K Number
K240051

Validate with FDA (Live)

Date Cleared
2024-02-29

(52 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs, Fentanyl Citrate, and select other drugs using ASTM D6978-05(2019)

Device Description

Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and select other drugs are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small, Small, Medium, Large, Extra Large and XXL. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs, Fentanyl Citrate, and select other drugs as per ASTM D6978-05(2019). The gloves are single use, disposable, and non-sterile.

AI/ML Overview

The provided document describes the acceptance criteria and performance of "Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and select other drugs" (K240051).

Here's the breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Test PerformedAcceptance CriteriaReported Device Performance
Physical Dimensions
LengthMinimum 220mm for size XS and S, 230mm for size M, L, XL, XXL (ASTM D6319-19)Pass
Palm Width (size)XS: 70±10 mm, S: 80±10 mm, M: 95±10 mm, L: 110±10 mm, XL: 120±10 mm, XXL: 130±10 mm (ASTM D6319-19)Pass
ThicknessFinger: 0.05mm (min), Palm: 0.05mm (min) (ASTM D6319-19)Pass
Physical Properties
Tensile Strength & Elongation (Before Aging)Tensile Strength: 14MPa, min; Elongation: 500%, min (ASTM D6319-19, ASTM D412-16(2021))Pass
Tensile Strength & Elongation (After Accelerated Aging)Tensile Strength: 14MPa, min; Elongation: 400%, min (ASTM D6319-19, ASTM D412-16(2021))Pass
Barrier Integrity
Watertight (1000ml)G-I, AQL 2.5 (21 CFR 800.20, ASTM D5151-19)Pass
Powder ResidueMax 2mg/glove (ASTM D6319-19, ASTM D6124-06(2017))Pass
Chemotherapy Drugs & Opioid Drug PermeationMinimum Breakthrough Detection Time (BDT) for various chemotherapy drugs and Fentanyl Citrate. Accepted values differ by drug, but generally, longer BDT indicates better performance. Specific acceptance criteria are not explicitly listed as a single value but are implied by the reported BDTs (e.g., >240 minutes for many drugs, with lower specific values for Carmustine and Thiotepa which are cautioned against).See specific BDT values in the tables below for each drug. Notably, Carmustine: 29.1 minutes, Thiotepa: 78.3 minutes. All other listed drugs exhibit >240 minutes BDT. Fentanyl Citrate: >240 minutes.
Biocompatibility
Irritation and Skin SensitizationNon-sensitization and Non-irritation (ISO 10993-10 & -23)Non-sensitization and Non-irritation
CytotoxicityCytotoxicity reactivity (ISO 10993-5:2009)Showed potential toxicity to L929 cells, but concern addressed by acute systemic toxicity testing.
Acute systemic toxicity studyNo adverse biological reaction (ISO 10993-11:2017)No evidence of acute systemic toxicity.

Chemotherapy, Fentanyl Citrate & other drugs Permeation Comparison (Proposed Device K240051):

