AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EVIS EXERA III BRONCHOVIDEOSCOPES OLYMPUS BF-XP190 is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

EVIS EXERA III BRONCHOVIDEOSCOPES OLYMPUS BF-P190 is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

EVIS EXERA III BRONCHOVIDEOSCOPES OLYMPUS BF-XT190 is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

BRONCHOVIDEOSCOPE OLYMPUS BF-H1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

Device Description

The EVIS EXERA III BRONCHOVIDEOSCOPES (OLYMPUS BF-XP190, OLYMPUS BF-P190, and BF-XT190) and BRONCHOVIDEOSCOPE BF-H1100 and BF-1TH1100 are used for endoscopic diagnosis and treatment within the respiratory organs. These endoscopes consist of three parts: the control section, the insertion section, and the connector section.

AI/ML Overview

The provided FDA 510(k) clearance documentation for the Olympus Bronchovideoscopes (K250862) primarily demonstrates substantial equivalence based on technological characteristics and bench testing. It does not contain information about comparative effectiveness studies (like MRMC studies for AI devices), expert ground truth establishment, or typical performance metrics associated with AI/software-as-a-medical-device (SaMD) clearances.

This submission focuses on:

  • Technological Equivalence: Showing that the devices (BF-XP190, BF-P190, BF-XT190, BF-H1100, and BF-1TH1100) are fundamentally the same as their predicates, with the primary change being compatibility with a new video system center (CV-1500) and updated labeling related to laser/high-frequency/APC systems.
  • Bench Testing: Verification of physical and imaging performance parameters (e.g., thermal safety, color performance, resolution, noise, video latency) to ensure they meet specifications, particularly when combined with the new video system.
  • Animal Testing: To assess the imaging modes (WLI, NBI, TXI, BAI-MAC) with the new video processor.

Therefore, the requested information about "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of AI/software performance (e.g., sensitivity, specificity, MRMC studies, ground truth establishment by experts) is not present in this document. The document explicitly states that "Software Testing and Cybersecurity" was "not performed" due to "no design, material, sterilization, reprocessing, packaging, shelf life, or software changes" (Page 29). This implies that the device itself is a hardware endoscope, and any software associated with it is considered an intrinsic part of its established functionality, not a new or significantly changed software component requiring a separate performance study with clinical endpoints or AI evaluation.

The "acceptance criteria" here relate to the successful completion of the listed bench (and limited animal) tests, demonstrating that the new combination (endoscope + CV-1500) functions as intended and safely, similar to the predicate combinations.

Below is a table summarizing the "acceptance criteria" and "reported device performance" as derived from the document's comparison tables and performance data section, which are primarily about technical specifications and functional verification, not software/AI performance metrics.


Acceptance Criteria and Reported Device Performance (Summary based on provided text)

Since this 510(k) is for existing endoscopes with a new video system and updated labeling, the "acceptance criteria" are implied by the extensive comparison tables (Tables 1-5) which show the subject devices having nearly identical technical specifications to their predicate devices, and the successful completion of specified bench and animal testing. The performance data section doesn't list specific quantitative acceptance criteria for each test but rather states that tests were conducted to "ensure that the subject device performs as intended and meet design specifications."

Acceptance Criteria CategorySpecific Acceptance Criteria (Implied/Defined by Predicate Equivalence)Reported Device Performance (as stated or implied by equivalence)
Indications for UseIdentical to predicate devices.The Indications for Use for the Subject Device is exactly the same as the Indications for Use for the equivalent Predicate Device (Page 9).
Material CompositionIdentical materials used for key components (e.g., Insertion Tube, Bending Section Rubber).Materials are identical to predicates (e.g., Fluoro Resin, Fluoro Rubber, Epoxy Glue, Polytetrafluoro-ethylene, Polysulfone, Glass/Sapphire) (Pages 12-13, 16-17, 20-21).
Physical DimensionsMaintain specific dimensions (e.g., Outer Diameter, Working Length, Instrument Channel inner diameter).Dimensions are identical to predicates across all models (e.g., BF-P190 Outer Diameter of Distal End Φ4.2mm, Working Length 600mm, Instrument Channel inner diameter φ2.0; BF-XP190 Outer Diameter of Distal End Φ3.1mm; BF-XT190 Outer Diameter of Distal End Φ6.1mm; BF-H1100 Distal end outer diameter ø 4.9 mm; BF-1TH1100 Distal end outer diameter ø 5.8 mm) (Pages 12, 16, 19-20, 23, 26).
Optical/Imaging PerformanceMaintain specific optical characteristics (e.g., Depth of Field, Field of View, Pixels, CCD type).Identical optical characteristics to predicates (e.g., Depth of Field 2-50mm/3-100mm, Field of View 110°/120°, Total Number of Pixels 50,621 / 465,588, Color CCD) (Pages 11-12, 15, 19, 22, 25).
Software FunctionalityIdentical software functions (e.g., pixel defect correction, scope info, NBI, Pre-freeze).All listed software functions are "Available" and match the predicate devices (Pages 13, 17, 21, 24, 27-28). The RDI, TXI, and BAI-MAC modes are explicitly stated as "Available" for the subject devices when previously "Not Available" for some predicates, implying this is the new feature enabled by the CV-1500 system, and their performance was assessed.
SafetyMeet safety standards for thermal, photobiological, and ignition factors (especially with laser/HF/APC compatibility).Bench testing conducted for: Thermal Safety, Photobiological Safety, Ignition Factors Safety Study (Laser, High-Frequency, APC, Scope Comparison) to ensure performance and safety (Page 28).
Functional VerificationPerformance as intended with the new CV-1500 video system center.Bench testing (Color Performance, Resolution, Noise and Dynamic Range, Image Intensity Uniformity, Video Latency, RDI, TXI, BAI-MAC, Automatic Brightness Adjustment, Pre-Freeze) was conducted to ensure intended performance (Page 28). Animal testing performed to assess imaging modes (WLI, NBI, TXI, BAI-MAC) with CV-1500 (Page 29).
Risk ManagementRisks identified and mitigated in accordance with ISO 14971:2019.Risk management performed, and design verification tests and acceptance criteria identified as a result (Page 29).

