K Number
K243149
Date Cleared
2024-12-20

(81 days)

Product Code
Regulation Number
880.5860
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Microliter Dosing Syringe is intended to inject fluid into, or withdraw fluid from the body. The Microliter Dosing Syringe is indicated for intravitreal use.

Device Description

The Microliter Dosing Syringe is a single-use, piston syringe-based device consisting of a syringe barrel, plunger stopper, syringe tip cap and incorporates a plunger rod subassembly. It is intended for use by healthcare professionals for general purpose fluid aspiration/injection. Its operation is manual. The Microliter Dosing Syringe is single use only, non-toxic, non-pyrogenic and sterilized by e-beam irradiation. The Microliter Dosing Syringe is suitable for ophthalmic use.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device called the "Microliter Dosing Syringe." This document focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving the device meets a set of performance criteria for an AI/ML algorithm.

Therefore, the document does not contain the information requested in the prompt regarding:

  • A table of acceptance criteria and reported device performance for an AI/ML algorithm.
  • Sample sizes, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, or ground truth for an AI/ML test or training set.

The document discusses the substantial equivalence of the Microliter Dosing Syringe to a predicate device (StaClear Syringe) based on:

  • Intended Use and Indications for Use: Both are intended to inject or withdraw fluid from the body, and are specifically indicated for intravitreal use.
  • Technological Characteristics: Comparisons are made across various characteristics like sterilization, shelf life, device materials, syringe volume, target dose volume, and connector type.
  • Non-clinical Testing: Bench testing according to ISO 7886-1 and USP standards (e.g., <788>, <789>, <85>) was performed to demonstrate performance equivalence.
  • Biocompatibility Testing: Tests were conducted according to ISO 10993 series standards (e.g., cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, hemolysis) to address material differences.

In summary, the provided text describes the regulatory clearance process for a physical medical device (syringe) by demonstrating substantial equivalence to a predicate, not the validation of an AI/ML-driven medical device against specific performance acceptance criteria.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

December 20, 2024

Congruence Medical Solutions, LLC Gautam Shetty Founder, CEO 3000 Falls Road Suite 200A Baltimore, Maryland 21211

Re: K243149

Trade/Device Name: Microliter Dosing Syringe (9, 20, 25, 37.5, 50 and 100 microliter models) Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QLY, FMF Dated: December 7, 2024 Received: December 9, 2024

Dear Gautam Shetty:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Shruti N. Mistry -S

Shruti Mistry Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K243149

Device Name

Microliter Dosing Syringe (9, 20, 25, 37.5, 50 and 100 microliter models)

Indications for Use (Describe)

The Microliter Dosing Syringe is intended to inject fluid into, or withdraw fluid from the body. The Microliter Dosing Syringe is indicated for intravitreal use.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K243149 510(k) Summary

Image /page/4/Picture/1 description: The image contains the logo for Congruence Medical Solutions, LLC. The logo features a stylized triangle shape at the top, with the word "CONGRUENCE" in a bold, sans-serif font below it. Underneath the word "CONGRUENCE", the text "MEDICAL SOLUTIONS, LLC" is written in a smaller font size.

