K Number
K243290
Device Name
BTL-785MJ
Date Cleared
2025-05-09

(203 days)

Product Code
Regulation Number
890.5850
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BTL-785MJ device with the controller and single-use applicators has the following indications for use:

  1. To relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain
  2. Muscle re-education
  3. Increasing blood flow
  4. Maintain or increase mandibular range of motion
Device Description

The BTL-785MJ is a state-of-the-art electrostimulation and radiofrequency platform with accessories that enables the application of therapy by electrical stimulation.
The control unit of the system is equipped with a large color touch screen that facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are set using the touch screen of the device. During the therapy the device displays information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.
The generated electromagnetic field is intended to interact with the tissues of the human body in order to relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain, muscle re-education, increase blood flow and maintain or increase mandibular range of motion.
The BTL-785MJ device consists of the following main components:

  • Main unit
  • Handpiece
  • Holding arm
  • Connection cables
  • Single-use applicators
  • Therapy discomfort button
AI/ML Overview

The provided FDA clearance letter (K243290) for the BTL-785MJ device does not contain explicit acceptance criteria or a detailed clinical study demonstrating the device meets such criteria. Instead, it relies on substantial equivalence to a predicate device.

Here's an analysis of the information available and what is missing relative to your requested points:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state acceptance criteria in terms of specific performance metrics or thresholds for the BTL-785MJ device's effectiveness in treating the listed indications. The "device performance" reported is primarily compliance with various electrical safety, EMC, usability, software, risk management, and biological evaluation standards.

Acceptance Criteria (Not explicitly stated for clinical performance)Reported Device Performance (Non-clinical)
(Not stated for clinical efficacy)Complies with IEC 60601-1 (General safety & essential performance)
(Not stated for clinical efficacy)Complies with IEC 60601-1-2 (EMC)
(Not stated for clinical efficacy)Complies with IEC 60601-2-10 (Nerve & muscle stimulators)
(Not stated for clinical efficacy)Complies with IEC 60601-1-6 (Usability)
(Not stated for clinical efficacy)Complies with IEC 62304 (Software life cycle processes)
(Not stated for clinical efficacy)Complies with ISO 14971 (Risk management)
(Not stated for clinical efficacy)Complies with ISO 10993-1, -5, -10 (Biological evaluation: general, cytotoxicity, irritation/sensitization)

2. Sample size used for the test set and the data provenance

  • Sample size: Not applicable. The document explicitly states: "A clinical study was not conducted." Therefore, there is no test set of patients for clinical efficacy.
  • Data provenance: Not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable, as no clinical study was conducted.

4. Adjudication method for the test set

Not applicable, as no clinical study was conducted.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The device is a physical muscle stimulator, not an AI-powered diagnostic or interpretive tool that would involve "human readers" or AI assistance in that context. Therefore, an MRMC comparative effectiveness study is not relevant to this device, and none was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical electrostimulation device, not an algorithm.

7. The type of ground truth used

Not applicable for clinical efficacy. The substantial equivalence relies on comparing the BTL-785MJ's technical characteristics and indications for use to a legally marketed predicate device (D function, K203525) that presumably established its effectiveness. The ground truth for safety and performance is based on compliance with international standards (IEC, ISO) through non-clinical testing.

8. The sample size for the training set

Not applicable, as this is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established

Not applicable.


Summary of Device Acceptance and Study:

The BTL-785MJ device received FDA clearance (K243290) based on substantial equivalence to an existing predicate device, the "D function" (K203525). This means its safety and effectiveness were demonstrated by showing that it has similar technological characteristics and the same intended use as the predicate device.

Specifically, the "study" demonstrating this involves:

  • Non-clinical testing: Compliance with various electrical safety, electromagnetic compatibility (EMC), usability, software lifecycle processes, risk management, and biological evaluation standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-6, IEC 62304, ISO 14971, ISO 10993 series). These tests verify the physical and operational safety of the device.
  • Comparison to a predicate device: A detailed table in the 510(k) summary (page 8-10) outlines the similarities in indications for use, principle of action, clinical use, energy source, electrical safety, energy type, accessory attachment, therapy time, pulse width, output frequency, output current, max output voltage, max current density, pulse shape, channel numbers, electrical safety class, features (timer, buzzer, display), and electrode properties. The applicant argues that "Any differences between the predicate device and BTL-785MJ device have no significant influence on safety or effectiveness of the BTL-785MJ device."