Study Details (based on provided text)

The document describes performance testing rather than a comparative clinical study for AI/software-as-a-medical-device.

  1. A table of acceptance criteria and the reported device performance:

    • See table above. The acceptance criteria are largely implied by the equivalence to the predicate devices in terms of physical, optical, and functional characteristics, and the successful completion of specified bench and animal tests. Quantitative metrics for these tests are not provided in this summary but would have been part of the full submission.
  2. Sample sizes used for the test set and the data provenance:

    • Test set sample size: Not specified. The document mentions "bench testing" and "animal testing." For bench tests, it typically refers to a small number of devices or engineered test setups. For animal testing, the number of animals or studies is not provided.
    • Data provenance: Not specified. It can be inferred that the testing was conducted by or on behalf of Olympus Medical Systems Corp. in Japan, given the manufacturing site and submitter location. It does not state if the data is retrospective or prospective, or from which country/region the "animal" data would originate.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable / Not specified. This type of information is typically provided for studies evaluating AI algorithms or diagnostic accuracy, where human experts establish a ground truth for imaging interpretation. The present submission is for a hardware endoscope system, and its performance evaluation relies on engineering specifications and functional testing, not expert interpretation of diagnostic images.
  4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable / Not specified. Adjudication methods are relevant for human reader studies or expert ground truth establishment, which are not detailed here.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, not performed / Not applicable. This submission is not for an AI companion diagnostic or an AI-software-as-a-medical-device. It's for an endoscope system. The document explicitly lists "Software Testing and Cybersecurity" and "Clinical" as "not performed" because there were no fundamental software changes beyond integration with a new video system, which itself does not constitute an AI component in the context of this 510(k) summary. The NBI, RDI, TXI, and BAI-MAC modes are imaging enhancements, not AI algorithms.
  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This relates to AI/software performance, which is not the focus of this 510(k).
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable. For this device (endoscope), ground truth typically relates to engineering specifications (e.g., accurate measurements, clear image quality as determined by reproducible test patterns, proper functionality of mechanical parts). For the animal testing of imaging modes, the "ground truth" would be the direct observation of the animal's internal anatomy via the endoscope itself and comparison to expected/known characteristics, not an independent "pathology" or "outcomes" ground truth in a diagnostic sense.
  8. The sample size for the training set:

    • Not applicable. This submission is not for an AI system that requires a "training set."
  9. How the ground truth for the training set was established:

    • Not applicable. As above, no AI training set is described.

FDA 510(k) Clearance Letter and Documentation

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 23, 2025

Olympus Medical Systems Corp.
℅ Teffany Hutto
Project Manager, Regulatory Affairs
Olympus Corporation of the Americas
800 West Park Drive
Westborough, Massachusetts 01581

Re: K250862

Trade/Device Name: Evis Exera IIi Bronchovideoscope (olympus Bf-xp190); Evis Exera IIi Bronchovideoscope (olympus Bf-p190); Evis Exera IIi Bronchovideoscope (olympus Bf-xt190); Bronchovideoscope (olympus Bf-h1100); Bronchovideoscope (olympus Bf-1th1100)

Regulation Number: 21 CFR 874.4680
Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories
Regulatory Class: Class II
Product Code: EOQ
Dated: May 29, 2025
Received: May 29, 2025

Dear Teffany Hutto:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250862 - Teffany Hutto Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250862 - Teffany Hutto Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shuchen Peng -S

Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.

K250862

Please provide the device trade name(s).

EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-XP190);
EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-P190);
EVIS EXERA III BRONCHOVIDEOSCOPE (OLYMPUS BF-XT190);
BRONCHOVIDEOSCOPE (OLYMPUS BF-H1100);
BRONCHOVIDEOSCOPE (OLYMPUS BF-1TH1100)

Please provide your Indications for Use below.