510(k) NumberK243149
Preparation DateDecember 20, 2024
SubmitterCongruence Medical Solutions, LLC3000 Falls RoadSuite 200ABaltimore, MD 21211Email: gs@congruencemed.comPhone: 216-272-5521
Primary ContactGautam Shetty, PhDFounder, CEOCongruence Medical Solutions, LLC3000 Falls RoadSuite 200ABaltimore, MD 21211
Subject DeviceTrade NameMicroliter Dosing Syringe (9, 20, 25, 37.5, 50 and 100microliter models)
Common nameOphthalmic Syringe,Piston syringe
Regulation nameSyringe, Piston
Regulation number21 CFR 880.5860
Device ClassClass II
Product CodesQLY, FMF
Regulation Medical SpecialtyGeneral Hospital
510(k) Review PanelGeneral Hospital
IntendedUse/Indications forUseThe Microliter Dosing Syringe is intended to inject fluid into, or withdraw fluid from the body.The Microliter Dosing Syringe is indicated for intravitreal use.
Device DescriptionThe Microliter Dosing Syringe is a single-use, piston syringe-based device consisting of asyringe barrel, plunger stopper, syringe tip cap and incorporates a plunger rod subassembly. It isintended for use by healthcare professionals for general purpose fluid aspiration/injection. Itsoperation is manual. The Microliter Dosing Syringe is single use only, non-toxic, non-pyrogenicand sterilized by e-beam irradiation. The Microliter Dosing Syringe is suitable for ophthalmicuse.
Predicate DeviceTrade NameStaClear Syringe
ApplicantTriboFilm Research, Inc.
510(k) NumberK200242
Clearance DateJuly 27, 2020
Common NameOphthalmic SyringeSyringe, PistonNeedle, Hypodermic, SingleLumen
Regulation NamePiston syringe;Hypodermic single lumen needle
Regulation Numbers21 CFR 880.5860,21 CFR 880.5570
Device ClassClass II
Product CodesQLY, FMF, FMI
Regulation Medical SpecialtyGeneral Hospital
510(k) Review PanelGeneral Hospital
Mechanism ofActionManual operation

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Congruence Medical Solutions, LLC. The logo features a stylized triangle above the company name. The triangle is divided into two sections, one dark blue and one light blue, with an orange circle in the center. The text "CONGRUENCE" is in a dark blue sans-serif font, and "MEDICAL SOLUTIONS, LLC" is in a smaller font below it.

