K Number
K242306
Date Cleared
2024-09-04

(30 days)

Product Code
Regulation Number
882.1400
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SignalNED Device is intended to record and display Quantitative EEG (qEEG) (relative band power, e.g., alpha, beta, delta, theta), which is intended to help the user analyze the EEG. The SignalNED does not provide any diagnostic conclusion about the patient's condition. The device is intended to be used on adults by qualified medical and clinical professionals.

The SignalNED is intended to be used in a professional healthcare environment.

Device Description

The SignalNED Model RE machine uses 10 patient electrodes (4 left, 4 right, 2 midh are used to form the 8 channels. The SignalNED machine requires the use of the SignalNED Sensor Cap, and the system includes the following components:

  • Portable EEG machine (Device)
  • I Battery & External Battery Charger
  • I SignalNED Sensor Cap
  • I SignalNED Sensor Cap Cable

The primary function of the SignalNED Model RE is to rapidly record EEG and derive the Quantitative EEG (qEEG) measurement of Relative Band Power for multiple bands (e.g., alpha, beta, theta) at each electrode. These measurements are intended to help the user analyze the underlying EEG. The SignalNED Model RE (client) achieves its intended without relying on wireless comectivity. The SignalNED RE does not provide any diagnostic conclusion about the patient's condition.

AI/ML Overview

This summary describes the acceptance criteria and the study that proves the SignalNED System (Model RE) meets those criteria, based on the provided FDA 510(k) summary.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Lead Off DetectionAbility to detect disconnected electrodes.All testing passed acceptance criteria.
Signal Acquisition Noise LevelsAcceptable noise levels in signal acquisition.All testing passed acceptance criteria.
Software ADC Conversion AccuracyAccuracy of software in Analog-to-Digital Converter (ADC) conversion.All testing passed acceptance criteria.
Quantitative Electroencephalogram (QEEG)Accuracy of the QEEG Relative Band Power calculation.All testing passed acceptance criteria.
EC12:2020 Electrical PerformanceCompliance with EC12:2020 electrical standards.All testing passed acceptance criteria.
Essential Performance Tests (IEC 80601-2-26)Compliance with IEC 80601-2-26 essential performance requirements.All testing passed acceptance criteria.
Electrical Performance (IEC 60601-1, IEC 60601-1-2)Compliance with IEC 60601-1 and IEC 60601-1-2.All testing passed.
Biocompatibility (ISO 10993-1, -5, -10, -23)Compliance with ISO 10993 for Cytotoxicity, Sensitization, and Irritation (for limited contact, intact skin).All testing passed.

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not specify the sample sizes (e.g., number of subjects, number of EEG recordings) used for the non-clinical performance testing. It only states that "All testing passed acceptance criteria and details are contained in the test report." The data provenance (e.g., country of origin, retrospective or prospective) is also not detailed in this summary.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The provided document describes non-clinical performance testing (lead-off detection, noise levels, ADC accuracy, QEEG calculation accuracy, electrical performance, biocompatibility). These tests do not typically involve human experts establishing ground truth in the way a clinical study for diagnostic accuracy would. The ground truth for these tests would be established through defined technical specifications, measurement standards, and validated testing protocols. Therefore, information about the number and qualifications of experts for establishing ground truth is not applicable in this context.

4. Adjudication Method for the Test Set

As the performance testing described is non-clinical and based on technical specifications and standards, an adjudication method (like 2+1 or 3+1) used in clinical studies for discrepancies in expert readings is not applicable here. The acceptance criteria for each test inherently define the "ground truth" to which the device's performance is compared.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. without AI Assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned or described in the provided document. The SignalNED System is intended to record and display QEEG, which "is intended to help the user analyze the EEG." It explicitly states, "The SignalNED does not provide any diagnostic conclusion about the patient's condition." This indicates that the device is a tool for professional analysis rather than an AI-driven diagnostic aid for human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is not described.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The document describes several standalone performance tests for the device's components and calculations (e.g., Lead Off Detection, Signal Acquisition Noise Levels, Software ADC Conversion Accuracy, Quantitative Electroencephalogram (QEEG) accuracy). These tests are conducted on the algorithm and hardware without human interpretation as part of the primary outcome assessment. For instance, the "Software ADC Conversion Accuracy" and "Quantitative Electroencephalogram (QEEG)" accuracy tests evaluate the algorithm's performance in generating calculated EEG measures.

