K Number
K250965
Date Cleared
2025-06-02

(63 days)

Product Code
Regulation Number
864.5425
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CN-Series (CN-6000) is a fully automated coagulation analyzer intended for in vitro diagnostic use in the clinical laboratory. The instrument analyzes citrated plasma samples (3.2 % sodium citrate) collected from venous blood, using clotting, chromogenic and immunoassay methods.

The performance of this device has not been established in neonate and pediatric patient populations.

Device Description

The CN-Series (CN-6000) coagulation analyzer is an automated blood coagulation instrument which can analyze samples using clotting, chromogenic and immunoassay methods. Analysis results are displayed on the IPU (Information Processing Unit) screen. They can be printed on external printers or transmitted to a host computer.

Sold separately from the instrument are the associated reagents, controls, calibrators, and consumable materials. The subject of this 510(k) notification is the analyzer together with the reagent applications which perform the coagulation tests:

  • Prothrombin Time (PT) seconds and PT INR with Dade® Innovin®
  • Activated Partial Thromboplastin Time (APTT) with Dade® Actin® FSL
  • Fibrinogen (Fbg) with Dade® Thrombin Reagent
  • Antithrombin (AT) with INNOVANCE® Antithrombin
  • D-dimer with INNOVANCE® D-Dimer.

The analysis principles used on the instrument are reflected by the reagent application testing provided in this submission and are described in the table below. These reagents were selected to show the analytical technology of the instrument while also selecting commonly used applications in coagulation laboratories in the United States.

AI/ML Overview

This document describes the acceptance criteria and the study that proves the device meets the acceptance criteria for the Sysmex CN-6000 Automated Blood Coagulation Analyzer.

The provided text is a 510(k) clearance letter for an in vitro diagnostic device, specifically an automated blood coagulation analyzer. The information focuses on demonstrating substantial equivalence to a predicate device (Sysmex CS-5100 K150678) rather than providing detailed acceptance criteria and proof of exceeding them in the format typically seen for novel AI/ML devices. Therefore, the information has been extracted and interpreted as closely as possible to the requested structure, with explanations where the requested information is not explicitly present due to the nature of the document.

1. Table of Acceptance Criteria and Reported Device Performance

For an in vitro diagnostic device like the CN-6000, acceptance criteria are generally defined as meeting performance specifications comparable to the predicate device and within clinically acceptable ranges. The "acceptance criteria" for each study are generally defined as the results meeting these pre-established standards. The reported device performance is presented as the outcomes of these studies.

Study TypeAcceptance Criteria (Implicit)Reported Device Performance (CN-6000)
Method Comparison (Against Predicate CS-5100)Results comparable to predicate device with high correlation (r-value close to 1.0) and minimal bias (slope close to 1.0, intercept close to 0).PT (seconds): y = 1.010x – 0.100, r = 0.999 (n=847) PT (INR): y = 1.010x – 0.010, r = 0.999 (n=813) APTT: y = 0.993x + 0.088, r = 0.996 (n=638) Fibrinogen: y = 1.062x – 4.092, r = 0.993 (n=456) Antithrombin: y = 0.984x – 0.998, r = 0.995 (n=450) D-dimer: y = 0.959x + 0.007, r = 0.997 (n=395) All results from each assay met the pre-established acceptance criteria.
PrecisionWithin-site precision (SD/CV) within acceptable clinical laboratory limits for each analyte.CVs generally low (e.g., PT seconds: 0.2-1.8%, PT INR: 0.4-2.2%, APTT: 0.3-2.6%, Fibrinogen: 1.0-6.2%, Antithrombin: 0.9-5.2%, D-Dimer: 1.7-5.8%), meeting pre-established criteria.
Linearity & Measuring RangeMeasured linear range should encompass or be equivalent to the Analytical Measurement Interval (AMI).Measured ranges effectively supported the claimed AMI for Fibrinogen (38.4-900.2 mg/dL vs. 50-860 mg/dL), Antithrombin (7.56-130.42% vs. 9.0-128.0%), and D-dimer (0.180-35.836 mg/L FEU vs. 0.19-35.20 mg/L FEU). All reagents met the predetermined acceptance criteria.
Interference StudiesNo significant interference from hemoglobin, bilirubin, triglycerides, and HES up to specified concentrations.All pre-established criteria were met, demonstrating substantial equivalent optical performance.
Reagent CarryoverNo significant carryover effects from one application to another.All results met the specified criteria.
Sample CarryoverNegligible carryover contamination between samples.All results met the specified criteria.
Limit of Blank/Detection/Quantitation (LoB/LoD/LoQ)Measured limits within acceptable performance for each assay.Fibrinogen LoQ: 36.1 mg/dL. Antithrombin LoB: 2.21%, LoD: 2.95%, LoQ: 8.31%. D-Dimer LoB: 0.085 mg/L FEU, LoD: 0.101 mg/L FEU, LoQ: 0.182 mg/L FEU.
Factor Sensitivity (PT, APTT)Reagent sensitivity to factors V, VII, VIII, IX within acceptable ranges.Factor V: 40.8% - 44.5%. Factor VII: 45.8% - 48.3%. Factor VIII: 40.2% - 42.8%. Factor IX: 33.2%.
Heparin Sensitivity (APTT)High correlation between CN-6000 and CS-5100 results for heparinized samples.Lot 1: n = 56, y = 1.000x - 0.200, r = 0.9993. Lot 2: n = 56, y = 0.981x - 0.313, r = 0.9995.
Lupus Anticoagulant (LA) SensitivityAcceptable performance with LA positive samples.Results for the study met the specified criteria.
Stability (Reagents & QC)Manufacturer's claim for onboard stability met for reagents and QC materials.Manufacturer's claim for onboard stability for all reagents and QC materials was met.
High Dose Hook EffectAppropriate instrument flagging ("antigen excess") and no erroneously low results up to 500 mg/L FEU.Acceptance criterion was met.
Matrix Comparison (Auto-Dilution vs Manual, Uncapped vs Capped, Frozen vs Fresh, Micro vs Normal Mode)Equivalence of results across different sample handling/processing methods.Pre-defined acceptance criteria were met for all matrix comparison studies.
Reference RangeEstablished adult reference intervals.PT (seconds): 9.9 – 12.3. PT (INR): 0.93 – 1.16. APTT: 23.8 – 32.0. Fibrinogen: 192 – 440 mg/dL. Antithrombin: 83.7 – 121.6%. D-Dimer: < 0.19 – 1.27 mg/L FEU.

