(257 days)
The Micro Catheter is intended for the delivery of interventional devices or contrast media into the vasculature of the peripheral and neuro anatomy.
The Micro Catheter is a sterile, single-use, single lumen, variable stiffness, composite catheter. The Micro Catheter is available in three inner diameters (0.017″, 0.021″ and 0.027″), and two working lengths (150cm and 155cm). All models are designed with a straight tip, and are steam shapeable by the user. Single or dual radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the catheter is coated with a hydrophilic coating to increase lubricity. The proximal end of Micro Catheter incorporates a standard luer adapter to facilitate the attachment of accessories. The catheter body has a semi-rigid proximal end which transitions into the flexible distal end to facilitate the advancement of the catheter in the tortuous vasculature. The Micro Catheter is compatible with ≤ 0.014″ guidewires and 5F or larger guide catheters.
The provided 510(k) clearance letter pertains to a Micro Catheter and describes its performance and testing to demonstrate substantial equivalence to a predicate device. This document does not discuss an AI/ML powered device, nor does it present data from a study involving human readers or the establishment of ground truth for AI model training or testing. Therefore, I cannot address most of your specific questions related to AI device evaluation.
However, I can extract the acceptance criteria and performance data for the Micro Catheter based on the provided text.
Acceptance Criteria and Device Performance for Micro Catheter
The document describes the testing performed on the Micro Catheter to demonstrate its substantial equivalence to a predicate device. The "acceptance criteria" are implied by the "Results" column in the tables, indicating whether the device met the required performance standards for each test.
1. Table of Acceptance Criteria and Reported Device Performance
Note: The FDA 510(k) summary for a medical device like a micro catheter focuses on engineering and biocompatibility performance rather than AI-specific metrics. The "acceptance criteria" are implicitly met if the "Results" indicate compliance, comparability to a predicate, or "met the acceptance criteria."
| Test (Category) | Specific Test Method Summary | Acceptance Criteria (Implied by Results) | Reported Device Performance |
|---|---|---|---|
| Bench Performance Testing | |||
| Dimensional Verification | Measured inner/outer diameter, effective length. | Met specified dimensions. | Micro Catheter and accessories met the acceptance criteria. |
| Radiopacity | Visualized under fluoroscopy. | Equivalent to predicate device. | Micro Catheter and the predicate device were imaged showing equivalence in terms of radiopacity. |
| Surface Inspection | Visual inspection under microscopy. | Met visual quality standards. | Micro Catheter met the acceptance criteria. |
| Corrosion Resistance | ISO 10555-1, Annex A. | No signs of corrosion. | Micro Catheter showed no signs of corrosion. |
| Peak Tensile Force/Bond Strength | Evaluated full system tensile force/bond strength. | Met minimum tensile strength requirement. | Micro Catheter met the acceptance criteria. |
| Liquid Leakage | ISO 10555-1, Annex C. | No leakage. | Micro Catheter showed no leakage. |
| Air Leakage | ISO 10555-1, Annex I. | No leakage. | Micro Catheter showed no leakage. |
| Hub Testing | ISO 80369-20. | Met hub standards. | Micro Catheter hub met the acceptance criteria. |
| Flowrate at Maximum Rated Infusion Pressure | Measured flow rate with saline, saline:contrast, contrast. | Met flow rate criteria; comparable to predicate. | Micro Catheter met the acceptance criteria. The mean flow rate values for the subject device and predicate device are comparable for the injectate media tested. |
| Dynamic Burst Pressure | ISO 10555-1, Annex G. | Met burst pressure criteria. | Micro Catheter met the acceptance criteria. |
| Static Burst Pressure | ISO 10555-1, Annex F. | Met burst pressure criteria. | Micro Catheter met the acceptance criteria. |
| Simulated Use | Evaluated in anatomical model for preparation, assembly, compatibility, trackability, lubricity, durability, kink resistance. | Met performance in simulated use. | Micro Catheter met the acceptance criteria. |
| Flexibility and Kink Test | Evaluated resistance to kinking in bends. | Met kink resistance criteria. | Micro Catheter met the acceptance criteria. |
