.
Technological Characteristic Comparison Table:
| Characteristics | Proposed device | Predicate device K212408 | Reference Device K250630 | Result |
|---|---|---|---|---|
| Dimension Reference standard | ASTM D6319-19 | Same | ||
| Length | Min 220mm for XS, S Min 230mm for M, L, XL, XXL | Min 220mm for XS, S Min 230mm for M, L, XL, XXL | Min 220mm for XS, S Min 230mm for M, L, XL, XXL | |
| Width | XS: 70±10 mm S: 80±10 mm M: 95±10 mm L:110±10 mm XL:120±10 mm XXL:130±10 mm | XS: 70±10 mm S: 80±10 mm M: 95±10 mm L: 110±10 mm XL:120±10 mm XXL:130±10 mm | XS: 70±10 mm S: 80±10 mm M: 95±10 mm L:110±10 mm XL:120±10 mm XXL:130±10 mm | |
| Thickness | Palm: Min 0.05 mm Finger: Min 0.05 mm | Palm: Min 0.05 mm Finger: Min 0.05 mm | Palm: Min 0.05 mm Finger: Min 0.05 mm | |
| Physical properties | ASTM D6319-19 ASTM D412-16 | Same | ||
| Before aging | ||||
| Tensile strength | Min 14MPa | Min 14MPa | Min 14MPa | |
| Ultimate elongation | Min 500% | Min 500% | Min 500% | |
| After aging | ||||
| Tensile strength | Min 14MPa | Min 14MPa | Min 14MPa | |
| Ultimate elongation | Min 400% | Min 400% | Min 400% | |
| Freedom from holes | ASTM D6319-19 ASTM D5151-19 | Same | ||
| G-I, AQL2.5 | G-I, AQL2.5 | G-I, AQL2.5 | ||
| Residual Powder | ASTM D6319-19 ASTM D6124-06 | Same | ||
| ≤ 2 mg per glove | ≤ 2 mg per glove | ≤ 2 mg per glove | ||
| Biocompatibility | Same | |||
| Irritation | Not a primary skin irritant | Not a primary skin irritant | Not a primary skin irritant |
Page 10
| Sensitization | Not a primary skin sensitizer | Not a primary skin sensitizer | Not a primary skin sensitizer | Same |
| Acute Systemic Toxicity Test | Not toxic | Not toxic | Not toxic | Same |
Chemotherapy Drugs Comparison Claim:
| Chemotherapy Drugs | Minimum Breakthrough Detection Time | Result of comparison | ||
|---|---|---|---|---|
| Proposed device | Predicate device K212408 | Reference Device K250630 | ||
| Carboplatin, 10 mg/ml | >240 min. | >240 min | >240 min | Same |
| Carmustine, 3.3 mg/ml | 15.3 min. | 58.3 min | 11.6 min | Different |
| Cisplatin, 1 mg/ml | >240 min. | >240 min | >240 min | Same |
| Cyclophosphamide, 20 mg/ml | >240 min. | >240 min | >240 min | Same |
| Cytarabine HCl, 100 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Dacarbazine, 10 mg/ml | >240 min. | >240 min | >240 min | Same |
| Daunorubicin HCl, 5 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Docetaxel, 10 mg/ml | / | >240 min | / | Different |
| Doxorubicin HCl, 2 mg/ml | >240 min. | >240 min | >240 min | Same |
| Etoposide, 20 mg/ml | >240 min. | >240 min | >240 min | Same |
| 5-Fluorouracil, 50 mg/ml | >240 min. | >240 min | >240 min | Same |
| Gemcitabine HCl, 38 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Idarubicin HCl, 1 mg/ml | / | / | >240 min | Different |
| Ifosfamide, 50 mg/ml | >240 min. | >240 min | >240 min | Same |
| Irinotecan HCl, 20 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Mechlorethamine HCl, 1 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Melphalan HCl, 5 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Methotrexate, 25 mg/ml | >240 min. | >240 min | >240 min | Same |
| Mitomycin-C, 0.5 mg/ml | >240 min. | >240 min | >240 min | Same |
| Mitoxantrone HCl, 2 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Pacilitaxel, 6 mg/ml | >240 min. | >240 min | >240 min | Same |
| Thiotepa, 10 mg/ml | 28.2 min. | 119min | 25.3 min | Different |
| Vincristine Sulfate,1 mg/ml | / | >240 min | >240 min | Different |
| Fentanyl, Gastric Acid and Xylazine | Minimum Breakthrough Detection Time | Result of comparison | ||
|---|---|---|---|---|
| Proposed device | Predicate device K212408 | Reference Device K250630 | ||
| Fentanyl Citrate Injection,50mg/ml | >240 min. | >240 min | >240 min. | Same |
| Simulated Gastric Acid | >240 min. | >240 min | >240 min. | Same |
| Fentanyl Citrate (50mcg/ml): Xylazine HCl(100mg/ml), 50:50 | >240 min. | / | >240 min. | Same as K250630 |
- Chemotherapy drugs and the minimum breakthrough time of subject device will be listed on labeling, so this difference does not raise questions of safety and effectiveness.
8. Summary of Non-Clinical Performance Data
Page 11
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device met the performance criteria with the following standards:
| Test performed | Methodology | Acceptance Criteria | Result |
|---|---|---|---|
| Freedom From Holes | ASTM D6319-19 ASTM D5151-19 | Meet requirement inspection level G- 1, AQL 2.5 (ISO2859-1) | Pass |
| Dimension-Length | ASTM D6319-19 | Minimum 220mm for size XS-S Minimum 230mm for size M-XXL | Pass |
| Dimension-Width | ASTM D6319-19 | XS: 70±10mm S: 80±10mm M: 95±10mm L:110±10mm XL:120±10mm XXL:130±10mm | Pass |
| Dimension-Thickness | ASTM D6319-19 | Finger: 0.05mm (min) Palm: 0.05mm (min) | Pass |
| Physical properties | ASTM D6319-19 ASTM D412-16 | Tensile Strength (Min14 Mpa) Elongation (Before Aging 500% and after aging 400%) Min | Pass |
| Powder Residue | ASTM D6319-19 ASTM D6124-06 | Not more than 2 mg per glove | Pass |
| Skin Irritation | ISO10993-23:2021 | Under the conditions of the study, not a primary skin irritant | Pass |
| Skin Sensitization | ISO10993-10:2021 | Under the conditions of the study, not a contact sensitizer | Pass |
| Acute Systemic Toxicity | ISO 10993-11:2017 | Under the conditions of the study, no signs of acute systemic toxicity were observed | Pass |
| Chemotherapy Drug | ASTM D6978-05 | Refer to the above table | Pass |
- ASTM D6319- 19, Standard Specification for Nitrile Examination Gloves for Medical Application.
- ASTM D5151- 19, Standard Test Method for Detection of Holes in Medical Gloves.
- ASTM D6124-06 (2017), Standard Test Method for Residual Powder on Medical Gloves
- ASTM D412- 16 (2021), Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers— Tension
- ASTM D6978-05(2023), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
Page 12
- ISO 10993-10: 2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization.
- ISO10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation.
- ISO 10993- 11:2017 Biological evaluation of medical devices — Part 11: Tests for systemic toxicity
9. Summary of Clinical Testing
Not applicable.
10. Conclusion
The conclusions drawn from the non-clinical tests demonstrate that the subject device, Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate, is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.