K Number
K243938

Validate with FDA (Live)

Date Cleared
2025-04-01

(102 days)

Product Code
Regulation Number
870.1330
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

The Aristotle 18 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

The Aristotle 24 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

The Zoom Wire 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

The Aristotle Colossus Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

Device Description

The Scientia Vascular's Aristotle 14, Aristotle 18, Aristotle 24, Zoom Wire 14, and Aristotle Colossus Guidewires ("the guidewires") are steerable guidewires with a shapeable tip to aid in accessing the neuro and peripheral vasculatures. The guidewires are supplied sterile, for single use only.

The distal portion of each guidewire's tip includes a radiopaque platinum wire marker coil to facilitate fluoroscopic visualization. All guidewires have a hydrophilic polymer coating on the distal portion and a polytetrafluoroethylene (PTFE) coating on the proximal portion to reduce friction during manipulation in vessels.

The guidewires are provided with an accessory kit consisting of an introducer (to aid with the insertion of the guidewire into a catheter hub and/or a hemostasis valve), a torque device (to attach to the proximal portion of the guidewire to facilitate gripping and manipulation of the guidewire during use), and a shaping mandrel (to aid in shaping the flexible tip of the guidewire). These accessory devices are included to facilitate use of the guidewires and are not intended to contact the patient's body.

AI/ML Overview

This document describes a 510(k) clearance for several guidewire devices, not an AI/ML medical device. Therefore, the information requested about acceptance criteria and studies related to AI/ML device performance (such as sample sizes for test/training sets, expert ground truth establishment, MRMC studies, etc.) is not applicable and cannot be extracted from this document.

The document primarily focuses on establishing substantial equivalence for physical medical devices—guidewires—by comparing their technological characteristics and performing non-clinical performance tests (biocompatibility, sterilization, bench performance).

Here's a breakdown of why this document does not contain the requested information about an AI/ML device:

  • Device Type: The clearance is for "Aristotle 14 Guidewire; Aristotle 18 Guidewire; Aristotle 24 Guidewire; Zoom Wire 14 Guidewire; Aristotle Colossus Guidewire" which are physical catheter guide wires. These are traditional medical devices, not software or AI/ML devices.
  • Study Types: The studies described are "Biocompatibility," "Sterilization," and "Bench Performance Testing" (Visual Inspection, Coating Integrity, Corrosion Resistance, Agent Compatibility, Simulated Use, Particulate). These are mechanical and biological tests relevant to physical devices, not algorithmic performance.
  • Missing Information: All points you requested regarding AI/ML performance are absent because they are not relevant to this type of device clearance:
    • No acceptance criteria for AI performance metrics (e.g., sensitivity, specificity, AUC).
    • No test or training set sample sizes for AI data.
    • No information about experts establishing ground truth for images or data used by an AI.
    • No adjudication methods.
    • No MRMC comparative effectiveness studies.
    • No standalone algorithm performance.
    • No ground truth type related to AI (e.g., imaging labels, pathology).

The document explicitly states:

  • "No animal testing was deemed necessary to support the substantial equivalence of the subject devices."
  • "No clinical testing was deemed necessary to support the substantial equivalence of the subject devices."

This indicates that the clearance relies on non-clinical data and comparison to predicate devices, which is common for guidewires, especially when the changes are minor (like a different PTFE coating).

FDA 510(k) Clearance Letter - K243938

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov

April 1, 2025

Scientia Vascular, Inc.
Bret Draper
Regulatory Affairs Specialist
2460 South 3270 West
West Valley City, Utah 84119

Re: K243938
Trade/Device Name: Aristotle 14 Guidewire; Aristotle 18 Guidewire; Aristotle 24 Guidewire; Zoom Wire 14 Guidewire; Aristotle Colossus Guidewire
Regulation Number: 21 CFR 870.1330
Regulation Name: Catheter Guide Wire
Regulatory Class: Class II
Product Code: MOF, DQX
Dated: March 7, 2025
Received: March 7, 2025

Dear Bret Draper:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243938 - Bret Draper Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243938 - Bret Draper Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan -S

Naira Muradyan, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K243938

Device Name: Aristotle 14 Guidewire; Aristotle 18 Guidewire; Aristotle 24 Guidewire; Zoom Wire 14 Guidewire; Aristotle Colossus Guidewire

Indications for Use (Describe)

