(28 days)
The TEG 6s Hemostasis System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a 3.2% citrated whole blood sample. The Citrated: K, KH, RT, FF Assay Cartridge, to be used with the TEG 6s analyzer, contains four independent assays (CK, CKH, CRT, and CFF), described below.
The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s System. Clotting characteristics are described by the functional parameters Clotting Time (R), Speed of Clot Formation (K and Alpha angle) and Maximum Clot Strength (MA).
The CKH assay monitors the effects of heparin in 3.2% citrated whole blood specimens on the TEG 6s System. CKH is used in conjunction with CK, and heparin influence is determined by comparing Clotting Times (R) between the two tests.
The CRT assay monitors the hemostasis process via both the intrinsic and extrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s System. Clotting characteristics are described by the functional parameter Maximum Clot Strength (MA). The CRT MA parameter is equivalent to the CK MA parameter but the final MA value is reached more quickly using the CRT assay.
The CFF assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s System after blocking platelet contributions to clot strength. Clotting characteristics are described by the functional parameters Maximum Clot Strength (MA) and the Estimated Functional Fibrinogen Level (FLEV).
Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests. The indication for TEG 6s System use is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure. The TEG 6s Hemostasis System can be used in the laboratory or at the point-of-care.
The TEG® 6s Hemostasis System (TEG® Hemostasis analyzer and TEG® 6s Assay Cartridges) is intended for in vitro diagnostic use to provide semi-quantitative indications of a blood sample's ability to form and maintain a clot. The TEG® 6s Hemostasis System records the kinetic changes in a sample of whole blood as the sample clots, retracts and/or lyses. The system output consists of a table of numerical values and graphs resulting from the hemostasis process over time. This information can be used by clinicians to aid in determining if a clotting dysfunction or coagulopathy is present.
To perform a test, a disposable TEG® 6s Assay Cartridge is inserted into the TEG® 6s Hemostasis analyzer. The instrument reads the bar code on the cartridge and identifies the type of cartridge for operator confirmation. Blood (collected in a 3.2% sodium citrate tube) or Quality Control (QC) material is added to the entry port on the cartridge and drawn into the cartridge under the TEG® 6s Hemostasis analyzer control. The amount of the sample drawn into the cartridge is determined by the pre-set volume of the blood chambers in the cartridge. Once in the cartridge, the sample is metered into as many as 4 separate analysis channels, depending upon the assays being performed. Reconstitution of reagents dried within the cartridge is accomplished by moving the sample back and forth through reagent chambers, under the control of microfluidic valves and bellows (pumps) within the cartridge. After each sample has been mixed with reagent, it is delivered to a test cell where it is monitored for viscoelastic changes due to coagulation. Excess sample material is moved under microfluidic control into an enclosed waste chamber within the cartridge.
The TEG® 6s technology is based on a disposable cartridge containing up to 4 independent measurement cells. Each cell consists of a short vertically-oriented injection molded tube (ring). Detection of clotting in the TEG® 6s Hemostasis System is performed optically. A piezoelectric actuator vibrates the measurement cell(s) through a motion profile composed of summed sinusoids at different frequencies. The movement of the measurement cells will induce motion in the sample meniscus, which will be detected by a photodiode. The resulting motion of the meniscus is monitored optically and analyzed by the instrument to calculate the resonant frequency and modulus of elasticity (stiffness) of the sample. By performing a Fast Fourier Transform (FFT) on meniscus motion data, the resonant frequencies can be determined. The analyzer monitors the harmonic motion of a hanging drop of blood in response to external vibration. As the sample transitions from a liquid state to a gel-like state during clotting, the modulus of elasticity (stiffness) and therefore resonant frequency increase. The TEG® 6s Hemostasis analyzer measures these variations in resonant frequency during clotting and lysis.
Resonance is the tendency of a material or structure to oscillate with greater amplitude at some frequencies than others. The exact frequencies at which resonance occurs will depend on the stiffness and mass of the sample. Stiffness, in turn, is a function of a material's modulus of elasticity and the boundary conditions to which the material is exposed, such as the geometry and materials of a test cell. By holding these boundary conditions and sample mass constant from sample to sample, the TEG® 6s Hemostasis System allows direct comparison of elasticity between samples. The output measurements are displayed in a table and on a graphical tracing that reflects the hemostasis profile of the clot formation.