Tested Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes
Bleomycin Sulfate 15mg/ml (15000 ppm)>240
Busulfan 6mg/ml (6,000 ppm)>240
Carboplatin 10mg/ml (10,000 ppm)>240
Carmustine 3.3 mg/ml (3,300 ppm)29.1
Cisplatin 1mg/ml (1,000 ppm)>240
Cyclophosphamide 20mg/ml (20,000 ppm)>240
Cytarabine, 100 mg/ml (100,000 ppm)>240
Dacarbazine 10 mg/ml (10,000 ppm)>240
Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240
Docetaxel , 10 mg/ml (10,000 ppm)>240
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
Epirubicin HCL, 2 mg/ml (2,000 ppm)>240
Etoposide, 20 mg/ml (20,000 ppm)>240
Fludarabine Phosphate, 25 mg/ml (25,000 ppm)>240
Fluorouracil, 50mg/ml (50,000ppm)>240
Gemcitabine HCL, 38mg/ml (38,000ppm)>240
Idarubicin HCL, 1mg/ml (1,000ppm)>240
Ifosfamide, 50mg/ml (50,000ppm)>240
Irinotecan HCL, 20mg/ml (20,000ppm)>240
Mechlorethamine HCI, 1mg/ml (1,000ppm)>240
Melphalan HCL, 5mg/ml (5,000ppm)>240
Methotrexate, 25mg/ml (25,000ppm)>240
Mitomycin C, 0.5mg/ml (500ppm)>240
Mitoxantrone HCL, 2mg/ml (2,000ppm)>240
Oxaliplatin, 5mg/ml (5,000ppm)>240
Paclitaxel, 6mg/ml (6,000ppm)>240
Paraplatin, 10mg/ml (10,000ppm)>240
Rituximab, 10mg/ml (10,000ppm)>240
Thiotepa, 10mg/ml (10,000ppm)78.3
Topotecan HCL, 1mg/ml (1,000ppm)>240
Trisenox (Arsenic Trioxide), 1mg/ml (1,000ppm)>240
Velcade (Bortezomib), 1mg/ml (1,000ppm)>240
Vincristine Sulfate, 1mg/ml (1,000ppm)>240
Fentanyl Citrate Injection (100 mcg/2ml)>240
Chloroquine 50mg/ml (50,000ppm)>240
Cyclosporin A 100 mg/ml (100,000 ppm)>240
Retrovir, 10mg/ml (10,000ppm)>240

Warning: Do not use with Carmustine and Thiotepa due to "extremely low permeation times" (29.1 minutes and 78.3 minutes, respectively).

2. Sample size used for the test set and the data provenance:

  • The document primarily refers to standard ASTM and ISO test methodologies for device performance (e.g., ASTM D6319-19, ASTM D6978-05(2019), ISO 10993 series). These standards typically define the sample sizes required for each specific test to ensure statistical validity.
  • Data Provenance: The document states "Non-clinical tests were conducted to verify that the proposed device met all design specifications." The tests were performed by Lingshi Hongruida Health Protection Technology Co., Ltd. located in China. The data is retrospective in the sense that the testing was performed and then submitted with the 510(k) application. Specific details on exact sample sizes (e.g., number of gloves tested for each characteristic) within each standard are not explicitly detailed in this summary for every test. However, the chemotherapy drug permeation testing likely involved multiple samples per drug as per ASTM D6978.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not applicable as the device is a physical medical device (examination glove) and the testing involved objective, quantitative, and standardized measurements according to established consensus standards (ASTM, ISO). There were no "experts" establishing a "ground truth" in the way they would for, say, image interpretation or clinical diagnosis. The ground truth for performance relied on the specified chemical and physical property measurements.

4. Adjudication method for the test set:

  • This is not applicable. As mentioned above, the testing involved objective physical and chemical measurements governed by established standards, not subjective interpretations requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • This is not applicable. This device is an examination glove, not an AI-powered diagnostic tool or imaging software. Therefore, no MRMC study or AI-related comparative effectiveness study was performed or is relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • This is not applicable. This device is an examination glove and does not involve any algorithm or AI for standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The ground truth for this device's performance is based on objective measurements against established performance standards and chemical permeation resistance thresholds.
    • For physical properties (length, thickness, tensile strength, elongation, watertightness, powder content): The ground truth is the quantitative measurement compared against the specified ASTM standard values.
    • For chemical permeation (chemotherapy drugs, Fentanyl Citrate): The ground truth is the experimentally determined Minimum Breakthrough Detection Time (BDT) in minutes, measured according to ASTM D6978-05(2019).
    • For biocompatibility (irritation, sensitization, cytotoxicity, acute systemic toxicity): The ground truth is the experimental findings and conclusions derived from tests conducted under ISO 10993 series standards.

8. The sample size for the training set:

  • This is not applicable. This device is a physical product, not a machine learning model, so there is no training set in the context of AI/ML.

9. How the ground truth for the training set was established:

  • This is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".