The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

The Aristotle 18 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

The Aristotle 24 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

The Zoom Wire 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

The Aristotle Colossus Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Scientia Vascular, Inc.
Special 510(k)
K243938
Pg. 1

510(k) SUMMARY

510(k) Sponsor: Scientia Vascular, Inc.
2460 South 3270 West
West Valley City, UT 84119
Tel (888) 385-9016

Contact Person(s): Bret Draper
Regulatory Affairs Specialist
Tel: (888) 385-9016
E-mail: regulatory@scientiavascular.com

Date Prepared: March 31, 2025

Page 6

Scientia Vascular, Inc.
Special 510(k)
K243938
Pg. 2

Table 1. Subject Device Information

DeviceInformation
Aristotle 14 Guidewire
Trade Name:Aristotle 14 Guidewire
Common Name:Guidewire
Classification NameCatheter Guide Wire per 21 CFR 870.1330
Primary Product Code:MOF
Subsequent Product CodeDQX
Predicate Device:Aristotle 14 Guidewire (K173235)
Aristotle 18 Guidewire
Trade Name:Aristotle 18 Guidewire
Common Name:Guidewire
Classification NameCatheter Guide Wire per 21 CFR 870.1330
Primary Product Code:MOF
Subsequent Product CodeDQX
Predicate Device:Aristotle 18 Guidewire (K231954)
Aristotle 24 Guidewire
Trade Name:Aristotle 24 Guidewire
Common Name:Guidewire
Classification NameCatheter Guide Wire per 21 CFR 870.1330
Primary Product Code:MOF
Subsequent Product CodeDQX
Predicate Device:Aristotle 24 Guidewire (K231954)
Zoom Wire 14 Guidewire
Trade Name:Zoom Wire 14 Guidewire
Common Name:Guidewire
Classification NameCatheter Guide Wire per 21 CFR 870.1330
Primary Product Code:MOF
Subsequent Product CodeDQX
Predicate Device:Zoom 14 Guidewire (K201760)
Aristotle Colossus Guidewire
Trade Name:Aristotle Colossus Guidewire
Common Name:Guidewire
Classification NameCatheter Guide Wire per 21 CFR 870.1330
Primary Product Code:MOF
Subsequent Product CodeDQX
Predicate Device:Aristotle Colossus Guidewire (K222437)

Page 7

Scientia Vascular, Inc.
Special 510(k)
K243938
Pg. 3

DEVICE DESCRIPTION

The Scientia Vascular's Aristotle 14, Aristotle 18, Aristotle 24, Zoom Wire 14, and Aristotle Colossus Guidewires ("the guidewires") are steerable guidewires with a shapeable tip to aid in accessing the neuro and peripheral vasculatures. The guidewires are supplied sterile, for single use only, in the following diameters, stiffness profiles, and lengths:

GuidewireSpecifications
Aristotle 14 Guidewire
Diameter0.014"
Stiffness ProfilesSoft, Standard
Lengths200 cm, 300 cm
Aristotle 18 Guidewire
Diameter0.018"
Stiffness ProfilesSoft, Standard, Support
Length200 cm, 300cm
Aristotle 24 Guidewire
Diameter0.024"
Stiffness ProfilesSoft, Standard, Support
Length200 cm, 300cm
Aristotle Colossus Guidewire
Diameter0.035"
Stiffness ProfilesStandard
Lengths150 cm to 300 cm
Zoom Wire 14 Guidewire
Diameter0.014"
Stiffness ProfilesSupport, Extra Support
Lengths200 cm, 300 cm

The distal portion of each guidewire's tip includes a radiopaque platinum wire marker coil to facilitate fluoroscopic visualization. All guidewires have a hydrophilic polymer coating on the distal portion and a polytetrafluoroethylene (PTFE) coating on the proximal portion to reduce friction during manipulation in vessels.

The guidewires are provided with an accessory kit consisting of an introducer (to aid with the insertion of the guidewire into a catheter hub and/or a hemostasis valve), a torque device (to attach to the proximal portion of the guidewire to facilitate gripping and manipulation of the guidewire during use), and a shaping mandrel (to aid in shaping the flexible tip of the guidewire). These accessory devices are included to facilitate use of the guidewires and are not intended to contact the patient's body.