EVIS EXERA III BRONCHOVIDEOSCOPES OLYMPUS BF-XP190 is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

EVIS EXERA III BRONCHOVIDEOSCOPES OLYMPUS BF-P190 is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

EVIS EXERA III BRONCHOVIDEOSCOPES OLYMPUS BF-XT190 is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

BRONCHOVIDEOSCOPE OLYMPUS BF-H1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, Endo Therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

Please select the types of uses (select one or both, as applicable).

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

Page 5

Traditional 510(k) Notification

EVIS EXERA III Bronchovideoscope BF-P190, BF-XP190, and BF-XT190
BF-H1100 and BF-1TH1100

510(k) Summary
Page 1 of 25

Date Prepared: April 21, 2025

510(k) Summary

1. GENERAL INFORMATION

■ 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP.
2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507, Japan

■ Contact Person: Teffany Hutto
Olympus Corporation of the Americas
3500 Corporate Parkway PO Box 610
Center Valley, PA 18034-0610, USA
Phone: 512-508-6550
Fax: 484-896-7128
Email: teffany.hutto@olympus.com

■ Manufacturing site: Aizu Olympus Co., Ltd.,
3-1-1 Niiderakita, Aizuwakamatsu-shi, Fukushima 965-8520, Japan

2. DEVICE IDENTIFICATION

Trade/Device NameEVIS EXERA III Bronchovideoscope BF-XP190
Model NameBF-XP190
Common NameBronchovideoscope
Regulation Number874.4680
Regulation NameBronchoscope (flexible or rigid) and accessories
Regulatory ClassII
Product CodeEOQ (Bronchoscope (Flexible or rigid))
NWB (Endoscope, Accessories, Narrow Band Spectrum)
Classification PanelEar, Nose and Throat

Page 6

Trade/Device NameEVIS EXERA III Bronchovideoscope BF-P190
Model NameBF-P190
Common NameBronchovideoscope
Regulation Number874.4680
Regulation NameBronchoscope (flexible or rigid) and accessories
Regulatory ClassII
Product CodeEOQ (Bronchoscope (Flexible or rigid))
NWB (Endoscope, Accessories, Narrow Band Spectrum)
Classification PanelEar, Nose and Throat
Trade/Device NameEVIS EXERA III Bronchovideoscope BF-XT190
Model NameBF-XT190
Common NameBronchovideoscope
Regulation Number874.4680
Regulation NameBronchoscope (flexible or rigid) and accessories
Regulatory ClassII
Product CodeEOQ (Bronchoscope (Flexible or rigid))
NWB (Endoscope, Accessories, Narrow Band Spectrum)
Classification PanelEar, Nose and Throat
Trade/Device NameBronchovideoscope BF-H1100
Model NameBF-H1100
Common NameBronchovideoscope
Regulation Number874.4680
Regulation NameBronchoscope (flexible or rigid) and accessories
Regulatory ClassII
Product CodeEOQ (Bronchoscope (Flexible or rigid))
NWB (Endoscope, Accessories, Narrow Band Spectrum)
Classification PanelEar, Nose and Throat

Page 7

Trade/Device NameBronchovideoscope BF-1TH1100
Model NameBF-1TH1100
Common NameBronchovideoscope
Product CodeEOQ (Bronchoscope (Flexible or rigid))
NWB (Endoscope, Accessories, Narrow Band Spectrum)
Regulation NameBronchoscope (flexible or rigid) and accessories
Regulatory ClassII
Product CodeEOQ (Bronchoscope (Flexible or rigid))
Classification PanelEar, Nose and Throat

3. PREDICATE DEVICES

• Predicate devices (BF-XP190, BF-P190 and BF-XT190)

Device name510(k) Submitter510(k) No.
EVIS EXERA III BRONCHOVIDEOSCOPE BF-P190, BF-XP190OLYMPUS MEDICAL SYSTEMS CORP.K201758
EVIS EXERA III BRONCHOVIDEOSCOPE BF-XT190OLYMPUS MEDICAL SYSTEMS CORPK183419

• Reference devices (BF-XP190, BF-P190 and BF-XT190)

Device name510(k) Submitter510(k) No.
BRONCHOVIDEOSCOPE OLYMPUS BF-H1100OLYMPUS MEDICAL SYSTEMS CORP.K222861
BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100OLYMPUS MEDICAL SYSTEMS CORPK222861

• Predicate devices (BF-H1100 and BF1TH1100)

Device name510(k) Submitter510(k) No.
BRONCHOVIDEOSCOPE OLYMPUS BF-H1100OLYMPUS MEDICAL SYSTEMS CORP.K222861
BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100OLYMPUS MEDICAL SYSTEMS CORPK222861

4. DEVICE DESCRIPTION

The EVIS EXERA III BRONCHOVIDEOSCOPES (OLYMPUS BF-XP190, OLYMPUS BF-P190, and BF-XT190) and BRONCHOVIDEOSCOPE BF-H1100 and BF-1TH1100 are used for endoscopic diagnosis and treatment within the respiratory organs. These endoscopes consist of three parts: the control section, the insertion section, and the connector section.