In a typical test, blood that has been delivered to the measurement cell will not clot for several minutes. During this time the sample has no inherent stiffness except that provided by surface tension, and since this remains constant the measured resonant frequencies will not change. Once clotting begins, however, the elastic modulus and thus the resonant frequencies increase rapidly. During fibrinolysis, the process is reversed, with elastic modulus and resonant frequencies decreasing. In tests where clotting does not occur, the resonant frequency of the sample will not change. During coagulation, however, a clot will bind to the ring contained in the cartridge and the resonant frequency will rise with increasing firmness of the clot. The TEG® 6s Hemostasis Analyzer collects meniscus motion data, tracks changing resonant frequencies and analyzes the frequency data to provide semi-quantitative parameters describing the clot.
The TEG® 6s Hemostasis System monitors the interaction of platelets within the fibrin mesh of the clot during clot formation and lysis, all in a whole-blood setting. The TEG® 6s Hemostasis System uses thromboelastography to provide continuous measurement of clot elasticity.
The provided document is a 510(k) clearance letter from the FDA for the TEG 6s Hemostasis System. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than providing the detailed acceptance criteria and study data often found in novel device approvals or publications on algorithmic performance.
Therefore, the document does not contain the level of detail requested for many of the points, especially those related to AI algorithm performance (e.g., sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth establishment for training). The device described is an in vitro diagnostic (IVD) system that measures blood coagulation properties, suggesting a direct measurement rather than an AI-driven interpretation of complex data like images.
Below is an attempt to extract and infer information based on the provided text, while also clearly indicating what information is not available in this document.
Acceptance Criteria and Device Performance Study
The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing specific acceptance criteria for a novel device or an AI algorithm. The core of the study described is a comparison of the modified device's performance against its predicate and within different sample matrices.
1. Table of Acceptance Criteria and Reported Device Performance
| Parameter Category | Acceptance Criteria (Inferred from "met clinically acceptable bias limits") | Reported Device Performance |
|---|---|---|
| Sample Matrix Equivalence | No statistical difference in agreement between TEG 6s and TEG 5000 by blood source. | Demonstrated. |
| Predicate Bias (Comparison between new and predicate cartridge types) | Bias met clinically acceptable bias limits across all parameters for TEG 6s Citrated: K, KH, RT, FF vs. TEG 6s Citrated: K, KH, RTH, FFH. | Demonstrated. |
Note: Specific numerical acceptance criteria (e.g., ±X% bias) are not provided in this document. The document states that the bias "met clinically acceptable bias limits," but the limits themselves are not specified.
2. Sample Sizes and Data Provenance for the Test Set
- Sample Size for Test Set: Not specified in the document.
- Data Provenance: The document mentions "an observational study" for the agreement analysis between TEG 6s and TEG 5000 by blood source. It does not specify the country of origin of the data or whether it was retrospective or prospective.
3. Number of Experts and Qualifications for Ground Truth
This information is not applicable/not provided for this type of IVD device clearance. The device measures physical properties of blood, and the "ground truth" for its performance would typically involve comparison to established laboratory methods or accepted clinical standards rather than expert consensus on subjective interpretations.
4. Adjudication Method for the Test Set
This information is not applicable/not provided. Adjudication methods are typically used in studies involving subjective interpretations (e.g., image reading) where multiple experts might disagree. This device measures quantifiable physiological parameters.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not performed/not applicable. An MRMC study is relevant to subjective interpretation tasks, often in medical imaging, where human readers interact with AI. This device directly measures blood coagulation parameters, so there is no "human reader" in the AI sense, nor is there an "AI assistance" component described.
6. Standalone Performance Study (Algorithm Only)
Not performed/not applicable. The TEG 6s Hemostasis System, as described, is a physical instrument that performs direct measurements. It is not an "algorithm only" device or an AI system in the typical sense that would have a standalone performance study independent of human interaction. The "algorithm" here refers to the internal processing of sensor data to derive hemostasis parameters. Its performance is intrinsically linked to the physical measurement system.
7. Type of Ground Truth Used
The ground truth used for this device would be the quantitative measurements of hemostatic parameters from established, validated laboratory methods or from the predicate device itself. The document mentions:
- Agreement analysis with TEG 5000 (a prior generation or related device).
