K Number
K243928
Device Name
ViaOne Epicardial Access System
Manufacturer
Date Cleared
2025-03-20

(90 days)

Product Code
Regulation Number
870.1340
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The ViaOne Epicardial Access System is intended to access the epicardial surface of the heart via a subxiphoid approach to deliver diagnostic and therapeutic interventions for ventricular tachycardia.
Device Description
ViaOne Epicardial Access System is a sterile, single-use access tool consisting of a concealed needle within a leading channel to facilitate guidewire access into the pericardium through a subxiphoid incision. The device consists of a radiopaque distal tip designed to retract a portion of the pericardial tissue into the device, and away from the heart muscle, where it is punctured by the concealed needle and a guidewire is introduced into the pericardial space.
More Information

Not Found

No
The summary describes a mechanical access tool and its performance studies, with no mention of AI or ML capabilities.

Yes
The device is described as facilitating "diagnostic and therapeutic interventions" for ventricular tachycardia, indicating its role in enabling therapeutic procedures.

Yes

The "Intended Use / Indications for Use" section explicitly states that the device is intended to "deliver diagnostic and therapeutic interventions". Although its primary role described in the "Device Description" is access, it enables diagnostic procedures by allowing access for diagnostic interventions.

No

The device description clearly outlines a physical, sterile, single-use access tool with mechanical components (needle, channel, radiopaque tip) used for a surgical procedure. It requires fluoroscopy for guidance, indicating a hardware dependency beyond software.

Based on the provided information, the ViaOne Epicardial Access System is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • Intended Use: The intended use is to "access the epicardial surface of the heart via a subxiphoid approach to deliver diagnostic and therapeutic interventions for ventricular tachycardia." This describes a procedure performed on the patient's body (in vivo), not a test performed on a sample taken from the body (in vitro).
  • Device Description: The device is described as an "access tool" consisting of a needle and channel to facilitate guidewire access into the pericardium. This is a surgical/interventional tool, not a device used to analyze biological samples.
  • Anatomical Site: The device is used on the "epicardial surface of the heart," which is within the patient's body.
  • Input Imaging Modality: Fluoroscopy is used to guide the procedure, which is an imaging technique used on the patient.

IVD devices are specifically designed to examine specimens derived from the human body (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening. The ViaOne Epicardial Access System is a tool used to perform a medical procedure directly on the patient.

N/A

Intended Use / Indications for Use

The ViaOne Epicardial Access System is intended to access the epicardial surface of the heart via a subxiphoid approach to deliver diagnostic and therapeutic interventions for ventricular tachycardia.

Product codes (comma separated list FDA assigned to the subject device)

DYB

Device Description

ViaOne Epicardial Access System is a sterile, single-use access tool consisting of a concealed needle within a leading channel to facilitate guidewire access into the pericardium through a subxiphoid incision. The device consists of a radiopaque distal tip designed to retract a portion of the pericardial tissue into the device, and away from the heart muscle, where it is punctured by the concealed needle and a guidewire is introduced into the pericardial space.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Fluoroscopy

Anatomical Site

epicardial surface of the heart, pericardium, pericardial space

Indicated Patient Age Range

Not Found

Intended User / Care Setting

OR with imaging equipment, including fluoroscopy under sterile technique.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The following performance data were conducted. The ViaOne Epicardial Access System met the predetermined acceptance criteria ensuring substantial equivalence to the predicate. No new safety or effectiveness issues were raised during testing:

  • Biocompatibility evaluation in compliance with ISO 10993-1 was performed. All tests were successfully completed and passed and the device met the defined acceptance criteria and was found to be biocompatible.
  • Sterilization, Packaging and Shelf Life Testing was performed to demonstrate compliance with ISO 11135-1. In addition, transportation simulation and environmental tests prior to real-time shelf life that included packaging testing and performance to support the one year labeled shelf life. All tests were successfully completed and passed. and the device met the defined acceptance criteria.
  • Performance Testing included Simulated Use, Dimensions, Visual inspection surface analysis, Spring force, Tensile Strength, Radiopacity, Corrosion, Engagement element functionality, Verification testing, following ISO 11070:2014 standards including: Outer tube surface analysis, Outer tube rim analysis, Laser-cut flower shaft & outer tube rim inspection. All tests were successfully completed and passed, and the device met the defined acceptance criteria.
  • Animal Study evaluated the ViaOne Epicardial Access System safety and performance. The devices performed as intended without malfunction. The Histopathology evaluation and pathological examination indicated that there were no clinically significant events following use of the ViaOne Epicardial Access System.
  • Clinical Information: CardioVia provided information on clinical experience from its multi-central clinical trial which included VT subjects. The primary endpoint of the clinical study focused on the occurrence and resolution of adverse events associated with the ViaOne device. The secondary endpoint of the clinical study evaluated the success rate of pericardial space access. The observed adverse events were consistent with those expected in similar epicardial procedures and were resolved without additional intervention. The collected clinical data demonstrated a safety profile consistent with established standards and demonstrates that the ViaOne Epicardial Access System supports its intended use in the indicated population.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K233959

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION" in blue.