Page 8

5. PRINCIPLE OF OPERATION

The endoscope consists of three parts: the control section, the insertion section, and the connector section. The basic principle including user interface and operation for the procedure of the endoscope is identical to that of the predicate device.

1) Control section

The UP/DOWN angulation control knob and the RIGHT/LEFT angulation control knob on the control section is connected to the tip of the bending section by a series of wires. By operating the UP/DOWN angulation control knob and the RIGHT/LEFT angulation control knob, the bending section at the distal end bends vertically or parallel to guide the distal end for insertion and observation.

The observation mode can be selected by focus switching function, "near focus mode" featuring ground-breaking resolving power for close observation or "normal focus mode" for normal observation. To realize the dual focus mechanism, Voice Coil Motor (VCM) is incorporated as an actuator.

The endoscope contains a cylinder to attach a suction valve for suction and air/water valve. Depressing the suction valve will allow the physician to use the endoscope to suction any fluids which are obscuring a good view of the tissue. Therapeutic instruments can be passed through the instrument channel for performing endoscopic biopsy and other therapies. Depressing the air/water valve will allow the doctor to feed water through the endoscope for lens washing. It also can be operated to feed air for removing any fluids or debris adhering to the objective lens.

2) Insertion section

The insertion section has main parts including the image guide, light guides that bring light from the video system center through the endoscope, and instrument channel where therapeutic tools can be pushed in and out (also the suction channel).

3) Connector section

The connector section connects the endoscope with the video system center (CV-1500) through the universal cord.

Page 9

6. INDICATIONS FOR USE

EVIS EXERA III BRONCHOVIDEOSCOPES OLYMPUS BF-XP190 is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

EVIS EXERA III BRONCHOVIDEOSCOPES OLYMPUS BF-P190 is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

EVIS EXERA III BRONCHOVIDEOSCOPES OLYMPUS BF-XT190 is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

BRONCHOVIDEOSCOPE OLYMPUS BF-H1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

INDICATIONS FOR USE COMPARISON

The Indications for Use for the Subject Device is exactly the same as the Indications for Use for the equivalent Predicate Device.

Page 10

7. COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject devices have the same technological characteristics and design as the applicable predicate devices. There have been no modifications from the prior clearances of these devices including:

  • Design
  • Materials
  • Sterilization
  • Shelf Life
  • Reprocessing
  • Packaging
  • Software

This 510(k) is to request clearance for use of the BF-XP190, BF-P190, and BF-XT-190 scopes with the recently cleared EVIS X1 VIDEO SYSTEM CENTER CV-1500 under K222861. The BF-XP190, BF-P190, and BF-XT-190 scopes will be used with imaging modes WLI, NBI, RDI, TXI, and BAI-MAC.

This 510(k) will also request clearance to update the labeling to include new instructions on proper use of these scopes, along with the other subject devices, with laser, high frequency, and APC systems for all Subject Devices (BF-P190, BF-XP190, BF-XT190, BF-H1100, and BF-1TH1100)

A side-by-side comparison of the subject device and the predicate device is provided below.

Table 1 – BF-P190

ItemSubject DevicePredicate Device
BF-P190 (combination with CV-1500)BF-P190 (combination with CV-190/CLV-190)
TBDK201758
Regulation Number874.4680874.4680
Regulatory ClassClass IIClass II
Product CodeEOQ; Bronchoscope (Flexible Or Rigid)EOQ; Bronchoscope (Flexible Or Rigid)
Review PanelEar Nose and ThroatEar Nose and Throat
Common nameBRONCHOSCOPEBRONCHOSCOPE
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.OLYMPUS MEDICAL SYSTEMS CORP.

Page 11

ItemSubject DevicePredicate Device
BF-P190 (combination with CV-1500)BF-P190 (combination with CV-190/CLV-190)
TBDK201758
Indications for UseThe EVIS EXERA III Bronchovideoscopes BF-P190 are intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190 This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.
Use EnvironmentHealthcare facility/hospitalHealthcare facility/hospital
Single-Use/ReuseReusableReusable
Duration and type of contactSurface-contacting device in contact with mucosal membranes. The contact duration is limited exposure (<24 hrs).Surface-contacting device in contact with mucosal membranes. The contact duration is limited exposure (<24 hrs).
Depth of Field2 - 50 mm2 - 50 mm
Direction of Forward View0゜ (Forward viewing)0゜ (Forward viewing)
Field of View(FOV)110°110°
Distortion
Magnification of Objective Lens at Optimum Working Distance0.0170.017
Pixels per square mm226,760226,760
Active Area of CCD Chip(H)477μm×(V)468μm(H)477μm×(V)468μm
Total Number of Pixels50,62150,621
Size of Pixel(H)2.1μm×(V)2.1μm(H)2.1μm×(V)2.1μm