Crucially, no clinical study was conducted to directly prove the clinical efficacy of the BTL-785MJ device itself. Its acceptance is predicated on its similarity to a device that was previously cleared and is presumed safe and effective for its stated indications.

FDA Clearance Letter for BTL-785MJ

Page 1

U.S. Food & Drug Administration

10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

May 9, 2025

BTL Industries, Inc.
David Chmel
CEO North America
362 Elm Street
Marlborough, Massachusetts 01752

Re: K243290
Trade/Device Name: BTL-785MJ
Regulation Number: 21 CFR 890.5850
Regulation Name: Powered Muscle Stimulator
Regulatory Class: Class II
Product Code: NUW, NFO
Dated: October 18, 2024
Received: April 10, 2025

Dear David Chmel:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243290 - David Chmel Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243290 - David Chmel Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K243290

Device Name
BTL-785MJ

Indications for Use (Describe)

The BTL-785MJ device with the controller and single-use applicators has the following indications for use:

  1. To relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain
  2. Muscle re-education
  3. Increasing blood flow
  4. Maintain or increase mandibular range of motion

Type of Use (Select one or both, as applicable)

  • Prescription Use (Part 21 CFR 801 Subpart D)
  • Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

Page 1 of 6

510(k) Summary – K243290

General Information

Sponsor: BTL Industries, Inc.
362 Elm Street
Marlborough, MA 01752
Tel: +1-866-285-1656
Fax: +1-888-499-2502

Applicant: BTL Industries, Inc.
362 Elm Street
Marlborough, MA 01752
Tel: +1-866-285-1656
Fax: +1-888-499-2502

Contact Person: David Chmel
BTL Industries, Inc.
chmel@btlnet.com

Summary Preparation Date: May 08, 2025

Device

Trade/Proprietary Name: BTL-785MJ
Common Name: Stimulator, muscle, powered, dental
Primary Classification Name: Powered muscle stimulator
Classification Regulation: 21 CFR 890.5850, Class II
Primary Product Code: NUW
Secondary Product Code: NFO

Page 6

510(k) Summary

Page 2 of 6

Legally Marketed Predicate Device

The BTL-785MJ is substantially equivalent to the following product(s) that are already cleared for distribution in the USA under the following 510(k) Premarket Notification numbers:

Primary predicate:
Device name: D function
Original 510(k) Sponsor: ITO CO., LTD.
510(k) Number: K203525

Reference device:
Device name: BTL-785BNF-E
Original 510(k) Sponsor: BTL Industries, Inc.
510(k) Number: K242532

Product Description

The BTL-785MJ is a state-of-the-art electrostimulation and radiofrequency platform with accessories that enables the application of therapy by electrical stimulation.

The control unit of the system is equipped with a large color touch screen that facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are set using the touch screen of the device. During the therapy the device displays information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.

The generated electromagnetic field is intended to interact with the tissues of the human body in order to relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain, muscle re-education, increase blood flow and maintain or increase mandibular range of motion.

The BTL-785MJ device consists of the following main components:

  • Main unit
  • Handpiece
  • Holding arm
  • Connection cables
  • Single-use applicators
  • Therapy discomfort button

Page 7

510(k) Summary

Page 3 of 6

Indications for Use

The BTL-785MJ device with the controller and single-use applicators has the following indications for use:

  1. To relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain
  2. Muscle re-education
  3. Increasing blood flow
  4. Maintain or increase mandibular range of motion

Non-clinical Testing (Performance Data)

The BTL-785MJ device has been thoroughly evaluated for electrical safety. The device has been found to comply with applicable medical device safety standards:

IEC 60601-1 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances – Requirements and tests

IEC 60601-2-10 Medical electrical equipment – Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators

IEC 60601-1-6 Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability

IEC 62304 Medical device software – Software life cycle processes

ISO 14971 Medical devices – Application of risk management to medical devices

ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process

ISO 10993-5 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity

ISO 10993-10 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization

Clinical performance data

A clinical study was not conducted.