2. Sample Size Used for the Test Set and Data Provenance

The "test set" in this context refers to the samples used in the performance evaluation studies.

  • Method Comparison:
    • Total Combined Sites Sample Sizes:
      • PT (seconds): N = 847
      • PT (INR): N = 813
      • APTT: N = 638
      • Fibrinogen: N = 456
      • Antithrombin: N = 450
      • D-dimer: N = 395
    • Data Provenance: Conducted at three clinical sites. The document does not specify the country of origin of the data, but the context of an FDA 510(k) submission strongly implies data from within the United States. The samples included "Normal, abnormal samples, and samples at the medical decision levels" which typically implies retrospective clinical samples (existing patient samples) or potentially prospectively collected samples for the purpose of the study. It's common in IVD submissions for these to be retrospective collections.
  • Precision: Commercial QC materials and plasma pools reflecting medical decision levels. No specific human patient sample size is given for precision, as it focuses on analytical performance using controlled materials.
  • Linearity & Measuring Range: Not specified as human samples, but rather analytical samples/dilutions.
  • Interference Studies: Not specified as human samples, but rather samples spiked with interfering substances.
  • Heparin Sensitivity: n = 56 (patients receiving unfractionated heparin (UFH) therapy).
  • Lupus Anticoagulant (LA) Sensitivity: 24 LA positive samples.
  • Reference Range: Specific sample size per site not detailed, but stated as "adult samples" collected at "three external clinical sites within the United States."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

For this type of in vitro diagnostic device, "ground truth" is established through laboratory reference methods and predicate device measurements, not typically by expert human readers' interpretations of images or clinical assessments in the same way as AI/ML imaging devices.

  • No "experts" in the sense of radiologists, clinicians, or similar domain specialists directly establishing "ground truth" through consensus or interpretation for the test sets are mentioned.
  • The "ground truth" for the method comparison and other performance studies is the result obtained from the predicate device (Sysmex CS-5100) or reference methods and established analytical specifications.
  • The studies were conducted at three clinical sites, implying performance by trained laboratory professionals using standard operating procedures. The qualifications of these professionals (e.g., licensed medical technologists) are implicit in CLSI guidelines for such studies but not explicitly stated.