Page 8

Scientia Vascular, Inc.
Special 510(k)
K243938
Pg. 4

INDICATIONS FOR USE

Aristotle 14 Guidewire

The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

Aristotle 18 Guidewire

The Aristotle 18 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

Aristotle 24 Guidewire

The Aristotle 24 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

Aristotle Colossus Guidewire

The Aristotle Colossus Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

Zoom Wire 14 Guidewire

The Zoom Wire 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

Page 9

Scientia Vascular, Inc.
Special 510(k)
K243938
Pg. 5

INTENDED USE

The subject guidewires are intended for use by a physician to help introduce and position catheters or other interventional devices within the neuro and peripheral vasculature.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The subject devices have the following similarities to the previously cleared predicate devices:

  • The same indications for use,
  • The same intended use,
  • The same operating principle,
  • The same basic guidewire design,
  • Similar materials, and
  • The same packaging materials and sterilization processes.

The tables (Table 2 - Table 6) below provide additional details on the technological characteristics of the subject devices compared to their respective predicate devices.

Table 2. Aristotle 14 Guidewire Technological Characteristics Comparison

CharacteristicSubject Device Aristotle 14 Guidewire (K243938)Predicate Aristotle 14 Guidewire (K173235)Comparison
Anatomical LocationNeuro and peripheral vasculatureNeuro and peripheral vasculatureSame
DimensionsO.D.: 0.014" (0.36 mm) Length: 200 cm, 300 cmO.D.: 0.014" (0.36 mm) Length: 200 cm, 300 cmSame
Core WireStainless SteelStainless SteelSame
Distal TipShapeable Length: 35 cm Material: NitinolShapeable Length: 35 cm Material: NitinolSame
Stiffness ProfilesSoft, standardSoft, standardSame
CoatingsDistal End: Hydrophilic Proximal End:Proposed PTFEDistal End: Hydrophilic Proximal End:PTFEDifferent

Page 10

Scientia Vascular, Inc.
Special 510(k)
K243938
Pg. 6

CharacteristicSubject Device Aristotle 14 Guidewire (K243938)Predicate Aristotle 14 Guidewire (K173235)Comparison
Radiopaque Marker1 radiopaque marker at distal tip1 radiopaque marker at distal tipSame
Centering Coil1 centering coil1 centering coilSame
Shaping Mandrel (Accessory)Provided with each guidewireNACleared in K220398
Guidewire Introducer (Accessory)Provided with each guidewireProvided with each guidewireSame
Torque Device (Accessory)Provided with each guidewireProvided with each guidewireSame
Sterilization Method100% Ethylene Oxide (EO)100% EOSame
Shelf Life1 year3 yearDifferent

Table 3. Aristotle 18 Guidewire Technological Characteristics Comparison

CharacteristicSubject Device Aristotle 18 Guidewire (K243938)Predicate Aristotle 18 Guidewire (K231954)Comparison
Anatomical LocationNeuro and peripheral vasculatureNeuro and peripheral vasculatureSame
DimensionsO.D.: 0.018" (0.46 mm) Length: 200 cm, 300 cmO.D.: 0.018" (0.46 mm) Length: 200 cm, 300 cmSame
Core WireStainless SteelStainless SteelSame
Distal TipShapeable Length: 35 cm Material: NitinolShapeable Length: 35 cm Material: NitinolSame
Stiffness ProfilesSupport, standard, softSupport, standard, softSame
CoatingsDistal End: Hydrophilic Proximal End:Proposed PTFEDistal End: Hydrophilic Proximal End: PTFEDifferent

Page 11

Scientia Vascular, Inc.
Special 510(k)
K243938
Pg. 7

CharacteristicSubject Device Aristotle 18 Guidewire (K243938)Predicate Aristotle 18 Guidewire (K231954)Comparison
Radiopaque Marker1 radiopaque marker at distal tip1 radiopaque marker at distal tipSame
Centering Coil1 centering coil1 centering coilSame
Shaping Mandrel (Accessory)Provided with each guidewireProvided with each guidewireSame
Guidewire Introducer (Accessory)Provided with each guidewireProvided with each guidewireSame
Torque Device (Accessory)Provided with each guidewireProvided with each guidewireSame
Sterilization Method100% EO100% EOSame
Shelf Life1 year3 yearDifferent