7. The Type of Ground Truth Used

The ground truth used for the reported performance tests is based on:

  • Defined Technical Specifications and Engineering Standards: For tests like Lead Off Detection, Signal Acquisition Noise Levels, Software ADC Conversion Accuracy, EC12:2020 Electrical Performance, and Essential Performance Tests (IEC 80601-2-26).
  • Validated Calculation Methods: For the Quantitative Electroencephalogram (QEEG) Relative Band Power calculation, the ground truth would be based on established mathematical and signal processing principles for deriving these metrics from raw EEG data.
  • International Biocompatibility Standards: For ISO 10993 series tests (Cytotoxicity, Sensitization, Irritation).

8. The Sample Size for the Training Set

The provided document describes performance testing for substantial equivalence, not the development or validation of a machine learning model with distinct training and test sets in the typical sense. While the device calculates QEEG, the details on how the underlying algorithms were developed or "trained" (if machine learning is involved beyond standard signal processing) are not provided. Therefore, a specific sample size for a "training set" is not mentioned.

9. How the Ground Truth for the Training Set Was Established

As information about a distinct "training set" for machine learning algorithms is not provided, the method for establishing ground truth for such a set is also not described. The document focuses on performance testing against established engineering, electrical, and biocompatibility standards.

{0}------------------------------------------------

September 4, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Forest Devices, Inc. % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059

Re: K242306

Trade/Device Name: SignalNED System (Model RE) Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OMC, OLT, GWQ Dated: August 5, 2024 Received: August 5, 2024

Dear Dave Yungvirt:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Image /page/2/Picture/3 description: The image shows the name "Patrick Antkowiak -S" in black text. The text is positioned to the right of a faded blue FDA logo. The name is split into two lines, with "Patrick" on the first line and "Antkowiak -S" on the second.

for

Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242306

Device Name SignalNED System (Model RE)

Indications for Use (Describe)

The SignalNED Device is intended to record and display Quantitative EEG (qEEG) (relative band power, e.g., alpha, beta, delta, theta), which is intended to help the user analyze the EEG. The SignalNED does not provide any diagnostic conclusion about the patient's condition. The device is intended to be used on adults by qualified medical and clinical professionals.

The SignalNED is intended to be used in a professional healthcare environment.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K242306

510(k) Summary

Traditional 510(k) Premarket Notification Summary of Safety and Effectiveness

This 510(k) summary was prepared on 5/28/2024 to provide an understanding of the basis for a determination of substantial equivalence in accordance with the requirements outlined in 21 CFR 807.92. The content is also contained with the eSTAR format within Submitter's eSTAR application.

Submitter Information (Contact Details)

Applicant Name:Forest Devices, Inc.
Applicant Address:544 Miltenberger St 1st Floor, #4 Pittsburgh PA 15219 United States:
Applicant Contact:Telephone: 412.522.0076
Applicant Contact:Email: carmelo.montalvo@forestdevices.com)

Date Prepared: September 3, 2024

Device Name

Device Trade Name:SignalNED System (Model RE)
Common Name:Electroencephalograph
Classification Name:Reduced-Montage Standard Electroencephalograph
Regulation Number:882.1400
Regulatory Class:Class II

Product Code(s)

Primary Product Code: OMC Secondary Product Codes: OLT and GXY

Legally Marketed Predicate Devices

PredicatePredicate Trade Name(Primary Predicate is listed first)Product Code
K170363Ceribell Pocket EEG DeviceOMC, Reduced-Montage Standard Electroencephalograph (Primary)
K143487Lifelines iEEG 2.0OLT, Non-Normalizing Quantitative Electroencephalograph Software (Secondary)
K171459Ceribell Instant EEG HeadbandGXY, Cutaneous Electrode (Secondary)

Device Description Summary

The SignalNED Model RE machine uses 10 patient electrodes (4 left, 4 right, 2 midh are used to form the 8 channels. The SignalNED machine requires the use of the SignalNED Sensor Cap, and the system includes the following components:

  • Portable EEG machine (Device)
  • I Battery & External Battery Charger
  • I SignalNED Sensor Cap
  • I SignalNED Sensor Cap Cable

The primary function of the SignalNED Model RE is to rapidly record EEG and derive the Quantitative EEG (qEEG) measurement of Relative Band Power for multiple bands (e.g., alpha, beta, theta) at each electrode. These measurements are intended to help the user analyze the underlying EEG. The SignalNED Model RE (client) achieves its intended without relying on wireless comectivity. The SignalNED RE does not provide any diagnostic conclusion about the patient's condition.

