(87 days)
The t:slim X2 Insulin Pump with Interoperable Technology (the pump) is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes melliting insulin. The Pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription. The Pump is indicated for use in individuals 2 years of age and greater.
The t:slim X2 Insulin Pump with Interoperable Technology is an ambulatory, battery operated, rate-programmable infusion pump designed for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.
The t:slim X2 insulin pump includes a disposable cartridge with a 300 unit reservoir which is filled with insulin by the user with the use of the syringe and needle. The cartridge needs to be changed every 48-72 hours depending on the type of insulin used.
The pump is motor driven to deliver patient programmed basal rates and boluses through an infusion set into subcutaneous tissue. The desired timing and quantity of insulin delivery (bolus or basal) is programmed by the patient based on their healthcare provider's recommendations.
The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, execute, and confirm commands from these devices.
The t:connect mobile app enables a user to connect a smartphone to the pump using Bluetooth® wireless technology to display pump information and perform some pump functions on the smartphone as well as display pump notifications. The t:connect mobile app can transmit pump and therapy data from the pump to the cloud as long as the user's smartphone is connected to the internet.
The pump is designed to be able to receive and display alerts and alarms to users based on information received from other interoperable devices.
The pump is compatible with Interoperable Automated Glycemic Controllers, such as Basal-IQ Technology (K193483) and Control IQ Technology (K200467) to aid in diabetes management. The latter is being submitted concurrently.
In addition, the Subject Device is compatible with iCGM systems cleared under K223931 21 CFR 862.1355.
The provided text describes the 510(k) summary for the Tandem Diabetes Care t:slim X2 Insulin Pump with Interoperable Technology (K232380). This submission focuses on a change to the age indication for the device, lowering it to individuals 2 years of age and greater.
Here's an analysis of the acceptance criteria and study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of specific quantitative acceptance criteria (e.g., minimum accuracy percentages, specific error rates) for the device's performance in relation to the expanded age indication. Instead, it makes a general statement about safety and effectiveness.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Continued assurance of safety and effectiveness for expanded age indication (2 years and greater) | Clinical data demonstrates continued assurance of safety and effectiveness for the expanded age indication. |
Adherence to Special Controls listed in 21 CFR 880.5730 | Evaluation and adherence to Special Controls ensures continued assurance of safety and effectiveness. |
Substantial equivalence to predicate device in safety and effectiveness for the expanded age indication | Clinical testing demonstrated that the t:slim X2 Insulin Pump with Interoperable Technology is the same as the predicate device in safety and effectiveness for the 2-5 year-old population. |
2. Sample Size Used for the Test Set and Data Provenance
The document mentions "clinical testing was performed to support the expanded age indication" and "the clinical testing provided in this 510(k) demonstrate that the t:slim X2 Insulin Pump with Interoperable Technology is the same as the predicate device in safety and effectiveness."
- Sample Size for Test Set: Not explicitly stated.
- Data Provenance: Not explicitly stated as retrospective or prospective, nor the country of origin. However, the nature of "clinical testing" for an expanded age indication typically implies a prospective clinical study involving the new patient population.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. The text does not describe a process of expert review or establishing ground truth in the context of the clinical testing.
4. Adjudication Method for the Test Set
This information is not provided in the document.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
This information is not provided in the document. The study described is a clinical evaluation of the device itself rather than a comparison of human readers with and without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The device is an insulin pump, which is a physical medical device that delivers insulin. While it has interoperable technology and can communicate with automated insulin dosing software, the "performance data" mentioned pertains to the overall device in a clinical setting with human interaction (patient programming, healthcare provider recommendations). Therefore, a "standalone algorithm only" performance study as typically understood for AI/CAD devices is not directly applicable or explicitly described in this context. The focus is on the device's safety and effectiveness for insulin delivery in the expanded age group.
7. The Type of Ground Truth Used
The "ground truth" for the clinical testing would be the actual physiological responses and outcomes of the pediatric patients using the insulin pump, as assessed by healthcare professionals. This would include parameters like blood glucose levels, incidence of hypoglycemia/hyperglycemia, and adverse events. The document refers to "safety and effectiveness," which points to clinical outcomes.
8. The Sample Size for the Training Set
The document describes a clinical study to support an expanded age indication for an existing device. This implies that the device's core algorithm and functionality were developed and "trained" (in a non-AI sense, meaning designed and validated) prior to this specific submission. The text does not provide information regarding a training set for an AI/algorithm, as this submission is not about the initial development or a new AI component. If the "interoperable technology" involves AI, the training data for that specific AI component is not detailed here.
9. How the Ground Truth for the Training Set Was Established
Given that this submission is for an expanded age indication of an existing device, and not a new AI model, the concept of a "training set" and its ground truth establishment, as typically understood for AI/ML, is not applicable or described in this document. The device's fundamental insulin delivery mechanism and safety profile would have been established through prior design, testing, and regulatory submissions. The current clinical testing validates the device's safety and effectiveness for a younger population.
§ 880.5730 Alternate controller enabled infusion pump.
(a)
Identification. An alternate controller enabled infusion pump (ACE pump) is a device intended for the infusion of drugs into a patient. The ACE pump may include basal and bolus drug delivery at set or variable rates. ACE pumps are designed to reliably and securely communicate with external devices, such as automated drug dosing systems, to allow drug delivery commands to be received, executed, and confirmed. ACE pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of drug delivery.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Evidence demonstrating that device infusion delivery accuracy conforms to defined user needs and intended uses and is validated to support safe use under actual use conditions.