Page 12

ItemSubject DevicePredicate Device
BF-P190 (combination with CV-1500)BF-P190 (combination with CV-190/CLV-190)
TBDK201758
Type of CCD ChipColorColor
Narrow Band Image (NBI) observationAvailableAvailable
Red Dichromatic Imaging (RDI) observationAvailableNot Available
Pre-freeze functionAvailableAvailable
Electronic zoom functionNot AvailableNot Available
Electronic shutter functionNot AvailableNot Available
Records of endoscope's informationAvailableAvailable
Outer Diameter of Distal EndΦ4.2mmΦ4.2mm
Outer Diameter of Insertion TubeΦ4.1mmΦ4.1mm
Bending Section AngulationUP:210°,DOWN:130°UP:210°,DOWN:130°
Working Length600mm600mm
Instrument Channel inner diameter [mm]φ2.0φ2.0
Distal end enlarged1 Objective lens2 Light guide lens3 Instrument channel outlet1 Objective lens2 Light guide lens3 Instrument channel outlet
Total Length880mm880mm
Minimum visible distance3mm3mm
Direction from which EndoTherapy accessories enter and exit the endoscopic image
Material
Insertion Tube (Outer layer)Fluoro Resin (F-COAT-KAI)Fluoro Resin (F-COAT-KAI)
Bending section rubberFluoro Rubber (GR1005)Fluoro Rubber (GR1005)

Page 13

ItemSubject DevicePredicate Device
BF-P190 (combination with CV-1500)BF-P190 (combination with CV-190/CLV-190)
TBDK201758
Adhesive on the bending rubberEpoxy Glue (ES-19MC)Epoxy Glue (ES-19MC)
Distal end adhesiveEpoxy Glue (ES-19MC)Epoxy Glue (ES-19MC)
Instrument channelPolytetrafluoro-ethylenePolytetrafluoro-ethylene
Distal endPolysulfone (P1700-216)Polysulfone (P1700-216)
Objective lensGlass (Sapphire)Glass (Sapphire)
Light guide lensGlass (OHARA S-LAH58)Glass (OHARA S-LAH58)

Control section

Configuration of Control section and location of scope switch
The insertion tube rotation angle by simply turning a ring on the control section of the scope.left or right up to 120°

Video connection section

Scope connectorWater resistant one-touch connectorWater resistant one-touch connector
Configuration of Video connection section and Light guide connector section

Software

Correction of pixel defectHaveHave
Scope information

Page 14

ItemSubject DevicePredicate Device
BF-P190 (combination with CV-1500)BF-P190 (combination with CV-190/CLV-190)
TBDK201758
Individual scope information (Scope ID)HaveHave

Compatibility of endotherapy accessories

High frequency treatment accessoriesAvailableAvailable
Laser treatment accessoriesNot AvailableNot Available

Other

Total weight840g840g

Table 2 – BF-XP190

ItemSubject DevicePredicate Device
BF-XP190 (combination with CV-1500)BF-XP190 (combination with CV-190/CLV-190)
TBDK201758
Regulation Number874.4680874.4680
Regulatory ClassClass IIClass II
Product CodeEOQ; Bronchoscope (Flexible Or Rigid)EOQ; Bronchoscope (Flexible Or Rigid)
Review PanelEar Nose and ThroatEar Nose and Throat
Common nameBRONCHOSCOPEBRONCHOSCOPE
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.OLYMPUS MEDICAL SYSTEMS CORP.
Indications for UseThe EVIS EXERA III Bronchovideoscopes BF-XP190 are intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190 This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.
Use EnvironmentHealthcare facility/hospitalHealthcare facility/hospital
Single-Use/ReuseReusableReusable

Page 15

ItemSubject DevicePredicate Device
BF-XP190 (combination with CV-1500)BF-XP190 (combination with CV-190/CLV-190)
TBDK201758
Duration and type of contactSurface-contacting device in contact with mucosal membranes. The contact duration is limited exposure (<24 hrs).Surface-contacting device in contact with mucosal membranes. The contact duration is limited exposure (<24 hrs).
Depth of Field2 - 50 mm2 - 50 mm
Direction of Forward View0゜ (Forward viewing)0゜ (Forward viewing)
Field of View(FOV)110°110°
Distortion
Magnification of Objective Lens at Optimum Working Distance0.0170.017
Pixels per square mm226,760226,760
Active Area of CCD Chip(H)477μm×(V)468μm(H)477μm×(V)468μm
Total Number of Pixels50,62150,621
Size of Pixel(H)2.1μm×(V)2.1μm(H)2.1μm×(V)2.1μm
Type of CCD ChipColorColor
Narrow Band Image (NBI) observationAvailableAvailable
Red Dichromatic Imaging (RDI) observationAvailableNot Available
Pre-freeze functionAvailableAvailable
Electronic zoom functionNot AvailableNot Available
Electronic shutter functionNot AvailableNot Available
Records of endoscope's informationAvailableAvailable