4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

Not applicable for this type of IVD device. Adjudication methods like 2+1 or 3+1 are typical for AI/ML imaging diagnostics where multiple human readers independently interpret data (e.g., images) to establish a consensus ground truth, which is then compared against the AI's performance. For an automated coagulation analyzer, the "ground truth" is typically the quantitative result from a validated method/predicate device.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an automated in vitro diagnostic analyzer, not an AI-assisted diagnostic imaging or interpretation tool that would involve human readers or MRMC studies. Its primary function is to perform laboratory tests on blood samples, not to assist human interpretation of complex medical cases in a subjective way.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the studies performed (method comparison, precision, linearity, interference, carryover, etc.) represent the standalone performance of the CN-6000 analyzer. The device functions automatically to analyze samples and output results based on its internal algorithms (for clotting, chromogenic, and immunoassay methods). The comparison is against a predicate device or established analytical criteria, not against human performance.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The ground truth for most of these studies is:

  • Predicate device measurements: For the method comparison studies, the results from the Sysmex CS-5100 (K150678) serve as the comparative "ground truth."
  • Defined analytical specifications/reference methods: For studies like precision, linearity, LoB/LoD/LoQ, and interference, the "ground truth" is adherence to established analytical limits, known concentrations of controls/calibrators, or the expected absence of interference, derived from CLSI guidelines and industry standards for chemical analyzers.
  • Clinical status: For studies like Heparin Sensitivity and Lupus Anticoagulant Sensitivity, the ground truth relates to the clinical status of the patient samples (e.g., patients on UFH therapy, LA positive samples).

8. The Sample Size for the Training Set

This document describes the validation of a laboratory instrument, not an AI/ML model for which "training set" is a common term. The device's underlying principles (clotting, chromogenic, immunoassay) are well-established analytical methods, not learned from data in the way an AI model is. Therefore, there is no "training set" in the context of machine learning. The device's "training" would be its initial design and calibration using scientific principles and engineering practices, which doesn't involve a data-driven training set in the AI sense.

9. How the Ground Truth for the Training Set Was Established

As explained in point 8, the concept of a "training set" and associated "ground truth" for a training set does not apply to this type of traditional in vitro diagnostic instrument where performance is based on established analytical principles rather than AI/ML model training. The instrument's design and calibration are based on physical and chemical principles of coagulation analysis.

FDA 510(k) Clearance Letter - Sysmex CN-6000 Automated Blood Coagulation Analyzer

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 2, 2025

Sysmex America, Inc.
Esther John-Olotu
Regulatory Affairs Manager
577 Aptakisic Road
Lincolnshire, Illinois 60069

Re: K250965
Trade/Device Name: Automated Blood Coagulation Analyzer CN-Series (CN-6000)
Regulation Number: 21 CFR 864.5425
Regulation Name: Multipurpose System For In Vitro Coagulation Studies
Regulatory Class: Class II
Product Code: JPA
Dated: March 28, 2025
Received: March 31, 2025

Dear Esther John-Olotu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

Sincerely,

MIN WU -S

Min Wu, Ph.D.
Branch Chief
Division of Immunology and Hematology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250965

Device Name: Automated Blood Coagulation Analyzer CN-Series (CN-6000)

Indications for Use (Describe):

The CN-Series (CN-6000) is a fully automated coagulation analyzer intended for in vitro diagnostic use in the clinical laboratory. The instrument analyzes citrated plasma samples (3.2 % sodium citrate) collected from venous blood, using clotting, chromogenic and immunoassay methods.

The performance of this device has not been established in neonate and pediatric patient populations

Type of Use (Select one or both, as applicable):
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

Sysmex America, Inc.
CN-Series (CN-6000) Automated Blood Coagulation Analyzer
Premarket Notification
Page 1 of 13
Classified as Confidential

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92

The assigned 510(k) number is: K250965

Submitter's name, address, telephone number, contact person, and date the summary was prepared:

Submitter's Name: Sysmex America, Inc.
Submitter's Address: 577 Aptakisic Road
Lincolnshire, IL 60069
Submitter's Telephone: (404)957-3627
Submitter's Contact: Esther John-Olotu
Date 510(k) Summary Prepared: March 28, 2025

Name of the device, including the trade or proprietary name, the common or usual name, and the classification name:

Proprietary name: Automated Blood Coagulation Analyzer CN-Series (CN-6000)
Common name: Automated Blood Coagulation Analyzer
Regulation description: Coagulation Instrument
Regulation Section: 21 CFR 864.5425
Device Class: II
Product Code: JPA

Predicate Device and 510(k) number:

Sysmex Automated Blood Coagulation Analyzer CS-5100 K150678

Description of the Device:

The CN-Series (CN-6000) coagulation analyzer is an automated blood coagulation instrument which can analyze samples using clotting, chromogenic and immunoassay methods. Analysis results are displayed on the IPU (Information Processing Unit) screen. They can be printed on external printers or transmitted to a host computer.