March 20, 2025

CardioVia Ltd. Orly Maor Company Regulatory Consultant Wadi El Haj, 13, P.O. Box 1252 Nazareth, 17111 Israel

Re: K243928

Trade/Device Name: ViaOne Epicardial Access System Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter introducer Regulatory Class: Class II Product Code: DYB Dated: December 16, 2024 Received: December 20, 2024

Dear Orly Maor:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

2

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Aneesh S. Deoras -S

Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

3

Indications for Use

510(k) Number (if known) K243928

Device Name ViaOne Epicardial Access System

Indications for Use (Describe)

The ViaOne Epicardial Access System is intended to access the epicardial surface of the heart via a subxiphoid approach to deliver diagnostic and therapeutic interventions for ventricular tachycardia.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

|× | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

4

510(k) Summary

Traditional Premarket Notification Submission - 510(k) ViaOne Epicardial Access System 510(k) Number K243928

Date Prepared: March 20, 2025

L SUBMITTER

CardioVia Ltd. Wadi El Haj, 13, P.O. Box 1252 1711 Nazareth, Israel M: +(972) 524807162 T: +(972) 46098609 W: www.cardiovia.com

Contact Person

Orly Maor 25 Sirkin Street Kfar Saba 44421 Israel Tel: +972-9-7453607 oram.ma@gmail.com

II. DEVICE

Name of Device: ViaOne Epicardial Access System Common or Usual Name: ViaOne Classification Name: Introducer, Catheter Regulatory Class: Class II, per 21 CFR 870.1340 Product Code: DYB. Classification Panel: Cardiovascular

III. PREDICATE DEVICE

CardioVia Ltd. believes that the ViaOne Epicardial Access System, which is the subject of this premarket notification, is substantially equivalent to the following predicate device:

  • AtriCure, Inc. EPi-Ease™ Epicardial Access System (EAS) cleared under K233959 (product . code DYB Regulation No. 21 CFR 870.1340).

DEVICE DESCRIPTION IV.

ViaOne Epicardial Access System is a sterile, single-use access tool consisting of a concealed needle within a leading channel to facilitate guidewire access into the pericardium through a subxiphoid incision. The device consists of a radiopaque distal tip designed to retract a portion of the pericardial tissue into the device, and away from the heart muscle, where it is punctured by the concealed needle and a guidewire is introduced into the pericardial space.

5

INDICATIONS FOR USE v.

The ViaOne Epicardial Access System is intended to access the epicardial surface of the heart via a subxiphoid approach to deliver diagnostic and therapeutic interventions for ventricular tachycardia.

TECHNOLOGICAL CHARACTERISTICS WITH THE VI. COMPARISON OF PREDICATE DEVICE

ViaOne Epicardial Access System has the same intended use as the predicate device. Its indications for use are identical to that of the predicate device. The ViaOne Epicardial Access System has similar technological characteristics to the predicate device as demonstrated in the table below.

