K Number
K251032
Date Cleared
2025-07-01

(89 days)

Product Code
Regulation Number
880.5730
Panel
CH
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MiniMed 780G insulin pump is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.

The MiniMed 780G insulin pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices.

The MiniMed 780G insulin pump contains a bolus calculator that calculates an insulin dose based on user-entered data.

The MiniMed 780G insulin pump is indicated for use in individuals 7 years of age and older.

The MiniMed 780G insulin pump is intended for single patient use and requires a prescription.

Device Description

The MiniMed 780G insulin pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with compatible digitally connected devices, including an integrated continuous glucose monitor (iCGM), interoperable Medtronic continuous glucose monitor (CGM), and interoperable automated glycemic controller (iAGC). The pump is intended to be used both alone and in conjunction with compatible, digitally connected medical devices for the purpose of drug delivery.

The MiniMed 780G insulin pump is an ambulatory, battery-operated, rate-programmable micro-infusion pump that contains pump software and houses electronics, a pumping mechanism, a user interface, and a medication reservoir within the same physical device. The pump also contains a bolus calculator that calculates an insulin dose based on user-entered data. It is comprised of several discrete external and internal components including a pump case made of a polycarbonate blend, an electronic printed circuit board assembly stacks and a drive motor system.

The MiniMed 780G insulin pump is an interoperable device that can communicate via a Bluetooth Low Energy (BLE) wireless electronic interface with digitally connected devices. The MiniMed 780G insulin pump is a host device for the iAGC and integrates iAGC algorithm into the pump firmware. The pump is then able to receive, execute, and confirm commands from an iAGC to adjust delivery of insulin. The pump receives sensor glucose (SG) data via BLE interface from a compatible iCGM or a compatible interoperable Medtronic CGM and transmits these CGM data to the embedded iAGCs.

The MiniMed 780G insulin pump can operate in one of two modes: Manual Mode or Auto Mode (also referred to as "SmartGuard Mode"). The pump provides the user with keypad pump controls, as well as a data screen for configuring therapy settings and viewing continuous real-time glucose values, glucose trends, alerts, alarms, and other information. The user interface and alerts provide the user with the ability to interact with the pump delivery system and digitally connected devices.

AI/ML Overview

The provided text is a 510(k) clearance letter and summary for a medical device, specifically an insulin pump. It details the device's characteristics, intended use, comparison to predicate devices, and a summary of non-clinical performance data.

However, it does not contain the information requested regarding acceptance criteria and a study proving a device meets these criteria in the context of an AI/Machine Learning device for image analysis or diagnostics.

The request asks for details like:

  • A table of acceptance criteria and reported device performance (which is present in a limited form for the pump's mechanical performance, but not for AI performance).
  • Sample size for the test set and data provenance.
  • Number of experts and their qualifications for ground truth.
  • Adjudication method for the test set.
  • MRMC study details and effect size.
  • Standalone performance.
  • Type of ground truth used.
  • Sample size for the training set.
  • How ground truth for the training set was established.

These specific points are highly relevant to the validation of AI/ML-enabled medical devices, particularly those that interpret or analyze data (like images) to aid in diagnosis or treatment decisions. The MiniMed 780G insulin pump, while having advanced software and interoperability, is primarily an infusion pump with a bolus calculator, not an AI/ML diagnostic or image analysis tool. Its software functions relate to pump control, communication, and basic dose calculation, not complex pattern recognition or inference typically associated with AI in medical devices that would require the detailed validation described in the prompt.

Therefore, I cannot extract the requested information from the provided text because the text describes a different type of medical device and its associated validation. The validation methods mentioned (Delivery Volume Accuracy, Occlusion Detection, Drug Stability, Cybersecurity, Human Factors) are appropriate for an insulin pump but do not align with the AI/ML-specific validation criteria outlined in your prompt.

§ 880.5730 Alternate controller enabled infusion pump.