- Comparison with predicate TEG 6s Citrated: K, KH, RT, FF (K243858).
8. Sample Size for the Training Set
The concept of a "training set" in the context of machine learning is not applicable to this device as described in the provided document. The device performs direct physical measurements rather than learning from a dataset via an AI algorithm. Therefore, no training set size is reported.
9. How Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" or AI algorithm in the described context.
FDA 510(k) Clearance Letter - TEG 6s Hemostasis System
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.07.05
April 30, 2025
Haemonetics
Erica Molaro
Senior Regulatory Affairs Specialist
125 Summer Street
Boston, Massachusetts 02110
Re: K251024
Trade/Device Name: TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay Cartridge
Regulation Number: 21 CFR 864.5425
Regulation Name: Multipurpose System For In Vitro Coagulation Studies
Regulatory Class: Class II
Product Code: JPA
Dated: April 2, 2025
Received: April 2, 2025
Dear Erica Molaro:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
Page 2
K251024 - Erica Molaro Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Page 3
K251024 - Erica Molaro Page 3
Sincerely,
Min Wu -S
Min Wu, Ph.D.
Branch Chief
Division of Immunology and Hematology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
FORM FDA 3881 (8/23) Page 1 of 2
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K251024
Device Name: TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay Cartridge
Indications for Use (Describe)
The TEG 6s Hemostasis System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a 3.2% citrated whole blood sample. The Citrated: K, KH, RT, FF Assay Cartridge, to be used with the TEG 6s analyzer, contains four independent assays (CK, CKH, CRT, and CFF), described below.
The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s System. Clotting characteristics are described by the functional parameters Clotting Time (R), Speed of Clot Formation (K and Alpha angle) and Maximum Clot Strength (MA).
The CKH assay monitors the effects of heparin in 3.2% citrated whole blood specimens on the TEG 6s System. CKH is used in conjunction with CK, and heparin influence is determined by comparing Clotting Times (R) between the two tests.
The CRT assay monitors the hemostasis process via both the intrinsic and extrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s System. Clotting characteristics are described by the functional parameter Maximum Clot Strength (MA). The CRT MA parameter is equivalent to the CK MA parameter but the final MA value is reached more quickly using the CRT assay.
The CFF assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s System after blocking platelet contributions to clot strength. Clotting characteristics are described by the functional parameters Maximum Clot Strength (MA) and the Estimated Functional Fibrinogen Level (FLEV).
Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests. The indication for TEG 6s System use is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure. The TEG 6s Hemostasis System can be used in the laboratory or at the point-of-care.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
Page 5
FORM FDA 3881 (8/23) Page 2 of 2
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 6
Attachment 13
© 2025 Haemonetics Corporation – Company Confidential
510(k) Summary
Submitter: Haemonetics Corporation
125 Summer Street
Boston, MA 02110, United States
Contact: Erica Molaro
Senior Regulatory Affairs Specialist
(201) 661-1493
emolaro@haemonetics.com
Date Prepared: March 31, 2025
I. Device Information
Device Trade Name: TEG® 6s Hemostasis System: Citrated K, KH, RT, FF Assay Cartridge
Common Name: Whole Blood Hemostasis System
Classification Name: System, Multipurpose For In Vitro Coagulation Studies
Regulatory Class: 2
Regulation Number: 21 CFR 864.5425
Product Code: JPA
II. Legally Marketed Predicate Device
| Predicate # | Predicate Trade Name | Product Code |
|---|---|---|
| K243858 | TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay Cartridge | JPA |
III. Device Description Summary
TEG® 6s System Description
The TEG® 6s Hemostasis System (TEG® Hemostasis analyzer and TEG® 6s Assay Cartridges) is intended for in vitro diagnostic use to provide semi-quantitative indications of a blood sample's ability to form and maintain a clot. The TEG® 6s Hemostasis System records the kinetic changes in a sample of whole blood as the sample clots, retracts and/or lyses. The system output consists of a table of numerical values and graphs resulting from the hemostasis process over time. This information can be used by clinicians to aid in determining if a clotting dysfunction or coagulopathy is present.