Page 16

ItemSubject DevicePredicate Device
BF-XP190 (combination with CV-1500)BF-XP190 (combination with CV-190/CLV-190)
TBDK201758
Outer Diameter of Distal EndΦ3.1 mm (Tapered part of distal tip: ø 2.9 mm)Φ3.1 mm (Tapered part of distal tip: ø 2.9 mm)
Outer Diameter of Insertion TubeΦ2.8 mmΦ2.8 mm
Bending Section AngulationUP:210°,DOWN:130°UP:210°,DOWN:130°
Working Length600mm600mm
Instrument Channel inner diameter [mm]φ1.2φ1.2
Distal end enlarged1 Objective lens2 Light guide lens3 Instrument channel outlet1 Objective lens2 Light guide lens3 Instrument channel outlet
Total Length880mm880mm
Minimum visible distance3mm3mm
Direction from which EndoTherapy accessories enter and exit the endoscopic image
Material
Insertion Tube (Outer layer)Fluoro Resin (F-COAT-KAI)Fluoro Resin (F-COAT-KAI)
Bending section rubberFluoro Rubber (GR1005)Fluoro Rubber (GR1005)
Adhesive on the bending rubberEpoxy Glue (ES-19MC)Epoxy Glue (ES-19MC)
Distal end adhesiveEpoxy Glue (ES-19MC)Epoxy Glue (ES-19MC)
Instrument channelPolytetrafluoro-ethylenePolytetrafluoro-ethylene
Distal endSteel (SUS303)Steel (SUS303)
Objective lensGlass (Sapphire)Glass (Sapphire)
Light guide lensGlass (OHARA S-LAH58)Glass (OHARA S-LAH58)

Page 17

ItemSubject DevicePredicate Device
BF-XP190 (combination with CV-1500)BF-XP190 (combination with CV-190/CLV-190)
TBDK201758

Control section

Configuration of Control section and location of scope switch
The insertion tube rotation angle by simply turning a ring on the control section of the scope.left or right up to 120°

Video connection section

Scope connectorWater resistant one-touch connectorWater resistant one-touch connector
Configuration of Video connection section and Light guide connector section

Software

Correction of pixel defectHaveHave

Scope information

Individual scope information (Scope ID)HaveHave

Compatibility of endotherapy accessories

High frequency treatment accessoriesNot AvailableNot Available

Page 18

ItemSubject DevicePredicate Device
BF-XP190 (combination with CV-1500)BF-XP190 (combination with CV-190/CLV-190)
TBDK201758
Laser treatment accessoriesNot AvailableNot Available

Other

Total weight840g840g

Table 3 – BF-XT190

ItemSubject DevicePredicate Device
BF-XT190 (combination with CV-1500)BF-XT190 (combination with CV-190/CLV-190)
TBDK183419
Regulation Number874.4680874.4680
Regulatory ClassClass IIClass II
Product CodeEOQ; Bronchoscope (Flexible Or Rigid)EOQ; Bronchoscope (Flexible Or Rigid)
Review PanelEar Nose and ThroatEar Nose and Throat
Common nameBRONCHOSCOPEBRONCHOSCOPE
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.OLYMPUS MEDICAL SYSTEMS CORP.
Indications for UseThe EVIS EXERA III Bronchovideoscopes BF-XT190 are intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XT190 This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.
Use EnvironmentHealthcare facility/hospitalHealthcare facility/hospital
Single-Use/ReuseReusableReusable
Duration and type of contactSurface-contacting device in contact with mucosal membranes. The contact duration is limited exposure (<24 hrs).Surface-contacting device in contact with mucosal membranes. The contact duration is limited exposure (<24 hrs).

Page 19

ItemSubject DevicePredicate Device
BF-XT190 (combination with CV-1500)BF-XT190 (combination with CV-190/CLV-190)
TBDK183419
Depth of Field2 - 50 mm2 - 50 mm
Direction of Forward View0゜ (Forward viewing)0゜ (Forward viewing)
Field of View(FOV)110°110°
Distortion
Magnification of Objective Lens at Optimum Working Distance0.0170.017
Pixels per square mm226,760226,760
Active Area of CCD Chip(H)477μm×(V)468μm(H)477μm×(V)468μm
Total Number of Pixels50,62150,621
Size of Pixel(H)2.1μm×(V)2.1μm(H)2.1μm×(V)2.1μm
Type of CCD ChipColorColor
Narrow Band Image (NBI) observationAvailableAvailable
Red Dichromatic Imaging (RDI) observationAvailableNot Available
Pre-freeze functionAvailableAvailable
Electronic zoom functionNot AvailableNot Available
Electronic shutter functionNot AvailableNot Available
Records of endoscope's informationAvailableAvailable
Outer Diameter of Distal EndΦ6.1mmΦ6.1mm

