AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EVIS X1 VIDEO SYSTEM CENTER OLYMPUS CV-1500 is intended to be used with Olympus ancillary equipment for endoscopic diagnosis, treatment, and video observation. This product is designed to process electronic signals transmitted from Olympus video endoscopes, output images to monitors, provide illumination to the endoscope, supply air through the endoscope while inside the body and control/monitor ancillary equipment. NBI (Narrow Band Imaging), RDI (Red Dichromatic Imaging), TXI (TeXture and color enhancement Imaging), and BAI-MAC (Brightness Adjustment Imaging with Maintenance of Contrast) are adjunctive tools for endoscopic examination which can be used to supplement Olympus white light imaging. NBI, RDI, TXI and BAI-MAC are not intended to replace histopathological sampling as a means of diagnosis. The CV-1500 Video System Center is compatible with scopes within the EVIS 190 and 1100 families.

The BRONCHOVIDEOSCOPE OLYMPUS BF-H1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

The BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

Device Description

The CV-1500 video system center is indicated to process electronic signals transmitted from Olympus video endoscopes, output images to monitors, and be used with Olympus ancillary equipment for endoscopic diagnosis, treatment, and video observation. This product also functions as a pump to supply air through the endoscope, a light source to the endoscope, and a controller/monitor of ancillary equipment. NBI (Narrow Band Imaging), RDI (Red Dichromatic Imaging), TXI (TeXture and color enhancement Imaging), and BAI-MAC (Brightness Adjustment Imaging with Maintenance of Contrast) are adjunctive tools for endoscopic examination which can be used to supplement Olympus white light imaging. NBI, RDI, TXI and BAI-MAC are not intended to replace histopathological sampling as a means of diagnosis.

RDI (Red Dichromatic Imaging) observation: RDI is optical-digital observation using red dichromatic narrow band light and green illumination light to enhance visibility of bleeding points in the endoscopic image due to the difference in light absorption.

TXI (TeXture and color enhancement Imaging): TXI emphasizes tonal changes, patterns, and image outlines. It also corrects the brightness of dark areas.

BAI-MAC (Brightness Adjustment Imaging with Maintenance of Contrast): BAI-MAC maintains the brightness of the bright part of the endoscopic image and corrects the brightness of the dark part of the endoscopic image.

The BF-H1100 and BF-1TH1100 endoscopes consist of three parts: the control section, the insertion section, and the connector section. The endoscope receives the illumination light from light guide connector connected to the video system center (CV-1500). The illumination light is transferred to the distal end through the optical fiber bundle inside of the endoscope and illuminates the inside of the patient body through the illumination lens at the distal end.

The endoscope receives the reflected light from the inner lumen of a patient by objective lens at the distal end. The built-in charge-coupled device (CCD) at the distal end converts the light to the electrical signal, and the signal is sent to the video system center via the electrical cable and the video connector of the endoscope. The endoscope transfers the image signal and displays the observation image on the screen.

AI/ML Overview

This FDA 510(k) summary focuses on establishing substantial equivalence for the EVIS X1 Video System Center Olympus CV-1500, Bronchovideoscope Olympus BF-H1100, and Bronchovideoscope Olympus BF-1TH1100 to their predicate devices. The listed performance data describes various bench tests and an animal study conducted. However, this document does not detail specific acceptance criteria or report device performance against those criteria in a tabular format. It also explicitly states that "No clinical study was performed to demonstrate substantial equivalence." Therefore, a comprehensive answer for all your requested points cannot be extracted from this particular document.

Here's what can be inferred and what information is missing:

1. A table of acceptance criteria and the reported device performance

This information is not explicitly provided in the document in a tabular format. The document states that "The design verification tests and the acceptance criteria were identified and performed as a result of the risk management," and lists various performance tests (e.g., Thermal Safety, Durability, Resolution, Depth of Field, RDI, TXI and BAI-MAC performance). However, it does not provide the specific acceptance thresholds for these tests or the quantitative results achieved by the device.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Bench Testing: The document does not specify sample sizes for the bench tests.
  • Animal Study: An "animal study" was performed for CV-1500 to confirm WLI, NBI, TXI, and BAI-MAC performance. The document does not specify the sample size (e.g., number of animals) or the country of origin for this study. It is implicitly a prospective study in an animal model.
  • Clinical Study: "No clinical study was performed to demonstrate substantial equivalence." Therefore, there is no human test set or associated provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided as there was no clinical study involving human readers or expert-established ground truth mentioned in this 510(k) summary for device performance evaluation. For the animal study, the ground truth establishment method and expert involvement are not detailed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided as there was no clinical study involving human readers discussed in this 510(k) summary. For the animal study, the adjudication method is not detailed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done involving human readers or AI assistance, as explicitly stated: "No clinical study was performed to demonstrate substantial equivalence." The device described is an endoscopic imaging system, not an AI-powered diagnostic tool. The imaging modes (NBI, RDI, TXI, BAI-MAC) are described as "adjunctive tools for endoscopic examination which can be used to supplement Olympus white light imaging" and "are not intended to replace histopathological sampling as a means of diagnosis."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The document does not describe the device as an standalone algorithm. The device is a video system center and bronchoscopes. The performance evaluation focuses on the technical characteristics and imaging capabilities of the hardware and integrated imaging modes.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the "Performance testing - Bench", the ground truth would be established by controlled experimental conditions and measurements against known physical standards. For the "Performance testing - Animal," the ground truth likely involved direct visual assessment and possibly histopathological examination of tissues within the animal model, though this is not explicitly stated.

8. The sample size for the training set

This is not applicable as the device is a hardware endoscopic system, not a machine learning algorithm that requires a training set in the typical sense. The "training set" concept does not apply to this type of device submission.

9. How the ground truth for the training set was established

This is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 15, 2023

Olympus Medical Systems Corporation % Brenda Geary Manager, Regulatory Affairs Olympus Corporation of the Americas 800 West Park Drive Westborough, Massachusetts 01581

Re: K222861

Trade/Device Name: EVIS XI Video System Center Olympus CV-1500, Bronchovideoscope Olympus BF-H1100, Bronchovideoscope Olympus BF-1TH1100 Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ, FET, NWB, NTN Dated: May 16, 2023 Received: May 16, 2023

Dear Brenda Geary:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shuchen Peng -S

Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222861

Device Name EVIS X1 VIDEO SYSTEM CENTER OLYMPUS CV-1500

Indications for Use (Describe)

The EVIS X1 VIDEO SYSTEM CENTER OLYMPUS CV-1500 is intended to be used with Olympus ancillary equipment for endoscopic diagnosis, treatment, and video observation. This product is designed to process electronic signals transmitted from Olympus video endoscopes, output images to monitors, provide illumination to the endoscope, supply air through the endoscope while inside the body and control/monitor ancillary equipment. NBI (Narrow Band Imaging), RDI (Red Dichromatic Imaging), TXI (TeXture and color enhancement Imaging), and BAI-MAC (Brightness Adjustment Imaging with Maintenance of Contrast) are adjunctive tools for endoscopic examination which can be used to supplement Olympus white light imaging. NBI, RDI, TXI and BAI-MAC are not intended to replace histopathological sampling as a means of diagnosis. The CV-1500 Video System Center is compatible with scopes within the EVIS 190 and 1100 families.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K222861

Device Name BRONCHOVIDEOSCOPE OLYMPUS BF-H1100

Indications for Use (Describe)

The BRONCHOVIDEOSCOPE OLYMPUS BF-H1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K222861

Device Name BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100

Indications for Use (Describe)

The BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are all capitalized and evenly spaced. There is a yellow line underneath the word. The registered trademark symbol is located to the right of the letter "S".

Date Prepared: June 15, 2023

510(k) Summary

SPONSOR INFORMATION

510(k) Submitter:OLYMPUS MEDICAL SYSTEMS CORP.2951 Ishikawa-choHachioji-shi, Tokyo, Japan 192-8507
--------------------------------------------------------------------------------------------------------------

Official Correspondent: Brenda Geary Olympus Corporation of the Americas 800 West Park Drive Westborough, MA 01581 Cell: (508) 641-0568 Email: brenda.geary@olympus.com