(A) Design input requirements must include delivery accuracy specifications under reasonably foreseeable use conditions, including ambient temperature changes, pressure changes (
e.g., head-height, backpressure, atmospheric), and, as appropriate, different drug fluidic properties.(B) Test results must demonstrate that the device meets the design input requirements for delivery accuracy under use conditions for the programmable range of delivery rates and volumes. Testing shall be conducted with a statistically valid number of devices to account for variation between devices.
(ii) Validation testing results demonstrating the ability of the pump to detect relevant hazards associated with drug delivery and the route of administration (
e.g., occlusions, air in line, etc.) within a clinically relevant timeframe across the range of programmable drug delivery rates and volumes. Hazard detection must be appropriate for the intended use of the device and testing must validate appropriate performance under the conditions of use for the device.(iii) Validation testing results demonstrating compatibility with drugs that may be used with the pump based on its labeling. Testing must include assessment of drug stability under reasonably foreseeable use conditions that may affect drug stability (
e.g., temperature, light exposure, or other factors as needed).(iv) The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible. This shall include chemical and particulate characterization on the final, finished, fluid contacting device components demonstrating that risk of harm from device-related residues is reasonably low.
(v) Evidence verifying and validating that the device is reliable over the ACE pump use life, as specified in the design file, in terms of all device functions and in terms of pump performance.
(vi) The device must be designed and tested for electrical safety, electromagnetic compatibility, and radio frequency wireless safety and availability consistent with patient safety requirements in the intended use environment.
(vii) For any device that is capable of delivering more than one drug, the risk of cross-channeling drugs must be adequately mitigated.
(viii) For any devices intended for multiple patient use, testing must demonstrate validation of reprocessing procedures and include verification that the device meets all functional and performance requirements after reprocessing.
(2) Design verification and validation activities must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following:
(i) Risk control measures shall be implemented to address device system hazards and the design decisions related to how the risk control measures impact essential performance shall be documented.
(ii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design.
(3) The device shall include validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following:
(i) Secure authentication (pairing) to external devices.
(ii) Secure, accurate, and reliable means of data transmission between the pump and connected devices.
(iii) Sharing of necessary state information between the pump and any digitally connected alternate controllers (
e.g., battery level, reservoir level, pump status, error conditions).(iv) Ensuring that the pump continues to operate safely when data is received in a manner outside the bounds of the parameters specified.
(v) A detailed process and procedure for sharing the pump interface specification with digitally connected devices and for validating the correct implementation of that protocol.
(4) The device must include appropriate measures to ensure that safe therapy is maintained when communications with digitally connected alternate controller devices is interrupted, lost, or re-established after an interruption (
e.g., reverting to a pre-programmed, safe drug delivery rate). Validation testing results must demonstrate that critical events that occur during a loss of communications (e.g., commands, device malfunctions, occlusions, etc.) are handled appropriately during and after the interruption.(5) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the system must, at a minimum, include:
(i) A record of all drug delivery
(ii) Commands issued to the pump and pump confirmations
(iii) Device malfunctions
(iv) Alarms and alerts and associated acknowledgements
(v) Connectivity events (
e.g., establishment or loss of communications)(6) Design verification and validation must include results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use.
(7) Device labeling must include the following:
(i) A prominent statement identifying the drugs that are compatible with the device, including the identity and concentration of those drugs as appropriate.
(ii) A description of the minimum and maximum basal rates, minimum and maximum bolus volumes, and the increment size for basal and bolus delivery, or other similarly applicable information about drug delivery parameters.
(iii) A description of the pump accuracy at minimum, intermediate, and maximum bolus delivery volumes and the method(s) used to establish bolus delivery accuracy. For each bolus volume, pump accuracy shall be described in terms of the number of bolus doses measured to be within a given range as compared to the commanded volume. An acceptable accuracy description (depending on the drug delivered and bolus volume) may be provided as follows for each bolus volume tested, as applicable: Number of bolus doses with volume that is 250 percent of the commanded amount.
(iv) A description of the pump accuracy at minimum, intermediate, and maximum basal delivery rates and the method(s) used to establish basal delivery accuracy. For each basal rate, pump accuracy shall be described in terms of the amount of drug delivered after the basal delivery was first commanded, without a warmup period, up to various time points. The information provided must include typical pump performance, as well as worst-case pump performance observed during testing in terms of both over-delivery and under-delivery. An acceptable accuracy description (depending on the drug delivered) may be provided as follows, as applicable: The total volume delivered 1 hour, 6 hours, and 12 hours after starting delivery for a typical pump tested, as well as for the pump that delivered the least and the pump that delivered the most at each time point.
(v) A description of delivery hazard alarm performance, as applicable. For occlusion alarms, performance shall be reported at minimum, intermediate, and maximum delivery rates and volumes. This description must include the specification for the longest time period that may elapse before an occlusion alarm is triggered under each delivery condition, as well as the typical results observed during performance testing of the pumps.
(vi) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device.
(vii) For any infusion pumps intended for multiple patient reuse, instructions for safely reprocessing the device between uses.