| Torque Strength | Rotated in anatomical model with distal tip fixed; recorded rotations to failure. | Similar rotations to failure as cleared comparator. | Micro Catheter and a cleared comparator showed a similar number of rotations to failure. |
| Coating Integrity | Inspected pre- and post-simulated use. | Met coating integrity standards. | Micro Catheter met the acceptance criteria. |
| Coating Lubricity | Evaluated frictional forces on universal testing machine. | Similar frictional forces to predicate. | Micro Catheter and the predicate showed similar frictional forces. |
| Particulate Evaluation | Evaluated particulate generation during simulated use. | Similar particle numbers to predicate. | Micro Catheter and the predicate showed similar particle numbers. |
| Tip Stiffness | Distal tip deflected on universal testing machine. | Similar tip stiffness to cleared comparator. | Micro Catheter and a cleared comparator showed a similar tip stiffness. |
| Distal Tip Inspection | Inspected for defects. | Met defect criteria. | Distal tip met the acceptance criteria. |
| Tip Shapeability | Shaped using shaping mandrel. | Met shapeability criteria. | Distal tip met the acceptance criteria. |
| Lumen Collapse | Measured force to collapse catheter. | Similar forces to collapse catheter as predicate. | Micro Catheter and the predicate showed similar forces to collapse the catheter. |
| Compatibility tests | Inspected for damage post-simulated use with compatible interventional devices. | Met compatibility criteria. | Micro Catheter met the acceptance criteria. |
| Biocompatibility Testing | |||
| ISO MEM Elution Test | ISO 10993-5 | Reactivity grade ≤2. | Non-cytotoxic (reactivity grade of ≤2). |
| ISO Guinea Pig Maximization Sensitization Test | ISO 10993-10 | No evidence of delayed dermal contact sensitization. | Non-sensitizer. |
| Intracutaneous Reactivity Test in Rabbits | ISO 10993-23 | Differences between test and control mean scores < 1.0. | Non-irritant. |
| Acute Systemic Toxicity Study in Mice | ISO 10993-11 | No mortality or evidence of acute systemic toxicity. | No evidence of acute systemic toxicity. |
| Material-Mediated Pyrogenicity Test in Rabbits | ISO 10993-11 | No temperature rise ≥0.5°C. | Non-pyrogenic. |
| ASTM Hemolysis Test | ISO 10993-4 | Hemolytic index < 2%. | Non-hemolytic. |
| Complement Activation SC5b-9 Assay | ISO 10993-4 | SC5b-9 concentration statistically lower than negative reference and comparator. | Not an activator. |
| Non-anticoagulated Venous Implant Study | ISO 10993-4 | Extent of thrombus formation not greater than comparator. | Non-thrombogenic. |
| Sterilization & Shelf Life | |||
| Sterilization | Ethylene Oxide. | Sterility Assurance Level (SAL) of 10⁻⁶. | Verified SAL of 10⁻⁶ in accordance with ISO 11135. |
| Shelf-Life | Aging studies. | Demonstrated 2-year shelf life. | Demonstrated a 2-year shelf-life. |
| Packaging Aging Tests | N/A | All acceptance criteria met. | Results met all acceptance criteria. |
| Packaging | |||
| Sterile Barrier System (SBS) Validation | ISO 11607-2. | Met requirements. | SBS validation conducted per ISO 11607-2. |
Based on the provided text, the following questions cannot be answered as they are relevant to AI/ML or image analysis devices, not the physical Micro Catheter:
- Sample sizes used for the test set and the data provenance: Not applicable. Performance is based on physical bench and biocompatibility testing, not a "test set" of data for an AI model.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. There is no "ground truth" establishment in the context of an AI model. Testing involves objective measurements and compliance with standards.
- Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a physical device, not an AI assistance tool.
- If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. The "ground truth" in this context is compliance with engineering standards and physiological responses (e.g., non-toxic, non-pyrogenic).
- The sample size for the training set: Not applicable. There is no AI model or "training set."
- How the ground truth for the training set was established: Not applicable.
This 510(k) clearance is for a conventional Class II medical device, and therefore the evaluation and acceptance criteria are based on established engineering, material science, and biological safety standards, not AI/ML performance metrics.