Page 20

ItemSubject DevicePredicate Device
BF-XT190 (combination with CV-1500)BF-XT190 (combination with CV-190/CLV-190)
TBDK183419
Outer Diameter of Insertion TubeΦ6.3mmΦ6.3mm
Bending Section AngulationUP:180°,DOWN:130°UP:180°,DOWN:130°
Working Length600mm600mm
Instrument Channel inner diameter [mm]φ3.2φ3.2
Distal end enlarged1 Objective lens2 Light guide lens3 Instrument channel outlet1 Objective lens2 Light guide lens3 Instrument channel outlet
Total Length880mm880mm
Minimum visible distance1.5mm1.5mm
Direction from which EndoTherapy accessories enter and exit the endoscopic image
Material
Insertion Tube (Outer layer)Fluoro Resin (F-COAT-KAI)Fluoro Resin (F-COAT-KAI)
Bending section rubberFluoro Rubber (GR1005)Fluoro Rubber (GR1005)
Adhesive on the bending rubberEpoxy Glue (ES-19MC)Epoxy Glue (ES-19MC)
Distal end adhesiveEpoxy Glue (ES-19MC)Epoxy Glue (ES-19MC)
Instrument channelPolytetrafluoro-ethylenePolytetrafluoro-ethylene
Distal endPolysulfone (P1700-216)Polysulfone (P1700-216)
Objective lensGlass (Sapphire)Glass (Sapphire)
Light guide lensGlass (Sapphire)Glass (Sapphire)

Page 21

ItemSubject DevicePredicate Device
BF-XT190 (combination with CV-1500)BF-XT190 (combination with CV-190/CLV-190)
TBDK183419
Configuration of Control section and location of scope switchThe insertion tube rotation angle by simply turning a ring on the control section of the scope.left or right up to 120°

Video connection section

Scope connectorWater resistant one-touch connectorWater resistant one-touch connector
Configuration of Video connection section and Light guide connector section

Software

Correction of pixel defectHaveHave

Scope information

Individual scope information (Scope ID)HaveHave

Compatibility of endotherapy accessories

High frequency treatment accessoriesAvailableAvailable
Laser treatment accessoriesAvailableAvailable

Other

Total weight840g840g

Page 22

Table 4 – BF-H1100

SUBJECT DEVICEPREDICATE DEVICE
BF-H1100 BronchoscopeBF-H1100 Bronchoscope (K222861)
ManufacturerOLYMPUS MEDICAL SYSTEMS CORPOLYMPUS MEDICAL SYSTEMS CORP
Classification & Regulation NumberClass II21 CFR §874.4680,21 CFR §874.4760,21 CFR §876.1500Class II21 CFR §874.4680,21 CFR §874.4760,21 CFR §876.1500
Product CodeEOQ: Bronchoscope (Flexible or rigid)EOB: Nasopharyngoscope (Flexible or rigid)NWB: endoscope, accessories, narrow band spectrumEOQ: Bronchoscope (Flexible or rigid)EOB: Nasopharyngoscope (Flexible or rigid)NWB: endoscope, accessories, narrow band spectrum
Indications for UseThe "BRONCHOVIDEOSCOPE OLYMPUS BF-H1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.The "BRONCHOVIDEOSCOPE OLYMPUS BF-H1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.
Type- CCD- CYM color filterSequential read image signal- CCD- CYM color filterSequential read image signal
Number of pixels465,588465,588
Pixel size1.5 μm (H) × 1.5 μm (V)1.5 μm (H) × 1.5 μm (V)

Optical System

Direction of View
Field of View120°120°
Depth of Field3 - 100 mm3 - 100 mm

Page 23

SUBJECT DEVICEPREDICATE DEVICE
BF-H1100 BronchoscopeBF-H1100 Bronchoscope (K222861)

Distortion

Insertion Section

Distal end outer diameterø 4.9 mmø 4.9 mm
Distal end enlarged1 Objective lens2 Light guide Lens3 Instrument channel outlet1 Objective lens2 Light guide Lens3 Instrument channel outlet
Insertion tube outer diameterø 4.9 mmø 4.9 mm
Insertion section working length600 mm600 mm

Instrument Channel

Channel inner diameterø 2.2 mmø 2.0 mm
Minimum visible distance3 mm3 mm
Direction from which EndoTherapy accessories enter and exit the endoscopic image

Page 24

SUBJECT DEVICEPREDICATE DEVICE
BF-H1100 BronchoscopeBF-H1100 Bronchoscope (K222861)

Configuration

The insertion tube rotation angle by simply turning a ring on the control section of the scopeleft or right up to 120°left or right up to 120°

Connector Section

Configuration

Bending Section

Angulation rangeUP 210°DOWN 130°UP 210°DOWN 130°

Total Length

Total length880 mm880 mm

Total Weight

Total Weight940 g940 g

Function

Pre-freeze functionAvailableAvailable
Electronic zoom functionAvailableAvailable
Electronic shutter functionAvailableAvailable
Records of endoscope's informationAvailableAvailable
NBI observationAvailableAvailable
RDI observationAvailableAvailable

Page 25

SUBJECT DEVICEPREDICATE DEVICE
BF-H1100 BronchoscopeBF-H1100 Bronchoscope (K222861)
High frequency treatmentAvailableAvailable
Laser treatmentAvailableAvailable
Scope IDAvailableAvailable