Table 4. Aristotle 24 Guidewire Technological Characteristics Comparison

CharacteristicSubject Device Aristotle 24 Guidewire (K243938)Predicate Aristotle 24 Guidewire (K231954)Comparison
Anatomical LocationNeuro and peripheral vasculatureNeuro and peripheral vasculatureSame
DimensionsO.D.: 0.024" (0.61 mm) Length: 200 cm, 300 cmO.D.: 0.024" (0.61 mm) Length: 200 cm, 300 cmSame
Core WireStainless SteelStainless SteelSame
Distal TipShapeable Length: 35 cm Material: NitinolShapeable Length: 35 cm Material: NitinolSame
Stiffness ProfilesSupport, standard, softSupport, standard, softSame
CoatingsDistal End: Hydrophilic Proximal End:Proposed PTFEDistal End: Hydrophilic Proximal End: PTFEDifferent

Page 12

Scientia Vascular, Inc.
Special 510(k)
K243938
Pg. 8

CharacteristicSubject Device Aristotle 24 Guidewire (K243938)Predicate Aristotle 24 Guidewire (K231954)Comparison
Radiopaque Marker1 radiopaque marker at distal tip1 radiopaque marker at distal tipSame
Centering Coil2 centering coils2 centering coilsSame
Bushing1 bushing1 bushingSame
Shaping Mandrel (Accessory)Provided with each guidewireProvided with each guidewireSame
Guidewire Introducer (Accessory)Provided with each guidewireProvided with each guidewireSame
Torque Device (Accessory)Provided with each guidewireProvided with each guidewireSame
Sterilization Method100% EO100% EOSame
Shelf Life1 year3 yearDifferent

Table 5. Zoom Wire 14 Guidewire Technological Characteristics Comparison

CharacteristicSubject Device Zoom Wire 14 Guidewire (K243938)Predicate Zoom 14 Guidewire (K201760)Comparison
Anatomical LocationNeuro and peripheral vasculatureNeuro and peripheral vasculatureSame
DimensionsMax O.D.: 0.014" (0.36 mm) Length: 200 cm, 300 cmMax O.D.: 0.014" (0.36 mm) Length: 200 cm, 300 cmSame
Core WireStainless SteelStainless SteelSame
Distal TipShapeable Length: 35 cm Material: NitinolShapeable Length: 35 cm Material: NitinolSame
Stiffness ProfilesSupport, extra supportSupport, extra supportSame
CoatingsDistal End: HydrophilicDistal End: HydrophilicSame

Page 13

Scientia Vascular, Inc.
Special 510(k)
K243938
Pg. 9

CharacteristicSubject Device Zoom Wire 14 Guidewire (K243938)Predicate Zoom 14 Guidewire (K201760)Comparison
Coatings (cont.)Proximal End:Proposed PTFEProximal End: PTFEDifferent
Radiopaque MarkerRadiopaque marker at distal tipRadiopaque marker at distal tipSame
Centering Coil1 centering coil1 centering coilSame
BushingNoneNoneSame
Shaping Mandrel (Accessory)Provided with each guidewireProvided with each guidewireSame
Guidewire Introducer (Accessory)Provided with each guidewireProvided with each guidewireSame
Torque Device (Accessory)Provided with each guidewireProvided with each guidewireSame
Sterilization Method100% EO100% EOSame
Shelf Life1 year3 yearDifferent

Table 6. Aristotle Colossus Guidewire Technological Characteristics Comparison

CharacteristicSubject Device Aristotle Colossus Guidewire (K243938)Predicate Aristotle Colossus Guidewire (K222437)Comparison
Anatomical LocationNeuro and peripheral vasculatureNeuro and peripheral vasculatureSame
DimensionsO.D.: 0.035" (0.88 mm) Length: 150 cm to 300 cmO.D.: 0.035" (0.88 mm) Length: 150 cm to 300 cmSame
Core WireStainlesssteelStainlesssteelSame
Distal TipShapeable Length: 35 cm Material: NitinolShapeable Length: 35 cm Material: NitinolSame
Stiffness ProfilesStandardStandardSame
CoatingsDistal End: HydrophilicDistal End: HydrophilicSame

Page 14

Scientia Vascular, Inc.
Special 510(k)
K243938
Pg. 10

CharacteristicSubject Device Aristotle Colossus Guidewire (K243938)Predicate Aristotle Colossus Guidewire (K222437)Comparison
Coatings (cont.)Proximal End:Proposed PTFEProximal End: PTFEDifferent
Radiopaque Marker1 radiopaque marker at distal tip1 radiopaque marker at distal tipSame
Centering Coil2 centering coils2 centering coilsSame
Bushing1 bushing1 bushingSame
Centering Washer1 washer1 washerSame
Shaping Mandrel (Accessory)Provided with each guidewireProvided with each guidewireSame
Guidewire Introducer (Accessory)Provided with each guidewireProvided with each guidewireSame
Torque Device (Accessory)Provided with each guidewireProvided with each guidewireSame
Sterilization Method100% EO100% EOSame
Shelf Life1 year3 yearDifferent