Intended Use/Indications for Use

The SignalNED Device is intended to record and store EEG signals and display Quantitative EEG (qEEG) (relative band power, e.g., alpha, beta, delta, theta), which is intended to help the EEG. The SignalNED does not provide any diagnostic conclusion about the patient's condition. The device is intended to be used on adults by qualified medical and clinical professionals. The SignalNED is intended to be used in a professional healthcare environment.

{5}------------------------------------------------

Substantial Equivalence Comparison Summary

AttributeSubject DeviceThe SignalNED Model RE SystemK PendingPrimary Predicate DeviceCeribell Pocket EEGK170363Comparison
Intended UseTo record and store EEG signalsTo record and store EEG signalsIdentical.
Presents EEG signals in visual formats in device real-timePresents EEG signals in visual formats in device real-timeIdentical.
Device does not have a diagnostic claimDevice does not have a diagnostic claimIdentical.
Device does not provide any diagnostic conclusions aboutthe patient's conditionDevice does not provide any diagnostic conclusions aboutthe patient's conditionIdentical.
Used on AdultsUsed on AdultsIdentical.
TechnologyRecords and stores EEG signalsRecords and stores EEG signalsIdentical.
Uses specially designed cutaneous electrodesUses specially designed cutaneous electrodesIdentical.
Uses specially designed cutaneous electrodesUses specially designed cutaneous electrodesIdentical.
Uses analog to digital converterUses analog to digital converterIdentical.
Uses frequency band of 0.5Hz - 100HzUses frequency band of 0.5Hz - 100HzIdentical.
Device samples at 250HzDevice samples at 250HzIdentical.
Displays calculated EEG measuresDisplays calculated EEG measuresIdentical.
Use 2.4 GHz IEEE 802.11 b/g/n WIFI frequencyUse 2.4 GHz IEEE 802.11 b/g/n WIFI frequencyIdentical.
Uses Integrated AntennaUses Integrated AntennaIdentical.
Features surface mount PCB moduleFeatures surface mount PCB moduleIdentical.
Battery poweredBattery poweredIdentical.

Intended Use/Indications for Use Comparison

Primary Predicate: K170363 Ceribell Pocket EEG

The SignalNED Model RE System (Subject Device) is identical to the Primary Predicate (Predicate) in Intended to record and store EEG signals and present them in real-time. Neither the Subject Device nor Primary Predicate provides any diagnostic conclusion about the subject's condition or automated alerts of a clinical event. Additionally, neither device has a diagnostic claim, nor do they provide any diagnostic conclusions about the patient's condition. Both the Primary Predicate are intended to be used on adults by qualified medical and clinical professionals.