To perform a test, a disposable TEG® 6s Assay Cartridge is inserted into the TEG® 6s Hemostasis analyzer. The instrument reads the bar code on the cartridge and identifies the type of cartridge for operator confirmation. Blood (collected in a 3.2% sodium citrate tube) or Quality
Page 7
Attachment 13
© 2025 Haemonetics Corporation – Company Confidential
Control (QC) material is added to the entry port on the cartridge and drawn into the cartridge under the TEG® 6s Hemostasis analyzer control. The amount of the sample drawn into the cartridge is determined by the pre-set volume of the blood chambers in the cartridge. Once in the cartridge, the sample is metered into as many as 4 separate analysis channels, depending upon the assays being performed. Reconstitution of reagents dried within the cartridge is accomplished by moving the sample back and forth through reagent chambers, under the control of microfluidic valves and bellows (pumps) within the cartridge. After each sample has been mixed with reagent, it is delivered to a test cell where it is monitored for viscoelastic changes due to coagulation. Excess sample material is moved under microfluidic control into an enclosed waste chamber within the cartridge.
TEG® 6s Measurement Technique
The TEG® 6s technology is based on a disposable cartridge containing up to 4 independent measurement cells. Each cell consists of a short vertically-oriented injection molded tube (ring). Detection of clotting in the TEG® 6s Hemostasis System is performed optically. A piezoelectric actuator vibrates the measurement cell(s) through a motion profile composed of summed sinusoids at different frequencies. The movement of the measurement cells will induce motion in the sample meniscus, which will be detected by a photodiode. The resulting motion of the meniscus is monitored optically and analyzed by the instrument to calculate the resonant frequency and modulus of elasticity (stiffness) of the sample. By performing a Fast Fourier Transform (FFT) on meniscus motion data, the resonant frequencies can be determined. The analyzer monitors the harmonic motion of a hanging drop of blood in response to external vibration. As the sample transitions from a liquid state to a gel-like state during clotting, the modulus of elasticity (stiffness) and therefore resonant frequency increase. The TEG® 6s Hemostasis analyzer measures these variations in resonant frequency during clotting and lysis.
Resonance is the tendency of a material or structure to oscillate with greater amplitude at some frequencies than others. The exact frequencies at which resonance occurs will depend on the stiffness and mass of the sample. Stiffness, in turn, is a function of a material's modulus of elasticity and the boundary conditions to which the material is exposed, such as the geometry and materials of a test cell. By holding these boundary conditions and sample mass constant from sample to sample, the TEG® 6s Hemostasis System allows direct comparison of elasticity between samples. The output measurements are displayed in a table and on a graphical tracing that reflects the hemostasis profile of the clot formation.
In a typical test, blood that has been delivered to the measurement cell will not clot for several minutes. During this time the sample has no inherent stiffness except that provided by surface tension, and since this remains constant the measured resonant frequencies will not change. Once clotting begins, however, the elastic modulus and thus the resonant frequencies increase rapidly. During fibrinolysis, the process is reversed, with elastic modulus and resonant frequencies decreasing. In tests where clotting does not occur, the resonant frequency of the sample will not change. During coagulation, however, a clot will bind to the ring contained in the
Page 8
Attachment 13
© 2025 Haemonetics Corporation – Company Confidential
cartridge and the resonant frequency will rise with increasing firmness of the clot. The TEG® 6s Hemostasis Analyzer collects meniscus motion data, tracks changing resonant frequencies and analyzes the frequency data to provide semi-quantitative parameters describing the clot.
The TEG® 6s Hemostasis System monitors the interaction of platelets within the fibrin mesh of the clot during clot formation and lysis, all in a whole-blood setting. The TEG® 6s Hemostasis System uses thromboelastography to provide continuous measurement of clot elasticity.
Table 1: TEG® 6s Parameters for TEG® 6s Citrated: K, KH, RT, FF Assay Cartridge
| TEG 6s Parameter | Definition | Parameter Relation to Hemostasis |
|---|---|---|
| R | R is the time from the start of the test until initial fibrin formation. This represents the enzymatic portion of coagulation | Clotting time – Normal / reduced / increased speed of coagulation initiation |
| K | K is the time after R needed to reach a certain level of clot strength. This represents clot kinetics | Normal / reduced / increased speed of coagulation amplification and propagation |
| Alpha α (Angle) | Alpha is the angle representing the rate of increase in the clot strength the rapidity of fibrin build-up and cross-linking | Normal / reduced / increased speed of coagulation amplification and propagation |
| MA | MA, or Maximum Amplitude, represents the maximum firmness of the clot during the test. | Normal / reduced /increased clot elasticity/strength |
| FLEV | FLEV is an estimate of the fibrinogen level in the blood sample. | Estimated Functional Fibrinogen Level |
Citrated Kaolin (CK) assay
The CK assay is a semi-quantitative in vitro diagnostic assay for monitoring the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG® 6s Hemostasis System. The CK assay consists of Kaolin which is used in the assay for activation of coagulation. It is combined with calcium chloride (CaCl2) to neutralize the sodium citrate in the blood.
