AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The glove is disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves were tested for use with Chemotherapy drugs, Fentanyl and Gastric acid in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.

Device Description

Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid (Light blue, Dark blue) is a Class I patient examination glove and specialty chemotherapy gloves, that made from synthetic nitrile latex. They are non-sterile, powder-free, fingertip textured, ambidextrous with beaded cuff, and single use only, and come in different sizes- XS, S, M, L, XL and XXL.

AI/ML Overview

The provided [K243441](https://510k.innolitics.com/search/K243441) document is a 510(k) premarket notification for Nitrile Powder-Free Exam Gloves. This type of device is a Class I medical device and does not involve AI or complex algorithms requiring human-in-the-loop studies or ground truth established by experts.

Therefore, most of the requested information regarding AI/algorithm performance, human reader studies, and expert-established ground truth is not applicable to this document. The "device" in this context is a physical examination glove, not an AI-powered diagnostic tool.

The acceptance criteria and performance are related to the physical properties and barrier efficacy of the gloves, not to analytical or diagnostic performance of an algorithm.

However, I can extract the relevant acceptance criteria and reported performance for the physical glove testing:


Acceptance Criteria and Reported Device Performance

As the device is a physical medical glove and not an AI/software device, the acceptance criteria and performance relate to its physical and chemical barrier properties.

Table of Acceptance Criteria and Reported Device Performance:

Test PerformedMethodologyAcceptance CriteriaReported Device Performance (Result)
Physical Properties
Freedom From HolesASTM D6319-19, ASTM D5151-19Meet requirement inspection level G-I, AQL 2.5 (ISO2859-1)Pass
Dimension - LengthASTM D6319-19Minimum 220mm for size XS-S; Minimum 230mm for size M-XXLPass
Dimension - WidthASTM D6319-19XS: 70±10mm; S: 80±10mm; M: 95±10mm; L: 110±10mm; XL: 120±10mm; XXL: 130±10mmPass
Dimension - ThicknessASTM D6319-19Finger: 0.05mm (min); Palm: 0.05mm (min)Pass
Physical properties (Tensile Strength)ASTM D6319-19, ASTM D412-16Min 14MPa (Before aging & After aging)Pass
Physical properties (Ultimate Elongation)ASTM D6319-19, ASTM D412-16Min 500% (Before aging); Min 400% (After aging)Pass
Powder ResidueASTM D6319-19, ASTM D6124-06Not more than 2 mg per glovePass
Biocompatibility
Skin IrritationISO10993-23:2021Not a primary skin irritantPass
Skin SensitizationISO10993-10:2021Not a contact sensitizerPass
Acute Systemic ToxicityISO 10993-11:2017No acute systemic toxicityPass
Permeation Testing
Chemotherapy Drug, Fentanyl, Gastric AcidASTM D6978-05Minimum Breakthrough Detection Time (minutes) - Specific values for each substanceSee detailed tables below

Detailed Permeation Performance (Minimum Breakthrough Detection Time in minutes):

(Light Blue Glove)

Chemotherapy DrugMinimum Breakthrough Detection Time (minutes)
Bendamustine HCl, 5 mg/ml>240
Bortezomib, 1 mg/ml>240
Busulfan, 6 mg/ml>240
Carboplatin, 10 mg/ml>240
Carmustine, 3.3 mg/ml16.6
Carfilzomib, 2 mg/ml>240
Cetuximab (Erbitux), 2 mg/ml>240
Chloroquine, 50mg/ml>240
Cisplatin, 1 mg/ml>240
Cladribine, 1.0 mg/ml>240
Cyclophosphamide, 20 mg/ml>240
Cyclosporin A,100mg/ml>240
Cytarabine HCl, 100 mg/ml>240
Cytovene, 10 mg/ml>240
Dacarbazine, 10 mg/ml>240
Daunorubicin HCl, 5 mg/ml>240
Decitabine, 5 mg/ml>240
Docetaxel, 10 mg/ml>240
Doxorubicin HCl, 2 mg/ml>240
Epirubicin HCl, 2 mg/ml>240
Etoposide, 20 mg/ml>240
Fludarabine, 25 mg/ml>240
Fluorouracil, 50 mg/ml>240
Gemcitabine HCl, 38 mg/ml>240
Idarubicin HCl, 1 mg/ml>240
Ifosfamide, 50 mg/ml>240
Irinotecan HCl, 20 mg/ml>240
Mechlorethamine HCl, 1 mg/ml>240
Melphalan HCl, 5 mg/ml>240
Methotrexate, 25 mg/ml>240
Mitomycin-C, 0.5 mg/ml>240
Mitoxantrone HCl, 2 mg/ml>240
Oxaliplatin, 2 mg/ml>240
Paclitaxel, 6 mg/ml>240
Retrovir, 10 mg/ml>240
ThioTEPA, 10 mg/ml44.9
Trisenox, 1 mg/ml>240
Vidaza (Azacitidine),25 mg/ml>240
Vincristine Sulfate, 1 mg/ml>240
Fentanyl Citrate Injection, 100mcg/2ml>240
Simulated Gastric Acid>240