February 29, 2024

Lingshi Hongruida Health Protection Technology Co., Ltd. Tinglong Chai General Manager Yangjiayuan, Liangdu Town, Lingshi County Jinzhong, Shanxi 031300 China

Re: K240051

Trade/Device Name: Powder Free Nitrile Examination Gloves (Blue), Tested For Use With Chemotherapy Drugs, Fentanyl Citrate, and select other drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, ODQ, OPJ Dated: January 6, 2024 Received: January 8, 2024

Dear Tinglong Chai:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Allan Guan -S

For Bifeng Qian, M.D., Ph.D. Assistant Director

{2}------------------------------------------------

DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240051

Device Name

Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and select other drugs

Indications for Use (Describe)

The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with chemotherapy drugs, Fentanyl Citrate, and select other drugs using ASTM D6978-05(2019)

The following chemicals have been tested with these gloves:

Chemotherapy DrugMinimum Breakthrough Detection Time in Minutes
Bleomycin Sulfate 15mg/ml (15000 ppm)>240
Busulfan 6mg/ml (6,000 ppm)>240
Carboplatin 10mg/ml (10,000 ppm)>240
Carmustine 3.3 mg/ml (3,300 ppm)29.1
Cisplatin 1mg/ml (1,000 ppm)>240
Cyclophosphamide 20mg/ml (20,000 ppm)>240
Cytarabine, 100 mg/ml (100,000 ppm)>240
Dacarbazine 10 mg/ml (10,000 ppm)>240
Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240
Docetaxel , 10 mg/ml (10,000 ppm)>240
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
Epirubicin HCL, 2 mg/ml (2,000 ppm)>240
Etoposide, 20 mg/ml (20,000 ppm)>240
Fludarabine Phosphate, 25 mg/ml (25,000 ppm)>240
Fluorouracil, 50mg/ml (50,000ppm)>240
Gemcitabine HCL, 38mg/ml (38,000ppm)>240
Idarubicin HCL, 1mg/ml (1,000ppm)>240
Ifosfamide, 50mg/ml (50,000ppm)>240
Irinotecan HCL, 20mg/ml (20,000ppm)>240
Mechlorethamine HCI, 1mg/ml (1,000ppm)>240
Melphalan HCL, 5mg/ml (5,000ppm)>240
Methotrexate, 25mg/ml (25,000ppm)>240
Mitomycin C, 0.5mg/ml (500ppm)>240
Mitoxantrone HCL, 2mg/ml (2,000ppm)>240
Oxaliplatin, 5mg/ml (5,000ppm)>240
Paclitaxel, 6mg/ml (6,000ppm)>240
Paraplatin, 10mg/ml (10,000ppm)>240
Rituximab, 10mg/ml (10,000ppm)>240
Thiotepa, 10mg/ml (10,000ppm)78.3
Topotecan HCL, 1mg/ml (1,000ppm)>240
Trisenox (Arsenic Trioxide), 1mg/ml (1,000ppm)>240
Velcade (Bortezomib), 1mg/ml (1,000ppm)>240
Vincristine Sulfate, 1mg/ml (1,000ppm)>240

Fentanyl Citrate & other drugs Fentanyl Citrate Injection, 100mcg/2mg

FORM FDA 3881 (6/20)

Minimum Breakthrough Detection Time in Minutes

240

{4}------------------------------------------------

Chloroquine 50mg/ml (50,000ppm)>240
Cyclosporin A 100 mg/ml (100,000 ppm)>240
Retrovir, 10mg/ml (10,000ppm)>240

*Please note that the following drugs have extremely low permeation times: Carmustine: 29.1 minutes, Thiotepa: 78.3 minutes Warning: Do not use with Carmustine and Thiotepa.

Type of Use (Select one or both, as applicable)

__ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

The assigned 510(K) numbers: K240051

Date Prepared: February 29, 2024

1. Owner's Identification:

Lingshi Hongruida Health Protection Technology Co., Ltd. Yangjiayuan. Liangdu Town, Lingshi County, Jinzhong City, Shanxi Province, 031300. China Tel:86-311-66179668 Contact: Mr. Chai Tinglong, General Manager Email: janicema(@hongrayusa.com or fdareg(@hongray.com.cn

2. Name of the Device:

Trade / Product Name: Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and select other drugs Common Name: Exam Gloves Classification Name: Patient Examination Glove Specialty Classification Regulation: 21 CFR 880.6250 Product Code: LZA, LZC, QDO, OPJ Classification Panel: General Hospital Device Class: Class I

3. Predicate Device Information:

Lingshi Hongruida Health Protection Technology Co., Ltd. Powder Free Nitrile Examination Gloves (Blue),Tested for Use with Chemotherapy Drugs and Fentany) Citrate (K223320)

4. Device Description:

Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and select other drugs are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small, Small, Medium, Large, Extra Large and XXL. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs, Fentanyl Citrate, and select other drugs as per ASTM D6978-05(2019). The gloves are single use, disposable, and non-sterile.