TechnologicalCharacteristicsThe technological characteristics of the subject device are substantially equivalent to thepredicate device with only minor differences observed in the device materials, syringe volume,connector type, biocompatibility tests completed, and performance testing completed.Biocompatibility testing and other non-clinical testing demonstrate these differences do notraise questions of safety and effectiveness when compared to the predicate device. Refer to Table1 below for a comparison oftechnological characteristics between the subject and predicate devices.
Table 1: Comparison of Technological Characteristics
CharacteristicSubject DeviceMicroliterDosing SyringePredicateDeviceStaClearSyringe(K200242)AssociatedStandard,TestingComment
Device NameMicroliter DosingSyringeStaClearSyringe--
ApplicantCongruenceMedicalSolutions, LLCTriboFilmResearch, Inc.--
510(k)NumberK243149K200242--
Intended UseThe MicroliterDosing Syringe isintended to injectfluid into, orwithdraw fluidfrom the body.The StaClearSyringe isintended toinject fluidsinto, orwithdrawfluids from,the body.PerformanceandBiocompatibilitytesting resultsprovided hereinsupport thesafety andeffectiveness ofthe device ascompared to thepredicate.Identical
Indications forUseThe MicroliterDosing Syringe isindicated forintravitreal use.The StaClearSyringe isindicated forintravitrealuse.ISO 10993-23,IntravitrealInjectionIrritationTesting, USP<788>, USP<789>Identical
Mechanism ofActionManualManualISO 7886-1Identical
SterilizationInformationProvided Sterile,single-useSterilizationMethod: Electron-beam irradiationProvidedSterile, single-useSterilizationMethod:EthyleneOxideISO 11137-1,ISO 11137-2,ISO 11737-1,ISO 11737-2,TIR13004Different
SterilityAssuranceLevel (SAL)10-610-6ISO 11137-1,ISO 11137-2,ISO 11737-1,ISO 11737-2,TIR13004Identical
Shelf Life2 years1 yearASTM F1980-21, ISO 7886-1, USP <71>Different
Device MaterialsBarrel - Cyclic olefin polymerPlunger Rod subassembly – Polycarbonate with glass fiber, Polyethylene terephthalate with glass fiber, stainless steelPlunger stopper – chlorinated butyl rubber with fluoropolymer coatingTip Cap – Chlorobutyl rubber with fluoropolymer coatingBarrel - PolypropylenePlunger - PolyethylenePlunger Stopper - Polyisoprene (Nipol IR2200, Zeon Chemicals)Needle – ASTM 304 Stainless SteelNeedle Shield - PolyethylenePlunger Cap - PolyethyleneBarrel Lubricant - Silicone Oil, crosslinked with inert argon gas plasmaISO 10993-1, ISO 10993-4, ISO 10993-5, ISO 10993-10, ISO 10993-11, ISO 10993-12, ISO 10993-23, ASTM F756Different
Syringe Volume0.35mL0.25mLISO 7886-1Different
Target Dose Volume$\pm$ 3µL (95% Confidence Level for all models)UnknownISO 7886-1Different
Connector TypeLuer lockPre-attached needleISO 7886-1, ISO 80369-7Different
CytotoxicityNo evidence of causing cell lysis or toxicityNo evidence of causing cell lysis or toxicityISO 10993-5Identical
SensitizationNo skin sensitizationNo skin sensitizationISO 10993-10Identical
Irritation or Intracutaneous ReactivityIntracutaneous Reactivity: No intracutaneous reactivityIntracutaneous Reactivity: No intracutaneous reactivityISO 10993-23Identical
Irritation, Ocular: Not considered irritants to the ocular tissueIrritation, Ocular: Not considered irritants to the ocular tissueISO 10993-23, Annex DIdentical
Irritation, Intravitreal Injection: Not considered cause intravitreal irritationIrritation, Intravitreal Injection: Not considered cause intravitreal irritationIntravitreal InjectionIrritation testingIdentical
AcuteSystemicToxicityNo evidence ofsystem toxicityNo evidenceof systemtoxicityISO 10993-11Identical
PyrogenicityNot pyrogenicNot pyrogenicISO 10993-11Identical
HemolysisNot hemolyticNot hemolyticASTM F756,ISO 10993-4Identical
ParticulateTestingConforms to USP<788> and USP<789>requirementsConforms toUSP <788>and USP<789>requirementsUSP <788>,USP <789>Identical
EndotoxinTesting≤ 0.2 EU/deviceor ≤ 0.2 EU/mLUnknown.The K200242510(k)Summary didnot containinformationaboutendotoxintestingUSP <85>No dataavailable forpredicate

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Congruence Medical Solutions, LLC. The logo features a stylized triangle shape at the top, with the left side in dark blue and the right side in light blue. There is a small orange circle in the center of the triangle. Below the triangle is the word "CONGRUENCE" in a dark blue sans-serif font, and below that is the text "MEDICAL SOLUTIONS, LLC" in a smaller font.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Congruence Medical Solutions, LLC. The logo features a stylized triangle divided into two sections, one dark blue and one light blue, with a small orange circle in the center. Below the triangle is the word "CONGRUENCE" in dark blue, and below that is the text "MEDICAL SOLUTIONS, LLC" in a smaller font.

Substantial Equivalence Discussion:

The subject device is substantially equivalent to the predicate device when evaluating intended use and technological characteristics.