Warning: Do not use Light Blue glove with Carmustine (16.6 minutes) and ThioTEPA (44.9 minutes).

(Dark Blue Glove)

Chemotherapy DrugMinimum Breakthrough Detection Time (minutes)
Bendamustine HCl, 5 mg/ml>240
Bortezomib, 1 mg/ml>240
Busulfan, 6 mg/ml>240
Carboplatin, 10 mg/ml>240
Carmustine, 3.3 mg/ml21.5
Carfilzomib, 2 mg/ml>240
Cetuximab (Erbitux), 2 mg/ml>240
Chloroquine, 50mg/ml>240
Cisplatin, 1 mg/ml>240
Cladribine, 1.0 mg/ml>240
Cyclophosphamide, 20 mg/ml>240
Cyclosporin A,100mg/ml>240
Cytarabine HCl, 100 mg/ml>240
Cytovene, 10 mg/ml>240
Dacarbazine, 10 mg/ml>240
Daunorubicin HCl, 5 mg/ml>240
Decitabine, 5 mg/ml>240
Docetaxel, 10 mg/ml>240
Doxorubicin HCl, 2 mg/ml>240
Epirubicin HCl, 2 mg/ml>240
Etoposide, 20 mg/ml>240
Fludarabine, 25 mg/ml>240
Fluorouracil, 50 mg/ml>240
Gemcitabine HCl, 38 mg/ml>240
Idarubicin HCl, 1 mg/ml>240
Ifosfamide, 50 mg/ml>240
Irinotecan HCl, 20 mg/ml>240
Mechlorethamine HCL, 1 mg/ml>240
Melphalan HCl, 5 mg/ml>240
Methotrexate, 25 mg/ml>240
Mitomycin-C, 0.5 mg/ml>240
Mitoxantrone HCl, 2 mg/ml>240
Oxaliplatin, 2 mg/ml>240
Paclitaxel, 6 mg/ml>240
Retrovir, 10 mg/ml>240
ThioTEPA, 10 mg/ml17.9
Trisenox, 1 mg/ml>240
Vidaza (Azacitidine),25 mg/ml>240
Vincristine Sulfate, 1 mg/ml>240
Fentanyl Citrate Injection, 100mcg/2ml>240
Simulated Gastric Acid>240

Warning: Do not use Dark Blue glove with Carmustine (21.5 minutes) and ThioTEPA (17.9 minutes).


Regarding the other requested information (which is not applicable to this physical device):

  • 2. Sample size(s) used for the test set and the data provenance: Not explicitly stated as "sample size" in a statistical sense for AI. Testing involves physical samples of gloves manufactured for the purpose of meeting ASTM standards. Data provenance is implied to be from laboratory testing according to the listed ASTM and ISO standards.
  • 3. Number of experts used to establish the ground truth...: Not applicable. Ground truth for physical properties and chemical permeation is established by standardized laboratory testing procedures (e.g., measuring dimensions, tensile strength, or breakthrough time), not by expert consensus on images or medical data.
  • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Adjudication is not relevant for objective laboratory measurements.
  • 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done...: Not applicable. This is a physical product, not an AI or imaging device that would involve human readers.
  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. There is no algorithm.
  • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): The "ground truth" for this device is based on objective measurements and established physical/chemical standards (e.g., ASTM, ISO), not expert medical interpretation or clinical outcomes data in the context of an AI device.
  • 8. The sample size for the training set: Not applicable. There is no training set as this is not an AI/ML device.
  • 9. How the ground truth for the training set was established: Not applicable. There is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 31, 2025