Sold separately from the instrument are the associated reagents, controls, calibrators, and consumable materials. The subject of this 510(k) notification is the analyzer together with the reagent applications which perform the coagulation tests:

  • Prothrombin Time (PT) seconds and PT INR with Dade® Innovin®
  • Activated Partial Thromboplastin Time (APTT) with Dade® Actin® FSL
  • Fibrinogen (Fbg) with Dade® Thrombin Reagent
  • Antithrombin (AT) with INNOVANCE® Antithrombin
  • D-dimer with INNOVANCE® D-Dimer.

Page 6

The analysis principles used on the instrument are reflected by the reagent application testing provided in this submission and are described in the table below. These reagents were selected to show the analytical technology of the instrument while also selecting commonly used applications in coagulation laboratories in the United States.

Reagent ApplicationMethodology
Dade® Innovin®PT, Prothrombin Time (seconds)
PT, Prothrombin Time (INR)
Dade® Actin® FSLAPTT, Activated Partial Thromboplastin Time
Dade® Thrombin ReagentFibrinogen quantitation
INNOVANCE® AntithrombinAntithrombin quantitation
INNOVANCE® D-DimerD-dimer quantitation

The intended environment of use is a clinical laboratory. The instrument and reagent applications are not labeled for home use nor for patient use.

Statement of Intended Use:

The CN-series (CN-6000) is a fully automated blood coagulation analyzer intended for in vitro diagnostic use in the clinical laboratory. The instrument analyzes citrated plasma samples (3.2% sodium citrate) collected from venous blood using clotting, chromogenic and immunoassay methods.

The performance of this device has not been established in neonate and pediatric patient populations.

Summary of Substantial Equivalence:

The CN-6000 uses the same principles of operation as the CS-5100 in using venous blood samples collected in 3.2% sodium citrate tubes to analyze clotting, chromogenic and immunoassay methods in the clinical laboratory.

The main similarities between the CN-6000 instrument and the CS-5100 are listed in the device comparison Table 1 and differences are in Table 2 below:

Page 7

Table 1: Instrument Comparison – Similarities with Predicate Device

Analyzer ComponentPredicate Device CS-5100 (K150678)Proposed Device CN-6000
Intended UseThe Sysmex CS-5100 is a fully automated blood coagulation analyzer intended for in vitro diagnostic use using plasma collected from venous blood samples in 3.2% sodium citrate tubes to analyze clotting, chromogenic and immunoassay methods in the clinical laboratory. For determination of: • Prothrombin Time (PT) seconds and PT INR with Dade® Innovin® • Activated Partial Thromboplastin Time (APTT) with Dade® Actin® FSL • Fibrinogen (Fbg) with Dade® Thrombin Reagent • Antithrombin (AT) with INNOVANCE® Antithrombin • D-dimer with INNOVANCE® D-Dimer The performance of this device has not been established in neonate and pediatric patient populations.The CN-6000 is a fully automated coagulation analyzer intended for in vitro diagnostic use in the clinical laboratory. The instrument analyzes citrated plasma samples (3.2 % sodium citrate) collected from venous blood, using clotting, chromogenic and immunoassay methods. The performance of this device has not been established in neonate and pediatric patient populations.
Sample typeHuman plasma, 3.2% sodium citrateSame
Application typeClotting Applications: • Prothrombin Time (PT) seconds and PT INR with Dade® Innovin® • Activated Partial Thromboplastin Time (APTT) with Dade® Actin® FSL • Fibrinogen (Clauss) with Dade® Thrombin ReagentSame
Chromogenic Application: Antithrombin with INNOVANCE® AntithrombinSame
Immunoassay Application: D-dimer with INNOVANCE® D-DimerSame

Page 8

Analyzer ComponentPredicate Device CS-5100 (K150678)Proposed Device CN-6000
Operating PrincipleClotting/Chromogenic/Immunoassay: Transmitted Light Detection (Absorbance) at 340, 405, 575, 660 or 800 nm.Same
Specimen ProcessingAutomatic Pipetting and DilutionSame
Random AccessYesSame
Liquid Level SensingYes – Reagent and SampleSame
Bar code readerReagent and SampleSame
Sample BarcodesITF, NW-7 (CODABAR), CODE-39, JAN-13, JAN-8, CODE-128 and ISBT128Same
STAT testingYesSame
Sampling CapabilitiesNormal and Micro ModeSame
Cap PiercingCap PiercerSame
Temperature ControlSample incubation well: 37°C ± 1.0°CSame
Pipetting CapabilitiesReagent probe: 20 - 200 μL Sample probe: 4 - 270 μLSame
Reagent MixingAutomaticSame