The subject devices have a different PTFE coating on the proximal end of the guidewire. This difference does not raise new questions of safety and effectiveness, nor does it result in new risks for the subject devices. Testing and evaluation (see below) of the subject devices have been performed with regards to this characteristic and demonstrated the subject devices perform as intended.

NON-CLINICAL PERFORMANCE TESTS

Biocompatibility

A biological risk assessment was performed in accordance with recommendations in the ISO 10993 series and ISO 14971:2019 to assess the impact of the alternative PTFE coating on the biocompatibility profile of the subject devices. The subject devices are considered externally communicating devices with circulating blood contact for a limited duration (≤ 24 hours) and the following tests were performed to assess the modification made to the indirect blood contacting region of the guidewires:

Table 7. Summary of Biocompatibility Testing

TestTest Method SummaryConclusion
CytotoxicityMEM elution cell culture observed for cytotoxic reactivity.Pass: Non-cytotoxic.
PyrogenicityStudy animals were observed for individual temperature rise.Pass: Non-pyrogenic.

Page 15

Scientia Vascular, Inc.
Special 510(k)
K243938
Pg. 11

TestTest Method SummaryConclusion
SensitizationStudy animals with subject device were observed for dermal sensitization.Pass: Non-sensitizing.
IrritationStudy animals with subject device were observed for dermal reaction.Pass: Non-irritant.
Acute Systemic ToxicityStudy animals with the subject device were observed for abnormal clinical signs indicative of toxicity.Pass: No evidence of acute systemic toxicity.
HemocompatibilityThe difference between the hemolytic indexes of the subject device and the negative control was evaluated (extract).Pass: Non-hemolytic.

Sterilization

The existing validated sterilization cycle uses 100% EO to achieve a sterilization assurance level (SAL) of 10-6. Testing for EO and ethylene chlorohydrin (ECH) residuals and bacterial endotoxin levels were performed.

Bench Performance Testing

Following review of the risk assessments conducted in accordance with ISO 14971:2019 performance testing was performed on the subject devices. Additional performance tests to evaluate continued compliance with ISO 11070:2014, "Sterile single-use intravascular introducers, dilators and guidewires," and the FDA guidance document, "Coronary, Peripheral and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019), were also performed. Table 8 summarizes these tests and their results below.

Table 8. Summary of Bench Performance Tests

TestTest Method SummaryResults
Visual Inspection and Dimensional VerificationTest per ISO 11070Acceptance criteria met

Page 16

Scientia Vascular, Inc.
Special 510(k)
K243938
Pg. 12

TestTest Method SummaryResults
Coating Integrity AssessmentTest per ISO 11070 and FDA guidance document "Coronary, Peripheral and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019)Acceptance criteria met
Corrosion ResistanceTest per ISO 11070Acceptance criteria met
Agent CompatibilityGuidewires were tested for compatibility with dimethyl sulfoxide (DMSO), heparinized saline, IsoVue 300 contrast, and IsoVue 370 contrast.Acceptance criteria met
Simulated UseGuidewires were tested for use with a microcatheter, the guidewire introducer, and torque device while navigating to target locations in a simulated use model.Acceptance criteria met
ParticulateParticulate measured and counted after guidewire use in a simulated pathway model with a microcatheter and the guidewire introducer.Acceptance criteria met

Animal Testing

No animal testing was deemed necessary to support the substantial equivalence of the subject devices.

Clinical Testing

No clinical testing was deemed necessary to support the substantial equivalence of the subject devices.

CONCLUSION

The Aristotle 14 Guidewire, Aristotle 18 Guidewire, Aristotle 24 Guidewire, Zoom Wire 14 Guidewire, and Aristotle Colossus Guidewire have the same intended use and indications for use statement as their respective predicate device. The identified technological difference does not raise new questions of safety or effectiveness regarding the use of the subject devices. Risk evaluation along with bench and biocompatibility testing, was completed for the subject devices. The testing and risk evaluation demonstrate that the subject devices are substantially equivalent to their respective predicate device.

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.