The clotting characteristics of the CK generated hemostasis profile are described by the functional parameters Clotting Time (R), Speed of Clot Formation (K and Alpha angle) and Maximum Clot strength (MA). Since it may take an or more for a non-activated whole blood sample to reach maximum amplitude MA, Kaolin is essential to reduce run time and variability associated with running non-activated whole blood samples.
Citrated Kaolin with Heparinase (CKH) assay
The CKH assay is a semi-quantitative in vitro diagnostic assay for monitoring the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG® 6s
Page 9
Attachment 13
© 2025 Haemonetics Corporation – Company Confidential
Hemostasis System. CKH is used in conjunction with CK, and heparin influence is determined by comparing Clotting Times (R) between the two tests. Calcium Chloride (CaCl2) is included to neutralize any sodium citrate in the blood.
The CKH assay monitors the effects of heparin, a commonly used anticoagulant in surgical procedures. Even very low concentrations of heparin, fractions of IU/mL of blood, can noticeably increase the R time and can even completely anticoagulate the blood, making it difficult if not impossible to monitor developing coagulopathies that are masked by high levels of therapeutic heparin.
Citrated RapidTEG (CRT) assay
The CRT assay is a semi-quantitative in vitro diagnostic assay for monitoring the hemostasis process after stimulation of both the intrinsic and extrinsic pathway in 3.2% citrated whole blood specimens on the TEG® 6s Hemostasis System. The CRT assay consists of a reagent containing kaolin and tissue factor. It is combined with Calcium Chloride (CaCl2) to neutralize sodium citrate in the blood sample. Clotting characteristics are described by the functional parameter Maximum Clot Strength (MA).
The CRT assay produces an accelerated clotting time which allows for an earlier MA results compared to the CK assay. Therefore, in the TEG® Hemostasis System, the CRT assay is simultaneously run along with the CK and CKH assays to provide a fast way to reach a stable value for MA (CRT) while still measuring the time- dependent parameters (CK).
Citrated Functional Fibrinogen (CFF) assay
The CFF assay is a semi-quantitative in vitro diagnostic assay for monitoring the hemostasis process after blocking platelet contributions to clot strength in 3.2% citrated whole blood specimens on the TEG® 6s Hemostasis System. The CFF assay consists of tissue factor and abciximab. It is combined with Calcium Chloride (CaCl2) to neutralize sodium citrate in the blood sample.
Tissue Factor is used for coagulation activation that would be classically described as extrinsic, with platelet aggregation inhibited by abciximab (a GPIIb/IIIa inhibitor), excluding its contribution to clot strength, and thereby measuring fibrinogen contribution to clot strength.
Clotting characteristics are described by the functional parameter Maximum Clot Strength (MA), the Estimated Functional Fibrinogen Level (FLEV).
IV. Intended Use/Indications for Use
The TEG 6s Hemostasis System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a 3.2% citrated whole blood sample. The Citrated: K, KH, RT, FF Assay Cartridge, to be used with the TEG 6s analyzer, contains four independent assays (CK, CKH, CRT, and CFF), described below.
The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s System. Clotting characteristics are described by the functional
Page 10
Attachment 13
© 2025 Haemonetics Corporation – Company Confidential
parameters Clotting Time (R), Speed of Clot Formation (K and Alpha angle) and Maximum Clot Strength (MA).
The CKH assay monitors the effects of heparin in 3.2% citrated whole blood specimens on the TEG 6s System. CKH is used in conjunction with CK, and heparin influence is determined by comparing Clotting Times (R) between the two tests.
The CRT assay monitors the hemostasis process via both the intrinsic and extrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s System. Clotting characteristics are described by the functional parameter Maximum Clot Strength (MA). The CRT MA parameter is equivalent to the CK MA parameter but the final MA value is reached more quickly using the CRT assay.