U.S. Food & Drug Administration 510(k) Clearance Letter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.00
July 30, 2025
Suzhou Zenith Vascular SciTech Limited
Jie Xia
Chief Executive Officer
Building 6, Block B, Phase 5, Biobay,
No.21 Dongyanli Road,
SIP Suzhou 215123
China
Re: K243534
Trade/Device Name: Micro Catheter
Regulation Number: 21 CFR 870.1210
Regulation Name: Continuous Flush Catheter
Regulatory Class: Class II
Product Code: KRA, QJP
Dated: June 26, 2025
Received: June 26, 2025
Dear Jie Xia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K243534 - Jie Xia
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K243534 - Jie Xia
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan -S
Naira Muradyan, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K243534
Device Name: Micro Catheter
Indications for Use (Describe):
The Micro Catheter is intended for the delivery of interventional devices or contrast media into the vasculature of the peripheral and neuro anatomy.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
510(k) Summary-K243534
Micro Catheter K243534
1 / 5
Device Description
The Micro Catheter is a sterile, single-use, single lumen, variable stiffness, composite catheter. The Micro Catheter is available in three inner diameters (0.017″, 0.021″ and 0.027″), and two working lengths (150cm and 155cm). All models are designed with a straight tip, and are steam shapeable by the user. Single or dual radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the catheter is coated with a hydrophilic coating to increase lubricity. The proximal end of Micro Catheter incorporates a standard luer adapter to facilitate the attachment of accessories. The catheter body has a semi-rigid proximal end which transitions into the flexible distal end to facilitate the advancement of the catheter in the tortuous vasculature. The Micro Catheter is compatible with ≤ 0.014″ guidewires and 5F or larger guide catheters.
| Applicant | Suzhou Zenith Vascular SciTech Limited |
|---|---|
| Building 6, Block B, Phase 5, Biobay, No.21 Dongyanli Road, SIP Suzhou, China 215123 | |
| Contact Person | Jie Xia |
| Contact Telephone | +86-18051093308 |
| Date Prepared | July 28, 2025 |
| Device Trade Name | Micro Catheter |
| Common Name | Percutaneous Catheter |
| Regulation Number | 21 CFR 870.1210, 21 CFR 870.1250 |
| Regulation Name | Continuous Flush Catheter, Percutaneous Catheter, Neurovasculature |
| Product Code | KRA, QJP |
| Device Classification | Class II |
| Predicate Device | Rebar™ Micro Catheter (K210114) |
Page 6
Micro Catheter K243534
2 / 5
Indications for Use
The Micro Catheter is intended for the delivery of interventional devices or contrast media into the vasculature of the peripheral and neuro anatomy.
Predicate Device Comparison
The predicate device for the Micro Catheter is the Rebar™ Micro Catheter (K210114). The table below describes the differences between the subject and predicate device.
Table 1. Comparison of the Subject Device and Predicate Device
| Characteristics | Subject Device- Micro Catheter(K243534) | Predicate Device- Rebar Micro Catheter(K210114) | Justification of Substantial Equivalence |
|---|---|---|---|
| Product Code | KRA, QJP | KRA, QJP | Same |
| Indications for Use | The Micro Catheter is intended for the delivery of interventional devices or contrast media into the vasculature of the peripheral and neuro anatomy. | The Rebar™ Micro Catheter is intended for the delivery of interventional devices or contrast media into the vasculature of the peripheral and neuro anatomy. | Same |
| Materials | |||
| Inner Layer | PTFE | PTFE | Same |
| Middle Layer | Stainless steel | Stainless steel | Same |
| Outer Jacket | Pebax | Pebax | Same |
| Coating | Hydrophilic coating | Hydrophilic coating | Same |
| Radiopaque Marker | Platinum-Iridium Alloy | Platinum-Iridium Alloy | Same |
| Strain Relief | Polyurethane/PU | Elvax and Dynaflex | Different. The subject device was evaluated via bench and biocompatibility testing with passing or acceptable results. |
| Hub | Polycarbonate | Polypropylene | |
| Specification | |||
| Model | 200017150/200017155, 200021150/200021155, 200027150/200027155 | Rebar 18, Rebar 27 | N/A |
| Proximal OD | 0.81 mm, 0.91 mm, 0.93 mm | 0.89 mm Max, 1.09 mm Max | Similar. The design validation testing has demonstrated the subject device performs as intended. |
| Distal OD | 0.66 mm, 0.81 mm, 0.93 mm | 0.93 mm Max, 1.09 mm Max | |
| Inner Diameter | 0.017 inch, 0.021 inch, 0.027 inch | 0.020 inch Min, 0.026 inch Min | |
| Effective Length | 150 cm and 155 cm | 130 cm and 153 cm, 130 cm and 145cm | |
| Coating Length | 80 cm, 100 cm, 95 cm | ||
| Tip Configuration | Straight, shapeable | Straight, shapeable | Same |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
| How Supplied | Sterile, non-pyrogenic. | Sterile, non-pyrogenic. | Same |
| Shelf-Life | 2 years | 2 years | Same |
Page 7
Micro Catheter K243534
3 / 5
Performance Data
The Micro Catheter was evaluated via the following non-clinical testing.