Page 8

510(k) Summary

Page 4 of 6

Comparison with the Predicate Device

510(k) numberK243290K203525K242532
Device nameBTL-785MJD functionBTL-785BNF-E
Company name:BTL Industries, Inc.ITO CO., LTD.BTL Industries, Inc.
TypeSubject devicePredicate deviceReference device
Product Code and RegulationPhysical medicineDental21 CFR 890.5850NUW – Stimulator, Muscle, Powered, DentalPhysical medicineDental21 CFR 890.5850NUW – Stimulator, Muscle, Powered, DentalNeurologyNeurology21 CFR 882.5890NFO – Stimulator, Transcutaneous Electrical, Aesthetic Purposes
Indications for UseThe BTL-785MJ device with the controller and single-use applicators has the following indications for use:1. To relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain2. Muscle re-education3. Increasing blood flow4. Maintain or increase mandibular range of motion1) To relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain2) Muscle re-education3) Increasing blood flow4) Maintain or increase mandibular range of motionThe BTL-785BNF-E device has the following indications for use:The device is indicated for aesthetic use including facial and neck or body skin stimulation.
Principle of ActionElectrostimulationElectrostimulationElectrostimulation
Clinical UsePrescription usePrescription usePrescription use

Page 9

510(k) Summary

Page 5 of 6

Energy Source100 – 120 V AC, 50/60 Hz200 – 240 V AC, 50/60 HzAC 100-240V, 50/60HzDC 12 V (AC adaptor)DC 7.4 (Lithium ion battery)100 – 120 V AC, 50/60 Hz200 – 240 V AC, 50/60 Hz
Electrical Safety & EMC TestingIEC 60601-1IEC 60601-1-2IEC 60601-2-10ANSI AAMI 60601-1IEC 60601-1-2IEC 60601-2-10IEC 60601-1IEC 60601-1-2IEC 60601-2-10
Energy TypeElectric stimulationElectric stimulationElectric stimulation
Accessory Attachment MethodPad – single-use applicatorPad – single-use applicatorPad – single-use applicator
Therapy time20-60 minutes60 minutes (Up to 480 min)20 minutes
Pulse WidthPreset 1: Up to 150 μsPreset 2: 488 μsPAIN: Up to 150μsCARE: 488 μsMCR: 150ms160 µs
Output Frequency (Hz)Preset 1: Up to 200Preset 2: 0.67PAIN: Up to 200CARE: 0.67MCR: 400250
Output CurrentPreset 1 : Up to 80 mAPreset 2: Up to 80 mAPAIN: Up to 80 mACARE: 24 mAMCR: 750 μAUp to 130 mA
Max. Output Voltage [V]BTL-785-7-2, -848,5V @ 500 Ω48,5V @ 2000 Ω0 @ 10 000 ΩNot publicly availableBTL-785-7-2, -848,5V @ 500 Ω48,5V @ 2000 Ω0 @ 10 000 Ω
Max. Current density at 500 Ω [µA/mm²]BTL-785-7-2 – 200 µA/mm²BTL-785-7-8 -276 µA/mm²Not publicly availableBTL-785-7-2 – 243 µA/mm²BTL-785-7-8 - 335 µA/mm²
Max average power density at 500 Ω [W/cm²]BTL-785-7-2 – 0,058 W/cm²BTL-785-7-8 -0.082 W/cm²Not publicly availableBTL-785-7-2 – 0,047 W/cm²BTL-785-7-8 - 0,065 W/cm²
Pulse ShapeRectangularRectangularRectangular
Channel Numbers22Up to 3

Page 10

510(k) Summary

Page 6 of 6

Electrical Safety ClassClass II, BFClass II, BFClass II, BF
TimerYesYesYes
BuzzerYesYesYes
DisplayYesYesYes
(a) Electrodes materials(b) Electroconductive media(c) electrode-to-skin impedance range(d) conductive surface areaa) Conductive ink (Silver)b) Conductive hydrogelc) 10-5000 Ohmd) BTL-785-7-2 - 4 cm²BTL-785-7-8 - 2.9 cm²a) Not publicly availableb) Not publicly availablec) Not publicly availabled) Not publicly availablea) Conductive ink (Silver)b) Conductive hydrogelc) 10-5000 Ohmd) BTL-785-7-2 - 4 cm²BTL-785-7-8 - 2.9 cm²
System Weight65 kg(143 lb)230 g(0.5 lb)65 kg(143 lb)
System Dimensions (W×H×D)53.94" x 26.38" x 26.38"3.3" x 0.93" x 5.9"53.94" x 26.38" x 26.38"

Substantial Equivalence

The BTL-785MJ device has similar technological characteristics and the same intended use compared to the primary predicate device and comparable mode of action and intended use compared to reference device. Any differences between the predicate device and BTL-785MJ device have no significant influence on safety or effectiveness of the BTL-785MJ device.

Therefore, the BTL-785MJ device is substantially equivalent to the predicate device.

Conclusion

Based upon the intended use, comparison of technical characteristics and performance testing provided in this premarket notification, the BTL-785MJ device has been shown to be substantially equivalent to the currently cleared predicate device for requested intended use.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).