The CFF assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s System after blocking platelet contributions to clot strength. Clotting characteristics are described by the functional parameters Maximum Clot Strength (MA) and the Estimated Functional Fibrinogen Level (FLEV).
Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests. The indication for TEG 6s System use is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure. The TEG 6s Hemostasis System can be used in the laboratory or at the point-of-care.
V. Comparison to Predicate Device
Indications for Use Comparison
The TEG® 6s system consists of the TEG® 6s analyzer including analyzer software and assay cartridges. The assay cartridges predicate device is the TEG® 6s Hemostasis System Citrated: K, KH, RT, FF (07-601-US; K243858). The indications for use are of the same intent with the following inclusion:
- Addition of arterial whole blood as a sample matrix
Technology Comparison
There is no change to the technology, design of the device, reported parameters, or mechanics of how the cartridge is run on the TEG® 6s analyzer. No additional product development of the TEG® 6s system was required.
Page 11
Attachment 13
© 2025 Haemonetics Corporation – Company Confidential
Table 2: Summary of Technological Characteristics for Substantial Equivalence: Similarities
| Characteristic | TEG® 6s Citrated: K, KH, RT, FF Cleared Device/Predicate [K243858] | TEG® 6s Citrated: K, KH, RT, FF Modified Device |
|---|---|---|
| Technological Purpose | Monitoring the physical response of a clot to low levels of applied strain (resonance frequency) | Same |
| Measurement | Changes in physical clot elasticity over time | Same |
| Initial Warm Up Time | 5 minutes | Same |
| Analyzer Hardware | Fully integrated Thromboelastography analyzer | Same |
| Analyzer Measuring Technique | Non-contact optical measurement of shear elasticity of a coagulating sample | Same |
| Measurement Output | Graphical tracings of resonant frequency per reagent type; table of parameters | Same |
| Physical Configuration Used for Sample Measurement | Vertically-oriented cylindrical container (ring or tube) containing sample with meniscus formed at bottom; non-contact measurement of meniscus amplitude of vibration | Same |
| Physical Principle Underlying Measuring Technique | Clotting process causes an increase in the modulus of elasticity, which increases stiffness and increases the force of the clot within the ring walls when moving up and down in the ring tube against the sample's own weight. This increases the resonant frequency, which increases the clot strength amplitude. During clot dissolution (lysis), the resonant frequency decreases, corresponding to decreasing clot strength amplitude. | Same |
| Boundary Conditions | Physical ring tube dimensions define boundary conditions that are help constant, allowing comparison of one sample to the next. | Same |
| Measuring Channel | Up to 4 | Same |
| Signal Generation | Multi-frequency harmonic oscillation of a small open-ended cylinder containing the sample | Same |
| Signal Transducer | Optical detection (silicon photodiode) of the motion of a free surface of the sample. | Same |
| Sensitivity to Applied Strains | Applied strains are very low, resulting in minimal interference with small changes in clot formation. | Same |
| Temperature Control | 20° to 50°C | Same |
| Sample Volume (per channel) | 20 μl | Same |
| Total Reaction Volume (single channel) | 20 μl | Same |
| Mains Supply Voltage | 100-240V, 50-60Hz (international power supply) | Same |
Page 12
Attachment 13
© 2025 Haemonetics Corporation – Company Confidential
| Characteristic | TEG® 6s Citrated: K, KH, RT, FF Cleared Device/Predicate [K243858] | TEG® 6s Citrated: K, KH, RT, FF Modified Device |
|---|---|---|
| Analyzer Input Voltage | 12 volts DC, 60 watts max | Same |
| Environment | Stable and level surface. Operating Temperature 10° to 32° C Storage Temperature: -20° to +50° C (instrument only) Relative humidity 20 to 80% (noncondensing) | Same |
| Operating Position | Level (adjustment not required) | Same |
| Sample Preparation | Performed under analyzer control within the disposable cartridge | Same |
| Pipetting | Unmetered transfer pipette or syringe; blood sample is added until it fills to a level above the line marked on the blood intake well of the cartridge. | Same |