  • Manufacturer for BF-H1100, BF-1TH1100 Manufacturing Site(s): Aizu Olympus Co., Ltd., 3-1-1 Niiderakita, Aizuwakamatsu-shi, Fukushima 965-8520, Japan
    Manufacturer for CV-1500 Shirakawa Olympus Co., Ltd. 3-1 Okamiyama, Odakura, Nishigo-mura, Nishishirakawa-gun, Fukushima 961-8061, Japan
■ Device Name(s):EVIS X1 VIDEO SYSTEM CENTER OLYMPUS CV-1500
■ Model Name(s):OLYMPUS CV-1500
■ Common Name:Endoscopic Video Imaging System
■ Regulation Number:874.4680
■ Regulation Name:Bronchoscope (flexible or rigid) and accessories
■ Regulatory Class:II
■ Product Code:EOQ: Bronchoscope (Flexible or Rigid)FET (Endoscopic Video Imaging System/Component,Gastroenterology-Urology)NWB (endoscope, accessories, narrow band spectrum)NTN (Led Light Source)
■ Classification Panel:Ear Nose & Throat (EOQ, EOB)Gastroenterology/Urology (FET, NWB, NTN)

DEVICE IDENTIFICATION

{6}------------------------------------------------

■ Device Name(s):BRONCHOVIDEOSCOPE OLYMPUS BF-H1100
■ Model Name(s):OLYMPUS BF-H1100
■ Common Name:Bronchoscope
■ Regulation Number:874.4680
■ Regulation Name:Bronchoscope (flexible or rigid) and accessories
■ Regulatory Class:II
■ Product Code:EOQ: Bronchoscope (Flexible or Rigid)NWB (endoscope, accessories, narrow band spectrum)
■ Classification Panel:Ear Nose & Throat (EOQ, EOB)Gastroenterology/Urology (NWB)
■ Device Name(s):BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100
■ Model Name(s):OLYMPUS BF-1TH1100
■ Common Name:Bronchoscope
■ Regulation Number:874.4680
■ Regulation Name:Bronchoscope (flexible or rigid) and accessories
■ Regulatory Class:II
■ Product Code:EOQ: Bronchoscope (Flexible or Rigid)NWB (Endoscope, accessories, narrow band spectrum)
■ Classification Panel:Ear Nose & Throat (EOQ, EOB)Gastroenterology/Urology (NWB)

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, dark blue letters. The letters are closely spaced and fill most of the frame. There is a registered trademark symbol to the right of the word. A thin gold line is visible below the word.

PREDICATE DEVICES

1 Predicate device for CV-1500

Device Name510(k) Submitter510(k) No.
PREDICATE
VIDEO SYSTEM CENTER OLYMPUS CV-190 &XENON LIGHT SOURCE OLYMPUS CLV-190Olympus MedicalSystems Corp.K121959,K131780

Predicate device for BF-H1100 .

Device Name510(k) Submitter510(k) No.
PREDICATE
EVIS EXERA III BRONCHOSCOPE OLYMPUS BF-H190Olympus MedicalSystems Corp..K121959

Predicate device for BF-1TH1100

Device Name510(k) Submitter510(k) No.
PREDICATE
EVIS EXERA III BRONCHOSCOPE OLYMPUS BF-1TH190Olympus MedicalSystems Corp..K121959

DEVICE DESCRIPTION

EVIS X1 VIDEO SYSTEM CENTER OLYMPUS CV-1500

The CV-1500 video system center is indicated to process electronic signals transmitted from Olympus video endoscopes, output images to monitors, and be used with Olympus ancillary equipment for endoscopic diagnosis, treatment, and video observation. This product also functions as a pump to supply air through the endoscope, a light source to the endoscope, and a controller/monitor of ancillary equipment. NBI (Narrow Band Imaging), RDI (Red Dichromatic Imaging), TXI (TeXture and color enhancement Imaging), and BAI-MAC (Brightness Adjustment Imaging with Maintenance of Contrast) are adjunctive tools for endoscopic examination which can be used to supplement Olympus white light imaging. NBI, RDI, TXI and BAI-MAC are not intended to replace histopathological sampling as a means of diagnosis.

RDI (Red Dichromatic Imaging) observation:

RDI is optical-digital observation using red dichromatic narrow band light and green illumination light to enhance visibility of bleeding points in the endoscopic image due to the difference in light absorption.

{8}------------------------------------------------

TXI (TeXture and color enhancement Imaging):

TXI emphasizes tonal changes, patterns, and image outlines. It also corrects the brightness of dark areas.

BAI-MAC (Brightness Adjustment Imaging with Maintenance of Contrast):

BAI-MAC maintains the brightness of the bright part of the endoscopic image and corrects the brightness of the dark part of the endoscopic image.