Table 2 - Instrument Comparison – Differences with Predicate Device

Analyzer ComponentPredicate Device CS-5100 (K150678)Proposed Device CN-6000
Cleaning Solutions On-board External• CA-CLEAN I • CA-CLEAN II • Dade® Owren's Buffer • Water• CN-COAGWASHER Same
Light SourceChromogenic & Immunochemical: Halogen LampLED

Page 9

Analyzer ComponentPredicate Device CS-5100 (K150678)Proposed Device CN-6000
Probes• 2 Sample probes • 3 Reagent probes• 2 Sample probes • 2 Reagent probes
Reagent Cooling• 10°C ± 2°C, when the ambient temperature is 20°C - 28°C. • 4°C - 15°C, when the ambient temperature is 15°C - 20°C or 28-30 °C• 4°C - 12°C, when the ambient temperature is 15°C - 28°C • 4°C - 15°C, when the ambient temperature is 28°C - 30°C
Sampler• 10 sample racks can be set simultaneously• 5 sample racks can be set simultaneously

There are minor differences between the analyzers such as rinse and buffer solutions, light source, probes and reagent cooling. These differences do not impact the safety and performance of CN-6000.

Summary of Performance Testing:

A series of studies were performed that evaluated traditional laboratory performance characteristics. A summary of each study follows:

Method comparison:

Method comparison studies designed according to EP09C, 3rd edition CLSI Guideline were conducted at three clinical sites. Normal, abnormal samples, and samples at the medical decision levels were enrolled in order to adequately capture full spectrum of the analytical measuring interval (AMI). Samples were measured on both the predicate device (Sysmex® CS-5100) and the investigational device, CN-Series (CN-6000). Results were compared by Passing-Bablok regression analysis as well as Bland-Altman analysis. Results from each assay met the pre-established acceptance criteria. The following table shows a summary of Passing-Bablok regression analysis.

Application (AMI)Site ASite BSite CTotal Combined Sites
Prothrombin Time with Dade® Innovin® (8.7 – 90.0 seconds)N = 490 y = 1.012x – 0.134 r = 1.000N = 168 y = 1.000x r = 0.999N = 189 y = 1.020x – 0.206 r = 0.997N = 847 y = 1.010x – 0.100 r = 0.999
Prothrombin Time (INR) with Dade® Innovin® (0.93 – 8.00 INR)N = 472 y = 1.013x – 0.012 r = 1.000N = 163 y = 1.000x r = 0.999N = 178 y = 1.024x – 0.023 r = 0.996N = 813 y = 1.010x – 0.010 r = 0.999

Page 10

Application (AMI)Site ASite BSite CTotal Combined Sites
Activated Partial Thromboplastin Time with Dade® Actin® FSL (20.0 – 139.0) seconds)N = 401 y = 1.000x – 0.100 r = 0.997N = 126 y = 0.961x + 1.093 r = 0.998N = 111 y = 0.988x + 0.290 r = 0.994N = 638 y = 0.993x + 0.088 r = 0.996
Fibrinogen quantitation with Dade® Thrombin Reagent (50 – 860 mg/dL)N = 225 y = 1.067x + 4.517 r = 0.996N = 121 y = 1.015x – 9.987 r = 0.998N = 110 y = 1.077x – 11.731 r = 0.995N = 456 y = 1.062x – 4.092 r = 0.993
Antithrombin quantitation with INNOVANCE® Antithrombin (9.0 – 128.0% of norm)N = 221 y = 0.949x + 1.364 r = 0.997N = 115 y = 0.988x – 2.119 r = 0.998N = 114 y = 1.028x – 1.879 r = 0.997N = 450 y = 0.984x – 0.998 r = 0.995
D-dimer quantitation with INNOVANCE® D-Dimer (0.19 – 35.20 mg/L FEU)N = 194 y = 0.976x – 0.006 r = 0.998N = 101 y = 0.897x + 0.030 r = 0.999N = 100 y = 1.015x – 0.005 r = 0.998N = 395 y = 0.959x + 0.007 r = 0.997

Precision:

Precision studies were conducted over twenty days, two runs per day, and two replicates per run for a total of eighty data points per assay. The study was conducted in accordance with recommendations of CLSI EP05-A3 using commercial QC materials and plasma pools reflecting medical decision levels. The precision data are summarized in the table below.