| Feature | Proposed:
ViaOne Epicardial
Access System | Predicate:
EPi-Ease™
Epicardial Access
System (EAS) | SE Justification |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K243928 | K233959 | |
| Manufacturer | CardioVia Ltd. | AtriCure, Inc. | |
| Product Code | DYB | DYB | Same |
| CFR | 21 CFR 870.1340 | 21 CFR 870.1340 | Same |
| Intended Use | The ViaOne
Epicardial Access
System is intended to
access the epicardial
surface of the heart via
a subxiphoid approach
to deliver diagnostic
and therapeutic
interventions for
ventricular
tachycardia. | The EPi-Ease™
Epicardial Access
System is intended
to access the
epicardial surface
of the heart via
subxiphoid
approach. | Same
Within the predicate
scope (The indication
was confirmed under
FDA's Breakthrough
Device Designation) |
| Access | Sub-xiphoid | Sub-xiphoid | Same |
| Needle size and
Device profile | 17.5 Ga. Needle
(similar to Tuohy - 17
Ga.)
11 Fr. device profile | 22 Ga. Tuohy
needle
18 Fr. device profile | Similar
ViaOne device profile
is 2/3 of the predicate
and the smaller needle
diameter has no impact
on safety as it is
concealed and intended
to puncture the
pericardium within the
tube |
| Visualization | Fluoroscopy required.
Endoscopic
visualization-Not
required | Fluoroscopy Required.
Endoscopic
visualization-Optional | Same |
| Feature | Proposed:
ViaOne Epicardial
Access System | Predicate:
EPi-Ease™
Epicardial Access
System (EAS) | SE Justification |
| Tissue retraction | Engagement
element enable
retraction | Vacuum enable
retraction | Different
Although capturing of
the tissue is performed
differently it brings the
same result of inserting
the pericardium into a
sealed compartment |
| Radiopaque marker | Yes-distal tip | Yes-distal tip | Same |
| Operating
Principles | Advancement
through adipose
tissue to reach
pericardial sac,
needle punctures
pericardial sac,
guidewire advances
through needle into
pericardial sac,
introducer advances
over guidewire and
pre- dilates
pericardial sac,
sheath advances
over guidewire into
pericardial sac to
access epicardial
surface of the heart.
Needle
advancement limits
to mitigate risks
associated with the
predicate. | Advancement
through adipose
tissue to reach
pericardial sac,
needle punctures
pericardial sac,
guidewire advances
through needle into
pericardial sac,
introducer advances
over guidewire and
pre- dilates
pericardial sac,
sheath advances
over guidewire into
pericardial sac to
access epicardial
surface of the heart.
Needle
advancement limits
to mitigate risks
associated with the
predicate. | Same |
| Reuse | Single Use. | Single Use. | Same |
| Environments of
Use | OR with imaging
equipment,
including
fluoroscopy under
sterile technique. | OR with imaging
equipment,
including
fluoroscopy under
sterile technique. | Same |
| Sterilization | Sterile for single
use (EtO). | Sterile for single
use (EtO). | Same |

6

7

VII. PERFORMANCE DATA

The following performance data were conducted. The ViaOne Epicardial Access System met the predetermined acceptance criteria ensuring substantial equivalence to the predicate. No new safety or effectiveness issues were raised during testing:

. Biocompatibility

Biocompatibility evaluation in compliance with ISO 10993-1 was performed. The following tests were conducted:

  • Cytotoxicity Study - ViaOne
  • Cytotoxicity Study fluid path ●
  • ISO Intracutaneous /irritation Study ●
  • Sensitization Test ●
  • . Acute Systemic Toxicity Study
  • . Pyrogen Study - Material Mediated Study

All tests were successfully completed and passed, and the device met the defined acceptance criteria and was found to be biocompatible.

● Sterilization, Packaging and Shelf Life Testing

Sterilization validation was performed to demonstrate compliance with ISO 11135-1. In addition, transportation simulation and environmental tests prior to real-time shelf life that included packaging testing and performance to support the one year labeled shelf life.

All tests were successfully completed and passed. and the device met the defined acceptance criteria.

● Performance Testing

Performance testing included the following:

  • Simulated Use O
  • o Dimensions
  • Visual inspection surface analysis O
  • Spring force O
  • Tensile Strength O
  • Radiopacity o
  • o Corrosion
  • Engagement element functionality о
  • · Verification testing, following ISO 11070:2014 standards including:
    • Outer tube surface analysis
    • Outer tube rim analysis
    • Laser-cut flower shaft & outer tube rim inspection

All tests were successfully completed and passed, and the device met the defined acceptance criteria.

8

● Animal Study

Animal Study evaluated the ViaOne Epicardial Access System safety and performance. The devices performed as intended without malfunction. The Histopathology evaluation and pathological examination indicated that there were or clinically significant events following use of the ViaOne Epicardial Access System.

● Clinical Information

CardioVia provided information on clinical experience from its multi-central clinical trial which included VT subjects. The primary endpoint of the clinical study focused on the occurrence and resolution of adverse events associated with the ViaOne device. The secondary endpoint of the clinical study evaluated the success rate of pericardial space access. The observed adverse events were consistent with those expected in similar epicardial procedures and were resolved without additional intervention. The collected clinical data demonstrated a safety profile consistent with established standards and demonstrates that the ViaOne Epicardial Access System supports its intended use in the indicated population.

VIII. CONCLUSIONS

CardioVia has demonstrated that the ViaOne Epicardial Access System is substantially equivalent in fundamental design, technology, function, device materials, packaging, sterilization, operating principal, and intended use/ indication for use to the predicate device, the EPi-Ease Epicardial Access System.