BRONCHOSCOPE OLYMPUS BF-H1100 & BF-1TH1100

The BF-H1100 and BF-1TH1100 endoscopes consist of three parts: the control section, the insertion section, and the connector section. The endoscope receives the illumination light from light guide connector connected to the video system center (CV-1500). The illumination light is transferred to the distal end through the optical fiber bundle inside of the endoscope and illuminates the inside of the patient body through the illumination lens at the distal end.

The endoscope receives the reflected light from the inner lumen of a patient by objective lens at the distal end. The built-in charge-coupled device (CCD) at the distal end converts the light to the electrical signal, and the signal is sent to the video system center via the electrical cable and the video connector of the endoscope. The endoscope transfers the image signal and displays the observation image on the screen.

INDICATIONS FOR USE

EVIS X1 VIDEO SYSTEM CENTER OLYMPUS CV-1500

The EVIS XI VIDEO SYSTEM CENTER OLYMPUS CV-1500 is intended to be used with Olympus ancillary equipment for endoscopic diagnosis, treatment, and video observation. This product is designed to process electronic signals transmitted from Olympus video endoscopes, output images to monitors, provide illumination to the endoscope, supply air through the endoscope while inside the body and control/monitor ancillary equipment. NBI (Narrow Band Imaging), RDI (Red Dichromatic Imaging), TXI (TeXture and color enhancement Imaging), and BAI-MAC (Brightness Adjustment Imaging with Maintenance of Contrast) are adjunctive tools for endoscopic examination which can be used to supplement Olympus white light imaging. NBI, RDI, TXI and BAI-MAC are not intended to replace histopathological sampling as a means of diagnosis. The CV-1500 Video System Center is compatible with scopes within the EVIS 190 and 1100 families.

BRONCHOSCOPES - OLYMPUS BF-H1100 & BF-1TH1100

The BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100 is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

The BRONCHOVIDEOSCOPE OLYMPUS BF-H1100 is intended to be used with an Olvmpus video system center, documentation equipment, monitor, EndoTherapy accessories

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are sans-serif and appear to be a custom font. A registered trademark symbol is located to the right of the letter "S".