Performance Testing-Bench
The following bench performance testing was performed to characterize the Micro Catheter.
Table 2. Bench Performance Testing
| Test | Test Method Summary | Results |
|---|---|---|
| Dimensional Verification | The dimensions (Inner Diameter, Outer Diameter, Effective Length) of the catheter were measured. The inner diameter of the introducer sheath was measured. The outer diameter of the shaping mandrel was measured. | Micro Catheter and accessories met the acceptance criteria. |
| Radiopacity | The catheter was visualized under fluoroscopy. | Micro Catheter and the predicate device were imaged showing equivalence in terms of radiopacity. |
| Surface Inspection | The catheter was visually inspected under microscopy. | Micro Catheter met the acceptance criteria. |
| Corrosion Resistance | The catheter was tested according to ISO 10555-1, Annex A. | Micro Catheter showed no signs of corrosion. |
| Peak Tensile Force/Bond Strength | The catheter was evaluated to verify the peak tensile force/bond strength of the full system meets the minimum tensile strength requirement. | Micro Catheter met the acceptance criteria. |
| Liquid Leakage | The catheter was tested according to ISO 10555-1, Annex C. | Micro Catheter showed no leakage. |
| Air Leakage | The catheter was tested according to ISO 10555-1, Annex I. | Micro Catheter showed no leakage. |
| Hub Testing | The hub was tested according to ISO 80369-20. | Micro Catheter hub met the acceptance criteria. |
| Flowrate at Maximum Rated Infusion Pressure | The flow rate at the maximum rated infusion pressure was measured using saline, 50:50 saline: contrast, and 100% contrast. | Micro Catheter met the acceptance criteria. The mean flow rate values for the subject device and predicate device are comparable for the injectate media tested. |
| Dynamic Burst Pressure | The catheter was tested according to ISO 10555-1, Annex G. | Micro Catheter met the acceptance criteria. |
| Static Burst Pressure | The catheter was tested according to ISO 10555-1, Annex F. | Micro Catheter met the acceptance criteria. |
| Simulated Use | The catheter was evaluated in a simulated anatomical model for the preparation and ease of assembly, introducer sheath compatibility and peel away, ancillary device compatibility with a guidewire and guide catheter, trackability, lubricity and durability of the hydrophilic coating, and kink resistance. | Micro Catheter met the acceptance criteria. |
| Flexibility and Kink Test | The catheter was evaluated for resistance to kinking around bends with clinically relevant radii. | Micro Catheter met the acceptance criteria. |
| Torque Strength | The catheter was evaluated for torque strength by rotating the test samples in an anatomical model with the distal tip fixed | Micro Catheter and a cleared comparator showed a similar number of rotations to |
Page 8
Micro Catheter K243534
4 / 5
| Test | Test Method Summary | Results |
|---|---|---|
| and recording the number of rotations to failure. | failure. | |
| Coating Integrity | The catheter coating integrity was inspected pre- and post-simulated use in an anatomical model. | Micro Catheter met the acceptance criteria. |
| Coating Lubricity | The catheter was evaluated for frictional forces on a universal testing machine. | Micro Catheter and the predicate showed similar frictional forces. |
| Particulate Evaluation | The catheter was evaluated for particulate generation during simulated use in an anatomical model. | Micro Catheter and the predicate showed similar particle numbers. |
| Tip Stiffness | The distal tip of the catheter was deflected on a universal testing machine. | Micro Catheter and a cleared comparator showed a similar tip stiffness. |
| Distal Tip Inspection | The distal tip of the catheter was inspected for defects. | Distal tip met the acceptance criteria. |
| Tip Shapeability | The distal tip of the catheter was shaped using the shaping mandrel supplied with Micro Catheter. | Distal tip met the acceptance criteria. |
| Lumen Collapse | The force needed to collapse the catheter at the tip and along the catheter shaft was measured using a universal testing machine. | Micro Catheter and the predicate showed similar forces to collapse the catheter. |
| Compatibility tests | The catheter was inspected for damage post simulated use testing with compatible interventional devices in a neurovascular model. | Micro Catheter met the acceptance criteria. |
The performance data demonstrate that the subject device is substantially equivalent to the predicate devices.