Precision within-site

RepeatabilityWithin-Device/Lab Precision
MeanSD [assay unit]
Prothrombin Time (seconds) with Dade® Innovin®
Ci-Trol 111.00.07
Ci-Trol 228.80.22
Ci-Trol 346.60.36
Plasma pool 110.30.07
Plasma pool 215.40.09
Plasma pool 319.40.1
Plasma pool 438.60.09
Plasma pool 549.60.18

Page 11

Precision within-site

RepeatabilityWithin-Device/Lab Precision
MeanSD [assay unit]
Plasma pool 668.50.27
Prothrombin Time (INR) with Dade® Innovin®
Ci-Trol 11.030.007
Ci-Trol 22.910.025
Ci-Trol 34.900.042
Plasma pool 11.010.005
Plasma pool 22.280.011
Plasma pool 33.180.011
Plasma pool 45.110.021
Plasma pool 57.410.042
Activated Partial Thromboplastin Time (APTT) (seconds) with Dade® Actin® FSL Activated PTT Reagent
Ci-Trol 126.60.11
Ci-Trol 249.20.24
Ci-Trol 361.20.39
Plasma pool 127.90.08
Plasma pool 244.40.29
Plasma pool 388.80.33
Plasma pool 4122.31.06
Fibrinogen (mg/dL) with Dade® Thrombin Reagent
Ci-Trol 12675.4
Data-Fi Abnormal Fibrinogen Control1013.2
Control Plasma N2385.6
Control Plasma P825
Plasma pool 1620.6
Plasma pool 22203
Plasma pool 34124.3
Plasma pool 47477.8
Antithrombin (%) with INNOVANCE® Antithrombin
Control Plasma N93.51.28
Control Plasma P31.50.66
Plasma pool 117.50.78

Page 12

Precision within-site

RepeatabilityWithin-Device/Lab Precision
MeanSD [assay unit]
Plasma pool 240.90.78
Plasma pool 3104.00.95
Plasma pool 4116.31.41
D-Dimer (mg/L FEU) with INNOVANCE® D-Dimer
INNOVANCE D-Dimer Control 10.270.007
INNOVANCE D-Dimer Control 22.460.046
Plasma pool 10.190.008
Plasma pool 20.480.013
Plasma pool 32.460.042
Plasma pool 428.620.524

Linearity & Measuring Range:

Linearity studies were performed for the following calibrated assays on the CN-series (CN-6000) coagulation analyzer: Fibrinogen with Dade® Thrombin Reagent, Antithrombin with INNOVANCE® Antithrombin, and D-dimer with INNOVANCE® D-Dimer. All reagents met the predetermined acceptance criteria and supported the Analytical Measuring Interval (AMI) claim. Studies were conducted as described in CLSI EP06 2nd Edition.

ApplicationMeasured Linear RangeAnalytical Measurement Interval
Prothrombin Time (seconds) with Dade® Innovin®Not applicable8.7 – 90.0 seconds
Prothrombin Time (INR) with Dade® Innovin®Not applicable0.93 – 8.00 INR
Activated Partial Thromboplastin Time (seconds) with Dade® Actin® FSLNot applicable20.0 – 139.0 seconds
Fibrinogen with Dade® Thrombin Reagent38.4 – 900.2 mg/dL50 – 860 mg/dL
Antithrombin with INNOVANCE® Antithrombin7.56 – 130.42 % of norm.9.0 – 128.0 % of norm.
D-dimer with INNOVANCE® D-Dimer0.180 – 35.836 mg/L FEU0.19 – 35.20 mg/L FEU

Page 13

Interference Studies:

The interference study was performed on the CN-series (CN-6000) analyzer. The study was conducted with one reagent lot and five replicates per sample. All pre-established criteria were met, and the study demonstrated the proposed device has substantially equivalent optical performance related to hemolytic, icteric, lipemic, and Hydroxyethyl Starch (HES) potential interference in samples.