(such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEVICE

EVIS XI VIDEO SYSTEM CENTER CV-1500

SUBJECT DEVICEPREDICATE
CV-1500 Video System CenterCV-190 Video System Center & CLV-190 LightSource (K121959, K131780)
ManufacturerOLYMPUS MEDICAL SYSTEMS CORPSAME
Classification &Regulation NumberClass II21 CFR §874.4680SAME
Product CodeEOQ - Bronchoscope (Flexible or Rigid)FET -endoscopic video imagingsystem/component, gastroenterology-urologyNWB - endoscope, accessories, narrow bandspectrumNTN - led light sourceEOQ - bronchoscope (flexible or rigid)EOB Nasopharyngoscope (Flexible or rigid)NWB - endoscope, accessories, narrow bandspectrum
Indications for UseThe EVIS X1 VIDEO SYSTEM CENTEROLYMPUS CV-1500 is intended to be usedwith Olympus ancillary equipment forendoscopic diagnosis, treatment, and videoobservation. This product is designed to processelectronic signals transmitted from Olympusvideo endoscopes, output images to monitors,provide illumination to the endoscope, supplyair through the endoscope while inside the bodyand control/monitor ancillary equipment. NBI(Narrow Band Imaging), RDI (Red DichromaticImaging), TXI (TeXture and color enhancementImaging), and BAI-MAC (BrightnessAdjustment Imaging with Maintenance ofContrast) are adjunctive tools for endoscopicexamination which can be used to supplementOlympus white light imaging. NBI, RDI, TXIand BAI-MAC are not intended to replacehistopathological sampling as a means ofdiagnosis.This video system center is intended to be usedwith OLYMPUS camera heads, endoscopes,light sources, monitors, EndoTherapyaccessories, and other ancillary equipment forendoscopic diagnosis, treatment, and videoobservation.This light source is intended to be used WithOlympus endoscopes, video system center, andother ancillary equipment for endoscopicdiagnosis, treatment and video observation.
Overall System Features
Rated voltage100 - 120V AC ±10 %Video Center: 100 - 120 V AC ± 10%Light Source: 100 - 120 V AC ± 10%
Rated frequency50/60 Hz ± 1 HzVideo Center: 50/60 Hz ± 1 HzLight Source: 50/60 Hz ± 1 Hz
Over-currentprotectionFuse type (Built-in type)Video Center: Fuse type (Built-in type)Light Source: Fuse type (Built-in type)
Rated input600 VAVideo Center: 150 VALight Source: 600 VA
Dimensions(W×H×D)370×198×488 mmVideo Center: 370×85×455 mmLight Source: 370×150×476 mm
Dimensions(maximum)(W×H×D)398×218×580 mmVideo Center: 382×91×489 mmLight Source: 382×162×551 mm
SUBJECT DEVICEPREDICATE
CV-1500 Video System CenterCV-190 Video System Center & CLV-190 LightSource (K121959, K131780)
Weight19.4 kgVideo Center: 10.7 kg
Light Source: 19 kg
Communicationterminals• Output socket• 1000BASE-T• Foot switch• Keyboard• Adaptor• Recorder• DF• Printer• CV-LINK• LINK OUT• UPD/PSCUVideo Center:• Video connector socket• 100BASE-TX• Foot switch• Keyboard• Option 1• Option 2• Adapter• Light source• Light source 2• Remote 1• Remote 2• Monitor remote 1• Monitor remote 2• EUS• CV-LINKLight Source:• Output socket• CV 1• CV 2• LINK - OUT
• UPD
Touch panelProvidedVideo Center: Not provided
(Brightness 10 steps)Light Source: Not provided
Observation modeWLI, NBI, RDIVideo Center: WLI, NBILight Source: WLI, NBI
Record to portablememoryProvidedVideo Center: ProvidedLight Source: Not provided
Cooling method ofinsideFan (Variable rotation)Video Center: Fan (Forced-air cooling)Light Source: Fan (Variable rotation)
Type of protectionagainst electricshockClass IVideo Center: Class ILight Source: Class I
Degree of protectionagainst electricshock of applied partType BF applied part(Depends on applied part)Video Center: Type BF or CF applied part(Depends on applied part)Light Source: Type BF or CF applied part(Depends on applied part)
Degree or protectionagainst explosionThe video system center should be kept awayfrom flammable gases.Video Center: The video system center shouldbe kept away from flammable gases.Light Source: The light source should be keptaway from flammable gases.
Light Source Features
Bulb typeLEDXenon short-arc lamp (ozone-free) 300W
Providing maximumlight intensityLess than 3.93 W3.21 W
Observation modeWLI, NBI, RDIWLI, NBI
Emergency LampNot providedHalogen Lamp 12V35W
Brightness adjustmentAutomatic (current control, 17 steps)• Automatic (current control, 17 steps)
SUBJECT DEVICEPREDICATE
CV-1500 Video System CenterCV-190 Video System Center & CLV-190 LightSource (K121959, K131780)
Video Center Features
Analog signal outputVBS compositeRGB component VBS composite and Y/C;simultaneous outputs possible.
Digital signal outputSDI:2SDI:2, DVI:1
User settingsThe function settings for up to 20 users can bestored.The function settings for up to 20 users can bestored.
White BalanceadjustmentBF-H190 and BF-1TH190Automaticallyadjusted using the white balance switch at thetime of connection with the scope with Scope IDBF-H1100 and BF-1TH1100Automaticallyadjusted without pressing white balance switchAutomatically adjusted using the white balanceswitch at the time of connection with the scopewith Scope ID
Standard color chartoutputColor bar imageColor bar image or the 50% white screen can bedisplayed
Color tone adjustment• Red adjustment ±8 steps• Blue adjustment ±8 steps• Chroma adjustment ±8 steps• Red adjustment ±8 steps• Blue adjustment ±8 steps• Chroma adjustment ±8 steps
ContrastNormal / High / LowNormal / High / Low
IrisAUTO/PEAK/AVEAUTO/PEAK/AVE
Image enhancementStructure enhancementType A: (8 steps)Type B: (8 steps)*User can preset three image enhancementsettingsStructure enhancementType A: (8 steps)Type B: (8 steps)Edge enhancement(8 steps)*User can preset three image enhancementsettings
TXI modesMode 1/Mode 2/Mode 3Not provided
Image size selectionThe size of the endoscopic image can beselected from 2 modes. (Except SDTV)The size of the endoscopic image can beselected from 2 modes. (Except SDTV)
Electric zoomSwitch between mode 1, mode 2, and mode 3.Switch between mode 1, mode 2, and mode 3.
PIP/POPProvidedProvided
Aspect ratioSwitch between 16:9 and 4:3. (Except SDTV)Switch between 16:9 and 4:3. (Except SDTV)
FreezeStill the endoscopic imageStill the endoscopic image
Pre-freezeAvailableAvailable
Custom switchAssign specific functions to the followingbuttons.• Remote switches (Up to 5)• Foot switches (Up to 2)• Keyboard custom key (Up to 4)• Touch panel custom button of basic functionsscreen (Up to 3)• Touch panel custom button of customAssign specific functions to the followingbuttons.• Remote switches (Up to 5)• Foot switches (Up to 2)• Keyboard custom key (Up to 4)

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are all capitalized and have a slightly rounded appearance. A registered trademark symbol is located to the right of the letter "S".