Table 5 – BF-1TH1100

Subject DevicePredicate Device
BF-1TH1100 BronchoscopeBF-1TH110 Bronchoscope (K222861)
ManufacturerOLYMPUS MEDICAL SYSTEMS CORPOLYMPUS MEDICAL SYSTEMS CORP
Classification & Regulation NumberClass II21 CFR §874.4680,21 CFR §874.4760,21 CFR §876.1500Class II21 CFR §874.4680,21 CFR §874.4760,21 CFR §876.1500
Product CodeEOQ: Bronchoscope (Flexible or rigid)EOB: Nasopharyngoscope (Flexible or rigid)NWB: endoscope, accessories, narrow band spectrumEOQ: Bronchoscope (Flexible or rigid)EOB: Nasopharyngoscope (Flexible or rigid)NWB: endoscope, accessories, narrow band spectrum
Indications for UseThe BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.The BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

Image Sensor

Type- CCD- CYM color filter- Sequential read image signal- CCD- CYM color filter- Sequential read image signal
Number of Pixels465,588465,588
Pixel Size1.5 μm (H) × 1.5 μm (V)1.5 μm (H) × 1.5 μm (V)

Page 26

Subject DevicePredicate Device
BF-1TH1100 BronchoscopeBF-1TH110 Bronchoscope (K222861)

Optical System

Direction of View
Field of View120°120°
Depth of Field3 - 100 mm3 - 100 mm

Distortion

Insertion Section

Distal end outer diameterø 5.8 mmø 5.8 mm
Distal end enlarged1 Objective lens2 Light guide lens3 Instrument channel outlet1 Objective lens2 Light guide lens3 Instrument channel outlet
Insertion tube outer diameterø 6.1 mmø 6.1mm
Insertion section working length600 mm600 mm

Instrument Channel

Channel inner diameterø 3.0 mmø 3.0 mm
Minimum visible distance3 mm3 mm

Page 27

Subject DevicePredicate Device
BF-1TH1100 BronchoscopeBF-1TH110 Bronchoscope (K222861)
Direction from which EndoTherapy accessories enter and exit the endoscopic image

Control Section

Configuration

The insertion tube rotation angle by simply turning a ring on the control section of the scope.left or right up to 120°left or right up to 120°

Connector Section

Configuration

Bending Section

Angulation rangeUP 180°DOWN 130°UP: 180°DOWN: 130°

Total Length

Total length880 mm880 mm

Total Weight

Total weight950 g950 g

Function

Pre-freeze functionAvailableAvailable
Electronic zoom functionAvailableAvailable
Electronic shutter functionAvailableAvailable

Page 28

Subject DevicePredicate Device
BF-1TH1100 BronchoscopeBF-1TH110 Bronchoscope (K222861)
Records of endoscope's informationAvailableAvailable
NBI observationAvailableAvailable
RDI observationAvailableAvailable
High frequency treatmentAvailableAvailable
Laser treatmentAvailableAvailable
Scope IDAvailableAvailable

8. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination for BF-XP190, BF-P190, and BF-XT-190:

1) Performance testing - Bench

Bench testing as listed below was conducted to ensure that the subject device performs as intended and meet design specifications.

  • Thermal Safety
  • Photobiological Safety
  • Color Performance
  • Resolution
  • Noise and Dynamic Range
  • Image Intensity Uniformity
  • Video Latency
  • RDI
  • TXI
  • BAI-MAC
  • Automatic Brightness Adjustment
  • Pre-Freeze
  • Ignition Factors Safety Study (Laser)*
  • Ignition Factors Safety Study (High-Frequency)*
  • Ignition Factors Safety Study (APC)*
  • Ignition Factors Safety Study (Scope Comparison)*

*Testing is also applicable to BF-H1100 and BF-1TH1100 to address the labeling change

Page 29

2) Performance testing - Animal

Animal testing was performed to assess the imaging modes of WLI, NBI, TXI, and BAI-MAC with video processor CV-1500.

3) Risk management

Risk management was performed in accordance with ISO 14971:2019. The design verification tests and their acceptance criteria were performed and identified as a result of this risk management.

4) Testing not completed

Since there were no design, material, sterilization, reprocessing, packaging, shelf life, or software changes, the following testing was not performed:

  • Biocompatibility Testing
  • Sterilization
  • Shelf Life
  • Stability Testing
  • Package Integrity Test
  • Software Testing and Cybersecurity
  • Electrical Safety and EMC
  • Residual Toxicity of Reprocessing Chemicals
  • Human Factors
  • Clinical

9. CONCLUSIONS

Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate device, EVIS EXERA III BRONCHOVIDEOSCOPES (OLYMPUS BF-XP190, OLYMPUS BF-P190, and BF-XT190) and BRONCHOVIDEOSCOPE BF-H1100 and BF-1TH1100 raise no new issue of safety and effectiveness and are substantially equivalent to the predicate device in terms of safety, effectiveness, and performance.

§ 874.4680 Bronchoscope (flexible or rigid) and accessories.

(a)
Identification. A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.