(a)
Identification. An alternate controller enabled infusion pump (ACE pump) is a device intended for the infusion of drugs into a patient. The ACE pump may include basal and bolus drug delivery at set or variable rates. ACE pumps are designed to reliably and securely communicate with external devices, such as automated drug dosing systems, to allow drug delivery commands to be received, executed, and confirmed. ACE pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of drug delivery.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Evidence demonstrating that device infusion delivery accuracy conforms to defined user needs and intended uses and is validated to support safe use under actual use conditions.
(A) Design input requirements must include delivery accuracy specifications under reasonably foreseeable use conditions, including ambient temperature changes, pressure changes (
e.g., head-height, backpressure, atmospheric), and, as appropriate, different drug fluidic properties.(B) Test results must demonstrate that the device meets the design input requirements for delivery accuracy under use conditions for the programmable range of delivery rates and volumes. Testing shall be conducted with a statistically valid number of devices to account for variation between devices.
(ii) Validation testing results demonstrating the ability of the pump to detect relevant hazards associated with drug delivery and the route of administration (
e.g., occlusions, air in line, etc.) within a clinically relevant timeframe across the range of programmable drug delivery rates and volumes. Hazard detection must be appropriate for the intended use of the device and testing must validate appropriate performance under the conditions of use for the device.(iii) Validation testing results demonstrating compatibility with drugs that may be used with the pump based on its labeling. Testing must include assessment of drug stability under reasonably foreseeable use conditions that may affect drug stability (
e.g., temperature, light exposure, or other factors as needed).(iv) The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible. This shall include chemical and particulate characterization on the final, finished, fluid contacting device components demonstrating that risk of harm from device-related residues is reasonably low.
(v) Evidence verifying and validating that the device is reliable over the ACE pump use life, as specified in the design file, in terms of all device functions and in terms of pump performance.
(vi) The device must be designed and tested for electrical safety, electromagnetic compatibility, and radio frequency wireless safety and availability consistent with patient safety requirements in the intended use environment.
(vii) For any device that is capable of delivering more than one drug, the risk of cross-channeling drugs must be adequately mitigated.
(viii) For any devices intended for multiple patient use, testing must demonstrate validation of reprocessing procedures and include verification that the device meets all functional and performance requirements after reprocessing.
(2) Design verification and validation activities must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following:
(i) Risk control measures shall be implemented to address device system hazards and the design decisions related to how the risk control measures impact essential performance shall be documented.
(ii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design.
(3) The device shall include validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following:
(i) Secure authentication (pairing) to external devices.
(ii) Secure, accurate, and reliable means of data transmission between the pump and connected devices.
(iii) Sharing of necessary state information between the pump and any digitally connected alternate controllers (
e.g., battery level, reservoir level, pump status, error conditions).(iv) Ensuring that the pump continues to operate safely when data is received in a manner outside the bounds of the parameters specified.
(v) A detailed process and procedure for sharing the pump interface specification with digitally connected devices and for validating the correct implementation of that protocol.
(4) The device must include appropriate measures to ensure that safe therapy is maintained when communications with digitally connected alternate controller devices is interrupted, lost, or re-established after an interruption (
e.g., reverting to a pre-programmed, safe drug delivery rate). Validation testing results must demonstrate that critical events that occur during a loss of communications (e.g., commands, device malfunctions, occlusions, etc.) are handled appropriately during and after the interruption.(5) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the system must, at a minimum, include:
(i) A record of all drug delivery
(ii) Commands issued to the pump and pump confirmations
(iii) Device malfunctions
(iv) Alarms and alerts and associated acknowledgements
(v) Connectivity events (
e.g., establishment or loss of communications)(6) Design verification and validation must include results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use.
(7) Device labeling must include the following:
(i) A prominent statement identifying the drugs that are compatible with the device, including the identity and concentration of those drugs as appropriate.
(ii) A description of the minimum and maximum basal rates, minimum and maximum bolus volumes, and the increment size for basal and bolus delivery, or other similarly applicable information about drug delivery parameters.
(iii) A description of the pump accuracy at minimum, intermediate, and maximum bolus delivery volumes and the method(s) used to establish bolus delivery accuracy. For each bolus volume, pump accuracy shall be described in terms of the number of bolus doses measured to be within a given range as compared to the commanded volume. An acceptable accuracy description (depending on the drug delivered and bolus volume) may be provided as follows for each bolus volume tested, as applicable: Number of bolus doses with volume that is 250 percent of the commanded amount.
(iv) A description of the pump accuracy at minimum, intermediate, and maximum basal delivery rates and the method(s) used to establish basal delivery accuracy. For each basal rate, pump accuracy shall be described in terms of the amount of drug delivered after the basal delivery was first commanded, without a warmup period, up to various time points. The information provided must include typical pump performance, as well as worst-case pump performance observed during testing in terms of both over-delivery and under-delivery. An acceptable accuracy description (depending on the drug delivered) may be provided as follows, as applicable: The total volume delivered 1 hour, 6 hours, and 12 hours after starting delivery for a typical pump tested, as well as for the pump that delivered the least and the pump that delivered the most at each time point.
(v) A description of delivery hazard alarm performance, as applicable. For occlusion alarms, performance shall be reported at minimum, intermediate, and maximum delivery rates and volumes. This description must include the specification for the longest time period that may elapse before an occlusion alarm is triggered under each delivery condition, as well as the typical results observed during performance testing of the pumps.
(vi) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device.
(vii) For any infusion pumps intended for multiple patient reuse, instructions for safely reprocessing the device between uses.