Basic Medical Technology Inc. John Zhao General Manager 5300 Concours Street Ontario, California 91764

Re: K243441

Trade/Device Name: Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid (Light blue, Dark blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO, OPJ Dated: November 6, 2024 Received: November 6, 2024

Dear John Zhao:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Allan Guan -S

For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K243441

Device Name

Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid (Light blue, Dark blue)

Indications for Use (Describe)

The glove is disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves were tested for use with Chemotherapy drugs, Fentanyl and Gastric acid in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs. Light Blue glove: Chemotherapy Drugs Minimum Breakthrough Detection Time (minutes) Bendamustine HCl. 5 mg/ml >240 Bortezomib, 1 mg/ml >240 Busulfan, 6 mg/ml >240 Carboplatin, 10 mg/ml >240 Carmustine, 3.3 mg/ml 16.6 >240 Carfilzomib, 2 mg/ml Cetuximab (Erbitux) 2 mg/ml >240 Chloroquine, 50mg/ml >240. >240 Cisplatin, 1 mg/ml >240 Cladribine, 1.0 mg/ml >240 Cyclophosphamide, 20 mg/ml Cyclosporin A, 100mg/ml >240 Cytarabine HCl, 100 mg/ml >240 Cytovene, 10 mg/ml >240 >240 Dacarbazine, 10 mg/ml Daunorubicin HCI, 5 mg/ml >240 >240 Decitabine, 5 mg/ml Docetaxel, 10 mg/ml >240 Doxorubicin HCl, 2 mg/ml >240 Epirubicin HCl, 2 mg/ml >240 Etoposide, 20 mg/ml >240 Fludarabine, 25 mg/ml >240 Fluorouracil, 50 mg/ml >240 Gemcitabine HCl., 38 mg/ml >240 Idarubicin HCl,, 1 mg/ml >240 Ifosfamide, 50 mg/ml >240 Irinotecan. 20 mg/ml >240 Mechlorethamine HCl, 1 mg/ml >240 Melphalan HCl. 5 mg/ml >240 Methotrexate, 25 mg/ml >240 Mitomycin-C, 0.5 mg/ml >240 Mitoxantrone HCl. 2 mg/ml >240 Oxaliplatin, 2 mg/ml >240 Paclitaxel, 6 mg/ml >240

{4}------------------------------------------------

Retrovir, 10 mg/ml>240
ThioTEPA, 10 mg/ml44.9
Trisenox, 1 mg/ml>240
Vidaza (Azacitidine),25 mg/ml>240
Vincristine Sulfate, 1 mg/ml>240
Fentanyl Citrate and Gastric AcidMinimum Breakthrough Detection Time, minutes
Fentanyl Citrate Injection, 100mcg/2ml>240
Simulated Gastric Acid>240
*Please note that the following drugs have extremely low permeation times:
Carmustine: 16.6 minutes, Thio Tepa: 44.9 minutes
Warning: Do not use with Carmustine and Thio Tepa.
Dark Blue glove:
Chemotherapy DrugsMinimum Breakthrough Detection Time (minutes)
Bendamustine HCI, 5 mg/ml>240
Bortezomib, 1 mg/ml>240
Busulfan, 6 mg/ml>240
Carboplatin, 10 mg/ml>240
Carmustine, 3.3 mg/ml21.5
Carfilzomib, 2 mg/ml>240
Cetuximab (Erbitux) 2 mg/ml>240
Chloroquine, 50mg/ml>240.
Cisplatin, 1 mg/ml>240
Cladribine, 1.0 mg/ml>240
Cyclophosphamide, 20 mg/ml>240
Cyclosporin A, 100mg/ml>240
Cytarabine HCI, 100 mg/ml>240
Cytovene, 10 mg/ml>240
Dacarbazine, 10 mg/ml>240
Daunorubicin HCI, 5 mg/ml>240
Decitabine, 5 mg/ml>240
Docetaxel, 10 mg/ml>240
Doxorubicin HCl, 2 mg/ml>240
Epirubicin HCl, 2 mg/ml>240
Etoposide, 20 mg/ml>240
Fludarabine, 25 mg/ml>240
Fluorouracil, 50 mg/ml>240
Gemcitabine HCl, 38 mg/ml>240
Idarubicin HCl, 1 mg/ml>240
Ifosfamide, 50 mg/ml>240
Irinotecan, 20 mg/ml>240
Mechlorethamine HCl, 1 mg/ml>240
Melphalan HCl, 5 mg/ml>240
Methotrexate, 25 mg/ml>240
Mitomycin-C, 0.5 mg/ml>240
Mitoxantrone HCl, 2 mg/ml>240
Oxaliplatin, 2 mg/ml>240
Paclitaxel, 6 mg/ml>240
Retrovir, 10 mg/ml>240
ThioTEPA, 10 mg/ml17.9