Technology Comparison

Device & Predicate Device(s):K242306K170363OMCPredicateK143487OLTPredicateK171459GWQPredicateDiscussion
General Device Characteristics
EEG Data Acquisition10 electrodes toform 8 recordingchannels10 electrodes toform 8 recordingchannelsSimilar
Data Acquisition MethodAnalog to digitalconverterAnalog to digitalconverterSimilar
Frequency band0.5 Hz - 100 Hz0.5 Hz - 100 HzSimilar
Sampling rate250 Hz250 HzSimilar
Displays calculated EEG measuresYesYesSimilar
WIFI Frequency Standard2.4 GHz IEEE802.11 b/g/n2.4 GHz IEEE802.11 b/g/nSimilar
Antenna TypeIntegratedIntegratedSimilar
Form Factor / DimensionsSurface mountPCB moduleSurface mountPCB moduleSimilar
Integrated Processor and Memory1.8GHz QuadCortexTM-A53STM32 ARMCortex-M3Similar; Thesubject device
Device & Predicate Device(s):K242306K170363OMCPredicateK143487OLTPredicateK171459GWQPredicateDiscussion
NXP i.MX 8MPlusprocessor with64 KB RAMincludesadditionalprocessing andmemorycapability
Device PowerBattery PoweredBattery PoweredSimilar
Battery100-240 V ACpower adapter tobattery charger100-240 V ACpower adapter tobattery chargerSimilar
Differential input impedance>2 Gohm>20 MohmSimilar;electricalspecificationexceedspredicate device.This differencedoes not impactsafety andeffectiveness.
Common mode input impedance>1 Gohm>100 MohmSimilar:electricalspecificationexceedspredicatedevice. Thisdifference doesnot impact safetyandeffectiveness.
Channel equivalent input noise1.97 μVpp<3.5 µVppSimilar;electricalspecificationexceedspredicatedevice. Thisdifference doesnot impact safetyandeffectiveness.
Frequency band0.5 Hz - 30 Hz0.16 Hz - 70 Hz(-6 dB)Similar; withinrange ofpredicate. Thisdifference doesnot impact safetyandeffectiveness.
Low filter0.5 Hz0.5 HzSame
High filter30 Hz10 Hz - 100 HzSimilar; withinrange ofpredicate. Thisdifference doesnot impact safetyandeffectiveness.
Sampling rate1 kHz200,256 HzSimilar;
Device & Predicate Device(s):K242306K170363OMCPredicateK143487OLTPredicateK171459GWQPredicateDiscussion
specificationexceedspredicatedevice. Thisdifference doesnot impact safetyandeffectiveness.
Displays calculated EEG measuresYesYesSame
Calculated EEG measures displayedPower SpectrumDensitySpectrum, PowerSpectrum,Density, bandpower, spectraledge, peakfrequencySimilar; subjectdevice is asubset to thepredicatedevice.
WIFI Frequency Standard2.4 GHz IEEE802.11 b/g/n2.4 GHz IEEE802.11 b/g/nSimilar
Antenna TypeIntegratedIntegratedSimilar
Where usedOn the headOn the headSimilar
Means of ConnectionA cableattached to theSensor Cap toallowconnection toSignalNEDModel RE deviceA cableattached tothe headbandto allowconnectionto an EEGacquisition/recording deviceSimilar
Electrode TypeSilver/silver-chloride(Ag/AgCl)electrodesSilver/silver-chloride(Ag/AgCl)electrodesSimilar
Number of Electrodes10 electrodes10 electrodesSimilar
Electrode FormReservoir isfilled withconductiveelectrolyte gelby separatedispensingcontainerReservoir isfilled withconductiveelectrolyte gelby separatedispensingcontainerSimilar
Electrode LocationsThe placement ofthe electrodes isaccording to theInternational 10-20 system ofelectrodeplacement or theAmericanElectroencephalographic Societypositioningsystem.The placementof the electrodesis according tothe International10- 20 system ofelectrodeplacement or theAmericanElectroencephalographic Societypositioningsystem. Thenumber of theelectrodes in useis according toSimilar; bothuse theInternational10-20 system ofelectrodeplacement orthe AmericanElectroencephalographicSocietypositioningsystem. TheSubject deviceuses 10electrodes
Device & Predicate Device(s):K242306K170363OMCPredicateK143487OLTPredicateK171459GWQPredicateDiscussion
whereas thePredicate devicehasconfigurationoption from 9-19electrodes.
Electrode MountsThermoplasticElastomerThermoplasticElastomerSame
Available SizesVarious sizes(overall headsize range 26 cm–66 cm)Various sizes(overall headsize range 26cm –66 cm)Same
Cap MaterialSpandex blend:Nylon KnittedFabric, 82%Nylon and 18%SpandexSpandex blend:Nylon KnittedFabric, 82%Nylon and 18%SpandexSame
Type of ConnectorCustom Touch-proof safetysocketTouch-proofsafety socketDIN 42- 802Similar; theelectricalperformancetesting completedfor both Subjectand Predicate metthe samerequirements perFDA guidance.document"CutaneousElectrodes forRecordingPurposes-PerformanceCriteria for Safetyand PerformanceBased Pathway,"August 2020.

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Intended Use/Indications for Use Comparison

Secondary Predicate: K143487 Lifelines iEEG 2.0

The Subject is identical to the Secondary Predicate (K143487, Lifelines iEEG 2.0) in Intended to acquire, display, and store EEG signals, using cutaneous electrodes. Neither the Subject Device nor Secondary Predicate provides any diagnostic conclusion about the subject's condition or automated alerts of a clinical event. Additionally, neither device claim, nor do they provide any diagnostic conclusions about the patient's condition.

