K Number
K250342
Date Cleared
2025-07-08

(152 days)

Product Code
Regulation Number
880.6250
Panel
HO
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on examiner's hand to prevent contamination between patient and examiner. In addition, the glove was evaluated using ASTM D6978 for permeation of chemotherapy drugs and fentanyl citrate. The glove is single use and non-sterile.

The following chemotherapy drugs have been tested on the glove:

Chemotherapy DrugsConcentrationBreakthrough Detection Time
Carmustine*3.3 mg/mL23.2 min
Cisplatin1.0 mg/mL>240 min
Cyclophosphamide20.0 mg/mL>240 min
Dacarbazine10.0 mg/mL>240 min
Doxorubicin HCL2.0 mg/mL>240 min
Etoposide20.0 mg/mL>240 min
Fluorouracil50.0 mg/mL>240 min
Methotrexate25.0 mg/mL>240 min
Paclitaxel6.0 mg/mL>240 min
ThioTepa*10.0 mg/mL13.5 min
Mitomycin C0.5 mg/mL>240 min
Vincristine Sulfate1.0 mg/mL>240 min
FentanylConcentrationBreakthrough Detection Time
Fentanyl Citrate100.0 mcg/2mL>240 min

*Please note that the following drugs have low permeation time.

  1. Carmustine
  2. ThioTepa

Warning: Do not use with Carmustine & ThioTepa

Device Description

The proposed Powder-Free Polychloroprene Examination Glove (Green Color) Tested for Use with Chemotherapy Drugs and Fentanyl is a disposable device intended for over-the-counter use and is provided powder-free and non-sterile. These patient examination gloves are formulated using Polychloroprene. The gloves are not made with natural rubber latex. General specifications of the glove are as below:

  1. Overall Length: 240 mm minimum
  2. Width: 95 ± 5 mm minimum (for medium glove)
  3. Palm Thickness: 0.05 mm minimum
  4. Finger Thickness: 0.05 mm minimum
  5. Cuff Thickness: 0.05 mm minimum
  6. Tensile Strength
    • a. Before Aging: 14 MPa minimum
    • b. After Aging: 14 MPa minimum
  7. Ultimate Elongation
    • a. Before Aging: 500 % minimum
    • b. After Aging: 400 % minimum
  8. Pinhole AQL: 2.5
AI/ML Overview

The provided FDA 510(k) clearance letter and associated documentation pertain to a Powder-Free Polychloroprene Examination Glove. This is a physical medical device, not an AI/software device. Therefore, the questions related to AI/software performance criteria (such as multi-reader multi-case studies, human-in-the-loop performance, training/test set ground truth, and expert adjudication) are not applicable to this documentation.

The acceptance criteria and performance data provided in the document relate to the physical and chemical properties of the gloves, and their resistance to permeation by chemotherapy drugs and fentanyl.

Here's an analysis of the acceptance criteria and study that proves the device meets them, based on the provided text, focusing on the relevant physical device aspects:

Acceptance Criteria and Reported Device Performance

1. Table of Acceptance Criteria and the Reported Device Performance

CharacteristicStandard / RequirementDevice Performance (Proposed Device K250342)
DimensionsASTM D6977-19
Length (XS)≥ 220 mm≥ 251 mm
Length (S)≥ 220 mm≥ 250 mm
Length (M)≥ 230 mm≥ 251 mm
Length (L)≥ 230 mm≥ 248 mm
Length (XL)≥ 230 mm≥ 250 mm
Length (XXL)≥ 230 mm≥ 250 mm
Width (XS)70 ± 10 mm79 - 80 mm
Width (S)80 ± 10 mm86 - 88 mm
Width (M)95 ± 10 mm98 - 100 mm
Width (L)110 ± 10 mm111 - 113 mm
Width (XL)120 ± 10 mm115 - 118 mm
Width (XXL)130 ± 10 mm127 - 129 mm
Thickness (Finger)≥ 0.05 mm≥ 0.11 mm
Thickness (Palm)≥ 0.05 mm≥ 0.08 mm
Thickness (Cuff)≥ 0.05 mm≥ 0.06 mm
Freedom from HolesASTM D6977-19, 21 CFR 800.20 (AQL 2.5)Passed (for all sizes XS-XXL)
Physical PropertiesASTM D6977-19 (ASTM D412-16, D573)
Before Aging (Tensile)≥ 14 MPa16.83 - 33.4 MPa (depending on size)
Before Aging (Elongation)≥ 500 %521 - 596 % (depending on size)
After Aging (Tensile)≥ 14 MPa23.8 - 40.2 MPa (depending on size)
After Aging (Elongation)≥ 400 %428 - 462 % (depending on size)
Residual PowderASTM D6977-19, ASTM D6124-06 (≤ 2.0 mg/glove)0.22 - 0.32 mg/glove (depending on size)
Biocompatibility
Skin IrritationISO 10993-10 (Score ≤ 0.4)Not considered an irritant
Skin SensitizationISO 10993-10 (Score 240 min
Cyclophosphamide->240 min
Dacarbazine->240 min
Doxorubicin HCL->240 min
Etoposide->240 min
Fluorouracil->240 min
Methotrexate->240 min
Paclitaxel->240 min
ThioTepa-13.5 min*
Mitomycin C->240 min
Vincristine Sulfate->240 min
Fentanyl PermeationASTM D6978-05 (Minimum Breakthrough Detection Time)
Fentanyl Citrate->240 min

*Note: The document explicitly warns that Carmustine and ThioTepa have low permeation times and should not be used with the glove. This is a crucial labeling instruction based on the test results.

2. Sample Size Used for the Test Set and Data Provenance

  • The document does not explicitly state the specific sample sizes used for each physical, chemical, and permeation test. However, tests are typically performed on a statistically relevant number of samples to ensure representativeness and robust results according to the specified ASTM and ISO standards.
  • Data Provenance: The tests appear to have been conducted by Encompass Industries Sdn. Bhd. or a contracted lab on their behalf. The company is located in Kertih Bio-Polymer Park, Kemaman, Terengganu, Malaysia. The data is prospective in the sense that the tests were specifically performed on the proposed device for the purpose of this 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • This question is not applicable as the document describes a physical medical device (examination glove) and its material/permeation performance, not a software/AI device that requires expert human interpretation of data for ground truth establishment. The "ground truth" for these tests is based on objective, standardized laboratory measurements (e.g., precise measurements of length, thickness, tensile strength; chemical analysis for powder residue; standardized biological assays for biocompatibility; and analytical chemistry for drug permeation breakthrough times).

4. Adjudication Method for the Test Set

  • This question is not applicable for the same reason as point 3. Testing of physical device characteristics and chemical permeation is typically performed according to established laboratory protocols with quantitative, measurable endpoints, not through a process of expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This question is not applicable as the document describes a physical medical device (examination glove), not an AI-assisted diagnostic or therapeutic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • This question is not applicable as the document describes a physical medical device (examination glove), not an AI algorithm.

7. The Type of Ground Truth Used

  • The "ground truth" for the performance of the examination gloves is based on objective, standardized laboratory measurements and adherence to established international and national standards (e.g., ASTM D6977-19, ASTM D6319-19, ASTM D5151-19, ASTM D412-16, ASTM D573, ASTM D6124-06, ASTM D6978-05, ISO 10993-10, ISO 10993-11, 21 CFR 800.20).
    • Examples include:
      • Physical measurements: Direct measurement of glove dimensions (length, width, thickness).
      • Mechanical properties: Tensile strength and elongation determined by standard material testing methods.
      • Integrity: Freedom from holes determined by water leak testing (ASTM D5151-19).
      • Chemical analysis: Measurement of residual powder.
      • Biocompatibility: In-vivo (skin irritation, sensitization, acute systemic toxicity) tests using animal models, following ISO standards.
      • Permeation: Breakthrough detection time of specific chemotherapy drugs and fentanyl using analytical methods (ASTM D6978-05).

8. The Sample Size for the Training Set

  • This question is not applicable as the document describes a physical medical device, which does not involve a "training set" in the context of machine learning or AI.

9. How the Ground Truth for the Training Set Was Established

  • This question is not applicable for the same reason as point 8.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.