{5}------------------------------------------------

Trisenox, 1 mg/ml>240
Vidaza (Azacitidine),25 mg/ml>240
Vincristine Sulfate, 1 mg/ml>240
Fentanyl Citrate and Gastric AcidMinimum Breakthrough Detection Time, minutes
Fentanyl Citrate Injection, 100mcg/2ml>240
Simulated Gastric Acid>240

*Please note that the following drugs have extremely low permeation times:

Carmustine: 21.5 minutes, Thio Tepa: 17.9 minutes

Warning: Do not use with Carmustine and Thio Tepa.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

5300 Concours Ontario, CA 91764

510(K) Summary K243441

1. Submission Information

Date Prepared: January 31, 2025 Submission Format: Traditional 510(k) 510(K)#: K243441

2. Submitter Information

Applicant Name: Basic Medical Technology Inc. Address: 5300 Concours Street, Ontario, CA 91764

Contact Person: John Zhao Tel : (909) 980-1678 Email: johnzhao@basicmedical.com

3. Device Information

Trade Name: Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid (Light blue, Dark blue)

Common Name: Nitrile Powder-Free Exam Gloves Classification Name: Non-Powdered Patient Examination Glove Regulation: 21 CFR 880.6250 Product Code: LZA, LZC, QDO, OPJ Classification Panel: General Hospital Device Class: Class I

4. Predicate Device Information

Lingshi Hongruida Health Protection Technology Co., Ltd.

Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and select other drugs (K240051)

Product code : LZA, LZC, QDO, OPJ.

5. Device Description:

Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid (Light blue, Dark blue) is a Class I patient examination glove and specialty chemotherapy gloves, that made from synthetic nitrile latex. They are non-sterile, powder-free, fingertip textured, ambidextrous with beaded cuff, and single use only, and come in different sizes- XS, S, M, L, XL and XXL.

{7}------------------------------------------------

5300 Concours Ontario, CA 91764

The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs, Fentanyl Citrate and Gastric Acid.

6. Indications for Use:

The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves were tested for use with Chemotherapy drugs, Fentanyl and Gastric acid in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.

Light Blue glove:

NO.Chemotherapy drugMinimum Breakthrough Detection Time, minutes
1Bendamustine HCl, 5 mg/ml>240
2Bortezomib, 1 mg/ml>240
3Busulfan, 6 mg/ml>240
4Carboplatin, 10 mg/ml>240
5Carmustine, 3.3 mg/ml16.6
6Carfilzomib, 2 mg/ml>240
7Cetuximab (Erbitux), 2 mg/ml>240
8Chloroquine, 50mg/ml>240
9Cisplatin, 1 mg/ml>240
10Cladribine, 1.0 mg/ml>240
11Cyclophosphamide, 20 mg/ml>240
12Cyclosporin A,100mg/ml>240
13Cytarabine HCl, 100 mg/ml>240
14Cytovene, 10 mg/ml>240
15Dacarbazine, 10 mg/ml>240
16Daunorubicin HCl, 5 mg/ml>240
17Decitabine, 5 mg/ml>240
18Docetaxel, 10 mg/ml>240
19Doxorubicin HCl, 2 mg/ml>240
20Epirubicin HCl, 2 mg/ml>240
21Etoposide, 20 mg/ml>240
22Fludarabine, 25 mg/ml>240
23Fluorouracil, 50 mg/ml>240
24Gemcitabine HCl, 38 mg/ml>240
25Idarubicin HCl, 1 mg/ml>240
26Ifosfamide, 50 mg/ml>240
27Irinotecan HCl, 20 mg/ml>240
28Mechlorethamine HCl, 1 mg/ml>240
29Melphalan HCl, 5 mg/ml>240
30Methotrexate, 25 mg/ml>240
31Mitomycin-C, 0.5 mg/ml>240
32Mitoxantrone HCl, 2 mg/ml>240