5. Device Modification

The proposed modification to the predicate device is adding additional Chemotherapy drugs. We have tested the proposed device (exactly same as predicate device) for use with chemotherapy drugs as per ASTM D6978 and gotten the testing report to support this claim. Please note that there are no differences with these two gloves from materials, manufacturing process and bench performance.

6. Indications for Use:

The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with chemotherapy drugs, Fentanyl Citrate, and select other drugs using ASTM D6978-05(2019)

The following chemicals have been tested with these gloves:

Chemotherapy DrugMinimum Breakthrough Detection Time in Minutes
Bleomycin Sulfate 15mg/ml (15000 ppm)>240
Busulfan 6mg/ml (6,000 ppm)>240
Carboplatin 10mg/ml (10,000 ppm)>240
Carmustine 3.3 mg/ml (3,300 ppm)29.1
Cisplatin 1mg/ml (1,000 ppm)>240
Cyclophosphamide 20mg/ml (20,000 ppm)>240
Cytarabine, 100 mg/ml (100,000 ppm)>240
Dacarbazine 10 mg/ml (10,000 ppm)>240
Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240
Docetaxel , 10 mg/ml (10,000 ppm)>240
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
Epirubicin HCL, 2 mg/ml (2,000 ppm)>240
Etoposide, 20 mg/ml (20,000 ppm)>240
Fludarabine Phosphate, 25 mg/ml (25,000 ppm)>240
Fluorouracil, 50mg/ml (50,000ppm)>240
Gemcitabine HCL, 38mg/ml (38,000ppm)>240
Idarubicin HCL, 1mg/ml (1,000ppm)>240
Ifosfamide, 50mg/ml (50,000ppm)>240
Irinotecan HCL, 20mg/ml (20,000ppm)>240
Mechlorethamine HCI, 1mg/ml (1,000ppm)>240
Melphalan HCL, 5mg/ml (5,000ppm)>240
Methotrexate, 25mg/ml (25,000ppm)>240
Mitomycin C, 0.5mg/ml (500ppm)>240
Mitoxantrone HCL, 2mg/ml (2,000ppm)>240
Oxaliplatin, 5mg/ml (5,000ppm)>240
Paclitaxel, 6mg/ml (6,000ppm)>240
Paraplatin, 10mg/ml (10,000ppm)>240
Rituximab, 10mg/ml (10,000ppm)>240
Thiotepa, 10mg/ml (10,000ppm)78.3
Topotecan HCL, 1mg/ml (1,000ppm)>240
Trisenox (Arsenic Trioxide), 1mg/ml (1,000ppm)>240
Velcade (Bortezomib), 1mg/ml (1,000ppm)>240
Vincristine Sulfate, 1mg/ml (1,000ppm)>240
Fentanyl Citrate & other drugsMinimum Breakthrough Detection Time in Minutes
Fentanyl Citrate Injection (100 mcg/2ml)>240
Chloroquine 50mg/ml (50,000ppm)>240
Cyclosporin A 100 mg/ml (100,000 ppm)>240
Retrovir, 10mg/ml (10,000ppm)>240

{6}------------------------------------------------

*Please note that the following drugs have extremely low permeation times: Carmustine: 29.1 minutes, Thiotepa: 78.3 minutes Warning: Do not use with Carmustine and Thiotepa.