Bronchoscope BF-H1100

SUBJECT DEVICEBF-H1100 BronchoscopePREDICATEBF-H190 Bronchoscope(K121959)
ManufacturerOLYMPUS MEDICAL SYSTEMS CORPSAME
Classification &Regulation NumberClass II21 CFR §874.4680,SAME
Product CodeEOQ: Bronchoscope (Flexible or rigid)NWB: endoscope, accessories, narrow band spectrumSAME
Indications for UseThe BRONCHOVIDEOSCOPE OLYMPUSBF-H1100 is intended to be used with anOlympus video system center, documentationequipment, monitor, EndoTherapy accessories(such as a biopsy forceps), and other ancillaryequipment for endoscopy and endoscopicsurgery. This instrument is indicated for usewithin the airways and tracheobronchial tree.This instrument is intended to be used with anOlympus video system center, light source,documentation equipment, monitor, EndoTherapyaccessories (such as a biopsy forceps), and otherancillary equipment for endoscopy and endoscopicsurgery. This instrument is indicated for use withinthe airways and tracheobronchial tree.
TypeCCDCYM color filterSequential read image signalCCDCYM color filterSequential read image signal
Direction of View
Field of View120°120°
Depth of Field3 - 100 mm3 - 100 mm
Distal end outerdiameterø 4.9 mmø 5.5mm
Distal end enlargedImage: Distal end enlarged diagram with labels 1, 2, 3, UP, DOWN, Objective lens, Light guide Lens, Instrument channel outletImage: Distal end enlarged diagram with labels 1, 2, 3, UP, DOWN, Objective lens, Light guide Lens, Instrument channel outlet
Insertion tube outerdiameterø 4.9 mmø 5.1 mm
Insertion sectionworking length600 mm600 mm
Channel innerdiameterø 2.2 mmø 2.0 mm
Channel length850 mm850 mm
Total length880 mm880 mm
Minimum visibledistance3 mm3 mm
Direction from whichEndoTherapyaccessories enter andexit the endoscopicimage
EVIS XI VIDEO SYSTEM CENTER WITH BRONCHOSCOPES
SUBJECT DEVICEPREDICATE
BF-H1100 BronchoscopeBF-H190 Bronchoscope(K121959)
Image: BF-H1100 BronchoscopeImage: BF-H190 Bronchoscope
ConfigurationImage: BF-H1100 BronchoscopeImage: BF-H190 Bronchoscope
ConfigurationImage: BF-H1100 BronchoscopeImage: BF-H190 Bronchoscope
Angulation rangeUP 210°DOWN 130°UP 210°DOWN 130°
Pre-freeze functionAvailableAvailable
Electronic zoomfunctionAvailableAvailable
Electronic shutterfunctionAvailableAvailable
Records of endoscope'sinformationAvailableAvailable
NBI observationAvailableAvailable
RDI observationAvailable (when using CV-1500)Available (when using CV-1500)
High frequencytreatmentAvailableAvailable
PREDICATE:
SUBJECT DEVICE:BF-1TH190 Bronchoscope
BF-1TH1100 Bronchoscope(K121959)
ManufacturerOLYMPUS MEDICAL SYSTEMS CORPSAME
Classification &Regulation NumberClass II21 CFR §874.4680,SAME
Product CodeEOQ: Bronchoscope (Flexible or rigid)NWB: endoscope, accessories, narrow bandspectrumSAME
Indications for UseThe BRONCHOVIDEOSCOPE OLYMPUSBF-1TH1100 is intended to be used with anOlympus video system center, documentationequipment, monitor, EndoTherapy accessories(such as a biopsy forceps), and other ancillaryequipment for endoscopy and endoscopicsurgery. This instrument is indicated for usewithin the airways and tracheobronchial tree.This instrument is intended to be used with anOlympus video system center, light source,documentation equipment, monitor, EndoTherapyaccessories (such as a biopsy forceps), and otherancillary equipment for endoscopy and endoscopicsurgery. This instrument is indicated for use withinthe airways and tracheobronchial tree.
Direction of View
Field of View120°120°
Depth of Field3 - 100 mm3 - 100 mm
Distal end outerdiameterø 5.8 mmø 6.2 mm
Distal end enlargedImage: Distal end enlarged BF-1TH1100Image: Distal end enlarged BF-1TH190
1 Objective lens2 Light guide lens3 Instrument channel outlet1 Objective lens2 Light guide lens3 Instrument channel outlet
Insertion tube outerdiameterø 6.1 mmø 6.0mm
Insertion sectionworking length600 mm600 mm
Channel innerdiameterø 3.0 mmø 2.8 mm
Channel length850 mm850 mm
Total length880 mm880 mm
Minimum visibledistance3 mm3 mm
Direction from whichEndoTherapyaccessories enter andexit the endoscopicimageImage: Direction from which EndoTherapy accessories enter and exit the endoscopic image BF-1TH1100Image: Direction from which EndoTherapy accessories enter and exit the endoscopic image BF-1TH190
Configuration
SUBJECT DEVICE:BF-1TH1100 BronchoscopePREDICATE:BF-1TH190 Bronchoscope(K121959)
Image: BF-1TH1100 BronchoscopeImage: BF-1TH190 Bronchoscope
ConfigurationImage: Configuration BF-1TH1100 BronchoscopeImage: Configuration BF-1TH190 Bronchoscope
Angulation rangeUP 180°DOWN 130°UP: 180°DOWN: 130°
Pre-freeze functionAvailableAvailable
Electronic zoom functionAvailableAvailable
Electronic shutter functionAvailableAvailable
Records of endoscope's informationAvailableAvailable
NBI observationAvailableAvailable
RDI observationAvailable (when using CV-1500)Available (when using CV-1500)
High frequency treatmentAvailableAvailable
Laser treatmentAvailableAvailable