{8}------------------------------------------------

5300 Concours Ontario, CA 91764

33Oxaliplatin, 2 mg/ml>240
34Paclitaxel, 6 mg/ml>240
35Retrovir, 10 mg/ml>240
36ThioTEPA, 10 mg/ml44.9
37Trisenox, 1 mg/ml>240
38Vidaza (Azacitidine),25 mg/ml>240
39Vincristine Sulfate,1 mg/ml>240
NO.Fentanyl Citrate and Gastric AcidMinimum Breakthrough Detection Time, minutes
1Fentanyl Citrate Injection, 100mcg/2ml>240
2Simulated Gastric Acid>240

*Please note that the following drugs have extremely low permeation times:

Carmustine: 16.6 minutes, Thio Tepa: 44.9 minutes

Warning: Do not use with Carmustine and Thio Tepa.

Dark Blue glove:

NO.Chemotherapy drugMinimum Breakthrough Detection Time, minutes
1Bendamustine HCl, 5 mg/ml>240
2Bortezomib, 1 mg/ml>240
3Busulfan, 6 mg/ml>240
4Carboplatin, 10 mg/ml>240
5Carmustine, 3.3 mg/ml21.5
6Carfilzomib, 2 mg/ml>240
7Cetuximab (Erbitux), 2 mg/ml>240
8Chloroquine, 50mg/ml>240
9Cisplatin, 1 mg/ml>240
10Cladribine, 1.0 mg/ml>240
11Cyclophosphamide, 20 mg/ml>240
12Cyclosporin A,100mg/ml>240
13Cytarabine HCl, 100 mg/ml>240
14Cytovene, 10 mg/ml>240
15Dacarbazine, 10 mg/ml>240
16Daunorubicin HCl, 5 mg/ml>240
17Decitabine, 5 mg/ml>240
18Docetaxel, 10 mg/ml>240
19Doxorubicin HCl, 2 mg/ml>240
20Epirubicin HCl, 2 mg/ml>240
21Etoposide, 20 mg/ml>240
22Fludarabine, 25 mg/ml>240
23Fluorouracil, 50 mg/ml>240
24Gemcitabine HCl, 38 mg/ml>240
25Idarubicin HCl, 1 mg/ml>240
26Ifosfamide, 50 mg/ml>240
27Irinotecan HCl, 20 mg/ml>240
28Mechlorethamine HCL, 1 mg/ml>240
29Melphalan HCl, 5 mg/ml>240

{9}------------------------------------------------

5300 Concours Ontario, CA 91764

30Methotrexate, 25 mg/ml>240
31Mitomycin-C, 0.5 mg/ml>240
32Mitoxantrone HCl, 2 mg/ml>240
33Oxaliplatin, 2 mg/ml>240
34Paclitaxel, 6 mg/ml>240
35Retrovir, 10 mg/ml>240
36ThioTEPA, 10 mg/ml17.9
37Trisenox, 1 mg/ml>240
38Vidaza (Azacitidine),25 mg/ml>240
39Vincristine Sulfate,1 mg/ml>240
NO.Fentanyl Citrate and Gastric AcidMinimum Breakthrough Detection Time, minutes
1Fentanyl Citrate Injection, 100mcg/2ml>240
2Simulated Gastric Acid>240

*Please note that the following drugs have extremely low permeation times:

Carmustine: 21.5 minutes, Thio Tepa:17.9 minutes

Warning: Do not use with Carmustine and Thio Tepa.