| Analyzer Software | Fully integrated Thromboelastography analyzer | Same |
| Consumables | Carrier (acrylic plastic) with microfluidics laminate and test rings (acrylic plastic) | Same |
| Assay | CK, CKH, CRT, CFF | Same |
| Assay Reagents | CK – kaolin and CaCl2 CKH – kaolin and CaCl2 with heparinase CRT – tissue factor, kaolin and CaCl2 CFF – abciximab (brand name ReoPro), tissue factor and CaCl2 | Same |
| Assay Parameters Reported | CK: R, K, Angle, MA CKH: R CRT: MA CFF: MA, FLEV | Same |
| Quality Controls | Cartridge Reagent QC - Level 1 Cartridge Reagent QC - Level 2 | Same |
| Use Location | Clinical laboratory or at the point-of-care. | Same |
Table 3: Summary of Clinical Values Comparisons: Similarities
| Clinical Value | TEG® 6s Citrated: K, KH, RT, FF Cleared Device/Predicate [K243858] | TEG® 6s Citrated: K, KH, RT, FF Modified Device |
|---|---|---|
| Clinical Values of Citrated: Kaolin (CK) Parameter R (min) | Initiation phase of coagulation triggered by enzymatic clotting factors and culminating with the initial fibrin formation. A prolonged R value is indicative of slow clot formation, due to coagulation factor deficiencies or heparin. | Same |
| Clinical Value of Citrated: Kaolin (CK) Parameter MA (mm) | MA, or Maximum Amplitude, Represents the maximum firmness of the clot during the test. The MA | Same |
Page 13
Attachment 13
© 2025 Haemonetics Corporation – Company Confidential
| Clinical Value | TEG® 6s Citrated: K, KH, RT, FF Cleared Device/Predicate [K243858] | TEG® 6s Citrated: K, KH, RT, FF Modified Device |
|---|---|---|
| provides information about the contribution of platelets/fibrin to the overall strength of the clot. | ||
| Clinical Value of Citrated: Kaolin with Heparinase (CKH) Parameter R (min) | Initiation phase of coagulation triggered by enzymatic coagulation factors and culminating with the initial fibrin formation. A prolonged R value is indicative of slow clot formation, due to coagulation factor deficiencies or heparin. Inclusion of heparinase in the blood chamber channel of the cartridge provides ability to compare R (min) without the effect of heparin on the clot. | Same |
| Clinical Value of Citrated: RapidTEG Parameter MA (mm) | RapidTEG™ MA is the point at which clot strength reaches its maximum and reflects the end result of minimal platelet-fibrin interaction via the GPIIb/IIIa receptors. Due to faster coagulation activation, clot strength is measured faster than Citrated: Kaolin (K) activated samples. Same results as CK maximum amplitude (CK-MA). The MA provides information about the contribution of platelets/fibrin to the overall strength of the clot. | Same |
| Clinical Value of Citrated: Functional Fibrinogen Parameter MA (mm) | The maximum amplitude of CFF provides the functional fibrinogen contribution to the clot strength. Provides the overall contribution of functional fibrinogen to clot strength. In conjunction with CRT-MA, this assay enables an assessment of the relative contributions of functional fibrinogen and platelets to clot strength. Results may be valuable for guiding fibrinogen supplementation or platelet transfusion. | Same |
| Patient Population from Intended Use | Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure. | Same |
Page 14
Attachment 13
© 2025 Haemonetics Corporation – Company Confidential
Table 4: Summary of Technological Characteristics for Substantial Equivalence: Differences
| Characteristic | TEG® 6s Citrated: K, KH, RT, FF Cleared Device/Predicate [K243858] | TEG® 6s Citrated: K, KH, RT, FF Modified Device |
|---|---|---|
| Matrix | 3.2% citrated venous whole blood | 3.2% citrated whole blood (venous and arterial) |
Non-Clinical Test Summary and Conclusions
Sample Matrix Equivalence
Agreement analysis between TEG 6s and TEG 5000 by blood source from an observational study determined no statistical difference in agreement between TEG 6s and TEG 5000.
Agreement analysis between TEG 6s Citrated: K, KH, RT, FF and TEG 6s Citrated: K, KH, RTH, FFH determined the predicated bias met clinically acceptable bias limits across all parameters.
Conclusion
The information provided in this submission support a substantial equivalence determination for the TEG® 6s Hemostasis System: Citrated: K, KH, RT, FF and the predicate device.
§ 864.5425 Multipurpose system for in vitro coagulation studies.
(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.