7. Comparison of Subject Device and Predicate Device:

{7}------------------------------------------------

General Comparison Table:

Proposed DevicePredicate DeviceResult of
comparison
K240051K223320
Trade NamePowder Free Nitrile ExaminationGloves (Blue), Tested for Use withChemotherapy Drugs. FentanylCitrate, and select other drugsPowder Free Nitrile ExaminationGloves (Blue), Tested for Use withChemotherapy Drugs and FentanylCitrateSame
Product CodeLZA, LZC, QDO, OPJLZA, LZC, QDOSame
Regulation Number21 CFR 880.625021 CFR 880.6250Same
ClassIISame
Indications for UseThe glove is a disposable deviceintended for medical purposes that isworn on the examiner's hand toprevent contamination betweenpatient and examiner. Gloves havebeen tested for use with chemotherapydrugs, Fentanyl Citrate, and selectother drugs using ASTM D6978-05(2019)The glove is a disposable deviceintended for medical purposes that isworn on the examiner's hand toprevent contamination betweenpatient and examiner. Gloves havebeen tested for use with chemotherapydrugs and Fentanyl Citrate usingASTM D6978-05(2019)Same
MaterialNitrileNitrileSame
Powdered or PowderFreePowder FreePowder FreeSame
ColorBlueBlueSame
Single useSingle useSingle useSame
Chemotherapy Drugsand Fentanyl Citrate andother drugs ClaimSee below comparison tableSee below comparison tableSee belowcomparisontable

Technological Characteristic Comparison Table:

TechnologicalCharacteristicsProposed DeviceK240051Predicate DeviceK223320Result ofcomparison
LengthMinimum 220mm for size XS and S, 230mm for size M, L, XL, XXLMinimum 220mm for size XS and S, 230mm for size M, L, XL, XXLSame
Palm Width (size) (mm)
XS70±1070±10Same
S80±1080±10Same
M95±1095±10Same
L110±10110±10Same
XL120±10120±10Same
XXL130±10130±10Same
Thickness(mm)
FingerMinimum 0.05Minimum 0.05Same

{8}------------------------------------------------

PalmMinimum 0.05Minimum 0.05Same
Tensile Strength, BeforeAging14MPa, min14MPa, minSame
Ultimate Elongation,Before Aging500%, min500%, min
Tensile Strength, AfterAccelerated Aging14MPa, min14MPa, minSame
Ultimate Elongation,After Accelerated Aging400%, min400%, minSame
Watertight (1000ml)21 CFR 800.20ASTM D5151G-1, AQL 2.521 CFR 800.20ASTM D5151G-1, AQL 2.5
Powder-Content≤ 2 mg per glove≤ 2 mg per gloveSame
10993-23:2021Skin Irritation StudyUnder the conditions of the study, not an irritantUnder the conditions of the study, not an irritantSame
10993-10:2021Sensitization StudyUnder the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizerSame
10993-5:2009Cytotoxicity TestUnder the conditions of this study, the test article extract showed potential toxicity to L929 cells.Cytotoxicity concern was addressed by acute systemic toxicity testing.Under the conditions of this study, the test article extract showed potential toxicity to L929 cells.Cytotoxicity concern was addressed by acute systemic toxicity testing.Same
ISO 10993-11:2017Acute Systemic toxicity studyUnder the conditions of this study, there was no evidence of acute systemic toxicity.Under the conditions of this study, there was no evidence of acute systemic toxicity.Same

Chemotherapy, Fentanyl Citrate & other drugs Permeation Comparison:

Tested Chemotherapy Drug andConcentrationMinimum BDT (Minutes)Result ofcomparison
Proposed DeviceK240051Predicate DeviceK223320
Bleomycin Sulfate 15mg/ml (15000 ppm)>240/Different*
Busulfan 6mg/ml (6,000 ppm)>240/Different*
Carboplatin 10mg/ml (10,000 ppm)>240/Different*
Carmustine 3.3 mg/ml (3,300 ppm)29.113.5Similar
Cisplatin 1mg/ml (1,000 ppm)>240>240Same
Cyclophosphamide 20mg/ml (20,000 ppm)>240>240Same
Cytarabine, 100 mg/ml (100,000 ppm)>240/Different*
Dacarbazine 10 mg/ml (10,000 ppm)>240>240Same
Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240/Different*
Docetaxel , 10 mg/ml (10,000 ppm)>240/Different*
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240>240Same
Epirubicin HCL, 2 mg/ml (2,000 ppm)>240/Different*
Etoposide, 20 mg/ml (20,000 ppm)>240>240Same
Fludarabine Phosphate, 25 mg/ml (25,000 ppm)>240/Different*