Technology Comparison

Secondary Predicate: K143487 Lifelines iEEG 2.0

Both the Subject Device and Secondary Predicate (K143487) are identical in that they display Quantitative Electroencephalograph (qEEG) measures that are derived from the acquired EEG. They are similar in differential input impedance, common mode input impedance, channel equivalent input noise, frequency band, low filter, sampling rate. The Subject exceeds the predicate in electrical specifications or is within the specified range of the Predicate.

Intended Use/Indications for Use Comparison

{9}------------------------------------------------

Secondary Predicates: K171459 Ceribell Instant EEG Headband

The Intended Use of the Subject Device and the Secondary Predicate (K171459, Cerbell Instant EEG Headband) is identical. Both use a disposable headpiece with an integrated array of 10 passive cutaneous electrodes that are applied to the patient's head to record EEG signals when connected to an EEG recording device.

Technology Comparison

Secondary Predicates: K171459 Ceribell Instant EEG Headband

The technology of the Subject Device and the Secondary Predicate (K171459, Cerbell Instant EEG Headband) is identical in type of patient contact, where used, means of connection, electrodes and electrode form. The Subject and Secondary Predicate both follow placement of the electrodes is according to the International 10-20 system of the American Electroencephalographic Society positioning system. The Subject uses 10 electrodes, and the Seconder 9-19 electrodes. The electrode mounts, sizes electrode cap materials. performance requirement, electrical compliance, designation for use are identical. Both are packaged non-sterile. Both demonstrate conformance to ISO 10993-1, ISO 10993-5, and ISO 10993-10.

TestingObjective and Outcome
Lead Off Detection■ Test is designed to validate the ability to detect disconnected electrodes.■ All testing passed acceptance criteria and details are contained in the test report.
Signal Acquisition Noise Levels■ Test is designed to assess noise levels in signal acquisition.■ All testing passed acceptance criteria and details are contained in the test report
Software ADC Conversion Accuracy■ Test is designed to verify accuracy of software in ADC conversion.■ All testing passed acceptance criteria and details are contained in the test report.
Quantitative Electroencephalogram(QEEG)■ Test is designed to ensure accuracy of the QEEG Relative Band Power calculation.■ All testing passed acceptance criteria and details are contained in the test report.
EC12:2020 Electrical Performance■ Test is designed to confirm compliance with EC12:2020 electrical standards.■ All testing passed acceptance criteria and details are contained in the test report.
Essential Performance Tests(IEC 80601-2-26)■ Test is designed to verify compliance with IEC 80601-2-26 essential performance requirements.■ All testing passed acceptance criteria and details are contained in the test report.

Summary of Non-clinical Performance Testing

Electrical Performance

Electrical testing was conducted on the subject device complies with IEC 60601-1, IEC 60601-1-2. All testing passed.

Biocompatible Performance

The Forest SignalNED System is an EEG device used by trained medical professionals. The EEG machine datient, but the Sensor Cap, which is part of the system, is placed on the patient's head (Limited Exposure less than 24 hours, Surface Contacting Medical Device that contacts intacts ). The SignalNED was assessed for biocompatibility under the requirements of ISO 10993 for the Biological Evaluation of Medical Devices for cytotoxicity, irritation within the Limited Contact duration (A), for Surface Medical Device for Intact Skin scope.

EN-ISO-10993-1:2020 the following test are the end points for our biocompatibility testing:

  • I ISO 10993-1: Risk Management
  • ISO 10993-5: Cytotoxicity
  • 트 ISO 10993-10: Sensitization
  • I ISO 10993-23: Irritation or intracutaneous reactivity

All testing passed.

Conclusion

In conclusion, Forest Devices, Inc. has successfully demonstrated that the SignalNED is equivalent in safety and efficacy to the predicate in the intended use, technological characteristics and performance testing. The Forest Devices SignalNED Model RE System raises no additional questions of safety or effectiveness and is substantially equivalent to the cleared with the ences in the Subject Device and predicates do not raise additional questions of safety or effectiveness or exceeds all conditions that are required for establishing substantial equivalence and obtaining FDA clearance to market our medical device, the SignalNED Model RE System.

§ 882.1400 Electroencephalograph.

(a)
Identification. An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.(b)
Classification. Class II (performance standards).