7. Comparison of Subject Device and Predicate Device:

General Comparison Table:

Proposed DevicePredicate DeviceK240051Result
Product NameNitrile Powder-Free Exam GlovesTested for Use with ChemotherapyDrugs, Opioid Fentanyl Citrate,Simulated Gastric Acid (Light blue, Darkblue)Powder Free Nitrile ExaminationGloves (Blue), Tested for Use withChemotherapy Drugs. Fentanyl Citrate,and select other drugs
Product CodeLZA, LZC, QDO, OPJLZA, LZC, QDO, OPJSame
RegulationNumber21 CFR 880.625021 CFR 880.6250Same
DeviceClassificationClass IClass ISame
Indications foruseThe glove is a disposable deviceintended for medical purposes that isworn on the examiner's hand toprevent contamination betweenpatient and examiner.Gloves were tested for use withChemotherapy drugs, Fentanyl andGastric acid in accordance with ASTMD6978-05 standards Practice forassessment of Medical Glove toPermeation by chemotherapy drugsThe glove is a disposable deviceintended for medical purposes that isworn on the examiner's hand toprevent contamination betweenpatient and examiner.Gloves have been tested for use withchemotherapy drugs, Fentanyl Citrate,and select other drugs using ASTMD6978-05(2019)Same
Powder freeYesYesSame
MaterialNitrileNitrileSame
ColorLight Blue, Dark blueBlueDifferent

{10}------------------------------------------------

5300 Concours Ontario, CA 91764

SizeXS, S, M, L, XL, XXLXS, S, M, L, XL, XXLSame
SterileNon-SterileNon-SterileSame
Single UseSingle useSingle useSame
Design featureAmbidextrousAmbidextrousSame
Chemo DrugsClaimSee below comparison tableSee below comparison table

Analysis: * The finished subject device has been tested with performance and Biocompatibility, all the test results meet the requirements, so the difference of color does not raise questions of safety and effectiveness.

Technological Characteristic Comparison Table:
----------------------------------------------------------
CharacteristicsProposed devicePredicate deviceK240051Result
DimensionReference standardASTM D6319-19LengthMin 220mm for XS, SMin 230mm for M, L,XL, XXLMin 220mm for XS, SMin 230mm for M, L, XL, XXLSame
WidthXS: 70±10 mmS: 80±10 mmM: 95±10 mmL: 110±10 mmXL: 120±10 mmXXL: 130±10 mmXS: 70±10 mmS: 80±10 mmM: 95±10 mmL: 110±10 mmXL: 120±10 mmXXL: 130±10 mm
ThicknessPalm: Min 0.05 mmFinger: Min 0.05 mmPalm: Min 0.05 mmFinger: Min 0.05 mm
Physical propertiesASTM D6319-19ASTM D412-16Before aging
Tensile strengthMin 14MPaMin 14MPaSame
Ultimate elongationMin 500%Min 500%
After aging
Tensile strengthMin 14MPaMin 14MPa
Ultimate elongationMin 400%Min 400%
Freedom from holesASTM D6319-19; ASTM D5151-19G-I, AQL2.5G-I, AQL2.5Same
Residual PowderASTM D6319-19; ASTM D6124-06≤ 2 mg per glove≤ 2 mg per gloveSame
BiocompatibilityIrritationISO 10993-23Under the conditions ofthe study, not anirritantUnder the conditions of thestudy, not an irritantSame
SensitizationISO 10993-10Under the conditions ofthe study, not asensitizerUnder the conditions of thestudy, not a sensitizerSame

{11}------------------------------------------------

5300 Concours Ontario, CA 91764

Acute SystemicToxicity TestISO 10993 - 11Under the conditions of this study, there was no evidence of acute systemic toxicity.Under the conditions of this study, there was no evidence of acute systemic toxicity.Same
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Chemotherapy Drugs Comparison Claim:

Minimum Breakthrough Detection Time, minutes
NO.Chemotherapy DrugsProposed devicePredicate deviceResult ofcomparison
Light BlueDark BlueK240051
1Bendamustine HCl, 5 mg/ml>240>240Different
2Bortezomib, 1 mg/ml>240>240>240Same
3Busulfan, 6 mg/ml>240>240>240Same
4Carboplatin, 10 mg/ml>240>240>240Same
5Carmustine, 3.3 mg/ml16.621.529.1Different
6Carfilzomib, 2 mg/ml>240>240Different
7Cetuximab (Erbitux), 2 mg/ml>240>240Different
8Chloroquine, 50mg/ml>240>240>240Same
9Cisplatin, 1 mg/ml>240>240>240Same
10Cladribine, 1.0 mg/ml>240>240Different
11Cyclophosphamide, 20 mg/ml>240>240>240Same
12Cyclosporin A,100mg/ml>240>240>240Same
13Cytarabine HCl, 100 mg/ml>240>240>240Same
14Cytovene, 10 mg/ml>240>240Different
15Dacarbazine, 10 mg/ml>240>240>240Same
16Daunorubicin HCl, 5 mg/ml>240>240>240Same
17Decitabine, 5 mg/ml>240>240Different
18Docetaxel, 10 mg/ml>240>240>240Same
19Doxorubicin HCl, 2 mg/ml>240>240>240Same
20Epirubicin HCl, 2 mg/ml>240>240>240Same
21Etoposide, 20 mg/ml>240>240>240Same
22Fludarabine, 25 mg/ml>240>240>240Same
23Fluorouracil, 50 mg/ml>240>240>240Same
24Gemcitabine HCl, 38 mg/ml>240>240>240Same
25Idarubicin HCl, 1 mg/ml>240>240>240Same
26Ifosfamide, 50 mg/ml>240>240>240Same
27Irinotecan HCl, 20 mg/ml>240>240>240Same
28Mechlorethamine HCL, 1 mg/ml>240>240>240Same
29Melphalan HCl, 5 mg/ml>240>240>240Same
30Methotrexate, 25 mg/ml>240>240>240Same
31Mitomycin-C, 0.5 mg/ml>240>240>240Same
32Mitoxantrone HCl, 2 mg/ml>240>240>240Same
33Oxaliplatin, 2 mg/ml>240>240>240Same
34Paclitaxel, 6 mg/ml>240>240>240Same
35Retrovir. 10 mg/ml>240>240>240Same

{12}------------------------------------------------

5300 Concours Ontario, CA 91764

36ThioTEPA, 10 mg/ml44.917.978.3Different
37Trisenox, 1 mg/ml>240>240>240Same
38Vidaza (Azacitidine),25 mg/ml>240>240Different
39Vincristine Sulfate,1 mg/ml>240>240>240Same
NO.Fentanyl Citrate and Gastric AcidMinimum Breakthrough Detection Time, minutesResult ofcomparison
Proposed devicePredicate deviceK240051
1Fentanyl CitrateInjection,100mcg/2ml>240>240>240Same
  • Chemotherapy drugs and the minimum breakthrough time of subject device will be listed on labeling, so this difference does not raise questions of safety and effectiveness.

8. Summary of Non-Clinical Performance Data

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device met the performance criteria with the following standards:

Test performedMethodologyAcceptance CriteriaResult
Freedom From HolesASTM D6319-19ASTM D5151-19Meet requirement inspection level G- 1, AQL 2.5(ISO2859-1)Pass
Dimension-LengthASTM D6319-19Minimum 220mm for size XS-SMinimum 230mm for size M-XXLPass
Dimension-WidthASTM D6319-19XS: 70±10mmS: 80±10mmM: 95±10mmL: 110±10mmXL: 120±10mmXXL: 130±10mmPass
Dimension-ThicknessASTM D6319-19Finger: 0.05mm (min)Palm: 0.05mm (min)Pass
Physical propertiesASTM D6319-19ASTM D412-16Tensile Strength (Min 14 Mpa)Elongation (Before Aging 500% and after aging400%) MinPass
Powder ResidueASTM D6319-19ASTM D6124-06Not more than 2 mg per glovePass
Skin IrritationISO10993-23:2021Not a primary skin irritantPass
Skin SensitizationISO10993-10:2021Not a contact sensitizerPass

{13}------------------------------------------------

5300 Concours Ontario, CA 91764

Acute Systemic ToxicityISO 10993-11:2017No acute systemic toxicityPass
Chemotherapy Drug,Fentanyl, and Gastric AcidASTM D6978-05Minimum Breakthrough Detection Time (minutes)Remark
  • · ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
  • . ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
  • · ASTM D6124-06 (2017), Standard Test Method for Residual Powder on Medical Gloves
  • ASTM D412-16 (2021), Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers— Tension
  • · ASTM D6978-05(2023), Assessment of Reissuance of Medical Gloves to Permeation by Chemotherapy Drugs.
  • · ISO 10993-10: 2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization.
  • · ISO10993-23:2021 Biological evaluation of medical devices Part 23: Tests for irritation.
  • · ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity

9. Summary of Clinical Testing:

Not applicable.

10. Conclusion:

The conclusions drawn from the non-clinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, K240051.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.