  • The intended use is equivalent to the predicate device.
  • The technological characteristics of the subject device are substantially equivalent to the . predicate device with only minor differences; these minor differences do not introduce or increase safety risk.
  • . Congruence completed performance testing according to the ISO 7886-1 standard and USP <789> for subvisible particulate testing. The Microliter Dosing Syringe met the applicable requirements of ISO 7886-1 and USP <789>. These standards were also used by the predicate device TriboFilm StaClear. In addition, endotoxin testing according to USP <85> was performed with the Microliter Dosing Syringe to establish conformance to the limits recommended for ophthalmic devices in FDA guidance. The Microliter Dosing Syringe test results demonstrate its performance characteristics are substantially equivalent to the predicate device. Refer to the Non-clinical Testing section below for a list of performance testing completed.
  • . Biocompatibility testing was conducted in accordance with ISO 10993-1:2018 to demonstrate that the Microliter Dosing Syringe is as safe and effective as the legally marketed predicate device, and that any minor differences in technological characteristics do not raise questions of safety and effectiveness as compared to the predicate device (K200242). Refer to the Biocompatibility Testing section below for a list of the biocompatibility testing completed.
  • . The differences between the predicate device (K200242) and the subject device for device volume, device materials, syringe volume and shelf-life do not introduce new or different questions of safety and effectiveness when compared to the predicate device based on successful completion of biocompatibility testing, non-clinical testing and performance testing conducted to support substantial equivalence.
  • Adequate performance testing has been completed to support that the differences in connector types do not raise new or different questions of safety and effectiveness between the subject device and the predicate device. Substantial equivalence is also claimed despite difference in sterilization method (electron beam irradiation for Microliter Dosing Syringe compared to ethylene oxide sterilization for predicate device (K200242)), based on having an identical Sterility Assurance Level (SAL) of 10-6 and based on successful completion of biocompatibility testing, non-clinical testing and performance testing conducted with devices treated with electron beam irradiation.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image is a logo for Congruence Medical Solutions, LLC. The logo features a stylized triangle shape at the top, with the left side in a dark blue color and the right side in a lighter blue color. There is a small orange circle in the center of the triangle. Below the triangle is the word "CONGRUENCE" in a dark blue sans-serif font, and below that is the text "MEDICAL SOLUTIONS, LLC" in a smaller font.

Non-clinical testingBench testing was performed to demonstrate the subject device is as safe and effective as thepredicate device. The following tests were completed:
ISO 7886-1 Cleanliness Acidity and Alkalinity Extractable Metals Tolerance on Graduations Dead Space Air and Liquid Leakage Past Plunger Plunger Force Fit of Stopper Particulate Testing: USP <788> Particulate Matter in Injections USP <789> Particulate Matter in Ophthalmic Solutions Endotoxin Testing: USP <85> Bacterial Endotoxins Test
Non-clinical Testing(Biocompatibility)A chemical characterization of the device was done to evaluate the subacute/subchronic andgenotoxicity endpoints. The toxicological risk assessment demonstrated an acceptable level ofrisk of systemic exposure to the extractable compounds when classifying the Microliter DosingSyringe as externally communicating device with prolonged contact (>24 hours to 30 days) withpatient tissue contact through the injectable fluid.The following biological safety tests were also completed: Cytotoxicity Sensitization Irritation, Ocular Irritation, Intravitreal Injection Intracutaneous Reactivity Acute Systemic Toxicity Pyrogenicity Hemolysis The test article extract did not show evidence of causing cell lysis or toxicity, was not considereda sensitizer, was not considered an irritant to the ocular tissue of the rabbit, was not consideredinflammatory to intraocular tissues of the rabbit, did not show evidence of erythema and edema,did not show evidence of mortality or systemic toxicity, was not considered pyrogenic, and wasnot considered hemolytic. Therefore, this biological safety testing demonstrates the subjectdevice is biocompatible for its intended use.
Clinical TestingNot applicable - "N/A"
ConclusionIn conclusion, the Microliter Dosing Syringe is biocompatible for its intended use, demonstratesequivalent performance as the predicate device, and conforms to limits for endotoxins in single-use ophthalmic devices recommended in FDA guidance.
The suitability of the device for intravitreal injection was evaluated through biocompatibility,intravitreal injection irritation testing, particulate and endotoxin testing.
Test results confirm that the differences between the predicate and the subject device do not raisenew questions of safety or effectiveness, supporting conclusion of substantial equivalence.
The Microliter Dosing Syringe is substantially equivalent to the predicate device - StaClearSyringe (K200242) with respect to the intended use, target populations, treatment method, andtechnological characteristics.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).