High Level without Interference

ReagentBase poolHemoglobinConjugated BilirubinUnconjugated BilirubinTriglyceridesHES
Dade® Innovin® (seconds)Normal1100 mg/dL44 mg/dL66 mg/dL825 mg/dL-
Pathological (Prolonged)1100 mg/dL44 mg/dL66 mg/dL825 mg/dL-
Dade® Innovin® (INR)Normal1100 mg/dL44 mg/dL66 mg/dL825 mg/dL-
Pathological (High)1100 mg/dL44 mg/dL66 mg/dL825 mg/dL-
Dade® Actin® FSL Activated PTT ReagentNormal1100 mg/dL44 mg/dL66 mg/dL825 mg/dL-
Pathological (Prolonged)1100 mg/dL44 mg/dL66 mg/dL825 mg/dL-
Dade® Thrombin ReagentPathological (Low)150 mg/dL20 mg/dL15 mg/dL605 mg/dL25.9 g/L
Normal561 mg/dL44 mg/dL66 mg/dL605 mg/dL31.4 g/L
Pathological (High)561 mg/dL44 mg/dL66 mg/dL605 mg/dL31.4 g/L
INNOVANCE® AntithrombinPathological (Low)1100 mg/dL44 mg/dL66 mg/dL495 mg/dL-
Normal1100 mg/dL44 mg/dL66 mg/dL495 mg/dL-
INNOVANCE® D-DimerNormal (Low)1100 mg/dL44 mg/dL66 mg/dL715 mg/dL-
Normal1100 mg/dL44 mg/dL66 mg/dL715 mg/dL-
Pathological (High)1100 mg/dL44 mg/dL66 mg/dL715 mg/dL-

Reagent Carryover:

This study has been designed in accordance with CLSI guidelines H57-A and EP10-A3-AMD. Two reagent lots were analyzed on the CN-series (CN-6000) analyzer. All results met the specified criteria, and the study showed that there was no reagent carryover effects from one application to another on the Automated Blood Coagulation Analyzer CN-6000.

Sample Carryover:

The sample carryover study evaluated the effects of possible contamination from one sample to the other and confirmed that the aspiration and washing of the sample probe on the CN series (CN-6000) reduces the likelihood of carryover contamination to negligible levels. The results of the sample carryover study met all specified criteria for PT, APTT, Fibrinogen, Antithrombin, and D-dimer quantitation assays.

Page 14

Limit of Blank/Detection/Quantitation

The Limit of Blank study was designed in accordance with CLSI guideline EP17-A2 and CLSI guideline EP39-ED1. The studies were carried out on one analyzer with three calibrated assays, two different reagent lots, five replicates per run per day over three days for a total of 15 measurements per analyte-free (LOB) and low analyte (LOD/LOQ) samples.

ApplicationMeasured LoB and LoDMeasured LoQ
Fibrinogen with Dade® Thrombin ReagentNot applicableLoQ – 36.1 mg/dL
Antithrombin with INNOVANCE® AntithrombinLoB – 2.21% of norm. LoD – 2.95% of norm.LoQ – 8.31% of norm.
D-Dimer with INNOVANCE® D-DimerLoB – 0.085 mg/L FEU LoD – 0.101 mg/L FEU0.182 mg/L FEU

Factor Sensitivity (PT, APTT):

The factor sensitivity study was conducted in accordance with CLSI guideline H47 Ed3. Reagent sensitivity to factors V and VII was performed using two lots of Dade® Innovin®, and reagent sensitivity to factors VIII and IX was performed using two lots of Dade® Actin® FSL. The calculated factor sensitivity results were as follows:

  • Factor V: 40.8% - 44.5%
  • Factor VII: 45.8% - 48.3%
  • Factor VIII: 40.2% - 42.8%
  • Factor IX: 33.2%

Heparin Sensitivity (APTT):

A Heparin sensitivity study was conducted for APTT assay to assess instrument performance. Samples from patients receiving unfractionated heparin (UFH) therapy were measured with both the CN-6000 and CS-5100 coagulation analyzers using two APTT reagent lots. Passing-Bablok analysis for two lots yielded the following correlation.

  • Lot 1: n = 56, y = 1.000x - 0.200, r = 0.9993
  • Lot 2: n = 56, y = 0.981x - 0.313, r = 0.9995

Lupus Anticoagulant (LA) Sensitivity Study:

The LA sensitivity study was performed in-house evaluating 2 lots of Dade® Actin® FSL reagent on a total of 24 LA positive samples. The results for the study met the specified criteria.

Page 15

Stability:

Reagents & Buffer Stability
Reagent stability was assessed for Dade® Innovin®, Dade® Actin® FSL, Dade® Thrombin, INNOVANCE® Antithrombin, INNOVANCE® D-Dimer, and Dade® Owren's Veronal Buffer. The manufacturer's claim for onboard stability for all reagents was met.

QC Stability
Studies were designed in accordance with CLSI guideline EP25. The manufacturer's claim for onboard stability for all QC materials was met.

High Dose Hook Effect

This study was designed in accordance with CLSI guideline EP34-ED1. The study was conducted with one reagent lot on one analyzer. The delta Optical Density, dOD, of each measurement, the presence of the antigen excess flag, and whether the instrument performed an automatic redilution (samples measuring 4.4mg/L FEU or less) were recorded for each sample.