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. A registered trademark symbol is located to the right of the word. A thin yellow line is located underneath the word.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are closely spaced together, creating a solid block of text. A registered trademark symbol is located to the right of the letter "S". A thin, horizontal, yellow line is located underneath the word.

Bronchoscope BF-1TH1100

{15}------------------------------------------------

{16}------------------------------------------------

PERFORMANCE DATA

> Software verification and validation

Software verification and validation testing were conducted for the CV-1500, BF-H1100 and BF-1TH1100 devices and documentations were provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".

> Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted for the CV-1500, BF-H1100 and BF-1TH1100 devices in accordance with the ANSI/AAMI ES 60601-1:2005/(R)2012 and A1:2012 and IEC 60601-2-18:2009 standards for safety and the IEC 60601-1-2:2014 standards for EMC.

> Performance testing - Bench

Bench testing as listed below were conducted for CV-1500, BF-H1100, and BF-1TH1100 to ensure that the subject device performs as intended and meet design specifications.

  • Thermal Safety -
  • Durability -
  • -Photobiological Safety
  • Color Performance -
  • Direction of View -
  • Field of View -
  • Distortion -
  • Resolution -
  • Depth of Field -
  • Noise and Dynamic Range -
  • Image Intensity Uniformity -
  • -Video Latency
  • RDI -
  • TXI and BAI-MAC -
  • Automatic Brightness Adjustment -
  • Pre-Freeze -

> Performance testing - Animal

An animal study was performed for CV-1500 to confirm the White Light Imaging (WLI) and Narrow Band Imaging (NBI) performance, TeXture and color enhancement Imaging (TXI) and Brightness Adjustment Imaging with MAintenance of Contrast (BAI-MAC).

> Performance testing - Clinical

No clinical study was performed to demonstrate substantial equivalence.

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image contains the word OLYMPUS in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. There is a registered trademark symbol to the right of the letter S. The background is white.

> Risk management

Risk management was performed for the CV-1500, BF-H1100 and BF-1TH1100 devices in accordance with ISO 14971:2007. The design verification tests and the acceptance criteria were identified and performed as a result of the risk management.

CONCLUSION

Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate device. the CV-1500, BF-H1100 and BF-1TH1100 devices raise no new issues of safety and effectiveness and are substantially equivalent to the predicate devices in terms of safety, effectiveness and performance.

§ 874.4680 Bronchoscope (flexible or rigid) and accessories.

(a)
Identification. A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.