Biocompatibility Testing
Biocompatibility tests were conducted with the Micro Catheter in accordance with ISO 10993-1:2018, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", and the FDA biological evaluation guidance. The performed tests are listed in Table 3.
Table 3. Biocompatibility Testing
| Test | Standards | Results | Conclusion |
|---|---|---|---|
| ISO MEM Elution Test with L-929 Cells | ISO 10993-5 | The test article had a reactivity grade of ≤2 under the conditions of this study. | Non-cytotoxic |
| ISO Guinea Pig Maximization Sensitization Test (2 Extracts) | ISO 10993-10 | The saline and sesame oil test article extracts did not show evidence of causing delayed dermal contact sensitization response in the guinea pigs. | Non-sensitizer |
| Intracutaneous Reactivity Test in Rabbits (2 Extracts) | ISO 10993-23 | The differences between the test extracts and their respective control mean scores were less than 1.0. | Non-irritant |
| Acute Systemic Toxicity Study in Mice (2 Extracts) | ISO 10993-11 | There was no mortality or evidence of acute systemic toxicity from the extracts. | No evidence of acute systemic toxicity |
| Material-Mediated Pyrogenicity Test in Rabbits (Saline Extract) | ISO 10993-11 | None of the test rabbits exhibited a temperature rise greater than or equal to 0.5°C. | Non-pyrogenic |
| ASTM Hemolysis Test - Direct Contact and Extract Methods | ISO 10993-4 | The test article hemolytic index was above the negative control and was less than 2% for both direct contact and extract methods under the conditions of this study. | Non-hemolytic |
Page 9
Micro Catheter K243534
5 / 5
| Test | Standards | Results | Conclusion |
|---|---|---|---|
| Complement Activation SC5b-9 Assay (with Comparator Control Article) | ISO 10993-4 | The concentration of SC5b-9 activated by the test article was statistically lower than that activated by the negative reference material and comparator control article. | Not an activator |
| Non-anticoagulated Venous Implant Study in Canines for Evaluation of Thromboresistance | ISO 10993-4 | The extent of thrombus formation on the test article was not greater than the comparator control article. | Non-thrombogenic |
Sterilization and Shelf Life
The Micro Catheter is sterilized using ethylene oxide. The sterilization cycle was verified to ensure a sterility assurance level (SAL) of 10⁻⁶ in accordance with ISO 11135.
Aging studies for subject device have been conducted and demonstrated a 2 year shelf-life. Packaging aging tests were also conducted and the results met all acceptance criteria.
Packaging
Each package consists of a Sterile Barrier System (SBS) and protective packaging. SBS validation was conducted per requirements of ISO 11607-2.
Animal Study
An animal study was not deemed necessary to support the substantial equivalence of the subject device to the predicate device. Non-clinical testing described above was determined to be sufficient to support substantial equivalence.
Clinical Study
Clinical testing was not deemed necessary to support the substantial equivalence of the subject device to the predicate device. Non-clinical testing described above was determined to be sufficient to support substantial equivalence.
Conclusion
The subject and predicate devices have the same intended use and the same or similar technological characteristics. The non-clinical testing demonstrates that the subject Micro Catheter performs as intended and supports substantial equivalence to the predicate device. The differences in technological characteristics do not raise new questions of safety and effectiveness.
§ 870.1210 Continuous flush catheter.
(a)
Identification. A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.(b)
Classification. Class II (performance standards).