The pre-established acceptance criterion of appropriate instrument flagging with "antigen excess", and no likelihood of reporting an erroneously low result due to High Dose Hook effect up to 500 mg/L FEU was confirmed. Therefore, the acceptance criterion was met for this study.

Matrix Comparison (Bridging Studies):

Auto-Dilution vs Manual Dilution:
The Auto-Dilution versus Manual dilution study performed on the CN-Series (CN-6000). Testing was performed for fibrinogen with Dade® Thrombin Reagent and D-Dimer with INNOVANCE® D-Dimer. A minimum of three samples per assay were tested, with five replicates per sample over three days. The pre-defined acceptance criteria were met for this study.

Uncapped Specimens vs Capped Specimens:
The objective of the uncapped versus capped specimens study was to determine equivalence of Prothrombin time (PT) with Dade® INNOVIN®, Activated partial thromboplastin time (APTT) with Dade® Actin® FSL, Fibrinogen with Dade® Thrombin Reagent, Antithrombin (AT) with INNOVANCE® Antithrombin and D-Dimer with INNOVANCE® D-Dimer results obtained from open (i.e., uncapped) blood collection tube samples versus closed (i.e., capped) blood collection tube samples. The pre-defined acceptance criteria were met.

Frozen Specimens vs Fresh Specimens:
The objective of the frozen versus fresh specimens study was to determine equivalency of

Page 16

Prothrombin time (PT) with Dade® INNOVIN®, Activated partial thromboplastin time (APTT) with Dade® Actin® FSL, Fibrinogen with Dade® Thrombin Reagent, Antithrombin (AT) with INNOVANCE® Antithrombin and D-Dimer with INNOVANCE® D-Dimer results obtained from frozen human citrated plasma versus fresh human citrated plasma on the CN-6000 analyzer. The results for this study were determined to be acceptable.

Micro Mode vs Normal Mode Analysis:
The objective of the micro mode analysis versus normal mode analysis study was to assess comparability of Prothrombin time (PT) with Dade® INNOVIN®, Activated partial thromboplastin time (APTT) with Dade® Actin® FSL, Fibrinogen with Dade® Thrombin Reagent, Antithrombin (AT) with INNOVANCE® Antithrombin and D-Dimer with INNOVANCE® D-Dimer results measured in the micro mode and normal mode sample processing on the CN-6000 analyzer. The pre-established acceptance criteria were met.

Reference Range:

The Reference Interval study was designed to establish the normal reference ranges on the CN-Series (CN-6000) analyzer for the adult population using Dade® Innovin®, Dade® Actin® FSL, Dade® Thrombin, INNOVANCE® Antithrombin, and INNOVANCE® D-Dimer reagents. The study was conducted at three external clinical sites within the United States using adult samples.

AssayReference Interval Sysmex® CN-6000
Prothrombin Time (seconds) using Dade® Innovin2.5th – 97.5th 9.9 – 12.3 (sec)
Prothrombin Time (INR) using Dade® Innovin2.5th – 97.5th 0.93 – 1.16 (INR)
Activated Partial Thromboplastin Time using Dade® Actin® FSL2.5th – 97.5th 23.8 – 32.0 (sec)
Fibrinogen quantitation using Dade® Thrombin Reagent2.5th – 97.5th 192 – 440 (mg/dL)
Antithrombin quantitation using Dade® Thrombin Reagent2.5th – 97.5th 83.7 – 121.6 (%)
D-Dimer quantitation using INNOVANCE® D-Dimerminimum – 97.5th < 0.19 – 1.27 (mg/L FEU)

Conclusion:

The CN-series (CN-6000) is similar to the CS-5100 in its intended use and system technological characteristics. There are minor differences in the number of probes, light source for the CN-series, rinse & buffer cleaning solutions, reagent cooling temperature and number of sample racks.

The list of assays used with the CN-series is the same as with the previously cleared assays in the CS-5100 (K150678).

Page 17

The technological characteristics and operating principles of the CN-series are essentially the same as the predicate CS-5100. In addition, performance testing did not raise any issues with safety and effectiveness of the CN-series when compared with the CS-5100 instrument.

The data from the analytical and clinical testing supports that the CN-series (CN-6000) instrument is substantially equivalent to the previously cleared CS-5100 instrument (K150678), and the device is safe and effective for its intended use.

§ 864.5425 Multipurpose system for in vitro coagulation studies.

(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.