{9}------------------------------------------------

Fluorouracil, 50mg/ml (50,000ppm)>240>240Same
Gemcitabine HCL, 38mg/ml (38,000ppm)>240/Different*
Idarubicin HCL, 1mg/ml (1,000ppm)>240/Different*
Ifosfamide, 50mg/ml (50,000ppm)>240/Different*
Irinotecan HCL, 20mg/ml (20,000ppm)>240/Different*
Mechlorethamine HCI, 1mg/ml (1,000ppm)>240/Different*
Melphalan HCL, 5mg/ml (5,000ppm)>240/Different*
Methotrexate, 25mg/ml (25,000ppm)>240>240Same
Mitomycin C, 0.5mg/ml (500ppm)>240/Different*
Mitoxantrone HCL, 2mg/ml (2,000ppm)>240/Different*
Oxaliplatin, 5mg/ml (5,000ppm)>240/Different*
Paclitaxel, 6mg/ml (6,000ppm)>240>240Same
Paraplatin, 10mg/ml (10,000ppm)>240/Different*
Rituximab, 10mg/ml (10,000ppm)>240/Different*
Thiotepa, 10mg/ml (10,000ppm)78.366.6Similar
Topotecan HCL, 1mg/ml (1,000ppm)>240/Different*
Trisenox (Arsenic Trioxide), 1mg/ml (1,000ppm)>240/Different*
Velcade (Bortezomib), 1mg/ml (1,000ppm)>240/Different*
Vincristine Sulfate, 1mg/ml (1,000ppm)>240/Different*
Fentanyl Citrate & other drugsMinimum BDT (Minutes)Result of
Subject DevicePredicate Devicecomparison
K240051K223320
Fentanyl Citrate Injection (100 mcg/2ml)>240>240Same
Chloroquine 50mg/ml (50,000ppm)>240/Different*
Cyclosporin A 100 mg/ml (100,000 ppm)>240/Different*
Retrovir, 10mg/ml (10,000ppm)>240/Different*

*This different is support with test report and will be indicated on labeling, so this different does not raise questions of safety and effectiveness of subject device.

8. Summary of Non-Clinical Performance Data

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device met the performance criteria with the following standards:

MethodologyTest PerformedAcceptance CriteriaResults
ASTM D6319- 19Physical DimensionsLengthMinimum 220mm for size XS and S,230mm for size M, L, XL, XXLPass
ASTM D6319- 19
S: $80\pm10$ mm
M: $95\pm10$ mm
L: $110\pm10$ mm
XL: $120\pm10$ mm
XXL: $130\pm10$ mm
ASTM D6319- 19Physical DimensionsThicknessFinger: 0.05mm (min)Palm: 0.05mm (min)Pass

{10}------------------------------------------------

ASTM D6319- 19ASTM D412-16(2021)Physical PropertiesTensile Strength (Min14 MPa) andElongation (Before Aging 500% andafter aging 400%) MinPass
ASTM D6319- 19ASTM D5151-19Water leak testG-I, AQL 2.5Pass
ASTM D6319- 19ASTM D6124-06(2017)Powder ResidueMax 2mg/glovePass
ASTM D6978-05(2019)Chemotherapy Drugs,Opioid Drug PermeationRefer the above tablesPass
ISO 10993-10&-23Irritation and SkinSensitizationSkin sensitization and Skin irritationnon-sensitizationand Non-irritation
ISO 10993-5:2009CytotoxicityCytotoxicity reactivityshowed potentialtoxicity to L929cells.
ISO 10993-11:2017Acute systemic toxicitystudySubject showed no adversebiological reactionno evidence ofacute systemictoxicity.

9. Summary of Clinical Information

No clinical testing was submitted for the subject device.

10. Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.